Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.

Slides:



Advertisements
Similar presentations
RADIOGRAPHIC GRIDS.
Advertisements

Producing an X-ray Exposure
Technical Aspects: the machine the image
Advanced Biomedical Imaging Lecture 3
[Radiography] Technique - Exposure Factors
Technique Guidance Systems
Control of Scatter Radiation
Chapter 11 Prime Factors.
Kilovolt Peak (kVp) kVp determine the quality of the beam.
Formation Characteristics
Quality Control in Diagnostic Radiology
Primary Exposure Factors IV
Chapter 18 Radiographic Exposure
IMAGE FORMATION  Introduction  The Invisible and Visible Image  Image Characteristics.
Radiographic Quality Chapter 5.
Topic 3 Selection of kV High kV scattered radiation is energetic. Angle of rays is same as those of primary beam and is not deflected very much. Follows.
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Radiographic Grids II By Professor Stelmark.
Resident Physics Lectures
Chapter 6, Fauber Supplement-Chapter 14, Bushong
The Grid Kyle Thornton DMI 50B.
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Chapter 19 & 20 Image Quality & Techniques
Analyzing the Image Density. Density Overall blackening of the image.
Advanced Biomedical Imaging
Technical Artifacts and Errors
Copyright © 2012, 2006, 2000, 1996 by Saunders, an imprint of Elsevier Inc. Chapter 8 Dental X-Ray Image Characteristics.
Radiographic Technique Evaluation
Conventional and Computed Tomography
Radiographic Grids I By Professor Stelmark.
Quality Control Rad T 110.
Grids, use of them and selection/problems, etc
Control of Scatter Radiation
1 Components of Image Quality & Radiographic Artifacts Radiologic Technology A SPRING 2012.
Quality Control.
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
IMAGE QUALITY REVIEW RT
Dr Mohamed El Safwany, MD.
RAD 354 Chapt. 11 Control of Scatter Break down into: Those that reduce pt. dose and those that are geometrical in nature and those NOT! 3 factors affecting.
Components of Image Quality & Radiographic Artifacts Radiologic Technology A Spring 2010 Final Final.
Image Quality Radiographic Resolution.
Components of Image Quality & Radiographic Artifacts
Prime Exposure Factors 1
Factors affecting CT image RAD
Exposure Factors or Prime Factors
Preparing Variable kVp Technique Charts By Prof. Stelmark.
PRINCIPLES OF TECHNIQUE AND EXPOSURE
Radiographic Quality Visibility and Sharpness
Radiographic Equipment
Image Quality Q&A. RAPHEX Diagnostic Question 2001 D9: The imaging system which is best for visualizing small high contrast objects is: A. Computed tomography.
Grids George David Associate Professor Department of Radiology
Radiographic Grids I By Professor Stelmark.
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
Grid Performance By Prof. Stelmark.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Tube Exposure Factors Math Technique Contrast and Density
Sergeo Guilbaud School of Radiologic Sciences
Exposure Factors or Prime Factors
Producing An Image RVT: Chapter 6. Learning Objectives: Chapter 6 Understand the 4 factors of radiographic exposure and how each impacts the production.
Chapters 21 & 22.
Resident Physics Lectures
Resident Physics Lectures (Year 1)
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures (Year 1)
What makes a Quality Image
Chapter 5, Fauber Supplement-Chapter 11, Bushong
Artifacts and Errors.
GRIDS PRESENTER-Dr.Vishwanath patil PG resident
Presentation transcript:

Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University

Radiographic Image Quality Image quality parameters (photographic properties) Density Density Contrast Contrast Resolution (Image details) Resolution (Image details) Signals / Noise (Grainy image) Signals / Noise (Grainy image) Lack of artifacts Lack of artifacts

1. Density Overall blackening of the image. overexposed. If the image is too dark, it has too much density, overexposed. If the image is too light, it lacks density, underexposed. underexposed.

2. Contrast Difference between densities of adjacent structures Controlled by energy of the beam. µα1/E 3 Depends on tissue and type of film Depends on tissue and type of film High contrast is too black & white. Low contrast is too gray.

Contrast & Density Contrast & density are controlled by exposure factors and film processing.

X ray control

X-ray Exposure Factors Settings controlled by the operator used to produce the radiograph. Will influence the amount of patient exposure Type of x-ray machine and screen film combination will impact the exposure factors.

