Progress report for the energy spectrometer test experiment at ESA at SLAC Bino Maiheu on behalf of the T474/T491 collaboration LCWS 2007, DESY Hamburg.

Slides:



Advertisements
Similar presentations
BPM Energy Spectrometry for ILC Bino Maiheu University College London for LC-ABD WP 4.2 LC-ABD Meeting IPPP Durham, 26 September 2006 LC-ABD Meeting, 26.
Advertisements

SLAC - End Station A ILC testrun – April '06 Bino Maiheu University College London Overview, status, results and future plans Dubna meeting May
LCABD WP 4.2 Review, 24 th September Mark Slater Motivation Final Results from 2006 Upgrades for 2007 Conclusions The Collaboration LCABD WP 4.2.
Energy spectrometer cavity BPM Royal Holloway, University of London S. Boogert, G. Boorman University of Cambridge M. Slater, M. Thomson & D. Ward University.
BPM Energy spectrometry, status & future plans at ESA in SLAC Bino Maiheu University College London Yerevan Collaboration Meeting October 2006 Yerevan.
An Energy Spectrometer for the International Linear Collider Bino Maiheu University College London Bino Maiheu, HEP Seminar, University.
LCABD WP 4.2 Review, 12 th April Mark Slater  Motivation  BPM and Electronics Development  BPM Simulation  Update on NanoBPM Progress  Update.
BPM Energy Spectrometry for the International Linear Collider Bino Maiheu for Alex, Stew, Fil, Steve, Matthew & David UCL – HEP group meeting 29 September.
Fast and Precise Beam Energy Measurement at the International Linear Collider Michele Viti.
SLAC, 5-Dec-2006 Magnets and Magnetic Measurements for Prototype Energy Spectrometer T-474.
Cavity BPM for Spectromety: progress and plans A. Lyapin.
March 22, 2011Mike Hildreth – ALCPG 2011, Beam Instrumentation Energy Spectrometer R&D; Plans for End Station A Test Beams Mike Hildreth University of.
October 1, 2009Mike Hildreth – ALCPG09 Energy Spectrometer R&D Progress and Plans Mike Hildreth University of Notre Dame representing SLAC T-474/491 &
Beam position monitors LCABD Plenary meeting Bristol, 24th March 2009 A. Aryshev, S. T. Boogert, G. Boorman, S. Molloy, N. Joshi JAI at Royal Holloway.
WP 4.2 review meeting Status of the ESA work... Wednesday, 13 September 2006, UCL,
Paul Emma LCLS FAC April 16, Initial Experience with Injector Commissioning P. Emma, et al. Facilities Advisory Committee.
News about magnetic field simulation and progress ‘Towards magnetic field measurement benches’ Nikolay Morozov JINR Dubna.
ATF2 FB/FF layout Javier Resta Lopez (JAI, Oxford University) for the FONT project group FONT meeting January 11, 2007.
March 17, 2011Mike Hildreth – SLAC ESA Testbeam Workshop Energy Spectrometer Tests at End Station A Mike Hildreth University of Notre Dame SLAC, March.
Bob Lill LCLS FAC June 16-18, 2008 RF BPM Status and Production Test Results.
Progress towards nanometre-level beam stabilisation at ATF2 N. Blaskovic, D. R. Bett, P. N. Burrows, G. B. Christian, C. Perry John Adams Institute, University.
High Resolution Cavity BPM for ILC final focal system (IP-BPM) ILC2007/LCWS 2007 BDS, 2007/6/1 The University of Tokyo, KEK, Tohoku Gakuin University,
Ground Motion + Vibration Transfer Function for Final QD0/SD0 Cryomodule System at ILC Glen White, SLAC ALCPG11, Eugene March 21, 2011.
12/12/04 1 Stewart Boogert (UCL) Second Mini-workshop Nano project at ATF UK plans for Energy spectrometer studies University College London Stewart Boogert.
___________________ Jürgen Schreiber, ECFA/DESY LC workshop, Amsterdam, April 1-4, 2003 BEAM ENERGY SPECTROMETER DESY – Dubna – TU Berlin Machine physicists,
Relative Error on Parameter Pessimistic Estimate Optimistic Estimate β function at the LW 3% 1% LW readout error 2% 1% Laser spot waist 10% Laser Pointing.
David Urner, Oxford University, RHUL – June StaFF Stabilization of Final Focus Motion Stabilization with Nano-Meter Precision David Urner Paul Coe.
BPMs and HOM-BPMs for the XFEL Linac N. Baboi for the BPM and the HOM teams (DESY, CEA-Saclay, SLAC, FNAL, Cockroft/Daresbury) XFEL Linac Review Meeting,
WP9 : Cavity BPM spectrometry Royal Holloway S. Boogert & G. Boorman University College London D. Attree, A. Lyapin, B. Maiheu & M. Wing Cambridge University.
