Electromyography (EMG)

Slides:



Advertisements
Similar presentations
Neuromuscular Adaptations to Conditioning
Advertisements

T.E.N.S. Trancutaneous Electrical Nerve Stimulation
7th Lecture Dimitar Stefanov. Recapping Three types electrodes are used for sensing of EMG signals: 1.indwelling (intramuscular) electrodes (single fiber.
Cryotherapy or ice therapy is the application of cold to the body tissues after injury. This practice is as old as medicine itself. Nowadays, local cold.
Differential Amplifier
ENGR 518 – Experimental TechniquesENGR 518 – Experimental Techniques Katlin LandersKatlin Landers ELECTROMYOGRAPHY.
Electromyography (EMG) Instrumentation
Biofeedback Jennifer L. Doherty-Restrepo Entry-level Athletic Training Education Program PET 4995: Therapeutic Modalities.
The electrocardiogram (ECG or EKG)
EMG ELECTRODES. Electromyography (EMG) is a technique for evaluating and recording the electrical activity produced by skeletal muscles. EMG is performed.
Traditional Electromyography (Electrodiagnosis) A Guide to Understanding Concepts of Electrodiagnostics Joseph S. Ferezy, D.C. © 1999.
ELECTROCARDIOGRAM (ECG)
Lecture 5 Dimitar Stefanov. Definitions: Purpose of the muscles is to produce tension Metabolic efficiency – the measure of a muscle’s ability to convert.
Electromyography: Physiology D. Gordon E. Robertson, Ph.D. Biomechanics Laboratory, School of Human Kinetics, University of Ottawa, Ottawa, CANADA.
Lecture 6 Dimitar Stefanov.
ELECTROMYOGRAPHY AND MOTOR NERVE CONDUCTION VELOCITY
Control of movement Four subsystems: Lower motor system- Grey matter of spinal cord and brainstem-contain lower motor neurons and lower circuit neurons-the.
EMG Theory of NCS/EMG.
Electromyography: Recording D. Gordon E. Robertson, Ph.D. Biomechanics Laboratory, School of Human Kinetics, University of Ottawa, Ottawa, CANADA.
ELECTRIC CURRENTS BY B.Nelson.
BIOFEEDBACK.
© 2004 Electromyographic Biofeedback Chapter 18. © 2004 Purpose To measure, process, and feedback biophysical information Biofeedback does not monitor.
Electromyography (EMG) Theory of Operation & Underlying Anatomical and Physiological Issues.
Chapter 37 Nervous System.
Applied Nerve & Muscle Physiology : Nerve Conduction Study ( NCS) )and Electromyography ( EMG) Dr Taha Sadig Ahmed Physiology Department, College of Medicine,
Ravi Soni MTech III Sem.  Amazing solution for those who had lost their voice but wish to speak over phone.  Developed at the Karlsruhe Institute of.
Chapter 9.2: Electrochemical Impulse Pages
Electromyography IE 665 Dr. Sengupta. Outline Muscle Moment – Moment Arm Review of Muscle Contraction Physiology Physiological Basis of EMG Methods of.
In The Name of Allah The Most Beneficent The Most Merciful 1.
OBJECTIVE Without reference, identify at least four out of six principles pertaining to the application of transducers related to patient care.
ECG How ECG is done? The electrical impulses originating from the heart can be transmitted to the body surface because the body contains fluids.
The Nuts and Bolts of Neurology
Electromyography (EMG)
Copyright © 2004 Pearson Education, Inc., publishing as Benjamin Cummings Human Anatomy & Physiology, Sixth Edition Elaine N. Marieb PowerPoint ® Lecture.
Dr. Aida Korish Assoc. Prof. Physiology KSU
Differential Amplifier. Surface Electromyography for Noninvasive Characterization of Muscle R Merletti, A Rainodi and D Farina.
Electromagnetic Spectrum
EE 4BD4 Lecture 10 Electromyography.
Electromyography (EMG)
Electromyography Tatiana Steinwarz.
Functional Organization of Nervous Tissue Chapter 11
Motor Unit & It ’ s Excitation By: Baljit Brar. What Is a Motor Unit? A Motor Unit is described as being a motor neuron plus the muscle fibres that it.
Electrocardiograph. History Italian scientist Carlo Matteucci realizes that electricity is associated with the heart beat Irish scientist.
PROSTHETIC LEG PRESENTED BY:-AWAIS IJAZ HASNAT AHMED KHAN.
Electromyography E.M.G..
Nerve Action potential L 21
Electricity within the body lect.8
Dr. Aida Korish Assoc. Prof. Physiology KSU
Three types electrodes are used for sensing of EMG signals:
Experimental methods in nerve muscle physiology
Objectives Define what is nerve conduction study (NCS) and electromyography ( emg) . Explain the procedure of NCS using Abductor Pollicicis Brevis muscle.
Lecture 1 Technological Principles of Medical Instrumentation
CRYOTHERAPY د. أحمد أبوالعينين
Dr. Salah Elmalik MBBS,PhD
Applied Nerve & Muscle Physiology : Nerve Conduction Study ( NCS) )and Electromyography ( EMG) د. طه صادق أحمد بكالوريوس الطب و الجراحة ( جامعة الخرطوم.
Physiologic signals Lecture 6.
Nerve Conduction Study (NCS) and Electromyography ( EMG)
Clinical neurophysiology exploration (cne01)
Electromyography E.M.G..
به نام خداوند جان و خرد کزین برتر اندیشه برنگذرد
Basics of Nerve Conduction Studies Review
Electromyography E.M.G..
27/11/2018.
Biofeedback Chapter 7.
Diathermy and its safe use
Applied Nerve & Muscle Physiology: Nerve Conduction Study ( NCS) and Electromyography ( EMG) Dr. Salah Elmalik.
Dr. Aida Korish Assoc. Prof. Physiology KSU
Dr. Salah Elmalik MBBS, PhD
Biomedical Electronics & Bioinstrumentation
Dr Moizuddin Khan Dr Beenish Mukhtar
Presentation transcript:

Electromyography (EMG) Mark David S. Basco, PTRP Faculty Department of Physical Therapy College of Allied Medical Professions University of the Philippines Manila

Objectives At the end of the session, you should be able to: Discuss the principles of EMG Describe the characteristics of normal EMG Describe the characteristics of abnormal EMG

Definition Study of muscle function through the examination of electrical signals ‘Electro’ – electric ‘Myo’ – muscle ‘Graphy’ – to graph / to measure Why use EMG? In vivo examination of muscle activity Quantifies muscle activities Clinical vs. Kinesiological

History 1791 Luiggi Galvani, depolarized frog legs with metal rods 1838 Carlo Matteucci, proved electrical currents originated in mm 1849 Du Bois-Reymond, designed a galvanometer, reduced impedance by popping blisters

History

Principles Involves detecting, amplifying, and displaying electrical changes that occur during mm contraction Electrical events are amplified electronically, visualized on a tv-like monitor or even transformed into sound TV like monitor: Oscilloscope

Principles TV like monitor: Oscilloscope

How can we detect electrical signals?

Electrodes Transducer; device converting one form of energy to another Different types Surface electrodes Fine – wire indwelling electrodes Needle electrodes

Electrodes Surface Electrodes Applied on the surface of the skin Measures signals from large muscles that lie close to the skin Often silver/silver chloride, coated with a suitable conducting gel, and can be taped on the skin Refer to Sullivan for pictures

Electrodes Fine-wire Indwelling Electrodes Two strands of 100micro-meter wires inserted into the muscle belly Used for monitoring activity from deep mm, small mm, or narrow mm May not be useful for large mm

Electrodes Needle Electrodes Used to record motor unit potentials from different parts of the muscle Fine needles with electrodes inserted into the muscle The bare tip of the needle serves as the recording electrode

