Aqua MODIS Cold FPA Performance and Operation MODIS Characterization Support Team (MCST) April 25, 2012.

Slides:



Advertisements
Similar presentations
22 March 2011: GSICS GRWG & GDWG Meeting Daejeon, Korea Tim Hewison SEVIRI-IASI Inter-calibration Uncertainty Evaluation.
Advertisements

Evaluating Calibration of MODIS Thermal Emissive Bands Using Infrared Atmospheric Sounding Interferometer Measurements Yonghong Li a, Aisheng Wu a, Xiaoxiong.
Summary of Terra and Aqua MODIS Long-term Performance Jack Xiong 1, Brian Wenny 2, Sri Madhaven 2, Amit Angal 3, William Barnes 4, and Vincent Salomonson.
SNPP VIIRS SDR RSR LUT Update: ADR 4971, CCR U. Wisconsin:Chris Moeller NASA VCST: Ning Lei, Shihyan Lee, Zhipeng Wang, Sam Anderson, Chengbo Sun,
Science Impact of MODIS Calibration Degradation and C6+ Improvements A. Lyapustin, Y. Wang, S. Korkin, G. Meister, B. Franz (+OBPG), X. Xiong (+MCST),
VIIRS Reflective Solar On-orbit Calibration and Performance Jack Xiong and Jim Butler Code 618.0, NASA/GSFC, Greenbelt, MD CLARREO SDT Meeting, NASA.
SABER Instrument Performance and Measurement Requirements.
RHESSI/GOES Observations of the Non-flaring Sun from 2002 to J. McTiernan SSL/UCB.
Chandra X-Ray Observatory CXC Paul Plucinsky EPIC Cal Nov ACIS Calibration: Planned Updates & Future Issues 2.ACIS Operations: Controlling the.
CIRSGSFC April 2004 dej-1 CIRS Instrument Overview CIRS Team Meeting Observatoire de Paris - Meudon April 26 – CIRS Instrument Group presented.
1 Calibration Adjustments for the MODIS Aqua 2015 Ocean Color Reprocessing Gerhard Meister, NASA Code 616 OBPG (Ocean Biology Processing Group) 5/18/2015.
6-1 EE/Ge 157b Week 6 EE/Ae 157 a Passive Microwave Sensing.
NIRSpec Operations Concept Michael Regan(STScI), Jeff Valenti (STScI) Wolfram Freduling(ECF), Harald Kuntschner(ECF), Robert Fosbury (ECF)
COB Basic Cooling Rate LWFPA MWFPA Time Degrees K 6Hrs to achieve operating temperature.
FTS Test Definition Summary Kaley Walker Test Readiness Review January 29, 2003.
ISPRS Gulfport Calibration Workshop December 3, 2003
MODIS Collection 6 MCST Proposed Changes to L1B. Page 2 Introduction MODIS Collection History –Collection 5 – Feb present –Collection 4 – Jan.
START NICMOS COOLING SYSTEM ACTIVITY SUMMARY. 1/25/01Ken Pulkkinen2 START NCS ACTIVITY Applicable SMOV Requirements: J NICMOS Cooling System.
Implementation and Processing outline Processing Framework of VIIRS instrument monitoring System Processing Framework of VIIRS EV data Monitoring SD/SDSM.
1 _________________________________________________________________________________________________________________________________________________________________.
TIPS - Oct 13, 2005 M. Sirianni Temperature change for ACS CCDs: initial study on scientific performance M. Sirianni, T. Wheeler, C.Cox, M. Mutchler, A.
VIIRS Product Evaluation at the Ocean PEATE Frederick S. Patt Gene C. Feldman IGARSS 2010 July 27, 2010.
Analysis of Nonlinearity Correction for CrIS SDR April 25, 2012 Chunming Wang NGAS Comparisons Between V32 and V33 Engineering Packets.
Recent Solar Irradiance Data From SBUV/2 and OMI Matthew DeLand and Sergey Marchenko Science Systems and Applications, Inc. (SSAI) SOLID WP-2 Workshop.
NPP ATMS Instrument Performance SDR Product Review, 23 Oct Kent Anderson Chief Engineer, Civil Space Programs NG Electronic Systems Azusa, CA.
11-Jun-04 1 Joseph Hora & the IRAC instrument team Harvard-Smithsonian Center for Astrophysics The Infrared Array Camera (IRAC) on the Spitzer Space Telescope.
IOP algorithm OOXX Jeremy Werdell & Bryan Franz NASA Ocean Biology Processing Group 25 Sep 2010, Anchorage, AK
IBB Calibration Update J. Rufus GIST21 Meeting Boulder, CO. 31 st March – 2 nd April 2004 © Imperial College London.