January 8, 2005Mike Hildreth – ILC-MDI Workshop A BPM-Based Energy Spectrometer Technical Considerations & Plans for the End Station A Test Beam Mike Hildreth.
ILC Feedback System Studies Nikolay Solyak Fermilab 1IWLC2010, Geneva, Oct.18-22, 2010 N.Solyak.
BPM-Based Energy Spectrometer Contributors: A. Lyapin, B. Maiheu, F. Gournaris, D. Attree, M. Wing, D. Miller University College London (UCL) M. Slater,
SPPS, Beam stability and pulse-to-pulse jitter Patrick Krejcik For the SPPS collaboration Zeuthen Workshop on Start-to-End Simulations of X-ray FEL’s August.
TBL as a 12 GHz power source, overview and first results from commissioning IWLC 2010 workshopSteffen Döbert, BE-RF  Current status of TBL  Status of.
Energy Spectrometer for the ILC Alexey Lyapin University College London.
Development of a Low-latency, High-precision, Intra-train Beam Feedback System Based on Cavity Beam Position Monitors N. Blaskovic Kraljevic, D. R. Bett,
1 Plan and Opportunities for Migration and Integration of the Photoinjector into New Muon Lab Mike Church AAC Review - 12/5/06.
July 19-22, 2006, Vancouver KIRTI RANJAN1 ILC Curved Linac Simulation Kirti Ranjan, Francois Ostiguy, Nikolay Solyak Fermilab + Peter Tenenbaum (PT) SLAC.
A Short History of Nearly Everything Michele Viti.
Production and Installation Policy of IP-BPM ATF2 Project Meeting, 2006/12/18 Y. Honda, Y. Inoue, T. Hino, T. Nakamura.
SABER, “maybe” a new facility in the South Arc (South Arc Beam Experimental Region) End Station A (ESA) in 2007 ILC Test Beams in 2008 Test Beams beyond.
Cesr-TA Simulations: Overview and Status G. Dugan, Cornell University LCWS-08.
1 ATF 2 Nanobpm (Q BPM) Electronics. Mark Slater: Cambridge Yury Kolomensky, Toyoko Orimoto: UCB Stewart Boogert, Steve Malton, Alexi Liapine: UCL Mike.
SLAC ESA T-474 ILC BPM energy spectrometer prototype Bino Maiheu University College London on behalf of T-474 Vancouver Linear Collider.
Simulations - Beam dynamics in low emittance transport (LET: From the exit of Damping Ring) K. Kubo
ATF status M. Ross October 15, 2004 The ATF is the largest test facility built exclusively for linear collider RD –Utility not reduced by the selection.
Summary of Tuning, Corrections, and Commissioning ( Short summary of ATF2 meeting at SLAC in March 2007 ) and Hardware Issues for beam Tuning Toshiyuki.
Diagnostics Workpackage G A Blair EUROTeV Final Meeting, DESY, Hamburg 30 th January 2009 Tasks Summary.
May 31, 2005Mike Hildreth – ATF 2005 Energy Spectrometry and ATF Components of the nano-BPM Test Program and Plans for Future Tests Mike Hildreth University.
Electron Spectrometer: Status July 14 Simon Jolly, Lawrence Deacon 1 st July 2014.
Development of a High Resolution Cavity BPM for the CLIC Main Beam
DRAFT: What have been done and what to do in ILC-LET beam dynamics Beam dynamics/Simulations Group Beijing.
Instrumentation at ATF / TTF Accelerator Test Facility (KEK) Tesla Test Facility – FLASH (DESY) ESA / LCLS (SLAC) Marc Ross, SLAC.
MLI Update Chris Adolphsen SLAC. CAVITY VESSEL T4CM QUAD T4CM BPM QUADS LEADS 80K BLOCK 4K BLOCK Quadrupole Package.
Wakefield effect in ATF2 Kiyoshi Kubo
Progress in CLIC DFS studies Juergen Pfingstner University of Oslo CLIC Workshop January.
TBL experimental program, evolution to RF power testing in TBL ACE 2011, February 2-3Steffen Döbert, BE-RF  Current status of TBL  Experimental program.
LCABD WP 4.2 – Spectrometer and BPM Studies
Steering algorithm experience at CTF3
For Discussion Possible Beam Dynamics Issues in ILC downstream of Damping Ring LCWS2015 K. Kubo.
CLIC Workshop 2016: Main Beam Cavity BPM
ATF nanoBPM system - nanoGrid ATF2 Movers TTF HOM – dipole / monopole
BPM-Based Energy Spectrometer
Beam Optics Set-Up at SLAC End Station A
Damping Ring Kicker Tests at AØ
Bubble Chamber – Beam Requirements
SLAC - End Station A ILC testrun – April '06 ...
ILC Beam Tests in End Station A
High Level Physics Applications for LCLS Commissioning
LCLS Injector/Diagnostics David H. Dowell, SLAC April 24, 2002
Presentation transcript:

Progress report for the energy spectrometer test experiment at ESA at SLAC Bino Maiheu on behalf of the T474/T491 collaboration LCWS 2007, DESY Hamburg LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session BPM BPM BPM

Outline Physics motivation Physics motivation Building an energy spectrometer based upon high resolution cavity BPMs Building an energy spectrometer based upon high resolution cavity BPMs The ESA test facility at SLAC, T474/T491, experimental setup The ESA test facility at SLAC, T474/T491, experimental setup Putting in the chicane, beam line modifications, progress since VLCWS '06 Putting in the chicane, beam line modifications, progress since VLCWS '06 Progress on results and data analysis, understanding systematics Progress on results and data analysis, understanding systematics Outlook Outlook

Physics motivation, project aim LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Uncertainty on beam energy measurement contributes directly to the uncertainty on the ILC physics output... Top quark mass scan Precision physics, constrain SM parameters, eg. top threshold properties Precision physics, constrain SM parameters, eg. top threshold properties Constrain/exclude SUSY scenarios Constrain/exclude SUSY scenarios Impact of beam energy on measurements, luminosity spectrum dL/dE Impact of beam energy on measurements, luminosity spectrum dL/dE See e.g. talks of F. Gournaris and S. Boogert in top QCD session...

Proposed ILC energy spectrometer LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session BPM BPM BPM Precision measurement : dE/E Precision measurement : dE/E Minimal impact on beam itself : allowed emittance growth from SR Minimal impact on beam itself : allowed emittance growth from SR Limited space budget in BDS ~ 60 m Limited space budget in BDS ~ 60 m Minimal impact on physics data taking for e.g. calibration runs Minimal impact on physics data taking for e.g. calibration runs Magnetic chicane with high resolution beam position monitors : Magnetic chicane with high resolution beam position monitors : cavity BPMs cavity BPMs Max 5 mm dispersion at center chicane : determines resolution Max 5 mm dispersion at center chicane : determines resolution Emittance growth determines chicane layout Emittance growth determines chicane layout Diagnostics needed: Diagnostics needed: Gain drifts : temperature Mechanical stability : interferometer Magnetic fields (∫B.dl ) : NMR, Hall, fluxgate magnetometers E ~ ∫B.dl/Θ

T474 test experiment at ESA, SLAC LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session ESA comparable repetition rate, bunch charge, energy spread as ILC Possibility to vary bunch length, energy, charge Easy steering with feedback system Build an energy spectrometer prototype, using a 4 magnet chicane Build an energy spectrometer prototype, using a 4 magnet chicane Goal is to demonstrate the stability of this type of energy measurement at level, and investigate how such a magnetic chicane can be operated most efficiently at the ILC Goal is to demonstrate the stability of this type of energy measurement at level, and investigate how such a magnetic chicane can be operated most efficiently at the ILC Operate at ~5 mm η X at center chicane as in current ILC design Need < 1 µm resolution on position measurement (BPM) With position measurement stability over multiple hours of ~ 100 nm