Electrodes Recording / Detection Electrode Either surface, needle, or fine-wire Ground Electrode Provides a mechanism for cancelling out the interference effect from external noise Surface electrode Attached near the recording electrode

Electrodes

Electrodes

Electrodes

Myoelectric Signal Variables that may affect MUAP Proximity of electrodes to the fibers that are firing Number and size of fibers in the motor unit Distance between the fibers Size of the electrodes Distance between the electrodes MUAP – Motor Unit Action Potential

Myoelectric Signal Cross – talk Electrical overflow Electrical activity from nearby contracting muscles reaches the electrodes Controlled by: Careful electrode placement, spacing, choice of size and type of electrode MUAP – Motor Unit Action Potential

Myoelectric Signal Artifacts Unwanted electrical activity Arises outside of the tissues beinng examined May distort output signals markedly Sources Movement Power line ECG MUAP – Motor Unit Action Potential

Myoelectric Signal Movement Artifact Skin Electrode Contracting muscle Electrode cables Produces high voltage, high frequency artifacts Controlled by: Firm fixation of electrodes, firmly taping cables, skin preparation Movement of skin, movement of electrode or pressure on electrodes, and movement of electrodes within contracting muscles usually produces artifacts below 20Hz causing minimal interference can easily be controlled.

Myoelectric Signal Power line Interference Human body attracts electromagnetic energy from power lines Loose electrode attachments, broken or frayed electrodes, broken electrode wires Diathermy, ES, cellphones, radios, vibrators

Myoelectric Signal ECG can occur if electrodes are placed on the trunk, upper arm, or upper thigh How to reduce this artifact? Correct application of ground electrode Use of amplifier with appropriate characteristics

Amplifier Signals (microvolts) recorded have to be amplified about a thousand times for them to be displayed on an oscilloscope and be heard through speakers or be recorded on a chart

Amplifier The electrical potential recorded by the electrodes is composed of EMG signal from mm contraction and unwanted noise EMGs use differential amplifiers Ground electrode and active electrode supply input to the amplifier Difference between the signals will be amplified and recorded If two electrodes receive equal signals, no activity is recorded

Amplifier

Amplifier

Amplifier Common Mode Rejection Ratio Measure of how much the desired signal is amplified relative to the unwanted signal The higher the value, the better A good differential amplifier should have a CMRR exceeding 100,000:1

Amplifier Signal-to-noise Ratio Reflects the ability of the amplifier to limit noise relative to amplified signal Ratio of wanted to unwanted signal Noise generated from Internal electrical components

Amplifier Gain Refers to amplifier’s ability to amplify signals Ratio of output signal to input signal level

Amplifier Input Impedance Resistive property observed in AC circuits Reduced by: Larger electrodes Skin preparation

NORMAL EMG

Insertional activity Caused by the needle breaking through muscle fiber membranes Also seen during needle repositioning Normally stops when the needle stops moving. May be described as Normal Reduced Prolonged Reduced absence; electrode not on excitable tissue / fibrotic mm tissue Prolonged - Hyperexcitable mm / during denervation / mm inflammation

Electrical Silence (-) electrical activity If needle is at the NMJ/motor end plate, very small spontaneous potentials may be observed Any other activity observed indicates pathology

Interference Pattern Normally seen with strong muscular contraction Individual potentials are summated with increasing number of motor units firing at higher frequencies

Abnormal EMG

Fibrillation Potentials Represents spontaneous, repetitive depolarization of a single mm fiber NOT visible May result from Denervation Metabolic dysfunction Inflammatory diseases (polio) Trauma Indicative of LMN lesions

Fibrillation Potentials

Positive sharp waves Represents asynchronous discharge of a number of denervated mm fibers Reflects an altered membrane excitability Usually accompanied by fibrillation potentials

Fasciculation Potential Represent spontaneous discharges from a group of mm Often visible; small mm twitches Seen with irritation or degeneration of ant horn cell, chronic pni, nerve root compression, cramps or spasms