MODIS Preprocessing (before L1B) Changes over the Last Year (and looking forward) Chris Moeller and others CIMSS; Univ. Wisconsin July 13, 2004 Thanks.
Validation of OMPS-LP Radiances P. K. Bhartia, Leslie Moy, Zhong Chen, Steve Taylor NASA Goddard Space Flight Center Greenbelt, Maryland, USA.
Polarization analysis in MODIS Gerhard Meister, Ewa Kwiatkowska, Bryan Franz, Chuck McClain Ocean Biology Processing Group 18 June 2008 Polarization Technology.
MODIS-Terra cross-calibration for ocean color bands Ewa Kwiatkowska Bryan Franz, Gerhard Meister, Gene Eplee OBPG 30 January 2008.
Ozone PEATE 2/20/20161 OMPS LP Release 2 - Status Matt DeLand (for the PEATE team) SSAI 5 December 2013.
CoRP Cal/Val Symposium July 13, 2005
Level 1.5 Processing Updates Martin Bates, RAL GIST th April, 2005.
Collect 5 Calibration Issues Chris Moeller and others Univ. Wisconsin March 22, 2005 Presented at MCST Calibration breakout meeting, March 22, 2005.
NASA OBPG Update OCRT Meeting 23 April 2012 Bryan Franz and the NASA Ocean Biology Processing Group.
AIRS – MODIS TEB Global Comparisons Chris Moeller Dave Tobin Univ. Wisconsin May 13, 2008 MODIS Calibration Mtg.
Collect 5 Calibration Issues Chris Moeller and others Univ. Wisconsin March 22, 2005.
# x pixels Geometry # Detector elements Detector Element Sizes Array Size Detector Element Sizes # Detector elements Pictorial diagram showing detector.
Denis Tremblay 1, Yong Han 2, Yong Chen 3, Likun Wang 3, Xin Jin 2, Xiaozhen Xiong 2, Lihang Zhou 2 1 Science Data Processing Inc., 2 NOAA/NESDIS/STAR,
Page 1 NPOESS Preparatory Project (NPP) VIIRS Calibration Maneuvers May 15, 2008 Ocean PEATE Team.
Bias analysis and correction for MetOp/AVHRR IR channel using AVHRR-IASI inter-comparison Tiejun Chang and Xiangqian Wu GSICS Joint Research and data Working.
MODIS-Terra cross-calibration for ocean color bands Ewa Kwiatkowska Bryan Franz, Gerhard Meister Ocean Biology Processing Group 13 May 2008 MODIS Science.
MsWG (Apr 22, 2015) Recent L1B LUT delivery since April 8, 2015
GSICS DCC calibration update
Pre-launch Characteristics and Calibration
NOAA VIIRS Team GIRO Implementation Updates
Lessons from MODIS Calibration and Characterization
GLAST Large Area Telescope:
Extending DCC to other bands and DCC ray-matching
MODIS and VIIRS Reflective Solar Bands Calibration Methodologies and
MODIS Instrument Operations Team (IOT): Bryan Breen & Tony Salerno
An Overview of MODIS Reflective Solar Bands Calibration and Performance Jack Xiong NASA / GSFC GRWG Web Meeting on Reference Instruments and Their Traceability.
Aqua MODIS Reflective Solar Bands (RSB)
On the use of Ray-Matching to transfer calibration
Calibration and Performance MODIS Characterization Support Team (MCST)
MODIS Lunar Calibration Data Preparation and Results for GIRO Testing
MODIS Characterization and Support Team Presented By Truman Wilson
Using the Moon for Sensor Calibration Inter- comparisons
Status of MODIS and VIIRS Reflective Solar Calibration
NanoBPM Status and Multibunch Mark Slater, Cambridge University
S-NPP Visible Infrared Imaging Radiometer Suite (VIIRS) Lunar Calibration using GSICS Implementation of the ROLO model (GIRO) for Reflective Solar Bands.
MODIS L1B Data Product Uncertainty Index Jack Xiong (Xiaoxiong
Launch and On-orbit Checkout
Early calibration results of FY-4A/GIIRS during in-orbit testing
Na Xu, Xiuqing Hu, Lin Chen, Min Min
Spectral Correction for Inter-comparion between VIIRS and MODIS TEB
goes-16/17 abi lunar calibration
G16 vs. G17 IR Inter-comparison: Some Experiences and Lessons from validation toward GEO-GEO Inter-calibration Fangfang Yu, Xiangqian Wu, Hyelim Yoo and.
Presentation transcript:

Aqua MODIS Cold FPA Performance and Operation MODIS Characterization Support Team (MCST) April 25, 2012

