How to Recognize and Eliminate Ghost- and Deformed Peaks in Gas Chromatography Jaap De Zeeuw, Restek Corporation, Middelburg, The Netherlands Summary The.

Slides:



Advertisements
Similar presentations
Chem. 231 – 3/4 Lecture. Announcements I Set 1 Lab Reports –Both Lab Reports due 3/13 –Only one of the lab reports needs to be detailed, the other can.
Advertisements

Gas Chromatography Introduction 1.) Gas Chromatography
Gas Chromatography 427 PHC.
Lecture 8b Gas Chromatography.
CAPILLARY GC INLETS.
GAS CHROMATOGRAPHY.
Gas Chromatography There an be many parts to a gas chromatography system but the basic components include: An injection system. A column (controllable.
Lab Methods Day June 25, 2014 Gas Chromatography
GC. Columns have a finite life, but they will last much longer if the gas chromatography system is operated with care. The major causes of reduced column.
The Foundations: Classical Split and Splitless Injection
Chromatography Integration and Trouble-shooting
Chromatography Dr.Tawfeq A. Al-Howiriny Associate Professor
Analysis of Residues. Laboratory Analysis of Debris and Other Samples 1. Preparation of Liquid Samples: Liquid samples are simply drawn into a special.
Lecture 8. GC/MS.
Chromatography.
In carbon-13 NMR, what do the number of peaks represent?
1 Chapter 24 GC Gas Chromatography. 2 GC Mechanism of separation is primarily volatility. Difference in boiling point, vapor pressure etc. What controls.
Intro to Chromatographic Separations Chap 26. Originally based on separation and identification by color Originally based on separation and identification.
Instant Notes Analytical Chemistry
AC Analytical Controls January 2001
Chromatography and Instrumentation. Invented by a Russian Botanist Mikhail Tswett in 1903 He used chromatography to separate the colour pigments in plants.
Chromatography Is a technique used to separate and identify the components of a mixture. Works by allowing the molecules present in the mixture to distribute.
Chromatography Year 12.
Chapter 6 - Chromatography
Introduction to High Performance Liquid Chromatography.
Chemical Ideas 7.6 Chromatography. The general principle. Use – to separate and identify components of mixtures. Several different types - paper, thin.
Gas Chromatography Carrier Gas Flow Control Injector Port Column Column Oven Detector Recorder The GC system consists of gas supplies for the mobile phase.
Chapter 27 Gas Chromatography 1. Principles
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 22 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Chromatography Chapter 6.
In gas-liquid chromatography (g.l.c.) a long tube contains the chromatography material. The tube is usually coiled so that it takes up less space.
CHROMATOGRAPHY. The general name given to methods by which two or more compounds in a mixture are physically separated by distributing themselves between.
PAS-Intro1 Dated 24th Aug Copyright © Siemens AG 1999 All Rights Reserved Siemens Advanced Engineering Pte Ltd Process Analytical Services Division.
Separation Science Differences in IMFs can be used to separate chemical substances for further analysis. Differences in IMFs can be used to separate chemical.
INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER 12 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
900-LC P. van Poppel – Field Support Specialist October 2009.
Experiment 2 DISTILLATION AND GAS CHROMATOGRAPHY OF ALKANES.
INTRODUCTION TO CHROMATOGRAPY
Gas Chromatography Experiment. Gas Chromatography - Gas Chromatography (GC) is a common technique used to separate and identify volatile organic compounds.
By: Arafath and Nick. What is it  Chromatography is a technique that is used to separate the substances present in a mixture.  It is widely used to.
Analytical Separations
1 Gas Chromatography Lecture Liquid Stationary Phases In general, the polarity of the stationary phase should match that of the sample constituents.
Introduction to Chromatography. Introduction Chromatography permit the scientist to separate closely related components of complex mixtures. In all chromatographic.
Created with MindGenius Business 2005® Instrumentation (1) Instrumentation (1) Origins Preparative column chromatography (Glass columns) Internal diameter:
HPLC.
PURIFCATION OF ORGANIC COMPOUNDS
Chem. 133 – 5/3 Lecture. Announcements Lab – Term Project Progress Report Due Today – Last Assignments: Term Project Poster and Peer Review Grading (Friday,
Introduction to Gas Chromatography
HPLC TROUBLESHOOTING. There are five major of symptoms to help you quickly identify the source of the problem(s): I.Pressure abnormalities II.Leaks III.Problems.
PREPARED BY VAISHNANI BHAVIN M.PHARMA-I (Q.A.). Qualitative error sources in HPLC: Poor separation power. Some special substances from the samples can.
Chromatography.
Evolution GC-MS/MS: Pesticide analysis in canola oil Evolution GC-MS/MS: Pesticide analysis in canola oil Vivian Watts 1, Ingo Christ 1, Mark Misunis 2.
Experiments in Analytical Chemistry
Reading assignment: section 26E(p781) Chapter 26 # 2, 3, 14, 15, 16 Chapter 27 # 7(a,d,f), 22, 23, 24, 25.
CHROMATOGRAPHY.
GC Trouble shooting.
Chem. 133 – 5/9 Lecture.
6850/6890 分流动画 Split/Splitless 进样口 – 部件
Overview Determining Identity Quantitative Analysis
Chemical Ideas 7.6 Chromatography.
Chem. 133 – 5/11 Lecture.
Gas Chromatography.
Chromatography Daheeya Alenazi.
Gas Chromatography.
5 week plan Each week you will have 2 single lessons and 1 double lesson. In those lessons you will learn aspects from Modern Analytical Techniques (MAT)
GAS CHROMATOGRAPHY.
Chromatography is a technique in which the components of a mixture are separated based on The differences in rates at which they are carried through a.
Chromatography is a technique in which the components of a mixture are separated based on The differences in rates at which they are carried through a.
32A-2 Sample Injection system
Presentation transcript:

How to Recognize and Eliminate Ghost- and Deformed Peaks in Gas Chromatography Jaap De Zeeuw, Restek Corporation, Middelburg, The Netherlands Summary The chromatogram is like a fingerprint. If you can read the chromatogram by looking at peak shapes, retention, base line and by comparing with “normal” situation, you have a good chance to solve problems and improve the analysis. A ghost peak is a peak that is showing up, but is not supposed to be there. Sometimes it is referred as a “system” peak. Ghost peaks can be created in many ways. It’s a component that is added/created somewhere in the system, it is injected/trapped/focused onto the column, and will elute. Problems will escalate if a ghost peak interferes with an analyte that has to be quantified. Sources for ghost peaks can be sample vials, gloves, syringes, reagents, carrier gas, tubing, the injection port, operation, memory effects and even the column-phase itself. “A ghost peak is a peak that appears at a position where we do not expect a peak”. It is a component that shows up in the system and it may show up everywhere in the chromatogram. Sometimes the peak is a sharp peak, sometimes it’s a broad deformed peak or a “hump”, sometime sit is a rising base line.. Some of the key origins for ghost or deformed peaks are: The purity of the carrier gas Using non-GC grade carrier gas tubing Memory effects due to back flash Contamination of injection port Injection port reactivity Septum, O-rings Analytical column bleed Column installation Sample contamination from syringe, vials, vial-septa, gloves What can cause a ghost or a deformed peak? Dirty tubing will release residual hydrocarbons, which will be trapped on the column and show up as “ghost” peaks. Use only GC grade copper tubing. Trace 1 in the figure above, shows the materials that elute from a “clean” batch of copper tubing. Specific cleaning procedures eliminate this risk. When a system is contaminated, a base line like above can develop. Here hydrocarbon oils were back-flashed in the gas lines. This was confirmed by the MS spectrum. Make sure that you do not inject more then the liner volume. Note that each solvent has a different expansion volume. To prevent, use larger diameter liners, inject less and know the liner and expansion volume of your solvent. Septa is a known source for ghost peaks. Septa is mostly 100% siloxane polymer. When heated it will form bleed products. These products are cyclic siloxanes that show as peaks. Intensity depends on septum type, age and temperature. To reduce: always use septum purge, use low injection temperatures, replace septa regularly. You can also consider to use a septum- less system like a Merlin Seal. Septa particles in the liner can be a big problem. As they are at high temperature, they form siloxanes that are injected on the column. Such particles will also “delay” sample transfer as the siloxane will also retain components. Liner condition upon injection is very important as here the chromatography starts. Use needles with more friendly (tapered/bended) top. Also a septum with a center guide, like the BTO, can be used. Replace liners systematically. If the precision liner is used the septa particles will remain on the top of the wool. Contamination of Split line can cause broad “blobby” peaks. When sample is splitted, the lower outlet temperature makes sample condense. This can built up a contamination and will be a source for ghost peaks. On the right: A: split line nearly blocked; B: after cleaning Regularly check the split line and clean it. There are also split line filters. Check also flow of split outlet with a flow meter. Decomposition of components in an active inlet. If components degrade in a hot inlet, extra peaks will be observed in the chromatogram. The reactivity can be reduced by using highly inert liners (for instance the latest sky-liners). Use lowest possible injection temperature; Often a good injection is possible at a lower liner temperature. Do not use wool; With splitless use a pressure pulse to make transfer faster; Consider to use a different technique that is “milder” for your components: This can be a PTV or the cold-on-column Other sources for Ghost peaks are shown above: Using the MS makes the identification easier. Besides siloxanes, septa also can contain phthalates. (m/z 149,167 and 279) The m/z 277 is indicator for a contaminant that is often released from the O- ring. Carry over of PAH is caused by contamination due to a previous injection, probably to back-flash. Sample contamination in the vial: Just by injecting 2 times through the vial – seal, all the peaks above were introduced. Ghost peaks can be caused by cross-contamination via the wash vial,. Also when the needle punctures through a vial septum, each time the outside will take material of the septum. Select sealing materials that are OK with your solvent. Also clean rinse vials regularly or use multiple rinse vials. Contamination via Gloves. In our aim to work as clean as possible, we sometimes overlook the obvious. In one of our trace analysis methods, we saw systematically a set of ghost peaks appear. After analysis a small piece of the glove, we found the cause.. Think about it: If you can “smell” something, it’s basically a head space, which just needs concentration to be visible.. Extra peaks showing up as large noise produced by the stationary phase upon fast cooling. High bleed in a column can focus when the oven is cooled down. Because oven will always cool more on one side, there will be a focusing effect for bleed products. In every column winding, bleed will focus. Next run, all these focused bleed-peaks, will elute. We see this especially with high bleed (cyano) phases. Solution is to use temperature controlled cool down. The right signal above was obtained with a -10ºC/min cool program. This effect will depend strongly on the GC type and the temperature gradient in the oven upon cooling. In column decomposition: Thermo labile components will start to decompose when the temperature reached a certain value. when the decomposition product has less retention, a typical peaks shape is observed as above. Here a brominated diphenylether, BDE-209 is measured. Important is to reduce elution temperature and column activity; this is realized by: 1 Use high flow rate; 2 Slow temperature program; 3 Use thin films; 4 Use short columns; 5 Use inert surface (Rtx-1614); Want to learn Chromatography,Tips & Tricks and troubleshooting? Visit ChromaBLOGraphy at Want to learn Chromatography,Tips & Tricks and troubleshooting? Visit ChromaBLOGraphy at