Performance of a Hybrid RF/DC Quadrupole-Linear Ion Trap Mass Spectrometer James W. Hager MDS SCIEX ASMS, 2002
Outline Linear ion trap mass spectrometer concept. Mass selective axial ion ejection. Marriage with triple quad ms. Trapping and extraction efficiencies Space charge effects Performance characteristics. Future directions.
Triple Quad. MS/MS --Tandem in Space Only 1 ion pair is stable at any one time. Poor scanning efficiency. Very efficient for MRM. Very selective scans available.
3D Ion Trap MS/MS --Tandem in Time Full mass spectrum for each pulse of ions. High scanning efficiency. Scanning instrument only.
Linear Ion Traps “Ion bottles” for optical spectroscopy. Minimize fringing fields to maximize performance. Ion accumulation for enhanced ms sensitivity. Mass analyzers: RCM, 2002, 16, 512-526. JASMS, 2002, 13, 659-669.
Linear Ion Trap Mass Spectrometer Ions are trapped by barriers: Radially: RF potential Axially: DC barriers Radial RF determines the ion frequencies of motion. Radial and axial ejection demonstrated. Drive RF & auxiliary AC
RF Fringing Fields Axial Ion Ejection: Exit fringing fields couple radial/axial degrees of freedom of the trapped ions. Excite ions radially to eject them axially.
Time Averaged Axial Electric Field Dehmelt approx.: radial amplitude fringing field F. Londry, ASMS, 2002 .
Hybrid Instrument with Q3 Linear Ion Trap MS N2 CAD Gas linear ion trap 3x10-5 Torr Aux AC IQ2 IQ3 Exit Skimmer Orifice LINAC IQ1 Q0 Q1 Q2 Q3
Q3 Trapping Efficiency vs. Chamber Pressure
Extraction Efficiency vs. Ejection q-Value (4x10-5 Torr)
Trapping & Extraction Efficiencies 5-inch long rod array: Trapping efficiency: ~45% 4x10-5 Torr background pressure Extraction efficiency: 16-20% q-value dependent Overall efficiency: ~8%
Hybrid Approach to Product Ion MS/MS RF/DC Q1 Q2 Q3 linear trap eV frags. Advantages: No time required to isolate the precursor ion. No loss for isolation of fragile precursor ions. The ion trap is filled with only precursor and fragment ions. Triple quad. fragmentation patterns. No inherent low mass cut-off.
Product Ion Scanning Precursor ions selection in Q1. N2 CAD Gas Aux AC Exit lens Ion accumulation Q0 Q1 Q2 Q3 Precursor ion selection Fragmentation linear ion trap 3x10-5 Torr Precursor ions selection in Q1. Fragmentation in Q2. Trap products in Q3. Mass scan. Concurrent trapping in Q0.
Sensitivity Enhancements Reserpine – Product Ion QqQ Q1 Open Resolution Enhanced Product Ion Scan > 350x
Sensitivity Enhancements—Reserpine Q1 Open Resolution QqQ Enhanced Product Ion Scan
Enhanced Prod Ion Scan – 75eV No Low Mass Cut-off Taurocholic Acid– Product Ion LCQ Deca – 45% CE 3 amu isolation width Enhanced Prod Ion Scan – 75eV
Resolution and Scan Speed: Reserpine 4000 amu/sec ~0.55 amu wide 1000 amu/sec ~0.35 amu wide 250 amu/sec ~0.15 amu wide
Mass Shift vs. Ejection q-Value 3.5x10-5 Torr 103X Change of Ion Current
Mass Shift vs. Input Ion Intensity q=0.86, m/z=609
Space Charge Effects Q3 Space Charge Effects (3x10-5 Torr): D~0.02 amu 104x > ion current D~0.02 amu
RF/DC Quadrupole-Linear Ion Trap Mass Spectrometer Ion trap and triple quad capabilities on a single platform. High sensitivity ion trap scans: Single MS survey scans Product ion scans MS3 in Q3 Triple quad. Functionality: MRM Precursor ion scans Neutral loss scans
Future Directions Tandem-in-space ion trap(s) approach allows for simultaneous ion processing while the primary mass analyzer is scanning. Increased sample utililization eff. and thus higher duty cycles. Very high resolution.
Acknowledgments Yves LeBlanc Frank Londry Bill Stott Bruce Collings John Vandermey