Use and Maintenance of Micropipets

Slides:



Advertisements
Similar presentations
Measuring Volume Biotech I pg. 301 to 323 in Basic Laboratory Methods.
Advertisements

Salk Institute Mobile Lab Gel Electrophoresis Student Instructions VWR Set up.
An Introduction to Microvolumetrics and Pipetting
Use and Maintenance of Micro-pipettes
Measuring Small Amounts Part 1: Using pipets and micropipets.
Pipetting Pipetting involves drawing a liquid into a pipet and allowing liquid to drain from the pipet in a controlled manner. Pipetting is used to quantitatively.
Cautions! 1. Set pipet volume only within the range specified for that micropipet 2. Do not attempt to set a volume beyond the pipet’s minimum or maximum.
Oct. 22, 2012 Generosity Ludicrous: causing or deserving laughter because of absurdity Do Now: What is a microliter? How many microliter are in one liter?
آزمايشگاه ايمونولوژي  كارشناس مسئول برنامه خانم قارداشي.
How to Use a Micropipettor A micropipettor is a precision instrument – expensive (about $200 each), easily broken. Good technique will let it work for.
What are Micropipettors?! We will be using 3 different sizes of micropipettors: ◦ P20 ◦ P200 ◦ P1000 Very expensive and very fragile – USE Responsibly!!!
Balances and Micropipettes By: Paola Restrepo and Ankita Desai.
Use and Functional Verification of Automatic and Micro-Pipettes Module 7: Quality Assurance of Analytical Phase.
Micropipette Tutorial
Solutions and Units of Measure. Today’s Laboratory Objectives To learn how to make solutions properly To learn how to make solutions properly To learn.
Glucose test Ms. Ibtisam alaswad Ms. Nour A. taim.
MLAB 1335 Immunology Serology Terry Kotrla, MS, MT(ASCP)BB
Measuring volumes which are what we scientists call “very, very small”
Introduction of CLS 333.  Course Description:  This is a required course in biochemistry for preparing a generalist in medical technology program. The.
Introduction to Equipment
Micropipette Tutorial From the Science Dept at SHS
MLAB 1335 Immunology Serology Terry Kotrla, MS, MT(ASCP)BB
Biochemistry Practical Dept.of Biochemistry Zhihong Li ( 李志红 )
Transformation is the process of introducing DNA into bacterial cells.
Biochemistry Practical Dept.of Biochemistry Zhihong Li ( 李志红 ) PhD.
Micropipettes and Centrifuges Bio 9. Micropipettes are essential equipment in a modern biology lab For moving liquids from container to container Can.
Mastering the micropipette. What is a micropipette? How do you use a micropipette? How can I determine if I’m pipetting correctly?
Micropipettes Step-by-Step to use Micropipettes 1) Check the volume 2) Attach disposable tip (different tips for each size micropipette) 3) Depress the.
CHAPTER 1: Some Tools of the Trade Lab Purpose of Lab 1.1 Become familiar with the small volumes of solutions used in molecular biology Introduce.
How to use a micropipette. When is a micropipette needed? Micropipettes are precision instruments designed to measure and transfer small volumes of liquids.
 Accurately measure and transfer small amounts of liquid  Volumes commonly range from 0.5 – 2000  L (1/millionth of a liter aka )  2-20  L.
1. Which tool would you use for picking up 15 microliter? Which tool for 25 ml? Why? 2. Why do the little test tubes look this way? 3. Answer “Micropipette.
Use of the Labnet Micropipettor. Purpose of the Pipettor Biotechnologists use a pipettor to transfer precise amounts of liquids from one container to.
Micropipettes Why they are useful How they work How to care for them.
Automatic pipettes use and calibration
Biotech 8/20/13 What pump would you use for a 1 or 2 ml pipet? What pump would you use for a 5 or 10 ml pipet?
Proper Micropipette Technique and Slide Preparation
Use and Maintenance of Micro-pipettes
1 REGIONAL BIOTECH AT THE UNIVERSITY OF TEXAS-PAN AMERICAN BIOSCIENCE EDUCATION FOR THE RIO GRANDE VALLEY.
Lab Safety And Skills SESEY 2006 Kelly Lab Group.
ABE Labs 1.1 and 1.2 Tools of the Trade. CHAPTER 1: Some Tools of the Trade Lab
An Introduction to Microvolumetrics and Pipetting
Biochemistry LABORATORY NOTEBOOK AND EXPERIMENT REPORTS
Automatic Pipettes Use And Calibration
Micropipette Tutorial
Techniques in Molecular Biology 2017 Fall Lecture -2-
Micropipet.
Measuring Volume.
مهارتك الأساسية في معامل التقنية الحيوية
Techniques in Molecular Biology 2017 Fall Lecture -2-
Measurement of Volume Chapter 20.
Use of the Labnet Micropipettor
Metric System and Lab Skills
Use and Maintenance of Micro-pipettes
Introduction To Medical Technology
Introduction of CLS 333 Demonstrated by: Daheeya AlEnazi 3rd floor,office#110 office hour: wed 9-10.
Micropipette Tutorial
Measuring volumes which are what we scientists call “very, very small”
Micropipettes.
How to Micropipette 1) Never adjust the volume below 5 uL or above 50uL 2) Never leave pipetter lying on bench with a tip attatched Basic Methods - Set.
Pipette Technique.
Restriction Enzymes.
Do Now How many ul (microliters) is in 1 liter?.
Using the micropipette
Grade Level: Middle School
Do Now How many ul (microliters) are in one liter?
Measurement & Micropipetting
© 2013 Elsevier, Inc. All rights reserved.
Laboratory Glassware.
Micropipette Tutorial
Presentation transcript:

Use and Maintenance of Micropipets

Introduction Automatic pipettes are used to accurately transfer small liquid volumes These are continuously adjustable digital pipettes Each pipette can be set to transfer any volume within its own volume range

Parts of the Pipette Pipette tips

Step-wise Operation of the MICRoPipet (1) Set the volume (2) Attach disposable tip (3) Depress the plunger to the first stop (4) Immerse tip in sample (5) Draw up the sample (6) Pause (7) Withdraw the tip (8) Dispense the sample (9) Withdraw the pipette (10) Release plunger (11) Discard the tip As you read the following steps, you should manipulate the automatic pipette, disposable tips, and the vials of colored solution TO LEARN HOW TO PROPERLY HANDLE THE PIPETS. STEP-WISE OPERATION OF THE AUTOMATIC PIPET 1) SET THE VOLUME: Hold the pipet in one hand while turning the volume adjustment knob until the correct volume shows on the digital indicator. Proper Grasp to Set Volume Turn 1/3 revolution above the desired setting and then slowly down to the desired setting for optimal accuracy. 2) ATTACH DISPOSABLE TIP: Press disposable tip of PROPER SIZE firmly to the shaft of the pipet to ensure a positive airtight seal. Tip Sizes Attaching Tip 3) DEPRESS THE PLUNGER TO THE FIRST STOP: Using your thumb, depress the plunger gently and smoothly until resistance is felt. The point of resistance is the first stop. It is that part of the stroke that extends the piston to the calibrated volume which is displayed on the digital indicator. Depress Plunger 1st Stop 4) IMMERSION INTO THE SAMPLE: With the pipet in a vertical position, immerse the disposable tip into the sample liquid to a depth of at least 3-4 mm (millimeters) below the liquid surface for the common pipets, Immerse Tip in Sample 5) DRAW UP THE SAMPLE: To aspirate the sample into the tip, allow the pushbutton to return slowly and smoothly to the fully extended UP POSITION. NEVER LET THE PLUNGER SNAP UP! This draws the exact calibrated volume into the tip if the tip remains below the liquid surface during withdrawal. 6) PAUSE: Wait a few seconds to ensure that the full volume of sample is drawn into the plastic tip. WAIT LONGER FOR LARGER VOLUMES. WAIT LONGER FOR MORE VISCOUS ("SYRUP-LIKE") SUBSTANCES. 7) WITHDRAW THE TIP: Remove the tip from the sample liquid. No liquid should remain on the OUTSIDE of the tip. Wipe away any droplets on the outside of the tip with a lint-free tissue, such as KIMWIPES, but only wipe droplets from the side of the tip. NEVER TOUCH THE TIP ORIFICE (opening) or you may absorb part of your sample. 8) DISPENSE THE SAMPLE: To dispense the sample from the pipet, a) Touch the tip end to the side wall of the receiving vessel and b) Depress the plunger to the FIRST STOP. c) Pause for at least one second-- or longer for viscous liquids. d) Press the plunger to the SECOND STOP (the second point, of greater resistance, at the bottom of the stroke) to expel any residual liquid in the tip (like "blowing out" a glass pipet). 