Proton Spin In absence of a magnetic field, protons spin at random

Slides:



Advertisements
Similar presentations
Pulse Timing parameters & Weighting
Advertisements

MRI Physics 2: Contrasts and Protocols
Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
Contrast T1 weighted – (MPRAGE-anatomical) T2 weighted – (fmri)
Magnetic Resonance Imaging
Imaging Sequences part I
Chapter 7: Gradient Echo Imaging Methods
Introduction to MRI Magnetic Resonance Imaging
PULSE SEQUENCES Emphasizing the differences among spin density, T1, and T2 relaxation time constants of the tissues is the key to the exquisite contrast.
Richard Wise FMRI Director +44(0)
M R I Pulse Sequences Jerry Allison Ph.D..
BE 581 Lecture 3- Intro to MRI.
Basis of the BOLD signal
PHYSICS OF MAGNETIC RESONANCE
Institute for Biomedical Engineering EXCITE Afternoon Hands-On MRI Sessions: fMRI & DTI.
Magnetic Resonance Imaging - MRI
MR Sequences and Techniques
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
Parameters and Trade-offs
Magnetic Resonance Imaging
Topics spatial encoding - part 2. Slice Selection  z y x 0 imaging plane    z gradient.
Equipment Magnetic resonance imaging (MRI) scan requires the use of a very strong magnetic field. Unlike other devices used in radiology, MR imaging.
Basics of Magnetic Resonance Imaging
Tissue Contrast intrinsic factors –relative quantity of protons tissue proton density –relaxation properties of tissues T1 & T2 relaxation secondary factors.
MRI: an Introduction By Mohammad Ali Ahmadi Pajouh Amirkabir University of Technology Biomedical Eng. Dep.
Magnetic Resonance Imaging 4
Physics of Magnetic Resonance Chapter 12
대구가톨릭대학병원 영상의학과 이 영 환 M M R R Basic Physics. MR Signal T1-, T2-weighted TR, TE MR Signal T1-, T2-weighted TR, TE.
Gradient echo pulse sequences
Principles of Magnetic Resonance
Imaging Sequences part II
Seminar October, 2008 j. brnjas-kraljević. Imaging (MRI)  tomography technique  tomography technique – the volume image is built up by images of thin.
Chapter 4 Mark D. Herbst, M.D., Ph.D.. Magnetization MRI depends on magnetization of tissues Temporary magnetization occurs when the hydrogen nuclei are.
Pulse sequences.
Pulse Sequences Types of Pulse Sequences: Functional Techniques
Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo.
RT 4912 Review (A) Rex T. Christensen MHA RT (R) (MR) (CT) (ARRT) CIIP.
Basic Concept of MRI Chun Yuan. Magnetic Moment Magnetic dipole and magnetic moment Nuclei with an odd number of protons or neutrons have a net magnetic.
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
September, 2003BME 1450 Introduction to NMR 1 Nuclear Magnetic Resonance (NMR) is a phenomenon discovered about 60 years ago.
G Practical MRI 1 Basic pulse sequences.
Nuclear Magnetic Resonance I Magnetization properties Generation and detection of signals.
Contrast Mechanism and Pulse Sequences
MRI Physics Dr Mohamed El Safwany, MD.. MRI Magnetic Resonance Imaging Magnetic Resonance Imaging.
BIOE 220/RAD 220 REVIEW SESSION 6 March 5, What We’ll Cover Today General questions? Spinal cord anatomy review Fat in images T2* vs T2 decay Review.
Magnetic Resonance Imaging
Protons (hydrogen nuclei act like little magnets) MRI Collective Magnetic Moment of Protons (M 0 ) Each pixel is a glass of protons B 0 = 3T (not to scale)
MRI Physics: Pulse Sequences
V.G.Wimalasena Principal School of Radiography
Magnetic Resonance Imaging (MRI). The Components: A magnet which produces a very powerful uniform magnetic field. A magnet which produces a very powerful.
Introducing cross- sectional imaging Dr.Ahmed S. Tawfeek FIBMS(Rad.)
MRI Physics: Spatial Encoding Anna Beaumont FRCR Part I Physics.
RT 4912 Review (C) Rex T. Christensen MHA RT (R) (MR) (CT) (ARRT) CIIP.
MRI: Contrast Mechanisms and Pulse Sequences
Spinning Nucleus Produces Magnetic Moment
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
بسم الله الرحمن الرحيم Dr. Maged Ali Hegazy Assistant Professor Alazhar University.
Lecture:2 INTRODUCTION TO RADIOLOGY.  Students should be able to understand different views of normal and abnormal x-rays  Students should understand.
Chapter 5 Mark D. Herbst, M.D., Ph.D.. The MR Imaging Process Two major functions –Acquisition of RF signals –Reconstruction of images.
10 spring fMRI: theory & practice
FMRI data acquisition.
Where Mt is the magnetization at time = t, the time after the 90o pulse, Mmax is the maximum magnetization at full recovery. At a time = one T1, the signal.
Magnetic Resonance Imaging: Physical Principles
Physics of Magnetic Resonance Imaging
Bioengineering 280A Principles of Biomedical Imaging Fall Quarter 2005 MRI Lecture 5 Thomas Liu, BE280A, UCSD, Fall 2005.
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
MRI: From Protons to Pixels
(4)ELECTRONIC SUPPORT SYSTEM
The echo time (TE) The echo time (TE) refers to the time between the application of the radiofrequency excitation pulse and the peak of the signal induced.
T2 Relaxation Time T2 relaxation time is defined as the time needed to dephase up to 37% of the original value. T2 relaxation refers to the progressive.
Presentation transcript:

Proton Spin In absence of a magnetic field, protons spin at random A magnetic field is used to align them (B0) Current MRI fields run at 0.5 and 3.0 Teslas 1 Tesla is equal to 10,000 Gauss Earth's magnetic field varies from between 0.3 - 0.7 Gauss. Refrigerator magnet ~10 -100 Gauss

MR visible nuclei Protons must be unbound to allow them to spin and change spins Bound protons are not MR visible as they don’t emit RF from energy changes Cortical bone contains bound H and thus always appears black on MR Fat and Water contain lots of unbound H and make soft tissue imaging so sensitive to MR

Precession In a magnetic field the proton precesses at a given frequency based on the Larmor equation This frequency is based on field strength and the specific nucleus – each has a different frequency In MRI the precession frequencies occur in the RF range and are detected by an antenna

Pulse sequence Tipping the spin into a different plane from B0 B1 is the field used to tip the spin measured in degrees from B0 90° RF pulse tips it into the transverse plane or x’ 180° RF pulse inverts it to a –z direction

Pulse Sequence Tipping the spin into another plane puts them into a high energy state once the new field is removed Decay happens rapidly back to the original state and the energy given off by the decay is in the radiofrequency range This energy is called the FID (Free Induction Decay) This is received by an antenna

Signal to Noise Ratio Signal to Noise Ratio The amount of useful signal compared to useless noise from RF Coherent signal is additive while incoherent noise tends to cancel out SNR is proportional to the square root of the number of averages. Doubling the # of slices doubles time but improves SNR by 1.4

T2 Having been tipped into the transverse plane, the net magnetisation begins to dephase (T2*) Rapid decay of unified spin based on tissue and magnetic field

T1 Once fully dephased the spins return to equilibrium (T1) Takes longer than T2 to reach equilibrium

TE and TR TE = Echo Time TR = Repetition time Time from 90° pulse to the time after the 180° that the echo from the tissue is detected TR = Repetition time Time between pulses (ie ms spacing between 90°-180°-180° pulses)

TE governs T2 Long TE allows for differences in T2 times to emerge. Water, with a long T2 time takes longer to decay and thus gives a stronger signal. T2 weighted = H20 bright Using a long TE interval, tissues with long TE becomes brighter. A short TE would make long TE tissues dark

TR governs T1 Tissue with a long TR (like water) takes a long time to recover to equilibrium Using a short TR interval, tissues with short TR (like fat) have given off more signal (brighter) than a slower decaying tissue

PD weighting Proton Density weighting Density of spins or number of protons in a tissue Short TE – decreases water signal Long TR – decreases fat signal Suppression of fat and water Better anatomic detail

MRI Contrast T2-weighting requires long TE, long TR T1-weighting requires short TE, short TR PD-weighting requires short TE, long TR

Four MRI images: Same slice

Figure 1. Representative axial brain images of the eight patients with ACA infarctions. Patient numbers appear in the upper left corner of each panel. Patient 1, T1-weighted axial MRI; patient 2, CT scan; patient 3, T2-weighted MRI; patient 4, T1-weighted MRI; patient 5, proton density MRI; patient 6, CT scan; patient 7, T2-weighted MRI; and patient 8, T2-weighted MRI.

