ELECTROCHEMISTRY CHEM 4700 CHAPTER 5 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.

Slides:



Advertisements
Similar presentations
Chapter 9 Chemical Change
Advertisements

CE 541 Complex Formation.
Review of Analytical Methods Part 2: Electrochemistry
CHEM 213 Instrumental Analysis
Potentiometric Methods A.) Introduction:
15-1 Potentiometry Potential measurements of electrochemical cells Ion selective methods §Reference electrode §Indicator electrode §Potential measuring.
Summary of Potentiometry:
Effect of Electrolytes on Chemical Equilibria
Introduction to Analytical Chemistry
Chemical equilibrium: electrochemistry 자연과학대학 화학과 박영동 교수.
MLAB 2401: C LINICAL C HEMISTRY K ERI B ROPHY -M ARTINEZ Designs in Instrumentation Electrochemistry.
Tracking Reaction Progress via Electrochemical Methods.
Measuring concentration using electrodes Indicator electrodes used with reference electrode to measure potential of unknown solution E cell = E indicator.
POTENTIOMETRY 9th lecture
Electrodes and Potentiometry
Section 11 Potentiometry. Potentiometric Electrodes Potentiometric electrodes measure: Activity not concentration Concepts to review: Activity and affect.
14-6 Cells as chemical probes
Potentiometry. pH is a Unit of Measurement  pH = Power of Hydrogen (H + )  Defined as the Negative Logarithm of Hydrogen Ion Activity  pH = log (1/H.
© Sauvé 2002 Sébastien Sauvé Department of Chemistry Université de Montréal Metal speciation using ion-selective electrodes.
Chem. 133 – 3/5 Lecture. Announcements Lab –Set 2 Period 2 Labs Set to Finish 3/10 –3/12 will be make up day (for both Period 1 and Period 2 labs) –Set.
Dr. Marc Madou, UCI, Winter 2012 Class V Potentiometric and Amperometric Sensors (I) Electrochemistry MAE-295.
Potentiometry and the Detection of Fluoride
Fall 2014, Prof. JB Lee Ion Sensitive FET (ISFET) - What and Why?
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 12 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Potentiometric Methods
Testing for Cations and Anions Determine the presence of a cation or anion by a chemical reaction Determine the identity of cations and anions in an unknown.
Ion-Pair Chromatography In addition to the aqueous buffer and an organic solvent that is typical for reversed-phase, the mobile phase contains a counter.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 15 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
THE HYDRONIUM ION The proton does not actually exist in aqueous solution as a bare H + ion. The proton exists as the hydronium ion (H 3 O + ). Consider.
Solubility Equilibrium In saturated solutions dynamic equilibrium exists between undissolved solids and ionic species in solutions Solids continue to dissolve.
Chapter 16: Applications of Aqueous Equilibria Renee Y. Becker Valencia Community College 1.
William L Masterton Cecile N. Hurley Edward J. Neth University of Connecticut Chapter 4 Reactions in Aqueous.
A Seminar on Potentiometric and Diazotization Titrations
Electroanalytical methods
Potentiometry and potentiometric measurements. potentiometer A device for measuring the potential of an electrochemical cell without drawing a current.
Copyright Sautter SOLUBILITY EQUILIBRIUM Solubility refers to the ability of a substance to dissolve. In the study of solubility equilibrium we.
1 Acid-Base Equilibria and Solubility Equilibria Chapter 16 Copyright © The McGraw-Hill Companies, Inc. Permission required for reproduction or display.
PRINCIPLES OF CHEMISTRY I CHEM 1211 CHAPTER 4 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state.
Announcements Course Evaluations Final is Wednesday Afternoon on May 9 th Homework 14-15, 14-26, 15-6, 16-3, 16-6,
Chapter 15 Electrode Measurements “Chem 7” test in hospital lab. Na +, K +, Cl -, CO 2, glucose, urea, & creatinine ion-selective electrodes.
Electrochemical Methods Dr M.Afroz Bakht. Potentiometry Potentiometry is a method of analysis used in the determination of concentration of ions or substances.
Precipitation Titrations Dr. Riham Ali Hazzaa Analytical chemistry Petrochemical Engineering.
CHAPTER 11 ELEMENTS OF ELECTROCHEMISTRY Introduction to Analytical Chemistry.
1 Titration Curve of a Weak Base with a Strong Acid.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 23 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Electrodes and Potentiometry Introduction 1.)Potentiometry  Use of Electrodes to Measure Voltages that Provide Chemical Information - Various electrodes.
Chem. 133 – 3/3 Lecture. Announcements Homework Set 2 (pass out) Grading –Working to get the Electronics labs and exam 1 graded by next Tuesday Lab –today.
POTENTIOMETRY.
POTENTIALS AND THERMODYNAMICS OF CELLS (2)
Home Reading Skoog et al. Fundamental of Analytical Chemistry. Chapters 18, 19, 21.
Art PowerPoints Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 14: ELECTRODES AND POTENTIOMETRY.
Ch. 21 Potentiometry 1. General principles
POTENTIALS AND THERMODYNAMICS OF CELLS (1) POTENTIALS AND THERMODYNAMICS OF CELLS (1)
Home Reading Skoog et al. Fundamentals of Analytical Chemistry. Chapter 10, 21.
SCI3023 ELECTROCHEMISTRY Chapter 8e: Potentiometry
Inovace bakalářského studijního oboru Aplikovaná chemie Reg. č.: CZ.1.07/2.2.00/
Instrumental Analysis Ion selective electrodes
Potentiometry and Ion-Selective Electrodes
Chem. 133 – 2/28 Lecture.
Electrochemistry MAE-212
Ion Selective Electrode
Potentiometry and potentiometric measurements
Ion Selective Electrode
Ion-Selective Electrode (I.S.E.)
Potentiometry Potential measurements of electrochemical cells
Electrode Measurements – Ch 15
Tracking Reaction Progress via Electrochemical Methods
The Solubility Product Constant, Ksp
Chapter 23 Potentiometry.
Presentation transcript:

ELECTROCHEMISTRY CHEM 4700 CHAPTER 5 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university

CHAPTER 5 POTENTIOMETRY

- Based on static (zero-current) measurements - Used to obtain information on the composition of an analyte - Potential between two electrodes is measured Applications - Environmental monitoring - Clinical diagnostics (blood testing, electrolytes in blood) - Control of reaction processes POTENTIOMETRY

- Also known as indicator electrodes - Respond directly to the analyte - Used for direct potentiometric measurements - Selectively binds and measures the activity of one ion (no redox chemistry) Examples pH electrode Calcium (Ca 2+ ) electrode Chloride (Cl - ) electrode ION-SELECTIVE ELECTRODES (ISE)

Advanteages - Exhibit wide response - Exhibit wide linear range - Low cost - Color or turbidity of analyte does not affect results - Come in different shapes and sizes ION-SELECTIVE ELECTRODES (ISE)

- Made from a permselective ion-conducting membrane (ion-exchange material that allows ions of one electrical sign to pass through) - Reference electrode is inbuilt - Internal solution (solution inside electrode) contains ion of interest with constant activity - Ion of interest is also mixed with membrane - Membrane is nonporous and water insoluble ION-SELECTIVE ELECTRODES (ISE)

- Responds preferentially to one species in solution Internal reference electrode Ion-selective membrane Internal (filling) solution ION-SELECTIVE ELECTRODES (ISE)

- Selective (preferential) ion is C + - Membrane is made of poly(vinyl chloride) (PVC) - Membrane is impregnated with nonpolar liquid - Membrane contains ligand L (ion-selective ionophore) - Membrane contains the complex LC + - Membrane contains hydrophobic anion R - (ion exchanger) ION-SELECTIVE ELECTRODES (ISE)

