Validation of screening methods (2002/657/EC)

Slides:



Advertisements
Similar presentations
A ERROR : FALSE POSITIVE ß ERROR : FALSE NEGATIVE ANALYTICAL LIMITS consumer risk producer risk non-compliant compliant Ref: H. De Brabander.
Advertisements

Quality is a Lousy Idea-
Dr. Birgit Schmauser, BfArM, Bonn
Calibration Techniques
Analytical Method Development and Validation
Instrumental Analysis
1 Method Selection and Development l Initial Considerations n What does the method need to do? 3 What analyte/s need to be assayed? 3 What range or concentration.
CHEMISTRY ANALYTICAL CHEMISTRY Fall
Chemical Analysis Qualitative Analysis Quantitative Analysis Determination “Analyze” a paint sample for lead “Determine” lead in a paint sample.
PHARM 462 PART / /31 Good Manufacturing Practices (GMP) VALIDATION of ANALYTICAL TEST METHODS.
World Health Organization
FEDERAL PUBLIC SERVICE (FPS) HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT 1 Validation of confirmatory methods for MRPL and MRL substances in matrices of.
Accreditation & Validation
Summary 1 l The Analytical Problem l Data Handling.
Limitations of Analytical Methods l The function of the analyst is to obtain a result as near to the true value as possible by the correct application.
Validation of Analytical Methods
Kyiv, TRAINING WORKSHOP ON PHARMACEUTICAL QUALITY, GOOD MANUFACTURING PRACTICE & BIOEQUIVALENCE Validation of Analytical Methods Used For Bioequivalence.
Chemometrics Method comparison
Instrumental Analysis
Validation: concept, & considerations
Validation of Analytical Method
Laboratory Validation of Analytical Methods
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
Analytical considerations
1 MARLAP SCIENCE ADVISORY BOARD CHAPTERS 6 & 7 Stan Morton DOE-RESL April 2002.
What is analytical chemistry? 5 subdisciplines of chemistry Some definitions In every day life… Analytical chemist versus chemical analyst.
Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.
Introduction to Analytical Chemistry Dr M. Abd-Elhakeem Faculty of Biotechnology General Chemistry Lecture 7.
Quality WHAT IS QUALITY
How to Select a Test Method Marlene Moore Advanced Systems, Inc. June 15, 2010.
Bias and Errors. Some Terms Used to Describe Analytical Methods Accuracy Precision LOD RDL LOQ Selectivity Sensitivity Linearity Ruggedness.
Chapter 5 Errors In Chemical Analyses Mean, arithmetic mean, and average (x) are synonyms for the quantity obtained by dividing the sum of replicate measurements.
Quality Control Lecture 5
Development and validation of Biacore methods for the detection of streptomycin/dihydrostreptomycin, gentamicin and neomycin in milk VIIth International.
Quality Assurance How do you know your results are correct? How confident are you?
5. Quality Assurance and Calibration Quality assurance is We do to get the right answer for our purpose. Have Sufficient accuracy and precision to support.
1 International Federation for Consulting GmbH IFC Validation on the basis of Commission Decision 2002/657/EC by Dr. Michael Scheutwinkel.
1 Exercise 7: Accuracy and precision. 2 Origin of the error : Accuracy and precision Systematic (not random) –bias –impossible to be corrected  accuracy.
Introduction The importance of method validation
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
Limit of detection, limit of quantification and limit of blank Elvar Theodorsson.
Wenclawiak, B.: Fit for Purpose – A Customers View© Springer-Verlag Berlin Heidelberg 2003 In: Wenclawiak, Koch, Hadjicostas (eds.) Quality Assurance in.
Industrial Technology Institute Test Method Validation & Verification H.P.P.S.Somasiri Principal Research Scientist / SDD-QAD /QM Industrial Technology.
Chapter 1: Introduction Analytical Chemistry Analytical Chemistry Qualitative analysis Qualitative analysis Quantitative analysis Quantitative analysis.
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
 Routine viral diagnostics: indirect and direct detection of viruses. ◦ Indirect detection: serological tests; ◦ Direct detection:  Viral antigens;
Lecture 10 ANALYTICAL METHOD DEVELOPMENT AND VALIDATION IN HPLC AND GC. Lecture 10 – Chromatography, Dr. Rasha Hanafi 1© Dr. Rasha Hanafi, GUC.
SEMINAR ON PRESENTED BY BRAHMABHATT BANSARI K. M. PHARM PART DEPARTMENT OF PHARMACEUTICS AND PHARMACEUTICAL TECHNOLGY L. M. COLLEGE OF PHARMACY.
Chapter 5 Quality Assurance and
EQUIPMENT and METHOD VALIDATION
means to “TO CHECK OR PROVE THE VALIDITY OF” According to FDA – “ The goal of validation is to establish a documented evidence which provides a high degree.
Art PowerPoints Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 05: QUALITY ASSURANCE AND CALIBRATION METHODS.
Method Validation-where do I start?
Multi-Analyte LC-MS/MS Methods – Best Practice.
Quality is a Lousy Idea-
Is My Method Fit For Purpose ?
World Health Organization
Dr. Birgit Schmauser, BfArM, Bonn
Validation of Screening Methods
Quality is a Lousy Idea-
Update on ASTM and Standard Methods method development activities
Analytical Method Validation
METHOD VALIDATION: AN ESSENTIAL COMPONENT OF THE MEASUREMENT PROCESS
Chapter 5 Quality Assurance and Calibration Methods
Environmental Laboratory Certification Program (ELCP)
ANALYTICAL METHOD VALIDATION
World Health Organization
Satish Pradhan Dnyanasadhana College, Thane. Department of Chemistry S
Introduction to Analytical Chemistry
Multi-laboratory Method Validation
Presentation transcript:

Validation of screening methods (2002/657/EC) N. Van Wouwe IPH AFSCA-FAVV

Definition (2002/657/EC) Screening method : used to detect the presence of a substance or class of substances at the level of interest. have the capability for a high sample throughput => are used to sift large numbers of samples for potential non-compliant results. Exemple: ELISA, plate test, biosensor, receptor test,…

Definition (2002/657/EC) Minimum criteria to use an analytical method as screening method: must be validated (traceability) must have a false compliant rate of <5% (β-error) at the level of interest

Performance characteristics for method validation (screening) Qualitative method: identifies a substance on basis of its chemical, biological or physical propriety (binary response: +/-, absence/presence) Quantitative method: determines the amount or mass fraction of a substance (response: numerical value of appropriate unit) + = determination is mandatory

Validation of screening test Definition of the scope of the method Analyte of group of analytes Range of concentration List of matrices Initial validation with the most often used matrice in national monitoring program Detection capacity (CCβ) Selectivity/Specificity Applicability/ Ruggedness/Stability Precision (only for semi-quantitative method) If possible: different sources of blank material, different technicians, different days on the same spiked sample

Validation of screening test Targeted test: for 1 compound validation for this compound Targeted test: for a family of compounds validation for 1 representative molecule of the family (antibody) Wide range test: for more than 50 different molecules Validation for at least a list of representative compounds Common pattern of activity on a specific bacteria? Common way of action (acting target)? Published reference data on validation available?

Proposition of the CRL for antimicrobials (in milk) Representative Compound Antimicrobial class Cefalonium/Cephapirin/Cefquinome CEPHALOSPORINS Penicillin G/Cloxacillin PENICILLINS Tetracycline/Doxcycline TETRACYCLINS Gentamicin/Streptomycin/Spectinomycin AMINOGLYCOSIDES Enrofloxacin/Flumequine QUINOLONES Sulfathiazole/sulfaguanidine/Sulfamerazine SULFONAMIDES Erythromycin/Tylosin MACROLIDES Lincomycin LINCOSAMIDES Thiamphenicol PHENICOLATED Trimethoprim/colistine MISCELLANEOUS

Performance characteristics Detection capacity Selectivity/Specificity Applicability/ Ruggedness/Stability Precision (only for semi-quantitative method)

Detection capability (CCβ) The smallest content of the substance that may be detected, identified and/or quantified in a sample with an error probability of β In case of MRPL, CCβ= lowest concentration at which the method is able to detect truly contaminated sample with a statistical certainty of 1-β In case of MRL, CCβ= concentration at which the method is able to detect the MRL concentrations with a statistical certainty of 1-β

Detection capability (CCβ) No permitted limit Analyse 20 blank materials => CCα = 3x signal/noise Analyse 20 blank materials fortified at CCα => CCβ = CCα + 1.64 x SDRW Calibration curve procedure (ISO 11843) Analyse of blank material fortified at 0 MRLP, 0.5 MRLP, 1 MRLP, 1.5 MRLP and 2 MRLP Plot analytical results (y-axis) vs concentration(x-axis) CCα = y-intercept (blank) + 2.33 x SDRW CCβ = CCα + 1.64 x SDRW

Detection capability (CCβ) No permitted limit If no quantitative results Analyse fortified blank samples at and above CCα (n ≥ 20 / concentration level) CCβ = concentration level where only ≤5% false compliant results remain

Detection capability (CCβ) CCa CCb Blank +2.33xSDblank +1.64xSDRW α=1% β=5% Signal or Concentration

Detection capability (CCβ) Permitted limit (MRL) Analyse 20 blank materials fortified at MRL => CCα = MRL + 1.64 x SDRW Analyse 20 blank materials fortified at CCα => CCβ = CCα + 1.64 x SDRW Calibration curve procedure (ISO 11843) Analyse of blank materials fortified at 0.5 MRL, 1 MRL, 1.5 MRL and 2 MRL Plot analytical results (y-axis) vs concentration(x-axis) CCα = MRL + 1.64 x SDRW CCβ = CCα + 1.64 x SDRW

