X-Ray Beam Restrictors

Slides:



Advertisements
Similar presentations
Producing an X-ray Exposure
Advertisements

Technical Aspects: the machine the image
Advanced Biomedical Imaging Lecture 3
Control of Scatter Radiation
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Quality Control in Diagnostic Radiology
Radiographic Quality Chapter 5.
Resident Physics Lectures
Chapter 6 Control of Secondary and Scattered Radiation.
Control of Scatter Radiation Beam Restricting Devices
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Radiographic Grids II By Professor Stelmark.
Resident Physics Lectures
Chapter 6, Fauber Supplement-Chapter 14, Bushong
Analyzing the Image Density. Density Overall blackening of the image.
X-ray Beam Composition and Collimation
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING  REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING  One meter is.
Resident Physics Lectures
Lecture 4 Beam restricting Devices, their Use and Maintenance
Factors affecting the X-Ray output
Radiographic Grids I By Professor Stelmark.
Beam Therapy Equipment 3 Patient Treatment and Accessories.
Quality Control Rad T 110.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Radiation Exposures Unit 2
Control of Scatter Radiation
Quality Control.
Radiation Monitoring and Protection
Spot Filming Cinefluorography Fluoro Room Shielding Radiation Areas.
Dr Mohamed El Safwany, MD.
Resident Physics Lectures
RAD 354 Chapt. 11 Control of Scatter Break down into: Those that reduce pt. dose and those that are geometrical in nature and those NOT! 3 factors affecting.
Image Quality Radiographic Resolution.
INVERSE SQUARE LAW. The picture above demonstrates the typical x-ray tube used to produce a point source of x-rays. Then as radiation exits the tube it.
PRINCIPLES OF TECHNIQUE AND EXPOSURE
Beam And Scatter Radiation Restricting Devices DMI 50B Kyle Thornton.
Beam Restricting Devices
Radiographic Equipment
Resident Physics Lectures Christensen, Chapter 4 Basic Interactions Between X-Rays and Matter, Grid Attenuation and Filtration George David Associate.
Part No...., Module No....Lesson No
Grids George David Associate Professor Department of Radiology
Radiographic Grids I By Professor Stelmark.
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
Exposure Factors or Prime Factors
Control of Scatter Radiation Beam Restricting Devices By Prof. Stelmark.
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
RADIATION PROTECTION 04/12/2016.
Diagnostic Equipment Quality Control
Ch 36 Radiation Protection Design
Chapters 21 & 22.
Resident Physics Lectures
Physics Case of the Day - Sunday
Resident Physics Lectures (year 1)
Protection of Patients
Resident Physics Lectures
Factors Effecting the Production
Resident Physics Lectures
Resident Physics Lectures (Year 1)
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures (year 1)
Resident Physics Lectures (Year 1)
Chapter 5, Fauber Supplement-Chapter 11, Bushong
Filters, Collimators and Grids
Fluoroscopy Static Image Recording Systems
Artifacts and Errors.
GRIDS PRESENTER-Dr.Vishwanath patil PG resident
Presentation transcript:

X-Ray Beam Restrictors George David Associate Professor Department of Radiology Medical College of Georgia

Restrictors Types Aperture Diaphragms Cones and Cylinders Collimators

Aperture Diaphragms Lead sheet with hole in center simple large penumbra (shadow) Lead close to focal spot Focal Spot Penumbra Film

Cones & Cylinders Focal Spot Focal Spot Film Film With cone angle as above, this functions as an aperture Reduced Penumbra

Collimators Advantages Configuration Provides adjustable rectangular field fluoro may also have circular field light beam indicates x-ray field Configuration two sets of shutters longitudinal transverse sets may actually have two pairs of shutters operating together One set; two pairs

Collimators Generally include filter & light Focal Spot Filter Lamp Mirror Filter Focal Spot Shutters (only 1 set shown)

Collimator Features Numerical field size indicators Field size is SID dependent 8 10 12 14 17 40” 8 10 12 14 17

Collimator Features Field center indicator (crosshairs) plastic sheet attached to bottom of collimator two black lines drawn on it light projects field center on patient Shutters Crosshairs Bottom of Collimator

Positive Beam Limitation (PBL or Automatic Collimation) senses film size, orientation, & SID source - image distance limits x-ray field to size of film expensive PBL requirement has been dropped for manufacturers PBL does not insure tube & bucky are aligned! Cassette

Collimator Checks X-Ray / Light Field alignment Alignment of x-ray field to bucky Use tubestand centering devices lateral detent longitudinal alignment light Field Size Indicator accuracy PBL accuracy

Collimation & Patient Dose Reduction 10 X 10 cm field has 1/4 the area of a 20 X 20 cm field small change in field size has large effect on volume of irradiated tissue scatter

Scatter Reduction Less scatter = improved image quality quantity of scatter depends on field size primary beam intensity independent of field size

Scatter Reduction as field size increases, so does fraction of scatter photons reaching film scatter peaks at ~ 30cm X 30cm field size smaller field size may require higher technique compensates for loss of density from scatter photon