Mobile phase is a gas! Stationary phase could be anything but a gas

Slides:



Advertisements
Similar presentations
Gas Chromatography.
Advertisements

Gas Chromatography Introduction 1.) Gas Chromatography
Gas Chromatography Lecture 36.
Gas Chromatography Vaporization of sample Gas-solid
Gas Chromatography 427 PHC.
Gas Chromatography & Gas-Liquid Chromatography
GAS CHROMATOGRAPHY ENVE 202 Dr. Aslıhan Kerç.
GAS CHROMATOGRAPHY.
Lab Methods Day June 25, 2014 Gas Chromatography
4. Advances in Gas Chromatography. Topics covered capillary columns headspace analysis solid phase micro-extraction.
The geometry of capillary columns is fairly simple, consisting of length, internal diameter, and stationary phase thickness. Nevertheless, there are endless.
Gas Chromatography.
Dehydration of 2-Butanol and Dehydrobromination of 2-Bromobutane;
Capillary Columns Nicholas H. Snow Department of Chemistry, Seton Hall University, 400 South Orange Avenue, South Orange, NJ 07079
Figure Figure 2 Chem 334 Expt. #3
Intro to Chromatographic Separations Chap 26. Originally based on separation and identification by color Originally based on separation and identification.
Gas Chromatography. Mobile phase: Inert gas such as N2 or He. Mobile phase: Inert gas such as N2 or He. Stationary phase: May be solid (GSC) or Stationary.
Introduction to Gas Chromatography Written by Bette Kreuz Produced by Ruth Dusenbery University of Michigan-Dearborn 2000.
Gas Liquid Chromatography
CHAPTER 29 Supercritical Fluid Chromatography The mobile phase is a supercritical fluid (a fluid above its critical T and critical pressure) Supercritical.
Types of Chromatography Liquid chromatography versus gas chromatography? Applications? Volatile  GC Non-volatile  LC.
Chromatography Year 12.
Effect of Packing Particle Size on Plate Height. Resolution between 2 adjacent peaks.
LECTURE 4: CHROMATOGRAPHY Content: - Chromatographic separation - classifying analytical separations; column chromatography, planar chromatography - gas.
Chemical Ideas 7.6 Chromatography. The general principle. Use – to separate and identify components of mixtures. Several different types - paper, thin.
Gas Chromatography Carrier Gas Flow Control Injector Port Column Column Oven Detector Recorder The GC system consists of gas supplies for the mobile phase.
GAS – LIQUID CHROMATOGRAPHY BY M. GLORY HEPSIBAH M.PHARM (PH. ANALYSIS) BHARAT COLLEGE OF PHARMACY.
Chapter 27 Gas Chromatography 1. Principles
Chromatography. Chromatography is a separation technique in which sample components distributes themselves between two immiscible phases (mobile phase.
SUPER CRITICAL FLUID CHROMATOGRAPHY
In gas-liquid chromatography (g.l.c.) a long tube contains the chromatography material. The tube is usually coiled so that it takes up less space.
PAS-Intro1 Dated 24th Aug Copyright © Siemens AG 1999 All Rights Reserved Siemens Advanced Engineering Pte Ltd Process Analytical Services Division.
Separation Science Differences in IMFs can be used to separate chemical substances for further analysis. Differences in IMFs can be used to separate chemical.
animation.php.
GC Advantages 1. Very Large N (Very Long Columns) 2. No Packing Material (A=0) 3. Simple Mobile Phase (Compressed Gas) 4. Universal Detectors (FID) 5.
INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER 12 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
An Introduction to Chromatography What is chromatography? The separation of a mixture by distribution of its components between a mobile and stationary.
History of Chromatography n Early LC carried out in glass columns n diameters: 1-5 cm n lengths: cm n Size of solid stationary phase n diameters:
GAS LIQUID CHROMATOGRAPHY
GAS CHROMATOGRAPHY (GC). GAS CHROMATOGRAPHY Sample is injected (using a syringe) into the injection port. Sample vaporizes and is forced into the column.
Gas Chromatography Lecture 38.
Gas Chromatography Experiment. Gas Chromatography - Gas Chromatography (GC) is a common technique used to separate and identify volatile organic compounds.
Lecture 7 Chromatography Intro!.
By: Arafath and Nick. What is it  Chromatography is a technique that is used to separate the substances present in a mixture.  It is widely used to.
Basic Gas Chromatography. History Separation of dyes by Runge Separation of plant pigments by Tswett Theoretical gc (Martin & Synge)
Created with MindGenius Business 2005® Instrumentation (1) Instrumentation (1) Origins Preparative column chromatography (Glass columns) Internal diameter:
GAS CHROMATOGRAPHY In gas chromatography (GC), the sample is injected onto the head of a chromatographic column and immediately vaporized. The components.
Gas Chromatography An Introduction. What is it? Gas chromatography – (gas-liquid chromatography) involves a sample being vaporized and injected onto the.
Chapter 32 Gas Chromatography. In gas chromatography, the components of a vaporized sample are separated by being distributed between a mobile gaseous.
Gas Chromatography. In Gas Chromatography (GC), a gaseous mobile phase transports a gaseous solute through a long, thin column containing solid or liquid.
Gas Chromatography Compiled by : Imanuelle Orchidea
Gas Chromatography Oleh : Rohayati, S.Pd SMK Negeri 13 Bandung.
Chromatography.
Gas Chromatography Gas Chromatography Effendy De Lux Putra.
Experiments in Analytical Chemistry
Reading assignment: section 26E(p781) Chapter 26 # 2, 3, 14, 15, 16 Chapter 27 # 7(a,d,f), 22, 23, 24, 25.
GAS CHROMATOGRAPHY Mr. Maywan Hariono.  One of chromatography technique that use a gas as a carrier (mobile phase)  The separation is based on the volatility.
Overview Determining Identity Quantitative Analysis
Chemical Ideas 7.6 Chromatography.
GAS CHROMATOGRAPH. GAS CHROMATOGRAPH Principle: Which type of compounds can analyse? Volatile Thermostable.
High Performance Liquid Chromatography (HPLC)
Chapter 27 Gas Chromatography
An Introduction to Chromatographic Separations
BASICS OF SUPERCRITICAL FLUID CHROMATOGRAPHY
VAPOUR PHASE CHROMATOGRAPHY
What is chromatography?
Chromatography Daheeya Alenazi.
GAS CHROMATOGRAPHY.
High performance liquid chromatography (HPLC)
32A-2 Sample Injection system
Presentation transcript:

