Theoretical Optical Performance of an Equal Conic Intraocular Lens and Comparison to Spherical and Aspheric IOLs Edwin J. Sarver, PhD.

Slides:



Advertisements
Similar presentations
Chapter 4 Companion site for Light and Video Microscopy Author: Wayne.
Advertisements

1 Comparison of Distance Image Quality with Accommodating and Multifocal Aspheric Presbyopia-Correcting IOLs John F. Doane, M.D., F.A.C.S. Discover Vision.
VisTor The new Toric IOL by Hanita Lenses
IN THE NAME OF GOD.
Consider Refraction at Spherical Surfaces:
Lenses in the Paraxial Limit
Issues to cover: Spherical vs. Aspherical Spherical vs. Aspherical Effect of Pupil Size Effect of Pupil Size Effect of IOL decentration Effect of IOL decentration.
Chapter 31 Images.
Lens and its forms Faculty Aravind School of Optometry.
Optics. Spherical Mirrors Spherical mirror – a section of a sphere of radius R and with a center of curvature C R C Mirror.
Chapter 26 Geometrical Optics. Units of Chapter 26 The Reflection of Light Forming Images with a Plane Mirror Spherical Mirrors Ray Tracing and the Mirror.
Chapter 23 Mirrors and Lenses. Notation for Mirrors and Lenses The object distance is the distance from the object to the mirror or lens Denoted by p.
Chapter 23 Mirrors and Lenses.
Light: Geometric Optics
Image Formation 2 Thin Lens Multi lens/mirror system
Correlation Of Corneal Contour With Higher Order Ocular Aberrations (HOA) In Indian Eyes Manish Chaudhary MS Amit Gupta MS Advanced Eye Centre, Postgraduate.
Chapter 23 Mirrors and Lenses.
Optics 1----by Dr.H.Huang, Department of Applied Physics
Chapter 33 Lenses and Optical Instruments
Visual outcome & subjective visual symptoms of the Tecnis ZM900 multifocal intraocular lens in Asian eyes Dr Colin S.H. Tan MBBS, MMed (Ophth), FRCSEd.
Multifocal Intraocular Lenses & Contrast Sensitivity
Aspheric IOL’s: clinical benefits and customizing cataract surgery Bojan Pajic, MD, PhD, FEBO Swiss Eye Research Foundation, ORASIS, Reinach, Switzerland.
1 OPTICAL QUALITY OF DUAL- OPTIC ACCOMMODATING INTRAOCULAR LENS Andrea Galvis, MD 1,3 a Ivan Ossma, MD MPH MSc 1,2, a Sanjeev Kasthurirangan, PhD 4, b.
Cataract Surgery Using Biaspheric IOLs in Patients With Corneal Irregularities James P. Gills, MD St. Luke’s Cataract & Laser Institute Tarpon Springs.
Ruth Lapid-Gortzak MD PhD 1,2, Jan Willem van der Linden BOpt 2, and Ivanka J. van der Meulen MD 1,2 1 Department of Ophthalmology, Academic Medical Center,
Chapter 3: Mirrors and Lenses. Lenses –Refraction –Converging rays –Diverging rays Converging Lens –Ray tracing rules –Image formation Diverging Lens.
Mitchell A Jackson MD Founder/Director, Jacksoneye Lake Villa, IL USA
Copyright © 2009 Pearson Education, Inc. Chapter 33 Lenses and Optical Instruments.
Human Image Formation What is the quality of the optics of the human eye? How can we measure it and describe it empirically?
R. Gary Lane, MD 1, J. Michael Jumper, MD 1, Yu Hyon Kim, MD 1, John Taboada, PhD 2 1 Wilford Hall Medical Center, San Antonio, TX, 2 Brooks AFB, San Antonio,
Lenses Chapter 30.
Refraction Snell’s Law: n1 sin(1) = n2 sin(2)
W. A. Maxwell, MD, PhD ASCRS 2008 Comparison of the Optical Image Quality for Presbyopia Correcting IOLs using Modulation Transfer Function Testing W.
Copyright © 2010 Pearson Education, Inc. Lecture Outline Chapter 26 Physics, 4 th Edition James S. Walker.
Fundamental of Optical Engineering Lecture 3.  Aberration happens when the rays do not converge to a point where it should be. We may distinguish the.
