Grayscale Calibration Grayscale Standard Display Function

Slides:



Advertisements
Similar presentations
Filtration based on Color distance
Advertisements

Digital Image Processing
Image Data Representations and Standards
Creation of a digital image from an analog signal. Analog-Digital Converter (ADC)
RAD 354 Chapt. 30 Digital QC Performance Assessment Standards Luminance Meter Digital Display QC QC by Technologist.
Quality Assurance and Digital Radiography
Radiology Understanding Display Characteristics What You Should Know When Viewing Images from PACS.
Digital Imaging and Image Analysis
 Image Characteristics  Image Digitization Spatial domain Intensity domain 1.
Image Processing IB Paper 8 – Part A Ognjen Arandjelović Ognjen Arandjelović
Imaging Science Fundamentals Chester F. Carlson Center for Imaging Science The Properties of Images and Imaging Devices Group II of the Imaging Chain.
Image Enhancement To process an image so that the result is more suitable than the original image for a specific application. Spatial domain methods and.
CS443: Digital Imaging and Multimedia Point Operations on Digital Images Spring 2008 Ahmed Elgammal Dept. of Computer Science Rutgers University Spring.
The PAC System By Andi Sherman CIS Spring Intro to Radiology Radiology is a fast growing area of the health profession in most hospitals and private.
Medical Image Display Bradley Hemminger School of Information and Library Science Department of Radiology, University of North Carolina, Chapel Hill
Graphics File Formats. 2 Graphics Data n Vector data –Lines –Polygons –Curves n Bitmap data –Array of pixels –Numerical values corresponding to gray-
Display Issues Ed Angel Professor of Computer Science, Electrical and Computer Engineering, and Media Arts University of New Mexico.
Computer Vision Lecture 3: Digital Images
Viewing the Digital Image
By : Alanoud Al Saleh. What is PACS quality control ? The PACS monitor quality control (QC) program objectives are:  to ensure consistent display performance.
Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry
Copyright © Texas Education Agency, All rights reserved.1 Introduction to Digital Cameras Principles of Information Technology.
 A monitor or display is an electronic visual display for computers.  The monitor consists of : o the display device o circuitry o enclosure The display.
9/14/04© University of Wisconsin, CS559 Spring 2004 Last Time Intensity perception – the importance of ratios Dynamic Range – what it means and some of.
Digital Image Characteristic
Colour Digital Multimedia, 2nd edition Nigel Chapman & Jenny Chapman
Digital Multimedia, 2nd edition Nigel Chapman & Jenny Chapman Chapter 6 This presentation © 2004, MacAvon Media Productions Colour.
Spectral contrast enhancement
Disturbing pixel faults Justification for not incorporation single pixels failure test procedure as part of type approval test within UN Reg.46 document.
DICOM Singapore Seminar:
February 8, 2005IHE Europe Educational Event 1 Integrating the Healthcare Enterprise Consistent Presentation of Images Integration Profile Integrating.
Page 18/30/2015 CSE 40373/60373: Multimedia Systems 4.2 Color Models in Images  Colors models and spaces used for stored, displayed, and printed images.
An automated image prescreening tool for a printer qualification process by † Du-Yong Ng and ‡ Jan P. Allebach † Lexmark International Inc. ‡ School of.
Any questions about the current assignment? (I’ll do my best to help!)
CSCI-235 Micro-Computers in Science Hardware Part II.