- [C + ] inside the electrode ≠ [C + ] outside the electrode - Results in a potential difference across the membrane Generally (at 25 o C) - 10-fold change in activity implies 59/z i mV change in E - z i is the charge on the selective ion (negative for anions) - z i = +1 for K +, z i = +2 for Ca 2+, z i = -2 for CO 3 2- ION-SELECTIVE ELECTRODES (ISE)

- Let c i = molarity of C + - Activity (a i ) rather than molarity is measured by ISEs - Activity is the effective (active) concentration of analyte (effective concentration decreases due to ionic interactions) - z i = ionic charge (±) a i = γ i c i where γ i = activity coefficient (between 0 and 1) ION-SELECTIVE ELECTRODES (ISE)

Debye-Hückel Equation - Relates activity coefficients to ionic strength (at 25 o C) α = size of ion in picometers (1 pm = m) µ = ionic strength ION-SELECTIVE ELECTRODES (ISE)

Ionic strength - A measure of the concentration of all ions in solution with their charges taken into account c i = the concentration of the ith species Ionic strength of electrolytes 1:1 electrolytes (NaCl) µ = molarity 2:1 electrolytes (CaCl 2 ) µ = 3 x molarity 3:1 electrolytes (AlCl 3 ) µ = 6 x molarity 2:2 electrolytes (MgSO 4 ) µ = 4 x molarity ION-SELECTIVE ELECTRODES (ISE)

- For very dilute solutions a i ≈ c i - Activity coefficient decreases as ionic strength increases For z i = mV change in potential implies 4% change in activity For z i = mV change in potential implies 8% change in activity - This is known as Nernstian behavior ION-SELECTIVE ELECTRODES (ISE)

Selectivity Coefficient (k) - A measure of the ability of ISE to discriminate against an interfering ion - It is assumed that ISEs respond only to ion of interest - In practice, no electrode responds to only one specific ion - The lower the value of k the more selective is the electrode - k = 0 for an ideal electrode (implies no interference) ION-SELECTIVE ELECTRODES (ISE)

Selectivity Coefficient (k) For k > 1 - ISE responds better to the interfering ion than to the target ion For k = 1 - ISE responds similarly to both ions For k < 1 - ISE responds more selectively to ion of interest ION-SELECTIVE ELECTRODES (ISE)

Empirical Calibration Plot Potential (mV) p[C + ] Slope = 59/z i mV z i = charge of ion Called Nernstian slope - Used to determine the unknown concentration of analytes - Departure from linearity is observed at low concentrations ION-SELECTIVE ELECTRODES (ISE)

Three groups of ISEs - Glass electrodes - Liquid electrodes - Solid electrodes ION-SELECTIVE ELECTRODES (ISE)

GLASS ELECTRODES - Responsive to univalent cations - Employs thin ion-selective glass membrane

pH GLASS ELECTRODE - The most widely used - For pH measurements (selective ion is H + ) - Response is fast, stable, and has broad range - pH changes by 1 when [H + ] changes by a factor of 10 - Potential difference is V when [H + ] changes by a factor of 10 For a change in pH from 3.00 to 6.00 (3.00 units) Potential difference = 3.00 x V = 0.177

pH GLASS ELECTRODE - Thin glass membrane (bulb) consists of SiO 4 - Most common composition is SiO 2, Na 2 O, and CaO Glass membrane contains - dilute HCl solution - inbuilt reference electrode (Ag wire coated with AgCl)

pH GLASS ELECTRODE Glass Electrode Response at 25 o C (potential across membrane with respect to H + ) ΔpH = pH difference between inside and outside of glass bulb β ≈ 1 (typically ~ 0.98) (measured by calibrating electrode in solutions of known pH) K = assymetry potential (system constant, varies with electrodes)

pH GLASS ELECTRODE - Equilibrium establishes across the glass membrane with respect to H + in inner and outer solutions - This produces the potential, E - Linearity between pH and potential - Calibration plot yields slope = 59 mV/pH units - Electrode is prevented from drying out by storing in aqueous solution when not in use

pH GLASS ELECTRODE Sources of Error - Standards used for calibration - Junction potential - Equilibration time - Alkaline (sodium error) - Temperature - Strong acids - Response to H + (hydration effect)

