Design of RP-HPLC 1 for Insulin Production

Slides:



Advertisements
Similar presentations
Chromatography Components stationary phase (eg., solid matrix) mobile phase (eg., solvent) solute Solutes which interact differently with the stationary.
Advertisements

A Grace company Avoiding Matrix Interferences in Ion Chromatography Using Solid Phase Extraction Alltech Associates, Inc. Brandstraat 12, 9160 Lokeren.
Gas Chromatography Introduction 1.) Gas Chromatography
Production of Insulin Reverse Phase – High Pressure Liquid Chromatography Unit (RP-HPLC) Presented by:Justin McComb Rachelle Bolton Young Chang.
Food Analysis Lecture 23 (4/19/2005) HPLC (2) and Gas Chromatography (1) Qingrong Huang Department of Food Science Read Material: Chapter 29, page 479.
1 HPLC Columns and Stationary Phases Lecture 2 Yuri Kazakevich Seton Hall University.
Size Exclusion Chromatography
Separation of molecules and determination of there molecular weight by gel filtration chromatography. Experiment 7 BCH 333.
Anion Exchanger Kristen Galea. Where Are We With in the Process? Entering stream is concentrated with, insulin ester, and denatured proteins This separation.
Reversed Phase HPLC Mechanisms Nicholas H. Snow Department of Chemistry Seton Hall University South Orange, NJ 07079
DR ZIAD W JARADAT PROTEIN BIOTECHNOLOGY BT 452 Chapter 3 HIGH PERFORMANCE LIQUID CHROMATOGRAPHY.
Chapter 32 HIGH-PERRORMANCE LIQUID CHROMATOGRAPHY High-performance liquid chromatography (HPLC) is the most versatile and widely used type of elution chromatography.
Standard Methods for the Examination of Water and Wastewater, 21st Ed
IEX Chromatography Presented by: Nikki Apostolakis Helen So Tiffany Yu CHEE450: Engineering Biology.
HPLC Systems. Column Chromatography HPLC Modes HPLC – System Components.
Chem. 133 – 5/5 Lecture. Announcements Lab Report 2.4 due Thursday – can turn in today for reduction of late penalties Term Project Progress Report –
Column Chromatography. Types of columns: 1- Gravity Columns: The mobile phase move through the stationary phase by gravity force. 2- Flash Columns (Air.
High-Performance Liquid Chromatography HPLC, when GC won’t cut it!!!
Lecture 18 High Preformance Liquid Chromatography I.
High Performance Liquid Chromatography. HPLC originally refered to: High Pressure Liquid Chromatography currently refers to: High Precision Liquid Chromatography.
HPLC when GC won’t cut it!!!. Types of HPLC Reverse-phase (water/MeOH-soluble) Normal Phase (very polar) Adsorption (very non-polar) Ion-Exchange (ionic)
Gel Filtration Chromatography.
B IOCHEMICAL INSTRUMENTAL ANALYSIS -11 Dr. Maha Al-Sedik.
Pn. Suryati Bt. Syafri 2009 High Performance Liquid Chromatography (HPLC)
Effect of Packing Particle Size on Plate Height. Resolution between 2 adjacent peaks.
Adsorption Liquid Chromatography. Open Column Chromatography Silica gel Glass Tube Eluent Vial for fraction collection.
High Performance Liquid Chromatography High Performance Liquid Chromatography Chem. 331.
High Performance Liquid Chromatography (HPLC)
High Performance Liquid Chromatography
High Performance Liquid Chromatography (HPLC) Jeannette Comeau CHEE April 2004.
GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5.
High Performance Liquid Chromatography Instrumentation.
Gihan Gawish.Dr High Performance Liquid Chromatography.
CHROMATOGRAPHY (DEMONSTRATION) Mrs. Chaitali Maitra
High Performance Liquid Chromatography High Performance Liquid Chromatography Presented by- MOHSIN NISAR KHAN MOHSIN NISAR KHAN.
HPLC – High Performance Liquid Chromatography
Introduction  High-performance liquid chromatography (HPLC) is a form of liquid chromatography.liquid chromatography  The main purpose is to separate.
1 HPLC Lecture Displacement pumps Displacement pumps, on the other hand, is composed of a one directional motor driven plunger that pushes the mobile.
Best Broken into four categories
Created with MindGenius Business 2005® Instrumentation (1) Instrumentation (1) Origins Preparative column chromatography (Glass columns) Internal diameter:
Chromatography High Performance Liquid Chromatography HPLC Chapter Dr Gihan Gawish.
HPLC.
High Performance Liquid Chromatography
Biochemical instrumental analysis - 11 Dr. Maha Al-Sedik 2015 CLS 332.
HPLC.
CHROMATOGRAPHY. Chromatography Chromatography basically involves the separation of mixtures due to differences in the distribution coefficient of sample.
Chem. 133 – 5/3 Lecture. Announcements Lab – Term Project Progress Report Due Today – Last Assignments: Term Project Poster and Peer Review Grading (Friday,
Principles of chromatography
Chromatography.
HPLC (High Performance Liquid Chromatography)
High Performance Liquid Chromatography. What is HPLC ? It is a separation technique that involves: Injection of small volume of liquid sample Into a tube.
Bioseparation II Chromatography Techniques. Chromatography Most widely used purification technique used for biomolecules. Most widely used purification.
High Performance Liquid Chromatography Presented by Dr. Kamal Modi 2 nd Year Resident.
9 Adsorption Chromatography: Normal-Phase Chromatography
High-Performance Liquid Chromatography HPLC, when GC won’t cut it!!!
Chem. 133 – 5/2 Lecture.
High Performance Liquid Chromatography (HPLC)
CHROMATOGRAPHY.
HPLC.
Size Exclusion Chromatography (SEC)
Horizon Chromatography
Chromatographic separation
Techniques of protein purification
Best Broken into four categories
HPLC.
High Performance Liquid Chromatography
Gel Filtration Chromatography.
Chromatography Daheeya Alenazi.
High Performance Liquid Chromatography
High performance liquid chromatography (HPLC)
Presentation transcript:

Design of RP-HPLC 1 for Insulin Production Presented by: Patricia Yee, Nicholas Irianto, Thomas Fang

Overview Background Information Significance of RP-HPLC1 Design Parameters Selected Design and Specifications Questions

Background Information What is HPLC? High Performance Liquid Chromatography - or - High-Pressure Liquid Chromatography Pat -used to separate components of a mixture based on chemical interactions Photo Credit: http://www.novasep.com/Technologies/img/Hipersep-units-illu.jpg

How HPLC works… Pat -generally separate analytes by polarity “Like attracts Like” -column is packed with a stationary phase, which usually consists of small round particles with a selected surface chemistry -a liquid (the mobile phase) is pumped through the column under high pressure while the sample to be analyzed and separated is injected in smaller amounts -the velocity of the components in the analyte are slowed down by different degrees due to the chemical/physical interactions taking place between the analyte, the stationary and mobile phases, which have different chemical properties Picture References: http://www.waters.com/WatersDivision/images/products/hplc_primer_fig_q.jpg http://www.waters.com/WatersDivision/images/products/hplc_primer_fig_r.jpg

Reverse-Phase HPLC Pat -reversed phase chromatography -in comparison to normal phase chromatography (NP-HPLC), which uses polar stationary phase and non-polar mobile phase, RP-HPLC uses non-polar stationary and polar mobile phase -RP-HPLC gives better reproducibility of retention times -“like attracts like” therefore if the analyte is non-polar: the more non-polar the stationary phase or the more polar the mobile phase, the longer the retention time

Insulin Production Process: RP-HPLC 1 Pat -after reactor and before diafilter -components in mixture that are not desired in end product (e.g. enzymes, denature and contaminated proteins), therefore, take them out and keep what you want with RP-HPLC -mobile phase: acetonitrile/water/acid (water-miscible organic acid and other acids) used to control the retention times -the more water, the more polar the mobile phase, and the more acetonitrile, the less polar the mobile phase Picture References: http://www.waters.com/WatersDivision/images/products/hplc_primer_fig_s_1.jpg http://www.waters.com/WatersDivision/images/products/hplc_primer_fig_s_2.jpg

Insulin Production Process: RP-HPLC 1 What comes out… 1) Contaminated protein, denatured insulin, enzyme and small amounts of insulin ester and WFI 2) Majority of desired insulin ester and WFI Pat -acetonitrile/acid/contaminated protein/denatured insulin/enzyme/and small amounts of insulin ester and WFI (water) get eluted first -then the majority of the desired insulin ester and WFI are eluted last

Insulin Production Process: RP-HPLC 1 Gradient Elution of mobile phase: Acetonitrile / Water / Acid Pat -achieved through gradient elution: different ratios of acetonitrile/water/acid put through the column -stationary phase polarity basically fixed -but can change the percentage of acetonitrile more acetonitrile: the more non-polar the wash, and non-polar components eluted earlier more water: the more polar the wash, and non-polar components eluted later -these ratios can be tweaked to optimize separation and processing time -the gradient concentrations and the times it changes can be automated

Main Design Parameters Column Size (Internal diameter) Material (Stainless Steels, Glass, etc) Column size - column that has a larger internal diameter has a larger loading capacity (the most common ID is 4-6mm these columns are called “analytical scale columns” for industrial scale or manufacturing purposes, the internal diameter is usually larger than 10mm) Column material - Usually made from stainless steels, due to the durability. We can also see glass columns for lab scale usage

Main Design Parameters Resin Particle size (10μm ~75μm ) Pore size (300 Á) Material (Silica, styrene-divinylbenzene, etc.) Smaller particle size - more effective at accommodating separation of the insulin polypeptide (larger surface area for adsorption.) but can cause problems with the fluid flow through the column (need a bigger pressure) Larger particle size - Easier to pack, lower pressure drop through the column. Pore size - The pore size must be large enough for the polypeptide to properly adsorb to the resin, most polypeptide separations are performed on columns with particles with pores of about 300 A (Carr D., 2002) http://www.vydac.com/vydacpubs/handbook/handbook_3rd_ed.pdf

Other Design Considerations Pump (pressure capacity) Solvent Acetonitrile Polar solvent (increase retention time) Hydrophobic solvent (decrease retention time) Pump - pumps vary in pressure capacity, modern HPLC systems has been improved to work at higher pressure, Pressure may reach as high as 400 atm. So therefore resin that has a smaller particle can be used in the column to get a better separation. Retention time - The retention time is longer for molecules which are more non-polar in nature, allowing polar molecules to elute more readily. The retention time can be increased by adding more polar solvent to the mobile phase, and decreased by addition of hydrophobic solvent.

