A Wireless, Implantable Intra- Ocular Pressure Sensor for the Management of Glaucoma Gabriel Simon, M.D. Ph.D. Sept 16, 2008 ESCRS.

Slides:



Advertisements
Similar presentations
Philosophies in Glaucoma
Advertisements

Il controllo circadiano della pressione intraoculare: implicazioni farmacologiche N. ORZALESI, P. FOGAGNOLO Università degli Studi di Milano Ospedale San.
V Glaucoma Implementing NICE guidance 2009 NICE clinical guideline 85.
Goldmann Applanation Tonometry
WHICH ANGLE OF HEAD-DOWN TILT BETTER SIMULATES THE INCREASE IN INTRA-OCULAR PRESSURE THAT OCCURS IN MICROGRAVITY? Luís Francisco B. Chotgues, MD Dario.
Scanning Tonometer for Eye Pressure (STEP™) Problem Measurement of eye pressure by Goldmann applanation tonometry is a technique first embodied in the.
Biomechanical Properties of the Cornea in Normal- Tension Glaucoma Authors: Leonidas Traipe Ines Cayuqueo Fabiola Cerfogli Claudia Goya Allister Gibbons.
Transpalpebral Tonometry H.v.Below, O.Riemer Common practice for ambulant and stationary ophthalmology Grimma/Sachsen.
Canadian Ophthalmological Society Evidence-based Clinical Practice Guidelines for the Management of Glaucoma in the Adult Eye.
Live tissue imaging (cornea) Optical imaging of tissue  cells Histology and Electron Microscopy can image ECM. Requires sectioning and staining. Stroma.
A Prospective Trial Evaluating Scleral Rebound Tonometry Sara Duke, MD, Andrew Logeman, Shuchi Patel, MD Loyola University Chicago, Department of Ophthalmology,
Applanation tonometry:
Recent Advances in IOP measurements
Comparison between Dynamic contour tonometry, Goldmann applanation tonometry and Ehlers-corrected Goldmann applanation tonometry in eyes after Laser In.
Sensing and Actuation in Miniaturized Systems presentation#3 Chia-Min Lin 2007/12/26 Parylene-based pressure sensor for IOP measurement Presenter : Chia-Min.
SEI IOP and its correlations in a population based study in Central India. The Central India Eye and Medical Study. Krishna Bhojwani, Vinay Nangia, Jost.
The Diabetic Retinopathy Clinical Research Network Repeated Intravitreous Ranibizumab Injections for DME and Risk of Sustained IOP Elevation or Need for.
CR The Glaucoma Tonometer. What sets it apart? Reichert’s 7CR Auto Tonometer + Corneal Response Technology takes corneal biomechanical properties into.
INTRAOCULAR PRESSURE LECTURE
OPEN ANGLE GLAUCOMA Frank J. Weinstock, MD, FACS Professor of Ophthalmology- NEOUCOM Canton, Ohio USA.
Comparison of the Diaton Transpalpebral Tonometer Versus Goldmann Applanation R. S. Davidson 1 ; N. Faberowski 2 ; R. J. Noecker 3 ; M. Y. Kahook 1 1.
Variations in corneal biomechanical parameters and central corneal thickness during the menstrual cycle. David Zadok, MD, Yakov Goldich, MD, Yaniv Barkana,
A Prospective Trial Comparing Scleral Pneumotonometry to Goldmann Applanation Tonometry Sara Duke, MD, Usiwoma Abugo, BS, Shuchi Patel MD Loyola University.
TRANSPALPEBRAL TONOMETER APPLICATION DURING INTRAOCULAR PRESSURE EVALUATION IN THE PATIENTS WITH REFRACTION ANOMALY BEFORE AND AFTER KERATOPHOTOREFRACTIVE.
Retinal nerve fiber layer thickness change in patients with wet AMD treated with ranibizumab, short term results Advantages: To determine the effect of.
Presents: Diaton Tonometer Tonometry through the Eyelid! September 2006.
Glaucoma Screening With a New Non-Contact Air Impulse Tonometer Pandelis A. Papadopoulos,1,2, Dimitrios P. Bessinis 3, Alexandros A. Papadopoulos 1, Konstantinos.
Glaucoma and Penetrating Keratoplasty : Incidence, Risk Factors, and Outcomes Sonika Gupta Consultant Ophthalmology Max Eye Care New Delhi, India Author.
Oliver Findl 1,2, Nino Hirnschall 1,2, Alja Crnej 2 1 VIROS Vienna Institute for Research in Ocular Surgery, Hanusch Hospital, Vienna, Austria 2 Moorfields.
OCULAR FEATURES OF VERNAL KERATOCONJUNCTIVITIS IN CHILDREN REQUIRING TRABECULECTOMY WITH MITOMYCIN-C FOR STEROID-INDUCED GLAUCOMA CORDELIA CHAN FRCSEd.