Fasciculation Potential May be seen in NORMAL individuals Pathological significance if seen with fibrillation and positive sharp waves

Myotonic / Complex Repetitive Discharge Seen with lesions of the anterior horn cell and peripheral nerves, myopathies Dive-bomber sound that occurs on needle insertion, voluntary contraction Probably triggered by movement of needle electrode within unstable mm fibers

Myotonic / Complex Repetitive Discharge

Nerve Conduction Velocity (NCV) Mark David S. Basco, PTRP Faculty Department of Physical Therapy College of Allied Medical Professions University of the Philippines Manila

Objectives At the end of the session, you should be able to: Discuss the principles of NCV Describe motor NCV testing Describe sensory NCV testing

Principles Involves direct stimulation to initiate an impulse in motor or sensory nerves (evoked potential) The conduction time is measured to determine the presence or absence of a lesion It provides data on the ability of the nerve to transmit impulses

Stimulating electrode Bipolar electrode with anode and cathode Pulse duration: 0.1 msec Frequency: 1Hz Contraindicated in patients with Indwelling cardiac catheter Central venous pressure line Contains a trigger mechanism that corresponds to a sweep on the screen Cathode stimulating electrode (black) Anode inactive electrode (red)

Stimulating electrode

Stimulating electrode

MOTOR NCV testing

Stimulation and Recording Evoked potential recorded from a distal muscle innervated by the nerve under study Small surface electrodes are used to record the evoked potential from muscles

Stimulation and Recording Recording electrode Placed over the belly of test muscle Reference electrode Over the tendon of the mm; distal to the active electrode Ground electrode Placed over a neutral area between the electrodes and the stimulation site

Stimulation and Recording

Stimulation and Recording

Stimulation and Recording Stimulus starts low and is slowly increased Intensity is increased until the evoked response no longer increases in size Supramaximal stimulus M-wave M-wave MAP (motor action potential) CMAP (compund motor action potential) Represents the summated activity of all motor units in the muscle that responded to the stimulation

Stimulation and Recording Conduction velocity Determined by dividing the distance between the two points by the difference between the two latencies Meters/second Proximal latency Distal Latency Latency time elapsed from the initial propagation of nerve impulse to the depolarization of mm fibers

Stimulation and Recording For example: Proximal latency: 7msec Distal latency: 2 msec Conduction distance: 300 mm or 30 cm CV = 30 cm / (7msec – 2 msec) = 30 cm / 5 msec = 60 m/s Latency time elapsed from the initial propagation of nerve impulse to the depolarization of mm fibers

Stimulation and Recording How to interpret? Upper Extremity: 50 – 70 m/s [60 m/s] Lower Extremity: 60 m/s

Sensory ncv testing

Stimulation and Recording Ring electrodes

Stimulation and Recording Techniques Orthodromic conduction Antidromic conduction

Stimulation and Recording Orthodromic conduction Normal way physiologic responses travel

Stimulation and Recording Antidromic conduction Stimulus applied to proximal sites and is recorded at the index or long finger

Stimulation and Recording How to interpret? Normal sensory NCV 40 – 75 m/s Sensory NCVs have been found to be faster compared with the motor NCVs

Factors that may affect motor and sensory conduction Increase in body temperature 5% increase in conduction velocity per degree Reduced latencies UE faster than LE Proximal segments faster than distal segments

OTHERS

H Reflex Hoffman Reflex Measures the integrity of both motor and sensory fibers For example Submaximal stimulus to the tibial n. in the popliteal fossa Motor response recorded from the medial soleus (rich in mm spindles) Norm: 29.8 msec

F Wave Elicited by UE: 30 seconds LE: less than 60 seconds Supramaximal stimulation of a peripheral nerve in a distal site resulting in both orthodromic and antidromic conduction. Antidromic conduction extends to the spinal cord Most helpful in conditions where the most proximal portion of the nerve is affected UE: 30 seconds LE: less than 60 seconds