T474/T491 collaboration, running LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Institutes involved SLAC, UC Berkeley/LBNL, Notre Dame, Dubna, DESY/Zeuthen, RHUL, UCL, Cambridge PI's M. Hildreth ( Notre Dame ), Y. Kolomensky ( Berkeley ) and S. Boogert ( RHUL ) FY06 running : January run : test run (4 days), commissioning of steering BPMs April run : commissioning of RF cavity BPMs outside of chicane ( old & new ) optimization of digitization and processing July run : commissioning of interferometer system on ILC linac prototype BPMs commissioning of energy BPM at high dispersion stability data taking with 10 BPMs, frequent calibrations FY07 running : March run : commissioning of second energy BPM installation and commissioning of magnetic chicane : first chicane data ! relocation of BPM/interferometer to center of chicane, cal tone system and new processors with remotely controllable attenuation Planned July run

Different BPM systems used LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Q XY Rectangular cavities, Q, x and y Rectangular cavities, Q, x and y Polarizations separated Polarizations separated GHz, high Q ~ GHz, high Q ~ mm aperture (0.8 “) 20 mm aperture (0.8 “) ILC linac prototype cavities ILC linac prototype cavities 36 mm aperture, GHz 36 mm aperture, GHz low Q (~ 500) low Q (~ 500) good monopole suppression good monopole suppression x/y polarizations in same cavity x/y polarizations in same cavity Middle BPM on x/y mover system Middle BPM on x/y mover system Referenced to downstream Q Referenced to downstream Q SLAC cavities ILC cold linac prototypes

New spectrometer BPM prototype LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Optimized design : A. Lyapin/UCL : high resolution : ~ 100 – 200 nm aperture monopole suppression own reference cavity developed by UCL/RHUL/MSSL mechanically rigid mover system installation planned in July Dipole cavity, 2878 MHz Reference cavity Processor electronics Reference cavity Mover system Use SLC MDL 2856 Digitize at 22 MHz

Magnetic measurements LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Simulation of magnets carried out by N. Morozov ( Dubna ) prepare for measurements in SLAC testlab (SLAC/Dubna/Zeuthen) Main simulation results : magnetic field integral uniformity region is ±15 mm magnetic field integral uniformity region is ±15 mm region for possible NMR probe use determined ( X*Z= ±7*±40 cm) region for possible NMR probe use determined ( X*Z= ±7*±40 cm) relative contribution of the fringe field to the total field integral is 22% relative contribution of the fringe field to the total field integral is 22% maximal level of the magnetic field in return yoke is no more 0.4 T maximal level of the magnetic field in return yoke is no more 0.4 T temperature factor for the magnetic field integral is 6.1×10 -5 ×1/C° temperature factor for the magnetic field integral is 6.1×10 -5 ×1/C° Screens to reduce fringe fields Screens to reduce fringe fields Model simulation using “Vector Fields” N. Morozov/Dubna

Magnetic measurements LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Results of magnetic measurements in SLAC lab, Nov. '06 (SLAC/Dubna/Zeuthen) Magnetic field integral RMS stability : 60 ppm (near working point – 150 A) Magnetic field integral RMS stability : 60 ppm (near working point – 150 A) Bdl relative RMS stability : ~ 100 ppm ( both at 150 A and 200 A) Bdl relative RMS stability : ~ 100 ppm ( both at 150 A and 200 A) measured temperature factor for the magnetic field integral is /C° measured temperature factor for the magnetic field integral is /C° in a good agreement with estimated one from magnetic field simulations /C° in a good agreement with estimated one from magnetic field simulations /C° ∫B.dl value (~ T.m when I ~ 150 A) is in agreement with simulations : T.m ∫B.dl value (~ T.m when I ~ 150 A) is in agreement with simulations : T.m Analytical dependence of ∫B.dl vs. Current obtained in the vicinity of the working point Analytical dependence of ∫B.dl vs. Current obtained in the vicinity of the working point ∫B.dl = x Current ∫B.dl = x Current Residual magnetic field along full chicane length (vertical component) Magnets in beam line S. Kostromin/Dubna

Interferometer LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Sub-nm resolution, installation itself is stable over 1 hour within 30 nm Sub-nm resolution, installation itself is stable over 1 hour within 30 nm with fixed mirrors with fixed mirrors Relocated for march '07 run ( previously on ILC cold linac prototype triplet ) Relocated for march '07 run ( previously on ILC cold linac prototype triplet ) Now monitor center of chicane + one head left for new UK BPM Now monitor center of chicane + one head left for new UK BPM 1 BPM in front of chicane, send laser beam down long pipe 1 BPM in front of chicane, send laser beam down long pipe Future location of new UK BPM Middle of chicane BPM3 BPM4