Medical Director - R & D Consultant, Ziemer Group AG, Port, Switzerland Speaker, Bausch & Lomb do Brasil Territory Manager.
W. Maxwell, MD, PhD California Eye Institute Fresno, California
Chapter 34 Lecture Eight: Images: II. Image Formed by a Thin Lens A thin lens is one whose thickness is small compared to the radii of curvature For a.
The authors have no financial interest in the subject matter of this poster Yinfei Xu, Peter S. Hersh, MD, David S. Chu, MD Institutional Affiliations.
Lesson 25 Lenses Eleanor Roosevelt High School Chin-Sung Lin.
Lenses and Imaging (Part I) Why is imaging necessary: Huygen’s principle – Spherical & parallel ray bundles, points at infinity Refraction at spherical.
Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,
ECEN 4616/5616 Optoelectronic Design Class website with past lectures, various files, and assignments: (The.
The Eye and Sight Chapter 12. Vision begins when light rays are reflected off an object and enter the eyes through the cornea, the transparent outer covering.
Prof. Charles A. DiMarzio Northeastern University Fall 2003 July 2003
AcrySof ® ReSTOR ® Aspheric IOL. Aspheric IOL AcrySof ® ReSTOR ® 2 AcrySof ® ReSTOR ® Aspheric IOL SN6AD3 Add Power: +4 D Spectacle Plane: 3.2 D Range:
Ophthalmology & Vision Sciences
Futoshi Taketani, MD,PhD,
Ocular functional optical zone following hyperopic LASIK/PRK: Analysis based on polychromatic retinal image quality Mitchell P. Weikert, MD Li Wang, MD,
LogMAR-Analysis of multifocal intraocular lenses: Clinical performance A. Mannsfeld, I.-J. Limberger, A. Ehmer, M.P. Holzer, G. U.Auffarth International.
AcrySof ® ReSTOR ® Apodized Diffractive IOL. What is the AcrySof ® ReSTOR ® IOL? The AcrySof ® ReSTOR ® IOL incorporates an apodized diffractive optic.
Chapter 19. Reflection The smooth surface of the lake reflects light rays so that the observer sees an inverted image of the landscape.
Financial Disclosure: Medical Director–Galilei R&D Consultant, Ziemer Group AG, Port, Switzerland Consultant & Territory Manager for.
Comparison of 2 Models of Aspheric Diffractive Multifocal IOL
Preliminary Results after Cataract Surgery with the Aspheric Acrysof ReSTOR IOL to Correct Presbyopia Meeting of the ASCRS Chicago 8-10 February 2007 R.M.M.A.
Warren Hill, MD, FACS East Valley Ophthalmology 5620 E. Broadway Road
Toric IOLs: wavefront aberrometry and quality of life Mencucci Rita Giordano Cristina, Stiko Ermelinda, Miranda Paolo, Eleonora Favuzza, Ugo Menchini Authors.
Refraction and Lenses. The most common application of refraction in science and technology is lenses. The kind of lenses we typically think of are made.
Geometrical Optics.
Chapter 23 Physics A First Course Light and Optics.
Vinohrady Teaching Hospital, Prague, Czech Republic Vinohrady Teaching Hospital, Prague, Czech Republic M. Vokrojova MD, M. Vokrojova MD, D. Sivekova MD,
Kavita Gala David Spalton Mayank Nanavaty St Thomas’ Hospital , London
July © Chuck DiMarzio, Northeastern University ECEG105/ECEU646 Optics for Engineers Course Notes Part 4: Apertures, Aberrations Prof.
Source: Figure 1. Image of colors with their different wavelengths refracted at different angles.
Eye clinic of the 3rd Faculty of Medicine, Prague, Czech Republic
Measurements of Multifocal Intraocular Lenses Using a Fluid-Filled Model Eye Charles Campbell This presentation is the result of work done with the assistance.
Carlos G. Arce, MD Associate Researcher and Ophthalmologist
Comparison of vision with an accommodating IOL versus a multifocal IOL
Michael Goodman, Alexandra Paul and Andrew Hsu
Presentation transcript:

Theoretical Optical Performance of an Equal Conic Intraocular Lens and Comparison to Spherical and Aspheric IOLs Edwin J. Sarver, PhD

The author(s) acknowledge financial interest in the subject matter of this presentation.

Acknowledgement… Contributors on this project…Contributors on this project… –Don Sanders, MD, PhD –John Clough, LensTec –Hayden Beatty, LensTec –Jim Simms, LensTec

Background… Recent studies have shown that aspheric IOLs can provide patients with significant optical benefits over traditional spherical surface IOLs.Recent studies have shown that aspheric IOLs can provide patients with significant optical benefits over traditional spherical surface IOLs. 1: Altmann GE, Nichamin LD, Lane SS, Pepose JS. Optical performance of 3 intraocular lens designs in the presence of decentration. J Cataract Refract Surg Mar;31(3): : Bellucci R, Morselli S, Piers P. Comparison of wavefront aberrations and optical quality of eyes implanted with five different intraocular lenses. J Refract Surg Jul- Aug;20(4): : Packer M, Fine IH, Hoffman RS, Piers PA. Improved functional vision with a modified prolate intraocular lens. J Cataract Refract Surg May;30(5): : Kershner RM. Retinal image contrast and functional visual performance with aspheric, silicone, and acrylic intraocular lenses. Prospective evaluation. J Cataract Refract Surg Sep;29(9):

Optical benefits… The optical benefits are due to a reduction in optical aberrations at the retina.The optical benefits are due to a reduction in optical aberrations at the retina. Primarily, spherical aberration is reduced.Primarily, spherical aberration is reduced.

Spherical aberration Spherical aberration occurs when rays away from the paraxial region do not intersect at the paraxial focus.Spherical aberration occurs when rays away from the paraxial region do not intersect at the paraxial focus.

Paraxial ray… Paraxial ray A paraxial ray is an optical ray traced “near” the optical axis.

Paraxial focus… Paraxial focus Paraxial ray The paraxial focus is where the paraxial ray crosses the optical axis after refraction by the lens.

Positive spherical aberration… Paraxial focus Paraxial ray Off axis ray (positive sa) When an off-axis ray is refracted by the lens and crosses the axis in FRONT of the paraxial focal point, the ray exhibits POSITIVE spherical aberration.

Paraxial focus Paraxial ray Off axis ray (positive sa) Off axis ray (negative sa) When an off-axis ray is refracted by the lens and crosses the axis in BACK of the paraxial focal point, the ray exhibits NEGATIVE spherical aberration. Negative spherical aberration…

Corneal spherical aberration… The mean corneal spherical aberration has been reported to be about microns 1The mean corneal spherical aberration has been reported to be about microns 1 About 90% of the population has positive corneal spherical aberration — About 10% of the population has negative corneal spherical aberration 2About 90% of the population has positive corneal spherical aberration — About 10% of the population has negative corneal spherical aberration 2 1 Holladay JT, et al, A new intraocular lens design to reduce spherical aberration of pseudophakic eyes. J Refract Surg., 2002 Nov-Dec;18(6): Krueger RR, et al, Wavefront Customized Visual Correction, Chapter 42, p. 368, 2004.

Approximate distribution of corneal spherical aberrations 0.27 μm 10% negative90% positive

Spherical IOLs A biconvex IOL with spherical surfaces exhibits positive spherical aberration.A biconvex IOL with spherical surfaces exhibits positive spherical aberration. Thus, usually, spherical IOLs ADD positive spherical aberration to the already positive corneal spherical aberrationThus, usually, spherical IOLs ADD positive spherical aberration to the already positive corneal spherical aberration

Aspheric IOLs Aspheric IOLs attempt to improve pseudophakic vision by controlling spherical aberrationsAspheric IOLs attempt to improve pseudophakic vision by controlling spherical aberrations One strategy is to design a lens with negative spherical aberrations to balance the normally positive corneal spherical aberrationsOne strategy is to design a lens with negative spherical aberrations to balance the normally positive corneal spherical aberrations Another strategy is to design a lens with minimum spherical aberrations so that no additional spherical aberration is added to the corneal spherical aberrationsAnother strategy is to design a lens with minimum spherical aberrations so that no additional spherical aberration is added to the corneal spherical aberrations –Could be an asymmetric design –Could be a symmetric design

Comparison of IOLs Given these IOL design strategies we want to investigate their potential strengths and weaknessesGiven these IOL design strategies we want to investigate their potential strengths and weaknesses First, we will describe the designs…First, we will describe the designs…

22 D IOL designs… Parameter Spherical surface IOL Negative spherical aberrations Asymmetric zero spherical aberrations Ref. index R K th & 6 th coef , R K Altmann, et al, Optical performance of 3 intraocular lens designs in the presence of decentration, J Cataract Refract Surg. 2005;31(3):