Mobile phase is a gas! Stationary phase could be anything but a gas GAS CHROMATOGRAPHY Mobile phase is a gas! Stationary phase could be anything but a gas

Gas Chromatography (GC) GC is currently one of the most popular methods for separating and analyzing compounds. This is due to its high resolution, low limits of detection, speed, accuracy and reproducibility. GC can be applied to the separation of any compound that is either naturally volatile (i.e., readily goes into the gas phase) or can be converted to a volatile derivative. This makes GC useful in the separation of a number of small organic and inorganic compounds (They can be big compounds if you can make them small before separation!) A simple GC system consists of: 1. Gas source (with pressure and flow regulators) 2. Injector or sample application system (sample inlet) 3. Chromatographic column (with oven for temperature control) 4. Detector & computer or recorder

Carrier gas: He (common), N2, H2 Pinlet 10-50 psig Flow = 25-150 mL/min packed column Flow = 1-25 mL/min open tubular column Column: 2-100 m coiled stainless steel/glass/Teflon/fused silica, packed vs. unpacked Oven: 0-400 °C ~ average boiling point of sample Accurate to <1 °C Detectors: FID, TCD, ECD, NPD, FPD, AED, PID, MSD. (SINGLE OR TANDEM)

Filling sampling loop Introduction of sample into column

Mobile Phase GC separates solutes based on their different interactions between mobile and stationary phases. solute’s retention is determined mostly by its vapor pressure and volatility solute’s retention is controlled by its interaction with the stationary phase Carrier gas – main purpose of the gas in GC is to move the solutes along the column, mobile phase is often referred to as carrier gas (MUST BE INERT!). Common carrier gas: include He, Ar, H2, N2 Carrier Gas or Mobile phase does not affect solute retention, but does affect: 1.) Desired efficiency for the GC System (Van Deemter!) - low molecular weight gases (He, H2)  larger diffusion coefficients - low molecular weight gases  faster, more efficient separations 2.) Stability of column and solutes - H2 or O2 can react with functional groups on solutes and stationary phase or with surfaces of the injector, connections and detector 3.) Response of the detector - thermal conductivity detector requires H2 or He - other detectors require specific carrier gas  compatibility

