Image Reconstruction.

Slides:



Advertisements
Similar presentations
Topic 8. Gamma Camera (II)
Advertisements

PRINCIPLES OF CT.
IMAGE RECONSTRUCTION.
ARTIFACTS IN COMPUTED TOMOGRAPHY
IMAGE GENERATION IN CT.
DIGITAL RADIOGRAPHY.
Technical Aspects: the machine the image
Advanced Biomedical Imaging Lecture 3
CT Scanning: Dosimetry and Artefacts
Computers and Computed Tomography
CT Scanning Dr. Craig Moore
Computed Tomography
Image Quality Bushong Chapter 7.
IMAGE QUALITY.
Computed Tomography Physics, Instrumentation, and Imaging
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Image Artifacts Chapter 8 Bushong.
X-ray diagnostics and Computed tomography. 15th century head examination.
Computed Tomography II
SCANCOMEDICAL Computed Tomography SCANCO User Meeting 2005 Dr. Bruno Koller SCANCO Medical AG
Advanced Biomedical Imaging
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods
CT Physics V.G.Wimalasena Principal School of radiography.
Computed Tomography III
Computed Tomography RAD309
tomos = slice, graphein = to write
X-Rays and CT Scans.
Computed Tomography
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
Conventional and Computed Tomography
…….CT Physics - Continued V.G.WimalasenaPrincipal School of radiography.
Seeram Chapter 11: Image Quality
COMPUTED TOMOGRAPHY (I) RAD 365 CT - Scan
Basic principles Geometry and historical development
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
Digital Image Characteristic
Without reference, identify principles relating to Computed Tomography Clinical Applications with at least 70 percent accuracy.
1 RT 255 C Cross Sectional Anatomy Week 1 FINAL
Quality Control.
Module B Computed Tomography Physics, Instrumentation, and Imaging.
Image Quality Radiographic Resolution.
Moscow-Bavarian Joint Advanced Student School March 2006, Moscow, Russia Correction and Preprocessing Methods Veselin Dikov Moscow-Bavarian Joint.
Seeram Chapter 7: Image Reconstruction
Computed Tomography Q & A
PRINCIPLE AND BASIC PHYSICS OF COMPUTED TOMOGRAPHY
X - RAYS IN DIAGNOSTICS D. Krilov
Factors affecting CT image RAD
Fluoroscopy Q&A Robert L. Metzger, Ph.D..
If F {f(x,y)} = F(u,v) or F( ,  ) then F( ,  ) = F { g  (R) } The Fourier Transform of a projection at angle  is a line in the Fourier transform.
Image Quality The capacity to define, measure, and assess image quality is a primarily responsibility of a CT Technologist.
CT IMAGE RECONSTRUCTION  Hounsfield envisioned dividing a slice into a matrix of 3-dimensional rectangular boxes (voxels) of material (tissue). Conventionally,
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
Module D Computed Tomography Physics, Instrumentation, and Imaging.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
Part No...., Module No....Lesson No
CT Chapter 4: Principles of Computed Tomography. Radiography vs. CT Both based on differential attenuation of x-rays passing through body Radiography.
Quality Assurance.
By Prof. Stelmark. Digital Imaging In digital imaging, the latent image is stored as digital data and must be processed by the computer for viewing on.
Chapter-4 Single-Photon emission computed tomography (SPECT)
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
CT Scanning Dr. Craig Moore
Computed Tomography Basics
Impact of Obesity on Medical Imaging and Image-Guided Intervention
Resident Physics Lectures (Year 1)
Basic principles Geometry and historical development
Resident Physics Lectures (Year 1)
CT ARTIFACTS.
Presentation transcript:

Image Reconstruction

Image Reconstruction II Generating an image from the acquired data involves determining the linear attenuation coefficients of the individual voxels. A mathematical algorithm takes the projection data and reconstructs the cross‑sectional CT image. There are three common ways to carry this out: 1: Backprojection 2: Iterative 3: Fourier methods

Backprojection The field of view is simulated as a 2-D array within the computer. 512 x 512 The attenuation for each voxel of a scan is 1: projected back along the direction of the beam 2: the attenuation is added to the value in each array element the beam passes through. At the end of the procedure a rough image is held in the array elements.