MRA Specific gradients are applied to phase and dephase the free H rapidly H that is flowing into an area that was just dephased has not yet been dephased, thus producing high signal This is repeated rapidly and flow can be monitored

Spin-Echo MRI This “family” includes T1, T2 & Proton Density RF pulses excite the protons to flip 90 degrees from the main magnetic field, then 180 degrees (to refocus the phase of precession). The first “echo” (RF from the tissue) is obtained. This first echo provides the data for either T1 or proton density contrast. To obtain the data for T2 contrast, a second 180 degree pulse is then applied to obtain a second “echo” This is termed the 90-180-180 pulse train. The times between the RF excitations (TR & TE) vary depending on the desired contrast. This produces various tissue signal intensity relationships which are referred to as “weighting” characteristics

Spin Echo (Gp) phase encoding gradient (y plane) (Gf) frequency encoding during signal detection (Gs) slice-selection gradient during RF pulse to spin up and refocus only the slice of interest

Inversion Recovery/ IR In this family the pulse train is 180-90-180 the time of inversion / TI (“tee eye”) = the time between the first 180 and the 90 pulse TI times can be tailored by suppressing signal from specific types of tissue. Greater contrast

STIR/ Short TI Inversion Recovery A short TI value suppresses the signal from fat and causes abnormal water destiny to “light up” brightly. one of the best ways to image bone marrow and spinal cord pathology. (T2 with fat saturation also works well.)

T2W fast-spin echo with fat-sat (saturation>suppression) shows normal meniscus, but increased signal at the posterior of the distal femoral metaphysis in teenager with a cortical desmoid

FLAIR= Fluid Attenuated Inversion Recovery Long TI inversion recovery produces bright signal from brain pathology but CSF signal is suppressed This allows better contrast of CNS lesions when they are located at CSF interfaces (common in MS)

True Inversion Recovery aka T1 FLAIR New pulse sequence on 3T systems Produces true anatomic contrast between grey vs. white brain matter

Gradient Echo or Field Echo In this technique, the protons are excited to flip angles of less than 90 degrees (which is faster than SE) instead of using a 180-degree refocusing RF pulse the precessions are “refocused” by reversing the polarity of the magnetic field gradient (The higher magnetic field side becomes the lower magnetic field side and vice versa.). This can be used to rapidly produce extremely bright water signal in the CSF (the Rapid Myelographic technique) or in synovial fluid.

T2* (tee 2 star) gradient echo/GE with fat saturation a variation of GE and FSE that suppresses the signal from fat while imaging fluids brightly but is extremely fast. another fine technique for imaging bone marrow pathology and pathology of articular cartilage.

Gradient Echo or Field Echo Rapid imaging produces myelographic effect Sensitive to magnetic susceptibility (Hgb)

3D Imaging This modification of Gradient/ Field Echo, and Fast Spin Echo/FSE, acquires data for imaging as a block or cube rather than as slices. The data can later be “cut up” into slices of any thickness in any desired plane. Allows for ultra thin slice images (1mm or less) when necessary.

69 yr old patient presented with loss of consciousness. A) CT demonstrates hypodensity consistent with infarction of basilar artery distribution in the left cerebellar lobe,vermis and medial portion of the right cerebellar lobe. B,C) MRI FLAIR and MRI T2WI demonstrate vasogenic edema in the same distribution. D) ADC map (b=1000) demonstrates diffusion restriction due to cytotoxic edema in the same distribution. E,F) DWI (b=500, b=1000) also demonstrates diffusion restriction due to cytotoxic edema. This case demonstrates acute infarction in the left and right cerebellar lobes and the vermis in a stage of intermixed vasogenic and cytotoxic edema.

What’s wrong here? Sagittal T2 weighted MRI showing a spindle shaped area of increased signal extending from C4 to T1. MADDISON P et al. J Neurol Neurosurg Psychiatry 2000;69:563-564 ©2000 by BMJ Publishing Group Ltd