Chiho Wang ATLAS TRT Duke University Hampton, May Xe-CO 2 Stability Chiho Wang Duke University
Chiho Wang ATLAS TRT Duke University Hampton, May Setup X-ray aperture limits active region ~ 6-7cm on straw. 0.002” Cu foil reduced X-ray intensity to ~0.1uA/cm Only 1 pair of ASD sockets are active to draw current. Scan along straw at 8cm intervals. At each location, raise HV from 1550V to 1750V at 25V step, 20 sec interval. Record current & # of Trips.
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09 ASD #1
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09 ASD #2
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #3
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #4
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #5
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #6
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #10
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #11
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #15
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #16
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #21
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #22
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #27
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #28
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #30
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #31
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #32
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD #33
Chiho Wang ATLAS TRT Duke University Hampton, May Observation Most ASD HV trip at mid wire support location. Some ASD HV trip at end wire support. After several hours of “burning” at the wire support, it gets more stable and less likely to trip u ASD were scanned many times with Ar-CO2 before taking Xe data. They are obviously more stable. u In the case of Ar-CO2, X-ray also triggers a self-sustained current draw that lasts for several hours even after X-ray is turned off. This contributes to an increase of current draw when X-ray passes through mid wire support. This current, however, also dies out after burning.
Chiho Wang ATLAS TRT Duke University Hampton, May Module 2.09, ASD 30, Ar-CO2
Chiho Wang ATLAS TRT Duke University Hampton, May Summary Xe-CO2 is stable at 1750V with 0.3uA/cm local rate Mid wire support tends to be unstable in the beginning of the chamber operation, but can be improved with “burning”. Some end wire supports also show the same instability but not as often as mid wire support.