Stability results, FY06 running LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Paper “Commissioning of Spectrometer BPMs in End Station A” in the make In depth analysis of 48 hour stability data taking with frequent calibrations Variation in ω and γ over 48 hours found negligible ( ~ 2 nm ) Variation in ω and γ over 48 hours found negligible ( ~ 2 nm ) justifies use fixed ω and γ in algorithm which extracts amplitude and phase justifies use fixed ω and γ in algorithm which extracts amplitude and phase Calibration with beam ( corrector scans ) induce quite large variations in Calibration with beam ( corrector scans ) induce quite large variations in IQ phase and scales : no feedback, beam jitter IQ phase and scales : no feedback, beam jitter mover calibrations + think of clever calibration scheme (helmholtz coils) mover calibrations + think of clever calibration scheme (helmholtz coils) Commissioned during march '07 run Fast pulse per pulse beam motion Fast pulse per pulse beam motion Less sensitive to beam jitter/drifts Less sensitive to beam jitter/drifts Average scales and IQ phases Average scales and IQ phases Automation : write status into ADC Automation : write status into ADC

Stability results, FY06 running LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session BPM3 BPM10 X Y XY Using SVD algorithm to predict Using SVD algorithm to predict position from spectator BPMs position from spectator BPMs Resolution given by RMS of residual Resolution given by RMS of residual Stability given by drift of residual Stability given by drift of residual Typical good Resolution : Typical good Resolution : ~ 700 nm for ILC cavities ~ 700 nm for ILC cavities ~ 300 nm for SLAC cavities ~ 300 nm for SLAC cavities ILC cold linac prototype SLAC cavities +/- 100 nm 1 hour Drifts can be seen over course of 1 hour Drifts can be seen over course of 1 hour no gain monitoring available yet no gain monitoring available yet correlate with temperature correlate with temperature Long term drifts : clearly degradation of Long term drifts : clearly degradation of resolution resolution Orbit stability shows to be +/- 1 μm over Orbit stability shows to be +/- 1 μm over the course of an hour and RMS on the the course of an hour and RMS on the level of 1 μm level of 1 μm Need to understand systematics way Need to understand systematics way better : calibration tone, more analysis !!! better : calibration tone, more analysis !!! M. Slater/Cambridge

Calibration tone data, March '07 run LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Send triggered CW tone down electronics and monitor gain / phase drifts UCL/RHUL/UC Berkeley Correlate with temperature Calibration tone test runs March '06 (e.g. BPM 41/42) Temperature monitoring on processors x42Amp/q42Amp x42Phase Clear correlation, phase seems fairly stable, need to analyze what effect this has on stability of energy measurements, data needed ! M.Christiakova/Berkeley

Preliminary spectrometer results LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Taking into account ∫B.dl and deflection at center of chicane, can compute correct beam energy Have to subtract incoming orbit in each event : prove we measure just energy ! Have to subtract incoming orbit in each event : prove we measure just energy ! Further detailed analysis, spectrometer stability studies underway... Further detailed analysis, spectrometer stability studies underway... More and better data to come in July... More and better data to come in July... PRELIMINARY !! S. Kostromin/Dubna

Outlook LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session Planned July run Installation of new BPM prototype at center of chicane on mover system Further data taking, full commissioning of calibration system will be first good run with complete chicane up and running also commissioning of synch stripe measurements (T475) : additional energy measurement systematics ! additional energy measurement systematics ! Future plans to install metrology grid (M. Hildreth et al.) Future plans to install metrology grid (M. Hildreth et al.) Understanding mechanical stability ~1m BPM Station Interferometer arms Metrology Grid : Crucial for Mechanical Stability Tests “straightness monitor” “reference bar” for triangulation. Contains internal interferometer for monitoring (LiCas) Publications in pipeline Publications in pipeline More systematic understanding of complete system needed ! More systematic understanding of complete system needed !

Backup Slides LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session

Stability/resolution over 48 hours LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session

Linked resolution LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session

Helmholtz calibration results LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session

Helmholtz calibration results LCWS DESY Hamburg, Saturday 02/06/2007 – MDI session