Detection capability (CCβ) CCa CCb MRL +1.64xSDMRL +1.64xSDRW α=5% β=5% Signal or Concentration

Performance characteristics Detection capacity Selectivity/Specificity Applicability/ Ruggedness/Stability Precision (only for semi-quantitative method)

Selectivity/specificity Specificity: ability of a method to distinguish between analyte being measured and other substances problem of interference? F(measuring technique, class of compounds, matrices,…)

Selectivity/specificity How to test specificity for qualitative screening method? Analyse 20 different blank samples and 20 positive samples (blind study, same or different days/technicians) Specificity= 100* NA/N- Other parameters: Accuracy= 100* (PA+NA)/(N- + N+) Sensitivity= 100* PA/N+ False positive= 100* FP/(N- + N+) False negative= 100* FN/(N- + N+) True positive (N+) True negative (N-) test result positive Positive agreement (PA) False positive (FP) test result negative False negative (FN) Negative agreement (NA)

Selectivity/specificity How to test specificity for semi-quantitative screening methods? Select potentially interfering substances (metabolites, derivatives,…) Analyse relevant blank samples (n ≥ 20) Analyse fortified blank samples with interfering substances at a relevant concentration Estimate the effect of the interferences False identification? Influence in quantification? Identification of the target analyte is hindered?

Performance characteristics Detection capacity Selectivity/Specificity Applicability/ Ruggedness/Stability Precision (only for semi-quantitative method)

Applicability Scope of the method must be define in term of : Matrix (solid/liquid matrix, type of tissue) Animal species To introduce a new matrix Analyse at least 10 different blank material fortified at level of interest for the new matrix (CCβ) + test of interferences If 10 positive results => method applicable for the new matrix If 1 negative result => 10 additional analyses If 1 negative result=> CCβ must be recalculated for the new matrix

Ruggedness Ruggedness: the susceptibility of an analytical method to changes in experimental conditions sample material analytes storage condition environmental condition sample preparation condition

Ruggedness How to test ruggedness? (during development) Identify possible factor that could influence the results (the analyst, solvents, pH, T°, rate of heating,…) Vary each factor slightly If one factor is found to influence results of the representative molecule, conduct further experiments => acceptability limits for this factor (in the method protocol) Recommendation of CRL: analyses of 10 blank and 10 spiked samples at the same concentration and with minor change of factor to detect influence on results

Stability Test are not necessary if stability data already exist (from other lab or from publication) To include in the validation report Stability test: the analyte in solution the analyte in matrix Aliquots of a fresh solution or sample stored under different conditions (T° and/or storing time)

Performance characteristics Detection capacity Selectivity/Specificity Applicability/ Ruggedness/Stability Precision (only for semi-quantitative method)

Precision (for quantitative screening) Precision: the closseness of agreement between independent test results obtained under predetermined conditions Expressed in terms of imprecision / standard deviation of test results

Precision (for quantitative screening) How to test precision? Repeatability test within-laboratory reproducibility test(or intermediate precision) Reproducibility test (between laboratories: interlaboratory studies) determination of RSD (%) < Precision criteria

Precision (for quantitative screening) Repeatability 3 concentrations: 1x; 1,5x; 2x MRPL 0,5; 1x; 1,5x MRL 6 replicates/level 3 times same conditions Within-laboratory Reproducibility 3 concentrations: 1x; 1,5x; 2x MRPL 0,5; 1x; 1,5x MRL 6 replicates/level 3 times different conditions (analyst, env. condition,…)

Precision (for quantitative screening) ANOVA treatment of data => RSDr & RSDRW Comparison with precision criteria: Horwitz equation: RSDR(%) = 2(1-0.5logC) Criteria for repeatability: RSDr = 1/2 to 2/3 RSDR Criteria for within-lab reproducibility: RSDRW = 2/3 to 1 RSDR ! For concentration < 100 µg/kg, RSDR becomes too high!

