Current Meter Maintenance Model OSSPC1 Hydrographic Training Bandar Seri Begawan Brunei August 2010.

Slides:



Advertisements
Similar presentations
72V - Recoil, Recoil IS, & Instinct New Vehicle Delivery Tips
Advertisements

Electricity.
MAGNETIC EFFECTS OF ELECTRIC CURRENT
Opening/Closing Rail Car Plug and Sliding Doors
USER MAINTENANCE.
Current Meter Maintenance Model OSS-B1 Hydrographic Basics Training Course No
Brio 2000 / Brio 2000-M E-Learning Service Manual Rev. 0.0.
K&W Permanent Metallic ® Block Seal
Current Meter Maintenance HS Model OSS-B1 Hydrological Services Pty Ltd Presented to HiQ Engineering Brunei JULY 2009.
Ice. Pure and Simple ™ Welcome to the Ice-O-Matic ® on-line training modules.
Safe Use and Operation of Vacuum Lines
Centrifuges. Topics  Principle of Operation  Diagrams  Applications  Types of Centrifuges  History  Safety  Operation  Preventive Maintenance.
Ford F Series Air Box Modification. Ford F series model year 2011 trucks have plastic air box assemblies. It has been discovered that a plastic tab on.
ATLIER FRANCOIS (COMPRESSOR). MAINTENANCE AND CHECKS DAILY CHECKS DAILY CHECKS INSPECT AND NOTE THE OIL LEVEL AND ALL THE PRESSURE AND TEMPERATURE LEVELS(FILLL.
A Manual for Repair and Maintenance Technicians 1ST STAGE.
FoamPro 1600 Series Fully automatic Class A foam ONLY
© Blue Graphics Concept Sauer-Danfoss START-UP PROCEDURE PROPEL PUMP Instruction Sauer-Danfoss Belgium – T. Van Mossevelde.
1 USDOC41 Rev 0 1/04 Training and Education Program ©2004 HemoCue, Inc. HemoCue ® Training and Education Program The HemoCue Hb System.
Cleaning H50/52 Blowers The condition of the blower motor has a direct affect on the operation of an H50/52 fryer.
CEILING FAN Object:- To study the part dismantling reassembling testing and repairing of ceiling fan. By:- Surya Lal TGT (WE) K.V.2,FCI GORAKHPUR.
If you use your computer more than a few hours per day, we suggest you not to turn off your computer at the end of the day. Instead turn off only the.
Engine Systems and Components
Section 4: Electric Motors
Power-Fill Operation Demonstration Power-Fill Operation Demonstration (See Installation and Operation Manual for detailed instructions)
Suction Pumps.
Introduction to FLUID COPLING.
Know You System Fire Extinguisher & Suppression Activation Training.
Inspection and Maintenance of Isotope Radiography Systems
Equipment Familiarization Amatrol 850 Basic Hydraulic Training System
AccuMax Multi-Point Injection Mechanics
Electrical Electricity Fuses Relays Solenoids Electrical Testing
Tru Count Air Clutch Maintenance
Preventative Maintenance & Troubleshooting
Bearings -Disassembly &Assembly Nizwa College of Technology.
© 2015 Cengage Learning. All Rights Reserved. May not be scanned, copied, duplicated, or posted to a publicly accessible website, in whole or in part.
Chapter Five Maintaining a Computer Part I: Cleaning a Computer.
By: Marideth Bradley. OBJECTIVES Identify problems that can occur if hardware is not properly maintained. Identify routine maintenance that can be performed.
Respirator Safety This Training Covers Respirator Inspection User Fit Checks Respirator Cleaning Respirator Storage.
Technical presentation SC350
1 Portable Powered Tools Use the Right Tool the Right Way.
Flat Saw Training. Flat Saw Sizes –Low Horsepower Blade diameters range from 8” (200mm) to 18” (450mm) Power ranges from 4 to 25 horsepower.
Tempest Blower SECTION: Tools and Equipment ISSUED: REVISED: ##-####
Lesson Using Pneumatic Systems. Interest Approach Can you explain how air makes a pneumatic hand tool work?
Why Maintenance Tasks are Important Produced by Joe Marchese.
Student Book © 2004 Propane Education & Research CouncilPage Performing Gas Distribution System Leak Checks Requirements for vapor distribution.
Electromagnetism Vanderbilt Student Volunteers for Science Training Presentation Spring 2016.
1  2008 Kaeser Compressors, Inc., USA (V2.2) “Specifications are subject to change without notice” Presenting... Kaeser Omega Blowers.
MAINTENANCE INTERVAL SCHEDULE D6R TRACK TYPE TRACTOR 7AR EVERY 10 SERVICE HOURS OR DAILY.
Creating a Clean and Hygienic Foodservice Operation
Procedure for Operating Leaf Blowers (Revision 1 – 17th November 2010) Toolbox Talk.
TROUBLESHOOTING ELECTRIC MOTORS
Universal Current Meter Training Model OSS-B1 Hydrological Services Pty Ltd Available for rent from Tech-Rentals.
Vanderbilt Student Volunteers for Science Training Presentation
Model : ASC Airswitch Installation Instructions
14 Refrigerant Recovery, Recycling, and Recharging.
E2 Motors and Motor Starting
Module 9: Microscope Maintenance
Know You System Fire Extinguisher & Suppression Activation Training
WASHERS.
Fuel Supply System Service
MAINTENEANCE AND REPAIR OF STARTING SYSTEMS
Engine Oil Changes.
Engine Oil Changes.
Fuel Supply System Chapter 21 Lesson 1.
How to set up the Flexmort CuddleCot™
Ford F Series Air Box Modification
TROUBLESHOOTING ELECTRIC MOTORS
TRENTON Casing Filler Presentation
Correct Procedures for Changing, Storing, and Maintaining Saw Blades
Fluid Power System Electrical Control
Presentation transcript:

Current Meter Maintenance Model OSSPC1 Hydrographic Training Bandar Seri Begawan Brunei August 2010

Ensures that induced errors in measurement from ‘mechanical’ sources are eliminated Prolongs life of current meter and components Why Is This Important?

Section 1 Prior to Gauging

Procedure

Important Points Dispose of used current meter oil in a thoughtful manner Place components on lint free cloth or paper Ensure components do not come into contact with dust Use ONLY approved HS current meter oil (PYG-17)

Spin Test Prior to spin test hold the current meter in the horizontal position, with fan facing directly to your mouth Blow air forcefully and allow fan to spin for approx. one minute (to minimise any hydraulic pressure ) DO NOT SPIN BY HAND

Spin Test Spin test the meter against time (as per table shown) Connect current meter to the counter and test operation Fan No.Minimum Time (secs) Fan No.Minimum Time (secs) Shaft Diameter 4.0mm Shaft Diameter 2.5mm

Section 2 During and After Discharge Measurements

General Rule After every discharge measurement, (or one hour of use), the current meter should be disassembled, cleaned, re-filled with oil and reassembled, if continuing a discharge measurement.

Procedure Remove fan from meter shaft and dispose of used oil from meter body Disassemble meter as shown in Section 1 Undo the cleaning jar cap and ‘¾ fill’ with white spirit

Place components on lint free cloth or paper (do not place in contact with dirt particles) Place shaft in cleaning jar and agitate in order to flush out used oil Fill body of meter with white spirit, place thumb over opening and shake Empty and dispose of white spirit from meter body

After Discharge Measurements Refill with oil if continuing a gauging Blow dry the shaft assembly, fan carrier and fan. Assemble ‘dry’ if meter going into storage

After Discharge Measurements Spin test the current meter against time as shown in Section 1 After the spin test is performed connect your current meter to the counter using the lead, to test the lead, reed switch and counter.

Section 3 Office Service

When? Once per month (if current meter is used regularly (daily?) Use same procedure described under Section 2 How?

Section 4 Calibration

Calibration Re-calibrate the current meter every 300 hours of use or once a year which ever comes first (see Hydrological Services calibration certificate for details) To minimise freight costs, send only the body and fans requiring re-calibration Note: Hydrological Services provides a re- calibration and repair service for all types of current meters.

Section 5 Troubleshooting

SYMPTONPOSSIBLE CAUSES ACTIONREMARK Failed spin testDirty bearings Faulty bearings Clean bearings Replace bearings if problem persists If fan stops suddenly towards end of spin test, likely cause is dirty bearings Failed spin testBent shaftIf shaft is bent, return to HS for repair or replace with new shaft Bent shaft can be observed by looking directly at point of fan during a spin test. If this point is turning ‘off centre’, shaft is bent Fan out of balanceFan squashed or dropped Remove the fan, rotate the shaft approx. 50° and re- fit the fan on to the shaft. If problem persists return the fan and body to HS for repair If fan stops and rotates in opposite direction at end of spin test, this indicates that the fan is out of balance MECHANICAL PROBLEMS

SYMPTONPOSSIBLE CAUSESACTION Fault occurs only when meter is immersed in water Water entering contact plug receptacle and earthing between body and plug Replace damaged plug (see manual) Circuit closed in any magnet position, but open with magnet removed Reed switch failedReplace reed switch assembly Circuit closed without magnet in position Reed damagedReplace reed switch assembly ELECTRICAL PROBLEMS (Closed Circuit Fault)

SYMPTONPOSSIBLE CAUSESACTION Open circuit between plug and insulated reed contact Faulty plug contactClean contact in plug Re-test Open circuit between plug and insulated reed contact Faulty insulated contact assembly Replace No closure of reed switch contact when magnet in position Damaged reedReplace ELECTRICAL PROBLEMS (Open Circuit Fault)

Questions?