Vector vs. Bitmap
Consistent Presentation of Images - Integration Profile Ellie Avraham Kodak Health Imaging IHE Planning and Technical Committees.
The Leeds Teaching Hospitals NHS Trust Is Grey the New Black? Image Display Optimisation & Consistency for Softcopy Reporting Mark Barnfield St James’s.
September 5, 2013Computer Vision Lecture 2: Digital Images 1 Computer Vision A simple two-stage model of computer vision: Image processing Scene analysis.
Consistent Presentation of Images
1 Introduction to Computer Graphics with WebGL Ed Angel Professor Emeritus of Computer Science Founding Director, Arts, Research, Technology and Science.
1 Chapter 2: Color Basics. 2 What is light?  EM wave, radiation  Visible light has a spectrum wavelength from 400 – 780 nm.  Light can be composed.
Lecture 3 The Digital Image – Part I - Single Channel Data 12 September
Image Display Systems Optimizing the Environment Stephen G Davies Consultant Radiologist Royal Glamorgan Hospital.
CSC361/ Digital Media Burg/Wong
June 28-29, 2005IHE Interoperability Workshop 1 Integrating the Healthcare Enterprise Consistent Presentation of Images Integrating the Healthcare Enterprise.
The Reason Tone Curves Are The Way They Are. Tone Curves in a common imaging chain.
Page 1 Visual calibration for (mobile) devices Lode De Paepe GOAL: Calibrate the luminance response of a display (transfer function of the display.
02-Gray Scale Control TTF. A TTF tells us how an imaging device relates the gray level of the input to the gray level of the output. P L Luminance, L.
More digital reading explaining LUT RT 244 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
Consistent Presentation of Images Profile IHE North America Webinar Series 2008 Chris Lindop IHE Radiology GE Healthcare.
IHE Workshop – June 2006What IHE Delivers 1 Ellie Avraham Kodak Health Group IHE Planning and Technical Committees Consistent Presentation of Images Profile.
Digital Reformatting of Photographic Formats -2- Aaron Choate Digital Library Production Services The University of Texas Libraries.
Applying Pixel Values to Digital Images
3-1 Chapter 3: Image Display The goodness of display of an image depends on (a) Image quality: i) Spatial resolution, ii) Quantization (b) Display device:
David Luebke 1 2/5/2016 Color CS 445/645 Introduction to Computer Graphics David Luebke, Spring 2003.
David Luebke2/23/2016 CS 551 / 645: Introductory Computer Graphics Color Continued Clipping in 3D.
More digital 244 wk 12 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of Medicine Atlanta, GA,
ECE 638: Principles of Digital Color Imaging Systems Lecture 12: Characterization of Illuminants and Nonlinear Response of Human Visual System.
Graphics and Image Data Representations 1. Q1 How images are represented in a computer system? 2.
Color Measurement and Reproduction Eric Dubois. How Can We Specify a Color Numerically? What measurements do we need to take of a colored light to uniquely.
Introduction to Digital Image Analysis Kurt Thorn NIC.
Click to edit Master title style Click to edit Master text styles Second level Third level Fourth level Fifth level 1 Integrated Color Solutions A presentation.
Consistent Presentation of Images Integration Profile
Initial Display Alternatives and Scientific Visualization
Ellie Avraham Kodak Health Imaging
Perception and Measurement of Light, Color, and Appearance
Video System TTFs Part (I): Basic Design Strategy.
Consistent Presentation of Images
Presentation transcript:

Grayscale Calibration Grayscale Standard Display Function accuGray™ Grayscale Calibration For Image Display Workstations DICOM Grayscale Standard Display Function accuGray

Imagine the following: Scenario 1 Imagine the following: A trauma patient is admitted at an outpatient clinic or rural clinic at a remote location. Several X-rays are taken and sent to a hospital for consultation.

Scenario 1 How can we make sure that what the physician sees at the clinic matches what is being displayed on the radiologist softcopy workstation at the hospital? ? ? ? ?

Scenario 1 The image is also sent to another radiologist for a second opinion. This radiologist has a diagnostic viewing station from another vendor - a different monitor and video display card. As a matter of fact, he has a new flat panel display instead of a conventional CRT. How do we make sure that the image again looks identical? ? ? ? ? ?

Scenario 2 Everyone misses something among the myriads of pixels that are tape archived for colleagues and lawyers to examine with 20/20 hindsight vision. Perhaps the missed findings were apparent in the image data, but a once-new PACS monitor had lost its ability to display those subtle differences. How can you know? What will you say when they ask what steps you took to assure that the display equipment you used was up to par?

Dicom Standard These issues are addressed the Digital Imaging and Communications in Medicine (DICOM) standard (PS 3.14-2003) Part 14 of this standard deals with grayscale consistency. Jointly developed by American College of Radiology (ACR) & National Electrical Manufacturers Association (NEMA) Standard method for transferring images and associated information Allows images to be displayed consistently on different softcopy and hardcopy devices devices manufactured by various vendors devices having different technologies (CRT, LCD) devices with different characteristics (max. Brightness and/or Density) different ambient light environments.

output luminance = input drive2.5 Display Gamma CRT monitors (and LCD monitors which emulate them) have a fixed power-law relationship between output luminance and input drive signal (gamma) of about 2.5. output luminance = input drive2.5 Standard Windows signals or video signals applied to monitors require an inverse gamma offset to counteract the display’s gamma. An exact inverse correction produces a linear relationship between the driving signal and the resulting light output.

Our eye-brain detection system is not linear Display Gamma Our eye-brain detection system is not linear much less sensitive in the blacks than in the whites why map all black values into luminance changes to display on the monitor when we can’t notice the differences between small black value changes? DDL* values of darker areas mapped into large luminance increments DDL* values of whiter areas mapped to smaller luminance increments Resulting curve is standardized as the DICOM Grayscale Standard Display Function (GSDF) *Digital Driving Level (DDL): digital value applied to a Display System to produce a luminance output

Perceptual Linearization DICOM Grayscale Standard Display Function (GSDF) generates a display function matching the perceptual characteristics of a human observer. Standard developed having human observer view monitor while luminance output of monitor changed in small steps from black to white and noting when they observed a Just Noticeable Difference (JND) in light output as the drive signal is changed. This standard display function creates a perceptually linearized display - equal changes in driving levels yield perceptually equivalent changes in luminance across entire luminance range. From the observer’s point of view, this means that equal steps in brightness sensation represent equal steps in the acquired image data. Smallest luminance change that average person can just perceive under ideal circumstances is approximately 256 shades of gray.

Grayscale Standard Display Function Perceptual Linearization Equal steps in perceived brightness represent equal steps in the acquired image data .01 .1 1 10 100 1000 200 400 600 800 Grayscale Standard Display Function JND Index Different change in absolute luminance Same number of Just Noticeable Difference == same perceived contrast

Perceptual Linearization Most digital radiologic images are windowed to attempt to display 256 different shades of gray at any one time, (256 is approximate JND that humans can perceive) The perceptual linearization implemented with the DICOM Grayscale Standard Display Function (GSDF) ensures that pixel values that are supposed to increase brightness in a linear fashion, say 6, 7, and 8, for example, actually appear that way to the human eye, despite the nonlinearity of our perception.

Perceptual Linearization The DICOM GSDF provides a mechanism to standardize the appearance of images on monitors of different inherent brightness and with different response curves. Also recalibrate monitors whose luminance and response curves have changed over time. Now it is a matter of calibrating the monitor and printers according to this curve.Calibration tools for softcopy display devices include a luminance meter that can be placed on the monitor and allows the light output (luminance) to be measured at multiple digital driving levels (DDLs). This output can be compared with the actual Grayscale Standard Display Function (GSDF) and any differences can be calculated and saved. Once the differences are calculated, a corrective transfer function can be downloaded directly to the video display controller.

Color Lookup Table Video Card Look Up Table Digitized images are made up of a matrix of pixels, each possessing at least three dimensions: two (or more) spatial and one intensity value. The quantized dimensions are stored on the computer as a file of binary numbers. In order to see the image on the computer monitor, the image pixel values must be mapped, one-to-one, to screen pixel values, via a Look Up Table (LUT). The LUT transfer function determines what screen values correspond to image pixel intensity values at all coordinates in the image.

Perceptually equalized Color Lookup Table If each image intensity value is mapped to its corresponding screen value, the LUT is a linear function and can be graphed as a 45° straight line. By modifying the transfer function’s slope & screen mapping, LUT image intensity values may be selectively increased or decreased to perceptually equalize image luminance steps. Linear Perceptually equalized 0.1 1 10 100 50 150 200 250 DDL or P-Values Standard Characteristic Curve

Guidelines for Calibrating Grayscale Monitors Calibrating Displays Guidelines for Calibrating Grayscale Monitors Calibrate each monitor regularly. Calibrate under regular viewing & lighting conditions. Adjust monitor contrast and brightness before/during calibration and never again. Calibrate with always the same background intensity (about 20%, or a value comparable to the average intensity emitted from the monitor). Calibrate to match the standard display function (DICOM).