Filtered Backprojection No commercial CT scanners use simple back projection The process produces artifacts where the attenuation changes rapidly. Most modern scanners use filtered back projection algorithms which have filter functions applied to reduce these effects. Different filters may be used, offering tradeoffs between spatial resolution and noise. Some filters permit reconstruction of fine detail but with increased noise in the image such as in bone algorithms. Algorithms such as those for soft tissue provide some smoothing, which decreases image noise but also decreases spatial resolution. The choice of the best filter to use with the reconstruction algorithm depends on the clinical task.

Iterative procedure

Fourier Techniques

Image display Reconstructed images are viewed on CRT monitors or printed onto film using a laser printer. Each pixel is normally represented by 12 bits, or 4096 grey levels The normal grey levels available on a display are 256 Window width and window level are used to optimise the appearance of CT images by determining the contrast and brightness levels assigned to the CT image data . Window level, or centre, is the CT number or HU value to be displayed as the median intensity in the image. The window level is normally chosen close to the average HU value of the tissue of interest (e.g., 10 to 50 HU for soft tissue). Window width is the range of CT numbers displayed around the selected centre. and it determines the contrast. A narrow window width provides higher contrast than a wide window width.

Image Quality Image quality may be characterised in terms of contrast noise spatial resolution Image quality involves tradeoffs between these three factors and patient radiation dose. Artifacts encountered during CT scanning can degrade image quality.

Contrast The difference in the HU values between tissues. Usually increases as kVp decreases but is not affected by mA or scan times. CT contrast may be artificially increased by adding a contrast medium such as iodine. Image noise may prevent detection of low‑contrast objects such as tumours with a density close to the adjacent tissue. The displayed image contrast is primarily determined by the CT window width and window level settings.

Noise CT noise is determined primarily by the number of photons used to make an image (quantum mottle). Quantum mottle decreases as the number of photons increases. The typical noise with a modern CT system is approximately 5 HU CT noise is generally reduced by increasing the ·        kVp, ·        mA ·        scan time ·        voxel size (i.e., by decreasing matrix size, increasing FOV, or increasing section thickness).

Resolution Typical resolution in CT scanning ranges from 0.7 to 1.5 lp/mm. If the CT Field Of View [FOV] is d and the matrix size is M, then the pixel size is d /M.   For a typical head scan with an FOV of 25 cm A matrix of 512 pix­els The pixel size is 0.5 mm. Factors that may also improve CT spatial resolution · smaller focal spots · smaller detectors · more projections. · smaller FOV · larger matrix size. Resolution perpendicular to the section is dependent on slice thickness

Radiation The dose profile in a CT scanner is not uniform along the patient axis and may vary within any irradiated section. Typical maximum doses for a single section are 40 mGy (4 rads) for a head scan 20 mGy (2 rads) for a body scan.

Radiation In head scans, the surface to centre ratio is approximately 1:1. In body scans, the surface to centre ratio is approximately 2:1.

Radiation Because of scattered x‑rays, the CT section dose profile is not perfectly square but has tails that extend beyond the section edges. Tissues beyond the section are thus exposed to radiation. When contiguous sections are scanned, the cumulative radiation dose in a section may be as high as twice the radiation dose associated with a single section.

Artifacts Partial volume artefact CT scanners may have artifacts in the reconstructed images. Partial volume artefact is the result of averaging the linear attenuation­ coefficient in a voxel that is not uniform in composition. Partial volume artefact increases with pixel size and section thickness.

Artifacts Streak Artifacts Random or unpredictable motion (e.g., if the patient sneezes) produces streak artifacts in the direction of motion. In high‑density structures, such as metal implants, the detector may record no transmission. The algorithm generates streaks adjacent to the high‑densities.

Artifacts Beam Hardening Artifacts Beam hardening artifacts or "cup" artifacts are caused by the polychromatic nature of the x‑ray beam (beam hardening). As the lower energy photons are preferentially absorbed, the beam becomes more penetrating and results in lower computed values of the attenuation coefficient (HU). Beam hardening artifacts are most marked at high‑contrast interfaces such as between dense bone in the skull and the brain

Artifacts Ring artifacts   Ring artifacts may arise in third‑generation systems if a single detector is faulty or miscalibrated. Artifacts caused by equipment defects are rare on modern CT systems

Ring Artifacts