Performance of a Hybrid RF/DC Quadrupole-Linear Ion Trap Mass Spectrometer James W. Hager MDS SCIEX ASMS, 2002.

Slides:



Advertisements
Similar presentations
What is the ISOLDE cooler RFQ CB - ISCOOL H. Frånberg.
Advertisements

Background and Theory of Electrospray and Tandem Mass Spectrometry
Orbitrap Mass Analyser - Overview and Applications in Proteomics
FC-MS from Teledyne Isco CombiFlash ® a Name You Can Rely On.
1 The world leader in serving science Proprietary & Confidential WiSIM DIA Method Set-up Orbitrap Fusion MS MS Instrument Control Software v.1.1 SP1.
ISOLDE WS 2007 Results with ISCOOL a new Radio Frequency Quadrupole Cooler and Buncher at ISOLDE P. Delahaye, H. Frånberg, AB-OP-PSB, ISCOOL, COLLAPS,
Biochem 523b: Advanced Physical Methods: Mass Spectrometry, X-ray Crystallography and NMR A. Mass Spectrometry Lecture 2 Mass analyzers Time of flight.
C d a b  =1/   Normalized Radial Position t/mt/m b The RF drive voltage induces an ion to move in a cyclic trajectory.
Method Building Essentials
MASS ANALYZERS AND IONIZATION METHODS
Sanja Risticevic Chem 323 Poster Presentation Quadrupole Ion Trap Mass Spectrometry.
Inductively Coupled Plasma Mass Spectrometry or ICP-MS
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Mass Spectrometry Inlet system Ion Source Mass Analyzer Ion Detector.
Mass Spectroscopy Mass Spectrometry ä Most useful tool for molecular structure determination if you can get it into gas phase ä Molecular weight of.
Simulation of Radio Frequency Quadrupole Mass Analyzer Stephen Jones EPS 109 Final Project 11/27/12.
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform.
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform.
Instrumental Chemistry Chapter 11 Atomic Mass Spectrometry.
Atomic Mass Spectrometry
Molecular Mass Spectroscopy Chem Introduction In Mass Spectroscopy (MS), atomic and molecular weights are generally expressed in terms of atomic.
Atomic Mass Spectrometry
Mass Spectrometry.
LC-MS Lecture 7.
Lecture 3 INFRARED SPECTROMETRY
Photoelectron Spectroscopy Lecture 7 – instrumental details –Photon sources –Experimental resolution and sensitivity –Electron kinetic energy and resolution.
De Novo Sequencing of MS Spectra
 The new instrument consists of a three-dimensional RF quadrupole ion trap (QIT), a linear RF quadrupole collision cell (q), and an orthogonal, reflectron.
Agilent 7500 ICP-MS Date: 28JUL2005.
Atomic Emission Spectroscopy
Mass Spectrometry Mass spectrometry (MS) is not true “spectroscopy” because it does not involve the absorption of electromagnetic radiation to form an.
Secondary ion mass spectrometry (SIMS)
Mass Analyzers III Quadrupoles Chem 5181 – Fall 2007 Joel Kimmel.
Collision-based methods: Electron-based methods: Primary methods for dissociating peptides Collision-based methods: Ion trap collisional activation.
Mass spectrometry Ions are analyzed on the basis of their m/z Chlorine has 2 isotopes, 35 Cl and 37 Cl, in the approximate ratio of 3 :1. Electrons are.
By P. Vainshtein and M. Shapiro Technion - Israel Institute of Technology Faculty of Mechanical Engineering An acoustic channel for aerosol particle focusing.
Hiden Analytical Excellence in plasma diagnostics An Introduction to the Hiden EQP Excellence in Plasma Diagnostics An Introduction to the Hiden EQP
Mass Analyzers IV Quadrupole Ion Traps Chem 5181 – Fall 2007 J. Kimmel.
Untargeted Metabolomics: Tandem LC-MSMS. Column and Flow Rate Selection Insert Barnes table for flow rates and sensitivity gain. Reverse Phase and Normal.
The Apparatus…. Ionic target studies…. Neutral target studies…. Queens University Belfast University College London.
MS Intro. MS requires gas-phase ions, why? MS uses magnetic and electric fields to control the path of a compound based on mass to charge ratio (m/z)
Temple University MASS SPECTROMETRY INTRODUCTION Ilyana Mushaeva and Amber Moscato Department of Electrical and Computer Engineering Temple University.
For all CHEM5161 students: The first day of class for CHEM5161 (Analytical Spectroscopy) will be on TUE Sept 4 (following Labor Day). There will be no.
Mass Analyzers: Quadrupole ion trap?  
A mass-purification method for REX beams
Mass Analyzers : Time-of-flight Mass Spectrometry CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Joel Kimmel Fall 2007.
Studies of Transient Neutral Molecules by Dissociative Photodetachment of Cooled Molecular Anions Christopher Johnson Continetti Research Lab University.
Threshold ionization mass spectroscopy of radicals in RF silane discharge Progress report 11/19/2003 Contributors: Alan Gallagher Peter Horvath Karoly.
Secondary Ion Mass Spectrometry A look at SIMS and Surface Analysis.
FLAIR meeting, GSI March Positron Ring for Antihydrogen Production A.Sidorin for LEPTA collaboration JINR, Dubna.
Mass Spectrometry Quantitative Mass Spectrometry
The world leader in serving science For Research Use Only. Not for use in diagnostic procedures Quantitative Analysis of 4 Immunosuppressant Drugs in Whole.
Objective  To develop methods for analysis of compounds in organic aerosol particles Why is this important?  Environmental impact  Alternative fuels.
Introduction to Liquid Phase Mass Spectrometry
Methodology of targeted mass spectrometry
Metabolomics Part 2 Mass Spectrometry
Components of Mass Spectrometer
Chem. 133 – 4/18 Lecture.
MoP-067 Michael Sudakov 1, R.Giles1 and Dimitris Papanastasiou2
Acknowledgements Slides and animations were made by Dr. Jon Karty Mass Spectrometry Facility Indiana University, Bloomington.
Michael Sudakov and Sumio Kumashiro
Time of Flight Analyzers
LC-MS Lectures Dr Mike Kinsella.
Mass Spectrometry Obaid M. Shaikh.
Metabolomics Part 2 Mass Spectrometry
Mass Spectroscopy. Mass Spectroscopy Mass Spectrometry Most useful tool for molecular structure determination if you can get it into gas phase Molecular.
Instrumental Chemistry
Molecular Mass Spectroscopy
Hybrid MS instruments including TOF analyzers
Improvement of a dc-to-pulse conversion efficiency of FRAC
Presentation transcript:

Performance of a Hybrid RF/DC Quadrupole-Linear Ion Trap Mass Spectrometer James W. Hager MDS SCIEX ASMS, 2002

Outline Linear ion trap mass spectrometer concept. Mass selective axial ion ejection. Marriage with triple quad ms. Trapping and extraction efficiencies Space charge effects Performance characteristics. Future directions.

Triple Quad. MS/MS --Tandem in Space Only 1 ion pair is stable at any one time. Poor scanning efficiency. Very efficient for MRM. Very selective scans available.

3D Ion Trap MS/MS --Tandem in Time Full mass spectrum for each pulse of ions. High scanning efficiency. Scanning instrument only.

Linear Ion Traps “Ion bottles” for optical spectroscopy. Minimize fringing fields to maximize performance. Ion accumulation for enhanced ms sensitivity. Mass analyzers: RCM, 2002, 16, 512-526. JASMS, 2002, 13, 659-669.

Linear Ion Trap Mass Spectrometer Ions are trapped by barriers: Radially: RF potential Axially: DC barriers Radial RF determines the ion frequencies of motion. Radial and axial ejection demonstrated. Drive RF & auxiliary AC

RF Fringing Fields Axial Ion Ejection: Exit fringing fields couple radial/axial degrees of freedom of the trapped ions. Excite ions radially to eject them axially.

Time Averaged Axial Electric Field Dehmelt approx.: radial amplitude fringing field F. Londry, ASMS, 2002 .

Hybrid Instrument with Q3 Linear Ion Trap MS N2 CAD Gas linear ion trap 3x10-5 Torr Aux AC IQ2 IQ3 Exit Skimmer Orifice LINAC IQ1 Q0 Q1 Q2 Q3

Q3 Trapping Efficiency vs. Chamber Pressure

Extraction Efficiency vs. Ejection q-Value (4x10-5 Torr)

Trapping & Extraction Efficiencies 5-inch long rod array: Trapping efficiency: ~45% 4x10-5 Torr background pressure Extraction efficiency: 16-20% q-value dependent Overall efficiency: ~8%

Hybrid Approach to Product Ion MS/MS RF/DC Q1 Q2 Q3 linear trap eV frags. Advantages: No time required to isolate the precursor ion. No loss for isolation of fragile precursor ions. The ion trap is filled with only precursor and fragment ions. Triple quad. fragmentation patterns. No inherent low mass cut-off.

Product Ion Scanning Precursor ions selection in Q1. N2 CAD Gas Aux AC Exit lens Ion accumulation Q0 Q1 Q2 Q3 Precursor ion selection Fragmentation linear ion trap 3x10-5 Torr Precursor ions selection in Q1. Fragmentation in Q2. Trap products in Q3. Mass scan. Concurrent trapping in Q0.

Sensitivity Enhancements Reserpine – Product Ion QqQ Q1 Open Resolution Enhanced Product Ion Scan > 350x

Sensitivity Enhancements—Reserpine Q1 Open Resolution QqQ Enhanced Product Ion Scan

Enhanced Prod Ion Scan – 75eV No Low Mass Cut-off Taurocholic Acid– Product Ion LCQ Deca – 45% CE 3 amu isolation width Enhanced Prod Ion Scan – 75eV

Resolution and Scan Speed: Reserpine 4000 amu/sec ~0.55 amu wide 1000 amu/sec ~0.35 amu wide 250 amu/sec ~0.15 amu wide

Mass Shift vs. Ejection q-Value 3.5x10-5 Torr 103X Change of Ion Current

Mass Shift vs. Input Ion Intensity q=0.86, m/z=609

Space Charge Effects Q3 Space Charge Effects (3x10-5 Torr): D~0.02 amu 104x > ion current D~0.02 amu

RF/DC Quadrupole-Linear Ion Trap Mass Spectrometer Ion trap and triple quad capabilities on a single platform. High sensitivity ion trap scans: Single MS survey scans Product ion scans MS3 in Q3 Triple quad. Functionality: MRM Precursor ion scans Neutral loss scans

Future Directions Tandem-in-space ion trap(s) approach allows for simultaneous ion processing while the primary mass analyzer is scanning. Increased sample utililization eff. and thus higher duty cycles. Very high resolution.

Acknowledgments Yves LeBlanc Frank Londry Bill Stott Bruce Collings John Vandermey