Mobile Phases Stationary phase in GC is the main factor determining the selectivity and retention of solutes. There are three types of stationary phases used in GC: Solid adsorbents Liquids coated on solid supports Bonded-phase supports 1.) Gas-solid chromatography (GSC) - same material is used as both the stationary phase and support material - common adsorbents include: alumina molecular sieves (crystalline aluminosilicates [zeolites] and clay) silica active carbon Magnified Pores in activated carbon

Gas-Solid Chromatography Advantages: - long column lifetimes - ability to retain and separate some compounds not easily resolved by other GC methods geometrical isomers permanent gases Disadvantages: - very strong retention of low volatility or polar solutes - catalytic changes that can occur on GSC supports - GSC supports have a range of chemical and physical environments different strength retention sites non-symmetrical peaks variable retention times

Gas-Liquid Chromatography 2.) Gas-liquid chromatography (GLC) LIKE DISSOLVES LIKE! - stationary phase is a liquid coated on a solid support - over 400 liquid stationary phases available for GLC - material range from polymers (polysiloxanes, polyesters, polyethylene glycols) to fluorocarbons, and liquid crystals etc. Based on polarity, of the 400 phases available only 6-12 are needed for most separations. The routinely recommended phases are listed below:

Gas-Liquid Chromatography Preparing a stationary phase for GLC: - slurry of the desired liquid phase and solvent is made with a solid support solid support is usually diatomaceous earth (fossilized shells of ancient aquatic algae (diatoms), silica-based material) - solvent is evaporated off, coating the liquid stationary phase on the support - the resulting material is then packed into the column Disadvantages: - liquid may slowly bleed off with time especially if high temperatures are used contribute to background change characteristics of the column with time

Cross-linked phases 3.) Bonded-Phase Gas chromatography - covalently attach stationary phase to the solid support material - avoids column bleeding in GLC - bonded phases are prepared by reacting the desired phase with a silica-surface reactions form an Si-O-Si bond between the stationary phase and support reactions form an Si-C-C-Si bond between the stationary phase and support - many bonded phases exist, but most separations can be formed with the following phases: Dimethylpolysiloxane Methyl(phenyl)polysiloxane Polyethylene glycol (Carbowax 20M) advantages: - more stable than coated liquid phases - can be placed on support with thinner and more uniform thickness than liquid phases

Column support large sample capacity smaller sample size There are two main types of supports used in GC: Packed columns large sample capacity preparative work Capillary (open-tubular) columns higher efficiency smaller sample size analytical applications

Pushing your way around A common problem to all chromatographic techniques is that in any one sample there may be many solutes present, each retained by the column to a different degree: Best separation and limits of detection are usually obtained with solutes with k’ values of 2-10 Difficult to find one condition that elutes all solutes in this k’ range  general elution problem Gradient elution - change column condition with time which changes retention of solutes to overcome general elution problem 1) Column 2) Temperature program 3) Flow

Pushing your way around Temperature Programming – changing the temperature on the column with time to simulate gradient elution in GC since a solute’s retention in GC is related to its volatility. ISOTHERMAL Column temp. 120oC Programmed temp. (30oC to 180oC) (5o/min) Temperature programming is usually done either in a stepwise change, a linear change or a combination of several linear changes. A single linear change or ramp is the most common

Detecting your peaks Thermal Conductivity Detector (TCD) The following devices are common types of GC detectors: 1. Thermal Conductivity Detector (TCD) 2. Flame Ionization Detector (FID) 3. Nitrogen-phosphorus Detector 4. Electron Capture Detector (ECD) 5. Mass Spectrometers The choice of detector will depend on the analyte and how the GC method is being used (i.e., analytical or preparative scale) Thermal Conductivity Detector (TCD) - hot-wire detector - first universal detector developed for GC Process - measures a bulk property of the mobile phase leaving the column. - measures ability to conduct heat away from a hot-wire (i.e., thermal conductivity) - thermal conductivity changes with presence of other components in the mobile phase

Design - based on electronic circuit known as a Wheatstone bridge. - circuit consists of an arrangement of four resistors with a fixed current applied to them. - thermal conductivity changes with presence of other components in the mobile phase. - the voltage between points (+) and (-) will be zero as long as the resistances in the different arms of the circuit are properly balanced as solute emerge from column: change in thermal conductivity  change in amount of heat removed from resistor  change in resistor’s temperature and resistance  change in voltage difference between points (+) and (-). one resistor in contact with mobile phase leaving column another in contact with reference stream of pure mobile phase

much higher then other GC detectors Considerations - mobile phase must have very different thermal conductivity then solutes being separated. - most compounds separated in GC have thermal conductivity of about 1-4X10-5. - H2 and He are carrier gases with significantly different thermal conductivity values (6-10 times greater than those of organic compounds). - H2 reacts with metal oxides present on the resistors, so not used. - Other carrier gases have a similar thermal conductivity as those of organic compounds. Advantages: - truly universal detector applicable to the detection of any compound in GC - non-destructive useful for detecting compounds from preparative-scale columns useful in combination with other types of GC detectors Disadvantages: - detect mobile phase impurities - sensitive to changes in flow-rates - limit of detection ~ 10-7 M much higher then other GC detectors