Outline Prior Meeting Summary TEB Introduction Updates on Aqua MODIS performance −Key instrument telemetry −TEB performance Observed impacts Mitigation Strategies 2

Prior Meeting Summary MCST held previous workshop in May 7, 2010 −Meeting was held to review the Aqua MODIS CFPA on-orbit performance, provide its possible implications on the L1B product, and give scenarios to stabilize the CFPA operating point. Aqua MODIS slow loss of cooler margin causing an increase and periodic variations (orbital & seasonal) of CFPA temperatures (up to 0.2K) Impacts observed – Orbital variation in TEB detector gain Captured predominantly by the scan-by-scan calibration Linear relationship between CFPA Temp and TEB detector gain PC bands more sensitive to CFPA temperatures – Increased fitting residuals from BB WUCD during a0/a2 derivation 3

Current Status MCST has continued to monitor telemetry and TEB performance −CFPA temperature continues to increase −CFPA orbital/seasonal oscillation grown larger – max/min difference of ~0.5 K −Radiative cooler margin almost entirely lost. Majority of impacts on L1B products occur during BB Warmup/Cooldown activities 4

TEB Calibration Requirements CW: center wavelength in  m; Ttyp: typical scene temperature in K; NEdT: noise equivalent temperature difference in K; UC: uncertainty in percentage (radiance) and in K 5

TEB Radiometric Calibration EV Radiance: RVS: Response Versus Scan-angle  : Emissivity L: Spectral band averaged radiance dn: Digital count with background corrected RSR: Relative Spectral Response Calibration Coefficients: WUCD T BB : 270 to 315K 6

MODIS Focal Plane Assemblies (FPA) Warm FPA: VIS and NIR; Cold FPA: SMIR and LWIR 7

BB Temp Trends Terra Aqua 8 Weekly average (Cal events excluded) 20 mK 10 mK

Instrument & Warm FPA Temperatures VIS NIR Terra Aqua ~3.5K increase over 11+ years ~2K increase over 9+ years 9 Weekly average (Cal events excluded)

Cold FPA Temperatures Terra Aqua Last CFPA Workshop SMIR LWIR extremely stable !!! SMIR LWIR 10 Weekly average (Cal events excluded) 0.2 K

TEB Gain Response 11 1 randomly selected granule per week (good detectors) PC Bands showing increased variation in b1

Aqua CFPA Heater Voltage Granule average

Aqua CFPA Heater Voltage Granule average

Heater Voltage & Duty Cycle Indicates complete loss in control 14 1 orbit average

Aqua CFPA Oscillations Short Term (1day) trend T SMIR T LWIR 15 Granule average

Aqua CFPA Oscillations Short Term (1day) trend T SMIR T LWIR 16 Granule average

Aqua CFPA Oscillations Long-term trend 17 Max/Min of 4 orbits, every 8 hours 0.5 K 0.6 K Solid vertical lines indicate calendar years, dashed vertical line time of last CFPA workshop.

Observed Impacts Gain variation with CFPA temperature – captured by scan-by-scan calibration – PC bands show largest effect – Linear relationship between gain and CFPA Temp BB Warmup-Cooldown Activities – Larger residuals for a2 derivation (a0 = 0 for PC bands) – Bands 33, 35 & 36 T BB Saturation FPA temperature dependent default b1 algorithm implemented for C6 Increased instances of EV saturation Band 27 – Temperature dependent filter bandpass shift – Not included in discussion here 18

Gain variation with CFPA Temperature 19 T SMIR T LWIR Granule average, top plot: detector 2 for each band

20 3 orbits, granule average, detector 2 for each band, normalized by b1 at CFPA = 82.8 K b1 vs CFPA Temp

21 3 orbits, granule average, normalized by b1 at CFPA = K

BB Warmup/Cooldown BB Warmup/Cooldown activities – Performed quarterly – Temperature range: K – Used for a0/a2 derivation/trending – Bands 33, 35 & 36 saturate for TBB > ~300 K Default b1 used in L1B Impact of CFPA Temperature – a0/a2 fitting residuals – Default b1 derivation – EV saturation 22

BB Warmup/Cooldown / /035 Fluctuation in CFPA Temp during WUCD influenced detector response to T BB Only bands 31 & 32 relied on the on-orbit derived a2 coefficients – all other bands from pre-launch

Aqua fitting residuals 2006/ / Residual T LWIR (K)

WUCD a2 trending 25 On-orbit derived coefficients show increased variability (flat line indicates LUT value derived as yearly average)

B33, 35 & 36 Saturation During WUCD activities bands 33, 35 & 36 saturate for T BB > ~300 K (known issue at launch) A constant ‘default b1’ used for scan-by-scan calibration during these periods. 26 T sat for Band 36 is 300K Band 36 b 1 for D1, M1; Constant b1 L1B methodology