In-vitro measurement of Intra- Ocular Pressure Alex Luce Undergraduate Research Symposium 2008.
Variations in corneal biomechanical parameters and central corneal thickness during the menstrual cycle. Yakov Goldich, MD, David Zadok MD, Yaniv Barkana,
Tonometry Objective measurement of IOP based, most commonly, on the forces required to flatten the cornea,or the degree of corneal indention produced.
Comparison of the Diaton Transpalpebral Tonometer Versus TonoPen- Applination Theodore H. Curtis, M.D.1, Douglas L Mackenzie, M.D.1, Robert J. Noecker.
A Prospective Trial Comparing Pneumotonometry of Cornea and Sclera Usiwoma Abugo BS 1, Sara Duke MD 2, Shuchi Patel, MD 2 1 Loyola University Chicago,
The effect of corneal edema on intraocular pressure measurements after phacoemulsification surgery Ilker Eser M.D. 1 Arzu Taskiran-Comez M.D. 1 Baris Komur.
GLAUCOMA داء الزرقاء.
Visual and IOP Outcomes after PRK in Pigment Dispersion Syndrome [Poster Number: P190] Kraig S. Bower, Denise A. Sediq, Charles D. Coe, Keith Wroblewski,
TONOMETER.
The recent advancement in the medical industry has given rise to the best healthcare and diagnostic devices. Today, Tonometer has created a revolution.
A WIRELESS PASSIVE SENSOR FOR TEMPERATURE COMPENSATED REMOTE PH MONITORING IEEE SENSORS JOURNAL VOLUME 13, NO.6, JUNE 2013 WEN-TSAI SUNG, YAO-CHI HSU Ching-Hong.
Mathematical Description of the Ocular Response Analyzer Applanation Signal in Post-LASIK Ectasia Mujtaba A. Qazi, M.D. 1,2 David A. Luce, Ph.D. 3 Jay.
Understanding GLAUCOMA… The Science Behind Current Testing and Therapy Mindy J. Dickinson, OD Midwest Eye Care, PC Omaha/Council Bluffs.
Date of download: 5/28/2016 Copyright © 2016 American Medical Association. All rights reserved. From: The Effect of Latanoprost, Brimonidine, and a Fixed.
9. Juni 2016HS-Präsentation1 Tonometer Prisms Disposable versus reusable.
BIOMECHANICAL PROPERTIES OF THE CORNEA IN PSEUDO-EXFOLIATION SYNDROME Aylin Kılıç Ertan, MD Özgür Bilen, MD Kudret Eye Hospital, Ankara, Turkey There is.
The Royal Victorian Eye and Ear Hospital 24-hour eye pressure and glaucoma Dr Simon Skalicky FRANZCO, MPhil, MMed (Ophthal Sci), MBBS (Hons 1) Visiting.
Concepts 3 Material 9 Method 10 Results/Discussion 11 Conclusion Questions.
D.r Nishant Nawani, MS Dr. Surinder Singh Pandav, MD Dr. Amit Gupta, MD Dr. Sushmita Kaushik, MD Advanced Eye Centre PGIMER, Chandigarh The authors have.
Date of download: 7/8/2016 Copyright © 2016 American Medical Association. All rights reserved. From: Dynamic Contour Tonometry: A Comparative Study on.
Use of the ICare Home Rebound Tonometer for home self-monitoring of intraocular pressure in a Singaporean population Jeanne Joyce OGLE1, Wai Cheng SOO.
CRF variations ( initial CRF – final CRF )
Ophthalmology – Raised IOP Referrals (Adults only)
Glaucoma Conference Intraocular pressure and tonometry
Effect of Primary Phacoviscocanalostomy / Viscocanalostomy On Intraocular Pressure Of Normal Tension Glaucoma Patients: 3-Year Results Derek Ho¹, Adesuwa.
Intraocular pressure & tonometry
Influence of UVA-Riboflavin corneal collagen cross-linking on biomechanical properties of keratoconic eyes David Zadok MD, Yakov Goldich MD, Yaniv Barkana.
Date of download: 10/26/2017 Copyright © ASME. All rights reserved.
IOP = intraocular pressure; SD = standard deviation.
Copyright © 2013 American Medical Association. All rights reserved.
R. S. Davidson1; N. Faberowski2; R. J. Noecker3; M. Y. Kahook1 1
2/24/2013.
Federal University of São Paulo, Brazil
Corneal Biomechanical Assessment using Dynamic Ultra High-Speed (UHS)
Yakov Goldich MD,Yaniv Barkana MD, Isaac Avni MD, David Zadok MD,
IOP is not affected by microfibril deficiency.
Jun Her MD. PhD, Sung-wook Kim MD, Jae-hwan Ahn MD
Percentages of eyes with BL IOP ≥21 mm Hg and
Box and whiskers plot for bIOP, Goldmann applanation tonometry IOP (GAT) and Goldmann adjusted (GATAdj) in ocular hypertension (OHT), primary open-angle.
Presentation transcript:

A Wireless, Implantable Intra- Ocular Pressure Sensor for the Management of Glaucoma Gabriel Simon, M.D. Ph.D. Sept 16, 2008 ESCRS

2 Current IOP Measurements Current Methods of IOP Measurement  Applanation Tonometer (Goldmann) Requires contact with eye surface Errors due to corneal thickness, past surgeries, etc.. Clinical & limited home use  Dynamic Contour Tonometry (PASCAL DCT) Relies on contour matching More accurate than applanation tonometers Clinical use only  Electronic Indentation Tonometer (Tono-Pen) Limited accuracy Home and clinical use No method for continuous, remote IOP monitoring currently exists  Continuous monitoring will allow complete glaucoma management RF Transmission of Data & Power  Current range exceeds 3 meters

3 Background – IOP Monitoring Current Approach – Reactive  Sample every 3-6 months  Tonometer measurement, clinic-based Corneal thickness-induced errors IOP Fluctuations can vary over 24-hour period  Diurnal variations in IOP (2008 Sit, et al.)  Circadian/Hourly fluctuations (2006 Barkana, et al.) Continuous IOP Monitoring – Managing IOP  Sample continuously, with daily upload  Accurate to 0.5 mmHg  Ophthamologist can monitor IOP trends daily  Requires: Wireless, high-sensitivity sensor Ultra-low power circuitry Miniature packaging (<5mm per side) for implantation External RF-charging and data collection device

4 Implantable Sensor Concept Implantable sensor for continuous patient monitoring  Prototyped at Purdue Brain-Computer Interface (BCI) Lab  Implanted in anterior chamber or suprachoroidal space 300µm overall thickness  Capacitive Sensor 0-50mmHg sensitivity 0.5mmHg accuracy  Amplifier and Telemetry Capture IOP every 5 minutes RF download and recharge  External Unit Stores & transmits all IOP data Held to eye for <10 seconds Provides significant improvement in quality of care

5 IOP Sensor Device Remote patient monitoring  Provides near-continuous IOP data on daily basis Disease progression & drug monitoring  Continuously sampled IOP data evaluated daily or internet interface, Bluetooth compatible  E-consultation if necessary  Minimize cost & time while improving quality of care Daily reports of IOP vs. sporadic visits Patient and Doctor Interface  Promotes patient compliance  Easy to use for physicians IOP trends and warnings based on relevant information

6 Intra-Ocular Pressure Sensor Design Research partners: Gabriel Simon 2, Babak Ziaie 3, SOLX 4 2 Professor of Ophthalmology, Boston University 3 Associate Professor, Department of Electrical Engineering, Purdue University 4 Boston University startup company Wireless data module LPF mixer LNA VCO antenna S1 S2 Voltage regulator Battery Biasing circuit Powering module External receiver with graphical user interface and power-coupling hardware 1) Amplifying module 2) Wireless data module 3) IOP sensing module 4) Powering module 5) External user interface IOP sensing module amp Amplifying module Clk

7 IOP Sensor Concept – Location Suprachoroidal Implant  Sensor surface protrudes into Anterior Chamber IOP measured in AC  15-minute implant procedure  Posterior Chamber: same as IOL procedure  Similar to gold shunt/suprachoroidal procedure  Minimal, transient complications Posterior Chamber Implant  Same as IOL procedure  No learning curve  5-10 minute procedure Candidate patients  Glaucoma patients  Cataract - IOL patients Does not preclude later IOL implant

8 Implant Prototypes and Materials Silicon substrate Implant materials:  Low-Temperature Co-fired Ceramic (LTCC)  Silicon  PMMA  Liquid-Crystal Polymer (LCP) LTTC substrate

9 Posterior Chamber Sensor Implant PMMA or Liquid-Crystal Polymer (LCP) substrate

10 Wireless Power Transmission Initial studies show that we can achieve low-power RF transmission from a miniature implantable device for ocular implant applications. In-vivo experiments show that the implant was measured to have a sufficient signal-to-noise ratio margin for high data-rate transmission, validating this approach for intra-ocular pressure telemetry.

11 Wireless Transmission - Testing The power received is at least 10 dB greater than the MDS (minimum detectable signal) and we can achieve successful wireless data transfer. LTCC based loop antennas provide less attenuation caused by tissues after implanting than the silicon based monopole antennas.

12 IOP Sensor Data Management Data Collection  Patient recharges & uploads once/day  Data Hub at patient’s residence  Relies on PC with modem, phone line, or cell phone Data Management  Multi-user, web-based database server  HIPAA compliant, with backup security  Allows multi-point access via internet Clinical Data Analysis  Patient information provided to clinician via internet UI IOP summary trends Medication compliance Capable of supporting additional physiological data capture  Presented to clinician in simple overview 1-5 minutes of review per patient  24-7 access to database with subscription

13 Summary and Future Studies Continuous, wireless IOP sensor will provide a new level of glaucoma management  Remote patient monitoring via internet  Improved treatment, compliance, and outcomes  Reduced office visits and cost-to-treat Future: Combination diagnostic and treatment  Passive suprachoroidal sensor with shunting capabilities flow channels incorporated into sensor package  Active suprachoroidal glaucoma manage- ment device Remote IOP monitoring capability Adjustable flow resistance –Wireless adjustment of outflow facility –Based on IOP signal from sensor –Remote adjustment via external charging device Aqueous Outflow