Flame Ionization Detector (FID) - most common type of GC detector - “universal” detector capable of measuring the presence of almost any organic Principle of operation: - measures the production of ions when a solute is burned in a flame. - ions are collected at an electrode to create a current Advantages: - universal detector for organics - doesn’t respond to common inorganic compounds - mobile phase impurities not detected - carrier gases not detected - limit of detection: FID is 1000x better than TCD - linear and dynamic range better than TCD Disadvantage: - destructive detector

Nitrogen-Phosphorus Detector (NPD) - used for detecting nitrogen- or phosphorus containing compounds - also known as alkali flame ionization detector or thermionic detector (TID) Principle of Operation - same basic principal as FID - measures production of ions when a solute is burned in a flame - ions are collected at an electrode to create a current - contains a small amount of alkali metal vapor in the flame - enhances the formation of ions from nitrogen- and phosphorus- containing compounds Alkali Bead

Nitrogen-Phosphorus Detector (NPD) Advantages: - useful for environmental testing detection of organophosphate pesticides - useful for drug analysis determination of amine-containing or basic drugs - Like FID, does not detect common mobile phase impurities or carrier gases - limit of detection: NPD is 500x better than FID in detecting nitrogen- and phosphorus- containing compounds - NPD more sensitive to other heterocompounds, such as sulfur-, halogen-, and arsenic- containing molecules Disadvantages: - destructive detector - NPD is less sensitive to organic compounds compared to FID

Electron Capture Detector (ECD) - radioactive decay-based detector - selective for compounds containing electronegative atoms, such as halogens Principle of Operation - based on the capture of electrons by electronegative atoms in a molecule - electrons are produced by ionization of the carrier gas with a radioactive source such as 63Ni ( emitter) - in absence of solute, steady stream of these electrons is produced - electrons go to collector electrode where they produce a current - compounds with electronegative atoms capture electrons, reducing current Advantages: - useful for environmental testing detection of chlorinated pesticides or herbicides detection of polynuclear aromatic carcinogens detection of organometallic compounds - selective for halogen- (I, Br, Cl, F), nitro-, and sulfur-containing compounds - detects polynuclear aromatic compounds, anhydrides and conjugated carbonyl compounds

Atomic emission detector (AED)

Flame Photometric Detector (FPD) - selective for compounds containing sulfur and phosphorus (could be used for other elements by suitable changes in flame condition and filter) Principle of Operation - Decompose of analytes using H2/Air flame - Products emit bands of radiation S2* → S2 + h (394 nm) Excited state S2* species could result from several two or three body collision reactions: H + H + S2 → S2* S + S → S2* S + S + M → S2* + M (third body)

Flame Photometric Detector (FPD) For P PO + H + M → HPO* + M PO + OH + H2 → HPO* + H2O 526 nm is given due to HPO*

Flame Photometric Detector (FPD)

Mass Spectrometry Detector

Concentration-sensitive vs Mass-sensitive Detector Concentration-sensitive detector responds to the concentration of the analyte in the mobile phase (the mobile phase is directly involved in the signal formation). (e.g. TCD and ECD). Mass-sensitive detector responds to the number of molecules or ions that contact with the detector (the mobile phase is not directly involved in the signal formation). (e.g. FID, AED, and FPD).

GC Column and Stationary Phase

Adsorption on column packing and capillary walls Silanol groups have strong affinity for polar organic molecules.

Support materials can be deactivated by silanization with dimethylchlorosilane (DMCS).

Stationary Phase Low volatility Thermal stability Chemical inertness Solvent characteristic (k’ and  etc)

Qualitative Analysis – Retention Index (I) I for an analyte is a measure of the rate at which it is carried through a column compared with the rate of movement of two normal alkanes one that moves faster than the analyte and the other that moves more slowly. I of alkanes, by definition, is 100 times the number of carbon atom they contain regardless of the column packing, temperature or other conditions e.g. butane, I = 400 pentane, I = 500

Quantitative Analysis Using GC