Default b1 Algorithm Modified for C6 Revised approach to compensate for CFPA temperature variations Algorithm details 1. For each WUCD cycle, select three continuous orbits data (~ 5 hrs) from a previous day. Calculate b1 & T LWIR 2. Calculate the rate (c1 in percent) of relative b1 change with T LWIR by linear fitting for each detector and mirror side. 3. Calculate default b1 (b1 Tsat ) and T LWIR,Tsat using scans with T BB close to T SAT 4. For each scan, when T BB > Tsat[Band]: LUT Inputs and Order: b1 Tsat = [Band, Detector, Mirror Side] c1 = [Band, Detector, Mirror Side] T baseline = [Band], Currently set at 83.0 T LWIR,Tsat = [Band, Mirror Side] T LWIR is the focal plane temperature for each scan and is read from TA_RC_LWIR_CFPAE 27

Default b1 Algorithm B33 B35 B36 28

Aqua TEB: C6 default b1 example 29 Black-Body TemperatureLWIR FPA Temperature Band 36 b 1 for D1, M1; Previous L1B methodology T sat for Band 36 is 300K Band 36 b 1 for D1, M1; Modified default b1 All Data shown are granule averaged

30 C6 default b1 impact assessment K K Impact will vary depending on CFPA temperature, scene temperature and the value of the ‘current’ default b1. Example shown here indicates the range of difference expected for the 2011/149 WUCD activity.

31 During WUCD when the T BB > T sat the DCR algorithm is effectively ‘off’ for bands 33, 35 & 36 DN BB = DN Ideal = 4095 DN for all other sectors (e.g. Space View) will begin to oscillate with the CFPA temperature due to temperature dependent gain change of detectors. EV sector will saturate periodically when DN EV + DN SV > DN(T sat ) during the T BB > T sat scans No fix possible at L1B level and length of periods with saturated EV will continue to increase in future. At most 5-6 hours of data lost every 3 months. EV Saturation

32 EV Saturation: 2006 case (Band 35) DN BB > DN(T sat ) DN SV fluctuates with CFPA temp DCR effectively off

33 EV Saturation: 2011 case (Band 35) DN SV fluctuation amplitude has increased

34 EV Saturation: 2012 case (Band 35) DN SV reaches saturation

31 Saturation Example: Band 36 T sat is conservatively determined to be 301 K Black-Body 302 K311 K315 K302 K LWIR Temp K83.03 K 315 K 313 K K83.02 K83.31 K83.00 K

B36 Saturation Trends The number of saturated pixels during the WUCD has increased All granules during WUCD

Mitigation Strategies Scan-by-scan calibration captures much of the impact of the CFPA variation. – CFPA temperature dependent default b1 algorithm implemented in C6 Options under consideration to address temperature fluctuations the same as previous meeting 1.No change – continue operations in current configuration 2.Reduce frequency of WUCD activities – Optional: Flag or do not use B33, 35 & 36 data when T BB > T sat 3.Change CFPA set point to 85K – Temporary: Set to 85K for limited period to assess impact before final decision – Permanent: Set to 85K for all future operations 4.Outgas – Permanent: Set to 83K 37

IOT Procedures IOT is prepared & ready to implement the agreed upon solution For any option involving an operational change the process going forward is identical – Command Authorization Meeting (CAM) at the FOT, approval from at least Mission Director, Flight Systems Manager, Instruments Engineer – Contact request for commanding – Ground Ops prepared and cleared 38

Option 2 – Change CFPA Set Point to 85K S/C needs to be in “nominal mode” for this activity Ensure that CFPA heater B is in the ON state Send the following commands: SET_RC_CFPA_TEMP TO 1/2/3 MOD_SET_PM_RC_CFPA_TEMP(‘T2’)‏ Monitor telemetry to ensure that the CFPA temperature adjusts accordingly (real time until it stabilizes) Concerns: – Majority of pre-launch LUT tables based on 83 K set point – DCR table for 85 K – Loss of dynamic range for TEB – Increased detector noise Advantages: – Gain stably controlled – EV saturation during WUCD removed 39

MCST has at its disposal the P.L. calibration coefficients for Aqua 85K Option 2 Cont’d... 40

Option 3 - Outgas An outgas without an anomaly requires transitioning MODIS from science mode to standby mode and then to outgas mode – Doors are closed, except SV door is moved to outgas position – This operation/action may require approval from HQ An outgas requires a set of commands and takes 2-3 days to complete (heaters are turned on in specific sequence) – CFPA will be back to ambient temperature during outgas operation Potential impact on SWIR (band 6 in particular) detector operability 41

Discussion Issues Actions Recommendations 42