Other recommendations False negative rate <5%: Analyses of 20 negative and 20 positive samples in order to test the screening method (see selectivity). One QC sample must be added in routine and results must be added to the validation file Method transfer/Commercial test Bibliographical survey to compil the evaluation of performance of the test Collection of data from supplier on validation study Experimental plan to test skillness of technician to perform the test Use of QC sample Participation to proficiency test

Exemple: analyse of PCDD/F by CALUX bioassay PCDD/F: 17 toxic congeners to analyse in various matrices (TCDD=most toxic dioxin) Results expressed in TEQ (=Sum (CCixTEFi)i=1-17) MRL for each matrix (milk, meat, egg, fish oil,…) MRL expressed in pg TEQ/g fat or ng TEQ/ kg Reference method: GC-HRMS Screening method: immunoassay, bioassay,…

Exemple: analyse of PCDD/F by CALUX bioassay CALUX bioassay= genetically modified cell-based bioassay (luciferase) Amount of light produced is proportional to the toxicity (TEQ) of extracts Gene expression LIGHT All substances fixing the Ah receptor

Analyse of PCDD/F by CALUX bioassay Advantage: Rapid Cheaper than GC-HRMS Time for analyses Disadvantage: Various compounds can fix the Ah receptor (PAH, PCB, PHDD/F,…) specificity!!!!

Analyse of PCDD/F by CALUX bioassay : protocol Extraction of fat Clean-up on silica acid + carbon columns Fraction with interfering compounds Fraction with PCBs Fraction with PCDD/F Evaporation Reading plate Dosing plate

Analyse of PCDD/F by CALUX bioassay : validation Selectivity/specificity Ruggedness/Stability Precision Detection capability

Analyse of PCDD/F by CALUX bioassay : selectivity Possible interfering compounds? PAH : mostly in environmental sample PCB: fractionation during clean-up Other compounds? (PHDD/F): dependant of the matrix? (matrix effect?) Results of the selectivity test: No interferences for feedstuff, milk, egg, fat Interferences for fish oil CALUX results = 2 x GC-HRMS results

Analyse of PCDD/F by CALUX bioassay : selectivity Matrix effect for fish oil

Analyse of PCDD/F by CALUX bioassay : ruggedness What are the critical point in the protocol? Carbon column (interferences) Solvent (interferences) Curve (results) Evaporation time (recovery) Age of CALUX cell line (RSD)

Analyse of PCDD/F by CALUX bioassay : ruggedness Carbon column: amount of carbon used (Rdt PCDD/F= 60%) (Rdt PCDD/F= 80%) DX fraction PCB fraction Not collected fraction (Rdt PCDD/F= 80%)

Analyse of PCDD/F by CALUX bioassay : ruggedness Evaporation time

Analyse of PCDD/F by CALUX bioassay : ruggedness Solvent: tested before use on a TCDD solution (antagonist/agonist effect) Curve: tested with an independant TCDD solution Age of CALUX cells: new cell every 2 months

Analyse of PCDD/F by CALUX bioassay : precision Validation protocol Day 1 Day 2 Day 3 Day 4 Day 5 Day 6 Blank solvent 1 Blank sample Sample at MRL/2 3 2 Sample at MRL 6 Sample at 2MRL Quality sample 10

Analyse of PCDD/F by CALUX bioassay : precision ANOVA results for the TEQ determination of PCDD/F in feedstuff by CALUX bioassay At MRL (0.75ng TEQ/kg) : XMRL= 0.751 ng TEQ/kg Sr= 0.063 => RSDr= 8.4% SRW=0.073 =>RSDRW= 9.7% At MRL/2 (0.376ng TEQ/kg) : XMRL/2= 0.464 ng TEQ/kg Sr= 0.051 => RSDr= 11% SRW=0.051 =>RSDRW= 11% At 2MRL (1.5ng TEQ/kg): X2MRL= 1.571 ng TEQ/kg Sr= 0.107 => RSDr= 6.8% SRW=0.115 =>RSDRW= 7.3% RSD < 30% (2002/70/EC)

Analyse of PCDD/F by CALUX bioassay : detection capacity CCβ for the TEQ determination of PCDD/F in feedstuff by CALUX bioassay CCα = MRL + 1.64 x SRW CCα = 0.75 + 1.64 x 0.073 = 0.87 ng TEQ/kg CCβ = CCα + 2.33 x SRW CCβ = 0.87 + 2.33 x 0.073 = 1.04 ng TEQ/kg 2002/70/EC: false negative rate < 1% ! => At a concentration of 1.04ng TEQ/kg, we are sure that the sample is a positive sample with 99% certainty

Analyse of PCDD/F by CALUX bioassay :confirmatory range ? COMPLIANT SUSPICIOUS NON COMPLIANT CC* MRL CCa CCb -2.33sMRL +1.64sMRL +2.33ssample Signal or Concentration * =1% β=5% α=5%

Analyse of PCDD/F by CALUX bioassay :confirmatory range Lower limit of the confirmatory range for the TEQ determination of PCDD/F in feedstuff by CALUX bioassay CC*= MRL-2.33 x SDRW CC*= 0.75- 2.33 x 0.073 = 0.58 ng TEQ/kg Conclusion Sample lower than 0.58 ng TEQ/kg are negative with 99% certainty (false negative rate < 1%) Sample above 0.58 ng TEQ/kg must be confirmed by GC-HRMS