PHYSICS OF MAGNETIC RESONANCE

Slides:



Advertisements
Similar presentations
Pulse Timing parameters & Weighting
Advertisements

Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
Contrast T1 weighted – (MPRAGE-anatomical) T2 weighted – (fmri)
Imaging Sequences part I
PULSE SEQUENCES Emphasizing the differences among spin density, T1, and T2 relaxation time constants of the tissues is the key to the exquisite contrast.
Proton Spin In absence of a magnetic field, protons spin at random
BE 581 Lecture 3- Intro to MRI.
Magnetic Resonance Imaging - MRI
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
Fysisk institutt - Rikshospitalet 1 Source: Courtesy of Warner Bros Science or black magic? Chap.12 (3) Medical imaging systems: MRI.
Parameters and Trade-offs
Magnetic Resonance Imaging
Magnetic Resonance Imaging
Magnetic Resonance Imaging
NMR Spectroscopy Part I. Origin of NMR. Nuclei in Magnetic Field Nucleus rotate about an axis -- spin Nucleus bears a charge, its spin gives rise to a.
MRI. Magnetic Resonance 1.Principle first observed in Used for spectroscopy and imaging 3.Imaging techniques are a form of tomography, where slices.
NMR SPECTROSCOPY.
ELEG 479 Lecture #9 Magnetic Resonance (MR) Imaging
Nuclear Magnetic Resonance Spectrometry Chap 19. Absorption in CW Experiments Energy of precessing particle E = -μ z B o = -μ B o cos θ When an RF photon.
Nuclear Magnetic Resonance (NMR) Magnetic Resonance Imaging (MRI)
Basics of Magnetic Resonance Imaging
MRI Physics I: Spins, Excitation, Relaxation
Magnetic Resonance Imaging
Magnetic Resonance Imaging
Magnetic Resonance Imaging Astounding. The MRI magnet and patient table The operator’s console.
Magnetic Resonance Imaging 4
Physics of Magnetic Resonance Chapter 12
대구가톨릭대학병원 영상의학과 이 영 환 M M R R Basic Physics. MR Signal T1-, T2-weighted TR, TE MR Signal T1-, T2-weighted TR, TE.
Principles of Magnetic Resonance
MRI Physics: Just the Basics
Determination of Spin-Lattice Relaxation Time using 13C NMR
Prof. Dr. Moustafa. M. Mohamed Dr. Eng. Safa Ahmed El-Askary
BE 581 Intro to MRI.
Magnetic Material Engineering. Chapter 6: Applications in Medical and Biology Magnetic Material Engineering.
Chapter 4 Mark D. Herbst, M.D., Ph.D.. Magnetization MRI depends on magnetization of tissues Temporary magnetization occurs when the hydrogen nuclei are.
Basic Physical Principles of MRI
Basic Concept of MRI Chun Yuan. Magnetic Moment Magnetic dipole and magnetic moment Nuclei with an odd number of protons or neutrons have a net magnetic.
September, 2003BME 1450 Introduction to NMR 1 Nuclear Magnetic Resonance (NMR) is a phenomenon discovered about 60 years ago.
fMRI Methods Lecture2 – MRI Physics
MR Signal Generation FMRI Undergraduate Course (PSY 181F)
NMR in Medicine and Biology
NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY Basics of …….. NMR phenomenonNMR phenomenon Chemical shiftChemical shift Spin-spin splittingSpin-spin splitting.
Variable Rate Selective Excitation Radio Frequency Pulse in Magnetic Resonance Imaging Stephen Stoyan.
Nuclear Magnetic Resonance I Magnetization properties Generation and detection of signals.
Magnetic Resonance Imaging – Basic Principles –
MRI Physics Dr Mohamed El Safwany, MD.. MRI Magnetic Resonance Imaging Magnetic Resonance Imaging.
MRI. Vector Review x y z Vector Review (2) The Dot Product The Cross Product (a scalar) (a vector) (a scalar)
BME1450 Intro to MRI February 2002 The Basics The Details – Physics The Details – Imaging.
Protons (hydrogen nuclei act like little magnets) MRI Collective Magnetic Moment of Protons (M 0 ) Each pixel is a glass of protons B 0 = 3T (not to scale)
V.G.Wimalasena Principal School of Radiography
Fund BioImag : Relaxation of nuclear magnetization 1.How is the MR signal detected ? 2.What is the quantum-mechanical equivalent of the rotating.
RT 4912 Review (C) Rex T. Christensen MHA RT (R) (MR) (CT) (ARRT) CIIP.
Magnetic Resonance Imaging Glenn Pierce, King’s College London, Department of Physics Introduction Edward Purcell and Felix Bloch were both awarded the.
Spinning Nucleus Produces Magnetic Moment
1 Nuclear Magnetic Resonance ANIMATED ILLUSTRATIONS MS Powerpoint Presentation Files Uses Animation Schemes as available in MS XP or MS 2003 versions A.
Charged particle. Moving charge = current Associated magnetic field - B.
NUCLEAR MAGNETIC RESONANCE
بسم الله الرحمن الرحيم Dr. Maged Ali Hegazy Assistant Professor Alazhar University.
Chap.12 (3) Medical imaging
MRI Physics in a Nutshell Christian Schwarzbauer
Where Mt is the magnetization at time = t, the time after the 90o pulse, Mmax is the maximum magnetization at full recovery. At a time = one T1, the signal.
Physics of Magnetic Resonance Imaging
Topics The concept of spin Precession of magnetic spin Relaxation
Basic MRI I Chapter 3 Notes.
10.4 Continuous Wave NMR Instrumentation coherent detection
Introduction to MR Robert Frost
(4)ELECTRONIC SUPPORT SYSTEM
T2 Relaxation Time T2 relaxation time is defined as the time needed to dephase up to 37% of the original value. T2 relaxation refers to the progressive.
Introduction to MR Robert Frost
Presentation transcript:

PHYSICS OF MAGNETIC RESONANCE RAJAN PANDYA & MOHIT GANGULY

Magnetic Resonance Imaging Provide high resolution anatomic structure (as with X-ray CT) Provide high contrast between different soft tissues (X-ray CT cannot) No exposure to radiation and hence safe More complicated instrumentation Takes longer to acquire a scan than CT, more susceptible to patient motion

Comparison X RAY PROJECTION MRI

Principle of MRI The hydrogen (1^H) atom inside body possess “spin”. In the absence of external magnetic field, the spin directions of all atoms are random and cancel each other. When placed in an external magnetic field, the spins align with the external field. By applying an rotating magnetic field in the direction orthogonal to the static field, the spins can be pulled away from the z-axis with an angle \alpha The bulk magnetization vector rotates around z at the Larmor frequency (presses). The precession relaxes gradually, with the xy-component reduces in time, z-component increases The xy component of the magnetization vector produces a voltage signal, which is the NMR signal we measure.

Spin Spin is a fundamental property of nature like electrical charge or mass. Spin comes in multiples of 1/2 and can be + or -. Protons, electrons, and neutrons possess spin. Individual unpaired electrons, protons, and neutrons each possesses a spin of ½ or - ½. Two or more particles with spins having opposite signs can pair up to eliminate the observable manifestations of spin. In nuclear magnetic resonance, it is unpaired nuclear spins that are of importance.