22 D IOL design shapes… Spherical surface IOL Sphere Propagation of light

22 D IOL design shapes… Spherical surface IOL Sphere Propagation of light Negative spherical aberrations IOL 6 th order asphereSphere

22 D IOL design shapes… Spherical surface IOL Sphere Propagation of light Negative spherical aberrations IOL 6 th order asphereSphere Asymmetric zero spherical aberrations IOL Conic

Equal conic, low spherical aberrations IOL… Want to use conic surface for both anterior and posteriorWant to use conic surface for both anterior and posterior Want both surfaces equalWant both surfaces equal Want low spherical aberrationsWant low spherical aberrations

Equal conic design strategy… First, we find the apical radius for the front and back surfaces to give the desired power. R-R n 0 = F=1336 / P Paraxial ray

Equal conic design strategy… Next, we find the conic K parameter so that off axis rays intersect the paraxial focus. KK n 0 = Off axis ray Paraxial ray

22 D IOL designs… Parameter Sphere / Sphere 6 th Order asphere / Sphere Conic 1 / Conic 2 Equal conic Ref. index R K th & 6 th coef , R K

Longitudinal aberrations… Negative spherical aberration Spherical Note: spherical aberration in opposite directions.

Longitudinal aberrations… Negative spherical aberration Spherical Negative spherical aberrations Positive spherical aberrations

Equal conic Unequal conic Longitudinal aberrations… Note: scale is 1000 x smaller than previous slide.

More important… Rather than just look at the performance of the IOL alone, it is more important to consider how it performs in the eye.Rather than just look at the performance of the IOL alone, it is more important to consider how it performs in the eye. To facilitate this analysis, we use a simple aspheric eye model.To facilitate this analysis, we use a simple aspheric eye model.

Choice of eye model… Negative spherical aberration IOL was optimized for anterior cornea K = Negative spherical aberration IOL was optimized for anterior cornea K = “Zero” spherical aberration IOLs work best with anterior cornea with K = -1/n 2 = -0.53“Zero” spherical aberration IOLs work best with anterior cornea with K = -1/n 2 = Mean cornea has K = -0.26Mean cornea has K = Kooijman 1 eye model has K = -0.25, (we use this model)Kooijman 1 eye model has K = -0.25, (we use this model) 1 Atchison and Smith, Optics of the human eye, Butterworth-Heinemann, 2000, p.255.

Kooijman/optical model R1=7.8, K1=-0.25 ELP=4.5 n=1.336 AL adjusted to give best focus for 3 mm pupil. R2=6.5, K2=-0.25 n=1.3771

Centered performance…

Centered – 3 mm Pupil All IOLs work pretty well here – MTF is limited by diffraction.

Centered – 5 mm Pupil, K= This is where negative spherical aberration IOL works best.

Centered, 5 mm Pupil, K=-0.25 As the eye model is adjusted, note how dramatically the performance is modified.

Centered, 5 mm Pupil, K=-0.53 When the cornea has spherical aberrations near zero, the “zero” spherical aberration IOLs perform best.

Centered IOL observations… Over this range of K values, the spherical IOL is has lowest performanceOver this range of K values, the spherical IOL is has lowest performance The best performer in the group of conic surface IOLs depends upon the K valueThe best performer in the group of conic surface IOLs depends upon the K value For the mean K of -0.25, the negative spherical aberration IOL performs bestFor the mean K of -0.25, the negative spherical aberration IOL performs best

Tilt…

10 deg tilt – 3 mm pupil, K= For this eye model, all IOLs perform about the same.

10 deg tilt, 3 mm pupil, K=-0.25 For mean corneal shape, the negative spherical aberration IOL performance starts to fall off.

10 deg tilt, 5 mm pupil, K= Performance for all IOLs close again…

10 deg tilt, 5 mm pupil, K=-0.25 “Zero” spherical aberration IOLs start to perform better for mean corneal shape.

Tilt observations… Depending upon the corneal eccentricity:Depending upon the corneal eccentricity: –The performance of the IOL designs are comparable –For some cases, the zero spherical aberration IOLs out perform the spherical surface and negative spherical aberration IOLs

Decentration…

Decentration 1 mm, 3 mm pupil, K= Clearly, the spherical surface and negative spherical aberrations IOLs have trouble with decentration.

Decentration 1 mm, 3 mm pupil, K=-0.25 This trend does not depend upon the corneal shape factor.