Grayscale Calibration Calibrating Displays accuGray™ Grayscale Calibration For Image Display Workstations

accuGray™ Grayscale Calibration Four Primary uses of accuGray™ Acceptance Testing of new displays. Calibration to Grayscale and Color standards. Visual Verification of proper operation. Conformance Tracking over time.

accuGray™ Grayscale Calibration Main Display screen

accuGray™ Grayscale Calibration Configuration screen

1. Acceptance Testing of new displays. Test new displays by evaluating critical test patterns and measuring luminance for white, black, and uniformity. You’ll select from these menu items

1. Acceptance Testing of new displays. Display SMPTE Pattern – analyze focus, geometry, smearing, luminance tracking.

1. Acceptance Testing of new displays. Display VeriLUM® Pattern – analyze 256 levels of gray, scan linearity, and focus.

1. Acceptance Testing of new displays. Display User Patterns – analyze display’s capabilities with standard or custom medical test images.

1. Acceptance Testing of new displays. Measure luminance for white, black, and uniformity.

1. Acceptance Testing of new displays. Measure White Level – tests display’s white output capability.

1. Acceptance Testing of new displays. Measure Black Level – tests display’s light output at a proper black level adjustment.

1. Acceptance Testing of new displays. Measure Luminance Uniformity – tests display’s capability to uniformly light screen from center to corners

1. Acceptance Testing of new displays. Measure Luminance – Perform general luminance & chromaticity measurements on a new display.

2. Calibration to Grayscale and Color standards. Calibrate display luminance and chrominance characteristics for accurate picture reproduction. You’ll select from these menu items

2. Calibration to Grayscale and Color standards. Measure Luminance Response for Gamma Correction – Calibrate display’s grayscale tracking to the DICOM standard.

2. Calibration to Grayscale and Color standards. Measure Luminance Response for Gamma Correction – Displays results compared to desired response curve.

2. Calibration to Grayscale and Color standards. Adjust Color Temperature – Calibrate display image white background color for accurate color rendition.

2. Calibration to Grayscale and Color standards. Adjust Color Temperature – View color calibration results in terms of 1931 or 1976 CIE Chromaticity Diagram.

3. Visual Verification of proper operation. Observe test patterns to confirm proper image geometry and optimum picture clarity. You’ll select from these menu items

3. Visual Verification of proper operation. Display SMPTE Pattern – analyze focus, geometry, smearing, luminance tracking.

3. Visual Verification of proper operation. Display VeriLUM® Pattern – analyze 256 levels of gray, scan linearity, and focus.

3. Visual Verification of proper operation. Display User Patterns – analyze display’s capabilities with standard or custom medical test images.

4. Conformance Tracking over time. Insure and document that a display continues to provide a properly calibrated image over its lifetime. You’ll select from these menu items

4. Conformance Tracking over time. Measure Luminance Response for Tracking - Measure a color display’s grayscale tracking and save to measurement history

4. Conformance Tracking over time. Display Tracking History - displays documentation of a color display’s white background color measurement history. Click for graphical Tracking and JND plots that show the display’s performance history

4. Conformance Tracking over time. History Plot – shows the current color Tracking measurement and past performance of the display.

4. Conformance Tracking over time. JND Plot - shows the current JND measurement and past performance of the display.

4. Conformance Tracking over time. Plot Results - provides on-screen and printable documentation of the display’s past and present performance.

4. Conformance Tracking over time. Gamma Correction Table - shows the current gamma curve and gamma history of the display.

4. Conformance Tracking over time. Calibration Luminance Response - shows the current luminance calibration curve and prior calibration curves.

4. Conformance Tracking over time. Tracking Luminance Response - shows the current luminance response and prior response curves.

Benefits of a Properly Calibrated Video Display: Displays a sharper-focused, full resolution image Shows full detail in both the darkest & brightest parts of all scenes Is properly matched to the viewing environment Produces a full range of accurate colors, including flesh tones Minimizes picture artifacts (distortions) Ensures a proper perceptually linearized display Satisfied user (Doctor) Properly diagnosed patient No litigation

Questions? 1-800-SENCORE www.sencore.com