9) WITHDRAW THE PIPET: With the plunger fully depressed, withdraw the pipet from the receiving vessel carefully, sliding the tip along the wall of the vessel. Holding the tip against the side of vessel is especially important when transferring small volumes of liquid. 10) RELEASE PLUNGER: Gently allow the plunger to return to the UP position. DO NOT allow it to SPRING BACK! 11) DISCARD THE TIP: Discard the tip by depressing the tip ejector button, A fresh tip should be used for each sample to prevent sample carryover. n handling the automatic pipettors. Also, be certain that you can BOTH READ AND SET THE PROPER VOLUMES. PIPETTING CONSIDERATIONS Accuracy and Precision: Accuracy means the closeness with which the dispensed volume approximates the volume set on the pipette. It is specified as mean error, the average deviation of replicate measurements from the expected set volume. Precision is the "scatter" or reproducibility of individual measurements of the same volume. It can also be expressed as standard deviation. Specifications are generally tighter for precision than for accuracy; and, in controlled experiments, the smaller precision specification is a good estimate of pipetting accuracy. The smaller the mean error and/or standard deviation of a measurement, the more accurate and/or precise it is. TO SUMMARIZE, relative accuracies are generally about 1% or less and precision is less than 0.5 % except when transferring the smallest recommended volume for a given pipet model. Us

Operating the Micropipette Step 1: Set the Volume Examples of Pipettors: Volume Adjustment Knob: Digital Volume Indicator:

Operating the Micropipette Step 2: Attach the Disposable Tip       Example of tip sizes: Attaching the disposable tip

Operating the Micropipette Step 4: Immerse Tip in Sample Step 3: Depress the Plunger to the First Stop Step 5: Draw up the sample Step 6: Pause Wait a few seconds to ensure that the full volume of sample is drawn into the plastic tip.

Operating the Micropipette Step 7: Withdraw the Tip Remove the tip from the sample liquid. No liquid should remain on the OUTSIDE of the tip.   Proper Droplet Removal WRONG Droplet Removal

Operating the Micropipette Step 8: Dispense the Sample a) Touch the tip end to the side wall of the receiving vessel b) Depress the plunger to the FIRST STOP. c) Pause d) Press the plunger to the SECOND STOP (the second point at the bottom of the stroke) to expel any residual liquid in the tip                                                                  (a) Start Dispensing  (b) 1st Stop = Dispense (c) 2nd Stop = Expel

Operating the Micropipette Step 9: Withdraw the Pipette With the plunger fully depressed, withdraw the pipette from the receiving vessel carefully, sliding the tip along the wall of the vessel. Gently allow the plunger to return to the UP position. DO NOT allow it to SPRING BACK! Step 10: Release the Plunger

Press ejector button to discard tip. Operating the Micropipette Step 11: Discard the Tip Discard the tip by depressing the tip ejector button, as shown below. A fresh tip should be used for each sample to prevent sample carryover. Press ejector button to discard tip.