Decentration 1 mm – 5 mm pupil, K= The same optical behavior is seen for the 3 and 5 mm pupils.

Decentration 1 mm, 5 mm pupil, K=-0.25 Again, the same trend that does not depend upon corneal eccentricity.

Decentration observations… For 1.0 mm decentration:For 1.0 mm decentration: –The spherical surface and negative spherical aberration IOLs do not perform as well as zero aberration IOL designs –The trends for decentration does not depend upon pupil size or corneal eccentricity

Defocus…

Defocus 0.5D, 3 mm Pupil, K= For a 3 mm pupil, the corneal eccentricity does not affect optical performance to a large degree – an seen in this and the next slide.

Defocus 0.5D, 3 mm Pupil, K=-0.25

Defocus 0.5D, 5 mm Pupil, K= The general performance of the IOLs for 0.5D of defocus and 5 mm pupil does not appear to depend upon corneal eccentricity.

Defocus 0.5D, 5 mm Pupil, K=-0.25 As a side issue, the large ripples corresponding to the negative spherical aberration IOL indicate regions of contrast reversal.

Defocus observations… For 0.5 D of defocus at 3.0 and 5.0 mm pupils, the performance of all IOLs are about equal.For 0.5 D of defocus at 3.0 and 5.0 mm pupils, the performance of all IOLs are about equal. The negative spherical aberration IOL shows more contrast for low frequency objects than the other IOLsThe negative spherical aberration IOL shows more contrast for low frequency objects than the other IOLs The negative spherical aberration IOL showed significant regions of contrast reversal at 5.0 mm pupilThe negative spherical aberration IOL showed significant regions of contrast reversal at 5.0 mm pupil

Closer look at EC & UC The equal conic IOLs and unequal conic IOL designs appear to perform about the sameThe equal conic IOLs and unequal conic IOL designs appear to perform about the same Want to consider variability in tangential and sagittal MTF components in more detailWant to consider variability in tangential and sagittal MTF components in more detail

Tilt of 10 deg, 5 mm pupil The tangential and sagittal MTF components indicate a greater variability for the unequal conic design compared to the equal conic design.

Tilt |T-S| graph The magnitude of the differences between the tangential and sagittal MTF components clearly show more variability for the unequal conic design.

Decentration It is more subtle which lens design is more variable.

Decentration |T-S| graph By comparison of the magnitude of the difference between tangential and sagittal MTF, we see that the equal conic design has less variability.

Discussion There are various conditions in which one IOL design will perform better than another, but generally…There are various conditions in which one IOL design will perform better than another, but generally… –Aspheric IOLs perform better than spherical surface IOLs –For the level of alignment errors investigated here, zero spherical aberration IOLs perform better than spherical surface IOLs and negative spherical aberration IOLs

Discussion… Recognizing the variability in corneal eccentricity, it may be prudent to decide upon the use of an aspheric IOL design as a function of measured corneal aberrations (not ocular aberrations)Recognizing the variability in corneal eccentricity, it may be prudent to decide upon the use of an aspheric IOL design as a function of measured corneal aberrations (not ocular aberrations) This IOL selection strategy was suggested by Krueger et al.This IOL selection strategy was suggested by Krueger et al. Krueger RR, et al, Wavefront Customized Visual Correction, Chapter 42, p. 368, 2004.

Summary Aspheric IOLs have optical advantages over spherical IOLsAspheric IOLs have optical advantages over spherical IOLs For small alignment errors and positive spherical aberration corneas, negative spherical aberration IOLs perform bestFor small alignment errors and positive spherical aberration corneas, negative spherical aberration IOLs perform best For larger alignment errors, “zero” spherical aberration IOLs perform bestFor larger alignment errors, “zero” spherical aberration IOLs perform best

Summary “Zero” spherical aberration IOLs perform well over a wider range of corneal shapes and alignment errors than negative spherical aberration IOLs“Zero” spherical aberration IOLs perform well over a wider range of corneal shapes and alignment errors than negative spherical aberration IOLs The equal and unequal conic IOL designs perform are very similarThe equal and unequal conic IOL designs perform are very similar The equal conic IOL design performs slightly better than the unequal conic IOL design in terms of smaller variability in tangential and sagittal MTF componentsThe equal conic IOL design performs slightly better than the unequal conic IOL design in terms of smaller variability in tangential and sagittal MTF components

Thank you!