Possible Design This shows the possible type of HPLC that can be used for insulin purification

Design Specifications 100% binding and 90% yield for insulin ester 60% binding and 5% yield for denatured insulin precursors Purification results in outlet stream consisting of 340 kg insulin ester/batch The following specs were used when running the simulation in SuperPro. The expected insulin ester production rate is 340 kg insulin ester/batch. Designing for length and diameter in SuperPro, gives out the column volume and number of units required to do the purification step. We kept other parameters as default values since we were unable to find these parameters from vendors. Playing around with the column length and max diameter would affect the required column volume and number of units. Since SuperPro determined the column volume based on the number of cycles per batch and the specified column length and maximum diameter. If the calculated diameter goes above the maximum diameter, this means that an additional RP-HPLC working in parallel configuration would be required to do the separation.

Design Selections HPLC Columns Novasep Prochrom® DAC Columns LC1600 H x D: 450 x 1600 mm HPLC Resins PRP-3 standard Poly(styrene-divinylbenzene) (PSDVB) Pore size 300 Å Diameter 10 μm After contacting several vendors, we were able to obtain information from the hamilton company. This company specifically suggested that PRP-3 resins would be an ideal resin to be used for insulin purification primarily due to the molecular weight of insulin itself (5800 MW) In order to meet the process specification of producing 188 kg insulin/batch, 23 RP-HPLC units with maximum bed volume of 900 L are required. We were unable to find out the exact resin binding capacity from the Hamilton Company. Similar issues were faced when trying to determine the number of cycles per batch, resin replacement frequency and linear flowrate. Therefore, we decided to keep these values to SuperPro default values. Notice that assuming that the batch requires 60 hours of operation with 36 hours turnover time between two batches, if we are operating 7920 hours in a year, we would make roughly 82 batches which reflects to about 15416 kg insulin crystal/year. This exceeds the operational target of 4000 kg insulin crystal/year. According to the PRP-3 Resins manufacturing, 1 gram of the PRP-3 resin is applicable for 2.5 cm3 of the column volumes. Thus for a 900 L column volumes, we would require 360 kg of PRP-3 resins. Cost from the Hamilton Company is $24/g PRP-3 resin. This would translate to $8 640 000 per column. For 21 columns, $ 181 440 000. 10 microns were selected because it would provide a better separation a greater peak. PRP-3 are 300 Angstorms pores polymer materials.

Design Justifications Stable from pH 1-13 Does not contain acidic silanol groups, which are present on silica supports. Highly cross-linked for exceptional durability and to prevent shrinking and Consistent performance 10μm particle diameter is selected for better solution and peak shape swelling.Compatible with 0-100% organic or aqueous solvents. Resins are highly cross-linked for exceptional durability and to prevent the resin from shrinking and swelling, which often occurs in other polymeric supports This eliminates irreversible, sample absorption on the support and thus greatly improves sample recovery and bioactivity. PRP columns are capable of providing sample recoveries of 95% Notice that assuming that the batch requires 60 hours of operation with 36 hours turnover time between two batches, if we are operating 7920 hours in a year, we would make roughly 82 batches which reflects to about 15416 kg insulin crystal/year. This exceeds the operational target of 4000 kg insulin crystal/year.

Design Details Twenty three LC 1600 HPLC column units with 0.91 hours process time would be required for the process according to SuperPro For one LC 1600 HPLC column with maximum bed volume of 900 L would require 360 000 grams of PRP-3 resins COSTS $24/g of PRP-3 resins This translates to $ 8 640 000 / LC 1600 HPLC column For 23 HPLC columns, it would cost $ 198 720 000 $28 000 for a Novasep Prochrom DAC LC 1600 column Total Cost $ 200 000 000 Assuming that 1 cycle per batch with resin binding capacity of 20 mg/ml and 200 cycles resin replacement frequency (default values from SuperPro) From Hamilton vendor, 1 gram of resin is approximately equivalent to 2.5 cc of column volumes Resins replacement frequency and beads replacement turnaround time cannot be found from vendor. Therefore we used SuperPro default values Cost for column is obtained from linear interpolation based on volume Notice that assuming that the batch requires 60 hours of operation with 36 hours turnover time between two batches, if we are operating 7920 hours in a year, we would make roughly 82 batches which reflects to about 15416 kg insulin crystal/year. This exceeds the operational target of 4000 kg insulin crystal/year.

Questions? THANK YOU!