26 companies have produced nearly 2000 models of pacemakers

Slides:



Advertisements
Similar presentations
Appendix E Pacemakers Gail Walraven, Basic Arrhythmias, Seventh Edition ©2011 by Pearson Education, Inc., Upper Saddle River, NJ.
Advertisements

Dual Chamber Temporary Pacing Operations & Troubleshooting
CARE OF PATIENT ON PACEMAKER. WHAT IS A PACEMAKER? - A cardiac pacemaker is an electronic device that delivers direct stimulation of the heart.
Fundamentals of Cardiac Rhythm Management Devices
EP Testing and Use of Devices in Heart Failure HFSA 2010 Recommendations.
Pacemakers & Pacing in the ED Albury Wodonga Education Program 2014.
ELECTROPHYSIOLOGY LAB
Department of Anesthesia and Perioperative Care advancing health worldwide TM Combined ICD/Pacing Devices What options are available? What’s new in the.
Subcutaneous Implantable Cardioverter Defibrillator (S-ICD)
Pacemaker Malfunctions Even less amusing!. Pacemaker Codes (NASPE/BPEG) Position I IIIII Category Chamber(s) Chamber(s) Response to paced sensed sensing.
The Very Basics of Pacing Glenn Estell Medtronic Pribcipal Clinical Specialist.
Saturday, May 09, 2015Saturday, May 09, 2015Saturday, May 09, 2015Saturday, May 09, Artificial pacemaker and anesthesia Presenter: Abraham sonny.
Pacemaker for beginners KITA yosuke Iizuka Hospital.
EKG Monitoring.
Jennifer Cohen, MD, Heather Costa, PhD, Robert Russo, MD, PhD, Division of Cardiovascular Diseases, Scripps Clinic, La Jolla, CA The MagnaSafe Registry:
Pacemakers and Implantable Cardioverter-Defibrillators
Pacemakers and Implantable Defibrillators
By Corey Thompson.  Thing I hope you learn  History  What they are  There importance to pt who have them  How they relate to us as EMS professionals.
Implantable Cardioverter Defibrillator Rebecca Boduch Biomedical Engineering University of Rhode Island.
Juan Camilo Diaz Cardiac Pacemakers.
By Ahmed M. Shehata Assistant lecturer. INTRODUCTION 30,000,000 patients worldwide have been implanted pacemakers while 3,000,000- 5,000,000 have a Implantable.
Anesthetic Management of Cardiac Rhythm Management Devices N746: Summer 2014 Jennifer Ranieri.
CARDIAC ARRHYTHMIA Charn Sriratanasathavorn, MD,FACC
Pacemakers and Implanted Defibrillators Mike Harlan.
Pacemaker Emergencies Arun Abbi MD Jan 21, Overview Initial approach Pocket Complications Acute complications with placement Nonarrythmic complications.
Single Chamber Temporary Pacing Operations & Troubleshooting.
Following the Path of Least Resistance: The Road Most Traveled
Elsevier items and derived items © 2006 by Elsevier Inc. Chapter 37 Interventions for Clients with Dysrhythmias.
Pacemakers and Implantable Cardioverter Defibrillators Chapter 10
IABP, VAD, ICD & Pacemaker Therapies Diane E. White RN CCRN PhD.
Wireless Cardiac Device Monitoring Presented by: Ashley D. Solomon, RN Nursing 457: Nursing Informatics Fall 2009.
Sick Sinus Syndrome. Description Your sinus node controls the rhythm of the heart. It sends electrical impulses across the atria to the ventricles, making.
Perioperative Care of Patients with Cardiac Rhythm Management Devices
Pacemakers.
Cardiac Pacemakers © D. J. McMahon 2012 rev
Device and Antiarrhythmic Drugs: Advantages and Pitfalls Teresa Menendez Hood, M.D.
ECG in Pacemaker Malfunction
Pacemakers and AICD ’ s. Pacemaker Basics Provides electrical stimuli to cause cardiac contraction when intrinsic cardiac activity is inappropriately.
Copyright © 2011 Delmar, Cengage Learning. ALL RIGHTS RESERVED. Chapter 40 Nursing Care of the Perioperative Client.
Implantable Cardioverter Defibrillator (ICD) Reprogramming Guidelines Lauren Butler.
THE HEART’S ELECTRICAL SYSTEM Marco Perez, MD Center for Inherited Cardiovascular Disease Inherited Cardiac Arrhythmia Clinic June 20, 2013.
Supparerk Prichayudh M.D
Cardiovascular Therapeutic Management 2013
Cardiology for Dr. Pelaez By Sai Kumar Reddy American International Medical University, St.Lucia.
Heart Blocks and Pacing
Staff Anesthesiologist – Allegheny Health Network
Bandar Al Ghamdi, MD,FACC,FHRS,CCDS.
Pacemaker for beginners
Temporary Pacemakers.
Cardiac Pacemakers new 10 Oct 12 D McMahon.
Deactivating the shocking component of an Implantable Cardioverter Defibrillator (ICD) (When no longer appropriate)
Pacemakers and Implantable Cardioverter-Defibrillators
PACEMAKER Yoga Yuniadi
Laurie Racenet, FNP, MSN, CCDS, CEPS, FHRS
Pacemakers and AICD’s.
Canadian Cardiovascular Society/Canadian Anesthesiologists' Society/Canadian Heart Rhythm Society Joint Position Statement on the Perioperative Management.
Pacemakers.
Dysrhythmias Disorders of formation or conduction (or both) of electrical impulses within heart Can cause disturbances of Rate Rhythm Both rate, rhythm.
Pacemakers and Devices – Interactive Session
EKGs and Pacemakers Cooper University Hospital
José L. Morales et al. JACEP 2017;j.jacep
Diathermy and its safe use
Volume 6, Issue 10, Pages (October 2009)
Cardiac Implantable Electronic Devices
David D. Berg et al. JACC 2018;71:
Cautery (ELECTRO MAGNETIC INTERFERENCE) & cardiac devices in endoscopy
Swimming pool saline chlorination units and implantable cardiac devices: A source for potentially fatal electromagnetic interference  John Wight, BS,
M.E. Stone, B. Salter, A. Fischer  British Journal of Anaesthesia 
Managing cross talk between a subcutaneous implantable cardioverter-defibrillator and a dual-chamber unipolar pacemaker system  Konstantinos Kossidas,
Pacemakers, Cardioverters, and Defibrillators
Presentation transcript:

“Just Put a Magnet On It” An Update on Cardiac Implantable Electronic Devices 26 companies have produced nearly 2000 models of pacemakers. (source: Rozner- pacemaker misinformation) Carolyn Boyle, RN, BSN, SRNA Goldfarb School of Nursing at Barnes-Jewish College

Objectives Review the perioperative implications of Cardiac Implantable Electronic Devices Discuss the risks of Electromagnetic Interference (EMI) Examine recommendations for preoperative assessment and preparation, intraoperative management, and postoperative recovery for patients with CIED’s

Outline Case Study CIED Review Magnet Mechanism Practice Recommendations The Future of CIED’s Conclusions

Case Study 65 y.o. male with a hx of SCC of the head and neck Presents with flap necrosis, osteonecrosis, orocutaneous fistula formation, & a complicated open wound s/p mandibulectomy, tracheostomy, bil neck dissection, split thickness skin grafting, and G tube placement ~ 8 mos prior. ~2 mos prior, pt had flap and trach revision presenting for removal of hardware with radial and scapular flap reconstruction PMH: SCC of head and neck, HTN, CHF (EF 30-40%), AFlutter, DM, COPD, Cardiac Arrest Torsades arrest following cardioversion in 2012, warranting AICD placement

Case Study con’t. AICD last interrogated preoperatively 2 months prior with recommendations for magnet placement. Uneventful induction of anesthesia, HOB turned 180 degrees, additional PIV’s and arterial line placed. Magnet placed over device and secured with 2 in silk tape.

Case Study con’t. Approximately 10 min after incision, surgeon using Bovie electrocautery while exposing the mandible Patient’s AICD fired No hemodynamic compromise No arrhythmias What happened? In order to explore why this patient’s device fired inappropriately, let’s review CIED’s and perioperative recommendations for their management. Bovie - monopolar