Nuclear Spin A nucleus consists of protons and neutrons. When the total number of protons and neutrons (=mass number A) is odd or the total number of protons is odd, a nucleus has an angular momentum (\phi) and hence spin – Ex. Hydrogen (1^H) (1 proton), 13^C The spin of a nucleus generates a magnetic filed, which has a magnetic moment (\mu) The spin causes the nucleus behave like a tiny magnet with a north and south pole

Bulk Magnetization at Equilibrium

Evolution of magnetization when a Time varying magnetic field is applied M(t) experiences a torque when an external magnetic field B(t) is applied Using the right hand rule, M will rotate around z if M is not aligned with z

Longitudinal and Transverse components

NMR Signal The rapidly rotating transverse magnetization (M_xy) creates a radio frequency excitation within the sample. If we put a coil of wire outside the sample, the RF excitation will induce a voltage signal. In MRI, we measure this voltage signal. Voltage produced is (Faraday’s Law of Induction)

RF excitation Circularly polarized (rotating) magnetic field B1(t) in the x-y plane with the same Larmor frequency forces the magnetization vector to tilt down to the x-y plan B1(t) has two orthogonal components, in x and y directions respectively, and is produced by using quadrature RF coil – Simplest envelop B1,e is a rectangular pulse

Motion of M(t) is spiral

Circularly Polarized Magnetic Field

Tip Angle If M is parallel to z-axis before the RF excitation pulse, the tip angle after the excitation is If B1^e(t) is rectangular

Tip angle Pulse that leads to alpha=pi/2 is called “pi over 2 pulse”, which elicits the largest transverse component Mxy, and hence largest NMR signal. Pulse that leads to alpha=pi is called “pi pulse” or inverse pulse, which is used to induce spin echo (later). The excitation pulse (envelop of B1(t)) is also called “an alpha pulse”.

Relaxation Magnetization cannot be presses forever . There are two relaxation processes working together cause the received signal vanish Transverse relaxation Longitudinal relaxation This properties defers as per tissue so MR images get rise to tissue contrast

Longitudinal Relaxation The magnetization vectors tend to return to equilibrium state (parallel to B0)

Transverse Relaxation The strength of the magnetic field in the immediate environment of a 1H nucleus is not homogeneous due to presence of other nucleus (and their interactions) Hence the Larmor frequencies of nearby nuclides are slightly different (some spins faster, some slower) – Spin-spin interactions This causes dephasing of the xy components of the magnetization vector

Transverse Relaxation

Transverse Relaxation T2 is called transverse relaxation time, which is the time for Mxy to decrease by 1/e. Also called spin-spin relaxation time T2 is much smaller than T1 – For tissue in body, T2: 25-250ms, T1: 250-2500 ms

Free Induction Decay The voltage signal (NMR signal) produced by decaying Mxy also decays. This is called free induction decay (FID), and is the signal we measure in MRI

T2 Star Decay Received signal actually decays faster than T2 (having a shorter relaxation time T2^*) Caused by fixed spatial variation of the static field B0 due to imperfection of the magnet – Accelerates the dephasing of magnetization vectors – T2 is caused by spatial variation of the static field due to interactions of nearby spins The initial decay rate is governed by T2^* , but the later decay by T2.

T2 Star Decay

Bloch Equations

Formation of Spin Echo By applying a 180 degree pulse, the dephased spins can recover their coherence, and form an echo signal

RF Pulse Sequence and Corresponding NMR Signal

MR Contrast Different tissues vary in T1, T2 and PD (proton density) The pulse sequence parameters can be designed so that the captured signal magnitude is mainly influenced by one of these parameters Pulse sequence parameters – Tip angle \alpha – Echo time TE – Pulse repetition time TR

Typical Brain Tissue Parameters

Spin density weighting Signal at equilibrium proportional to PD Long TR: – Minimizes effects of different degrees of saturation (T1 contrast) – Maximizes signal (all return to equilibrium) Short TE: – Minimizes T2 contrast – Maximizes signal

Spin density weighting

T2 weighting Long TR: – Minimizes influence of different T1 Long TE: – Maximizes T2 contrast – Relatively poor SNR

T2 weighting

T1-weighting Short TR: – Maximizes T1 contrast due to different degrees of saturation – If TR too long, tissues with different T1 all return equilibrium already Short TE: – Minimizes T2 influence, Maximizes signal

T1-weighting