OTHEER GLASS ELECTRODES Glass Electrodes For Other Cations K + -, NH 4 + -, Na + -selective electrodes - Mechanism is complex - Employs aluminosilicate glasses (Na 2 O, Al 2 O 3, SiO 2 ) - Minimizes interference from H + when solution pH > 5 pH Nonglass Electrodes - Quinhydrone electrode (quinone – hydroquinone couple) - Antimony electrode

LIQUID MEMBRANE ELECTRODES - Employs water-immiscible substances impregnated in a polymeric membrane (PVC) - For direct measurement of polyvalent cations and some anions - The inner solution is a saturated solution of the target ion - Hydrophilic complexing agents (e.g. EDTA) are added to inner solutions to improve detection limits - Inner wire is Ag/AgCl

Ion-Exchange Electrodes - The basis is the ability of phosphate ions to form stable complexes with calcium ions - Selective towards calcium - Employs cation-exchanger that has high affinity for calcium ions (diester of phosphoric acid) - Inner solution is a saturated solution of calcium chloride - Cell potential is given by LIQUID MEMBRANE ELECTRODES

Other Ion-Exchange Electrodes - Have poor selectivity and are limited to pharmaceutical formulations Examples - IEE for polycationic species (polyarginine, protamine) - IEE for polyanionic species (DNA) - IEE for detection of commonly abused drugs (large organic species) LIQUID MEMBRANE ELECTRODES

Neutral Carrier Electrodes - Employs neutral carriers such as crown ethers and cyclic polyesters - Carriers envelope target ions in their pockets Used for clinical analysis - detection of blood electrolytes - detection alkali and alkaline earth metal cations LIQUID MEMBRANE ELECTRODES

Neutral Carrier Electrodes Examples of Carriers - Monensin for sodium - Macrocyclic thioethers for Hg and Ag - Valinomycin for potassium ions - Calixarene derivatives for lead - 14-crown-4-ether for lithium LIQUID MEMBRANE ELECTRODES

Anion-Selective Electrodes - For sensing organic and inorganic anions Examples of Anions - Phosphate - Salicylate - Thiocyanate - Carbonate LIQUID MEMBRANE ELECTRODES

SOLID-STATE ELECTRODES - Solid membranes that are selective primarily to anions Solid-state membrane may be - single crystals - polycrystalline pellets or - mixed crystals

SOLID-STATE ELECTRODES Examples - Most common is fluoride-ion-selective electrode (limited pH range of 0-8.5) (OH - is the only interfering ion due to similar size and charge) - Iodide electrode (high selectivity over Br - and Cl - ) Chloride electrode (suffers interference from Br - and I - ) Thiocynate (SCN - ) and cyanide (CN - ) electrodes

OTHER ELECTRODES - Coated-wire electrodes (CWE) - Solid-state electrodes without inner solutions - Made up of metallic wire or disk conductor (Cu, Ag, Pt) - Mechanism is not well understood due to lack of internal reference - Usually not reproducible For detection of amino acids, cocaine, methadone, sodium

APPLICATIONS OF ISEs - Used as detectors for automated flow analyzers (flow injection systems) - High-speed determination of blood electrolytes in hospitals (H +, K +, Cl -, Ca 2+, Na + ) - For measuring soil samples (NO 3 -, Cl -, Li +, Ca 2+, Mg 2+ ) - Coupling ion chromatography with potentiometric detection - Micro ISEs as probe tips for SECM - Column detectors for capillary-zone electrophoresis