ILS and alignment revisited David Griffith, Voltaire Velazco University of Wollongong with very significant inputs from John Robinson (Lauder) Frank Hase.

Slides:



Advertisements
Similar presentations
Microscopes SC.912.L14.4.
Advertisements

The EZ-SAT.
JED Microprocessors Pty Ltd Presenting the JED T430 low-cost Projector Controllers Nov 22nd, 2009.
Lab 1 Paraxial Optics Lab in 106A. Look at paraxial optics rules Use a bi-convex singlet at 1:1 conjugates Do it double pass so can see image Lateral.
Elementary Science Cart. 8 Laptops 1 Projector Computer to Projector Connect to laptop.
Bellwork Why do scientists use Microscopes?
Basic Illuminating Light Paths and Proper Microscope Alignment E. D. Salmon University of North Carolina at Chapel Hill.
Telescope Tear-Down Anatomy of a 114mm f/8 Newtonian Reflector.
Use of a commercial laser tracker for optical alignment James H. Burge, Peng Su, Chunyu Zhao, Tom Zobrist College of Optical Sciences Steward Observatory.
NGAO Alignment Plan See KAON 719 P. Wizinowich. 2 Introduction KAON 719 is intended to define & describe the alignments that will need to be performed.
Physics 1809 Optics 2: Spherical Lenses and Optical Instruments Purpose of this Minilab Use lens formula to determine focal length of a lens. Learn about.
Modern Optics Lab Lab 3: Experiments with the Telemicroscope  Measuring: Focal lengths of thin lenses Real image magnification Virtual image magnification.
RET Optics Research Workshop Workshop #4 Basic Optics Dr. Mike Nofziger Professor College of Optical Sciences University of Arizona Dr. Mike Nofziger 2013.
Conventional and Computed Tomography
Integration and Alignment of Optical Subsystem Roy W. Esplin Dave McLain.
Absolute Photon Calibration of Zeiss Observer Z1 Microscope Shawn Miller Department of Optical Sciences University of Arizona, Tucson, AZ Optics.
Microscopes.
COMPOUND MICROSCOPE Dr. Jamba Gyeltshen 7/9/07.
The microscope practical NO (2)
Visual Angle How large an object appears, and how much detail we can see on it, depends on the size of the image it makes on the retina. This, in turns,
Advanced Optics Lab at San Jose State University Ramen Bahuguna Department of Physics.
Chapter 12 Optical Instruments Physics Beyond 2000.
A diffraction grating with 10,000 lines/cm will exhibit the first order maximum for light of wavelength 510 nm at what angle? (1 nm = 10-9 m) 0.51° 0.62°
DIC Differential Interference Contrast Kim Peifley 08/14/15.
PACS IIDR 01/02 Mar 2001 FPFPU Alignment1 D. Kampf KAYSER-THREDE.
Aligning the Cantilever, Laser and Photodetector.
干涉仪实验 Michelson interferometer Sliver mirror Sliver mirror with piezo Sliver mirror HeNe source.
Lyot Stop Focal Plane Mask OAP3 Out of plane spherical mirror.
Science 8 – Unit C: Light and Optical Systems Topic 5: Extending Human Vision.
YOU BREAK IT YOU BUY IT The Microscope Parts of, Use of, and Care.

The Microscope. There are 2 types of microscopes: 1. Simple- contains one lens 2. Compound- contains 2 or more lenses.
FTS Test Definition Summary Kaley Walker Test Readiness Review January 29, 2003.
Types of Microscopes 0.1 mm is the smallest object that is seen with the human eye. When we view with a microscope, we are either viewing in micrometer.
Introduction to the Microscope
 string beans  plastic  paper towel  Velcro  push pins  mushrooms  a bone  a tree branch  a strand of hair.
Microscope Eyepiece, ocular lens
Structure of a microscope. coarse adjustment knob fine adjustment knob arm stage base eyepiece objective clip mirror.
Charge Couple Device (CCD) – A dimesized component of the camera head into which light enters and is converted into an electronic signal. The video signal.
Reflection and refraction of light
EYEPIECE: 10x BODY/EXTENSION TUBE COARSE ADJUSTMENT FINE ADJUSTMENT
Introduction to the Microscope. Compound Microscope Dissection Microscope Scanning Electron Microscope (SEM)
J. H. Burgea,b, W. Davisona, H. M. Martina, C. Zhaob
Compressor Helium filled optional equivalent plane imaging laser beam dump hole in mirror for e-beam SSA and more.
Plane Reflection – Learning Outcomes  Demonstrate the laws of reflection.  Discuss images formed in plane mirrors.  Differentiate between real and virtual.
MICROSCOPE A compound microscope is an expensive precision instrument that requires special care and handling. let’s learn the parts of a compound microscope,
IOPtiMate™ OT-135P2 Technical Training – Part I System Description, Installation & Operation IOP (07)
What you need: In order to use these programs you need a program that sends out OSC messages in TUIO format. There are a few options in programs that.
Introduction to the Microscope
 Imagine a clear evening when a full moon is just starting to rise. Even though the Moon might seem large and close, it is still too far away for you.
Microscope Parts and Functions
Prepared By: Ms. Carolyn R. Magat
Digital Light Sources First introduced in 2001.
Photography Lab.
Lab 1 Paraxial Optics Lab in 106A.
Pre-launch Characteristics and Calibration
ACD Airborne FTS Spectrometer History Acquired in 1974 for measurements of stratospheric composition Flown hours on NCAR Sabreliner.
Microscope  An optical instrument that uses a lens or a combination of lenses to produce magnified images of small objects, especially of objects too small.
Setting up a trinocular compound microscope for best viewing
PRACTICAL 1 BY NANA ABOAGYE ACHEAMPONG
The Microscope Parts of, Use, and Care.
Using A Compound Light Microscope
Types of Microscopes 0.1 mm is the smallest object that is seen with the human eye. When we view with a microscope, we are either viewing in micrometer.
FLIP CAMERA BASICS MMS Tech Academy 2010.
Light Microscope: – light passes through one or more lenses to produce an enlarged image of a specimen compound light microscope – light microscope that.
Light Microscope: – light passes through one or more lenses to produce an enlarged image of a specimen compound light microscope – light microscope that.
The Microscope.
Purpose of this Minilab
Light Microscope: – light passes through one or more lenses to produce an enlarged image of a specimen compound light microscope – light microscope that.
The Microscope.
Presentation transcript:

ILS and alignment revisited David Griffith, Voltaire Velazco University of Wollongong with very significant inputs from John Robinson (Lauder) Frank Hase (KIT) Graham Kettlewell (UoW)

IFS125 FT spectrometer

IFS125 interferometer IR beam Laser Moving CC Beamsplitter Fixed CC Entrance stop (aperture) Exit stop (aperture) OAPs

The IR beam path HeNe Beam- expanded HeNe laser at input aperture Haidinger fringes at exit aperture 4mm

Some important points u The IR and laser alignment are independent of eachother l There are no shared adjustments u Laser affects wavenumber scale, does not affect ILS u IR alignment affects the ILS u There are two alignment methods on the wiki 1. Caltech/JPL method (Blavier, Washenfelder, Wunch) 1. Align the laser for max modulation (=> on interferometer axis) 2. Locate IR axis (input & exit apertures) on the laser axis 3. Iteratively adjust the apertures and fixed CC for best ILS 2. Karlsruhe method (Hase, Blumenstock) 1. Align IR axis (ie apertures) to centre on the Haidinger fringes at large OPD 2. Align fixed corner cube to centre fringes near ZPD 3. Align laser for either max modulation or centre on IR axis

Align 5.4 Perfect alignment

FOV 3mrad off axis

Fixed CC shear 0.05mm

Align the IR 1: fringes u Illuminate entrance stop with beam expanded HeNe laser l Vellum on aperture to diffuse beam u Fold exit beam upwards and view image at exit stop focus with eyepiece or USB microscope l Alternative: remove exit OAP and view through external telescope u Observe fringes and image of entrance stop while adjusting entrance stop position at mid – long OPD u Observe fringes around ZPD while adjusting fixed CC position

Viewing fringes Small OPD: shear offset aligned 50 cm OPD: FOV offset aligned

Align the IR 2: exit stop u Mount a telescope in the scanner arm and view the entrance and exit stops l First focus it to infinity => focuses parallel beam u Adjust exit stop to centre on the entrance stop l It is normally one setting larger u After laser alignment, can also check alignment of laser relative to stops Misaligned Co-aligned, with laser

Align the laser u Make laser path parallel to the interferometer axis l Adjust input mirrors/prisms, and detector collection OAP u Two methods 1. Maximise modulation (laser interferogram) at max OPD 2. Coalign to the IR beam after IR alignment l Via Haidinger fringes or apertures.

Equipment A.Telescope and improvised holder B.Rail to hold the telescope above the scanner arm of the IFS-125 C.USB microscope D.Periscope eyepiece E.Periscope mount and flat reflecting mirror F.HeNe laser with beam expander and mount G.HeNe Laser power supply H.Vellum and paper for blocking laser beams

Laser and fringe viewer HeNe laser Folding mirror and eyepiece/microscope holer

Telescope mounted in scanner arm

USB microscope US Microscope: typical specs: Main Features: - Adjustable LEDs for object illumination Megapixel lens - Measurement function using enclosed software - Snapshot and video recording function - Alloy stand and stand alone capture button - Tacton rubber touch housing - 20x or 200x magnification - Fine slider adjustment for brightness, hue, saturation and sharpness, plus a black and white display mode - VMS-001 USB Microscope is supplied with Windows and Mac drivers. For software for side-loading CD drives just visit the download centre - The 1.3 megapixel CMOS lens can be adjusted to point in any direction using the movable arm - Simply install the drivers, plug the microscope into your computer via the USB and you're ready to go

The result

But wait, we have a problem... u 3 Wollongong, 3 different ILSs ! u Absolute retrieved ILS / ME depends on: l Pressure HCl in cell l Pressure air in cell l Hitran linewidths and strengths u Measurement of one cell does NOT define the ILS / ME u Refer to Frank Hase’s telecon paper from May 2013 l Some cells appear to have air in them (up to 1 mb?) l Retrieved ME depends on choice of P total, P HCl in linefit l Frank uses C2H2 cell to define ILS and ME u Sensitivity:  P HCl =1 mb =>  ME = 5%  ME = 1%=>  X CO2 = 0.01% = 0.04 ppm

The end u We need to agree on how to manage this u Await Frank’s analysis of all cells and updating his telecon paper u Thanks especially to John Robinson, NIWA, for the tips on implementing Frank and Thomas’s alignment procedure.

IFS125 interferometer IR beam Laser

IFS125 interferometer