Increasing voltage (kV) increases maximal energy of x-rays produced, & ↑ penetration of beam,film exposure, making the film darker, ↓ pt dose & ↓ exp time but ↓ contrast. (obese pt) Increasing voltage (kV) increases maximal energy of x-rays produced, & ↑ penetration of beam, film exposure, making the film darker, ↓ pt dose & ↓ exp time but ↓ contrast. (obese pt) Increasing mAs increases the amount of x-rays produced, exposure, and signals, making the film darker, ↓ exp time. Increasing mAs increases the amount of x-rays produced, exposure, and signals, making the film darker, ↓ exp time. Film dose is α Kv 4 mAS. Film dose is α Kv 4 mAS. Exposure factors

A small focal spot produces sharper images. A small focal spot produces sharper images. A large focal spot can tolerate more heat but increase image unsharpness. (see lecture 3). A large focal spot can tolerate more heat but increase image unsharpness. (see lecture 3). Exp time: ↓ by ↑ Kv & large focal spot (↑blurring) Exp time: ↓ by ↑ Kv & large focal spot (↑blurring)

mAs Product of mA times the exposure time 100 seconds = 10 mAs seconds = 10 mAs As long as the product of the mA and time are the same, the exposure should be the same if the Ma and timer calibration is accurate.

quantity mAs determines the quantity of x-ray photons. density mAs establishes the density on the image. Density is the blackness of the image. It takes a change of 25% to 30% in mAs to make a significant change in the density of the image.

Change the mAs and Change the Density of the Image decreased. The center image has the mAs decreased by 25%. Density is decreased. increased Image on the right has the mAs increased by 25% above the image on the left. Density is increased.

Wrong mAs under exposed This image is under exposed. Not enough mAs was used & density reduced overexposed This image is overexposed for the patella. Too much mAs was used & density increased.

30-50 Rule Relatively large changes in mAs are needed to change the density of the image. It takes a change of 20 to 30% to make any change in the density of the image. It needs a change by 50% to make a significant change. This is the Rule

Changing mAs mAs As mAs increased; the density of the image changes from under exposed to over exposed.

Over exposed This was taken with 60 mAs. To correct the density of the image, 30 mAs would be used. So mAs reduced by 50%.

Under exposed This image was taken at 30 mAs. To darken the image, 60 mAs would be used. To correct the density of the films double the mAs.

Grid Ratio and Frequency The higher the ratio and frequency, the more radiation is needed to produce the image. Low grid ratio is limited to low kVp. A 5:1 grid is limited to below 90 kVp. High ratio grids can be used up to 125 kVp.

Grid Ratio and Frequency Low frequency grids have prominent grid lines on image. They are used in true Buckies. The grid moves during the exposure to blur the lines. High frequency grids produce almost invisible lines and do not need to move during the exposure.

For imaging of lumbar spine without the grid: – 8 74 kVp used – kVp used to produce the image with the 5:1 grid. (doubling mAs) – The conversion factor would be 2 for adjusting the mAs for the 5:1 grid Grid Ratio and Technique

The 5:1 grid required doubling the mAs. The conversion factor is determined by dividing the new mAs by the old mAs. What would be the conversion factor for the 10:1 grid? Old mAs = 8 New mAs = 40 What would be the conversion factor for the 10:1 grid? Old mAs = 8 New mAs = 40 Grid Ratio and Technique

No Grid Used Lumbar Spine needs the use of a grid. why? There is too much scatter radiation due to the size of the body part Exposure factors needed to be adjusted to visualize this area.

Low Ratio & Frequency Grid For the A-P Lumbar Spine, a 5:1 with 80 grid lines will remove much of the scatter radiation. Going from Non-Grid to a 5:1 ratio requires doubling the mAs. Grid lines are very noticeable.

High Ratio & Frequency Grid A 10:1 grid with 100 lines will provide excellent clean-up of scatter radiation. Grid lines are nearly invisible. But, requires 5 times more radiation.

Grid Cut-off When high frequency and ratio focused grids are used, the lead strips must be perfectly aligned with the primary beam. A minor misalignment will result in the grid removing primary radiation. Misalignment of more than 2° will result in grid cutoff.

Grid Cut-off Is the density of the image of both knees the same? This is an example of grid cut-off. Some of the primary beam is being removed by the grid.

Grid Cut-off causes If grid lines are not parallel to primary radiation due to – tube being too close – too far from the grid,

– If the tube is angled against the grid lines. not perpendicular to the beam (mos – If grid is not perpendicular to the beam (most common problem).

backwards – If grid is backwards, only center of beam will pass though the grid. So proper alignment must be maintained. So proper alignment must be maintained.

Air-gap Vs Grids Using air gap not grid, Radiation exposure reduced five folds With no loss of image quality.

3. Resolution Resolution (image details): Minimum distance between two points that can be recognized as being separate.

Noise (grainy image): variation in no of x ray photons absorbed 4. Noise

Artifact is the appearance of signal in an image location not representative of actual properties of object. 5. Lack of artifacts

Questions 1.What are main parameters that control image quality? 2. Enumerate causes of grid cut-off? 3. Define contrast, density and artifact?