Pacemakers In 2001, standardized programming codes were developed: Can be single chamber, dual chamber (A-V), or multi-chamber (bi-V) The leads can be either monopolar or bipolar Bipolar is most common today – reduces the risk of electromagnetic interference (EMI) In 2001, standardized programming codes were developed: PACING SENSING RESPONSE RATE MODULATION MULTISITE PACING A = Atrium I = Inhibited R = Rate-modulating V = Ventricle T = Triggered O = none D = Dual (A&V) D = Dual (I and/or T) O = None Bipolar leads have the anode and cathode in the tip of the lead, only centimeters apart, whereas monopolar leads have the cathode in the tip and the device or pulse generator functions as the anode. The reduced distance between the two in the bipolar leads reduces the likelihood of EMI. Standardized programming codes developed by the north american society of pacing and electrophysiology and the british pacing and electrophysiology group in 2001 Rate Modulation: allows the pacemaker to increase the paced heart rate in response to certain physiologic situations such as exercise. This can be sensed using different technologies including an accelerometer, thoracic bioimpedance, or even minute ventilation. Multisite Pacing or Cardiac Resynchronization Therapy: mostly used to notate Bi-ventricular pacing. This optimizes the timing of LV and RV systole in patients with severe HF (currently indicated NYHA class 3 or 4, and for LVEF<35%). CRT can improve cardiac output, hemodynamics, HF symptoms, and QOL.

Internal Cardioverter Defibrillators (ICD) Detect & treat ventricular arrhythmias Today, incorporate pacemakers in case defibrillation results in bradycardia or asystole Can be single chamber (RV lead only), dual chamber (A&V leads), or triple chamber (atrial, RV, LV leads). Pre-programmed ‘zones’ based on heart rates and chamber of origin. Based on rate, the device can initiate anti-tachycardia pacing or deliver a defibrillatory shock Programmed zones - allows the device to differentiate ventricular arrhythmias from SVT

Potential Perioperative Problem: Electromagnetic Interference (EMI) Classic Causes: Surgical electrocautery Nerve stimulator Evoked potential monitoring Radiofrequency Ablation Potential Causes: Fasciculations Shivering Large tidal volumes Lithotripsy Pacemaker: Inhibition of pacing due to oversensing Brief exposure generally not a problem Prolonged exposure can lead to conversion to asynchronous pacing mode Defibrillator: inappropriate defibrillation EMI is the most common problem in the perioperative period for patients with devices. EMI: Radiofrequency waves interfering with proper device function

Magnet Mechanism: The Reed Switch A reed switch consists of two metal strips made of magnetic material in a glass capsule. The strips are open and are not in contact. Application of a magnet causes the strips to come in contact (closed) which triggers the pulse generator to switch to perform programmed functions such as asynchronous pacing or the suspension of anti-tachycardia therapies in ICD’s. This mechanism can be effected by the body habitus of the patient (some manufacturers recommend 2 magnets for super morbidly obese patients), manufacturer, programmed settings, & battery life. Sometimes the reed switch can malfunction by staying in the closed or open position regardless of magnet presence. This is known as a sticky reed switch. Malfunction can lead to anti-tachycardia therapy failure and accelerated battery depletion. As an example of this - Guidant has had problems with this mechanism in a number of ICD’s used in patients with severe cardiomyopathies and they recommended that magnet function be permanently disabled by programming Source: Sony et al 2011 ***Magnet effect on CIED is extremely variable depending on the device, body habitus, manufacturer, programmed settings, and battery life.

Magnet + PACEMAKER Usually, a magnet will convert a pacemaker to asynchronous mode Device response to magnet can be programmed Rate depends on the manufacturer and the battery life Asynchronous pacing mode depends on prior settings DDD  DOO VVI  VOO AAI  AOO Caution: Asynchronous rate may not always meet the physiologic demands of the patient Upon removal, device should revert to originally programmed pacing mode Ie- Decreased SVR and hypovolemia may require increased HR

Source: Sony et al 2011 BOL – beginning of life ERI – elective replacement interval EOL – end of life Source: Sony et al 2011

Magnet + DEFIBRILLATOR Usually, a magnet will prevent antitachycardic pacing and defibrillation In order to prevent oversensing of EMI Modern AICD’s are also pacemakers – a magnet will not have any effect on the pacemaker function!! For patients with AICD’s who are pacemaker dependent  preoperative interrogation and reprogramming is recommended

-magnet application will have NO effect on pacing. -Boston Scientific and St. Jude devices can be programmed to ignore the magnet! -some (but not all) will may audible noises to indicate that a magnet has been successfully placed -most return to baseline upon magnet reapplication -as always, when in doubt - interrogate Source: Sony et al 2011

February, 2011

Focused Pre-Operative Evaluation Presence of device H&P, medical record review, CXR, EKG, physical exam Type of device Manufacturer ID card, CXR, supplemental records, consult cardiology, year placed Dependency on pacemaker function Verbal history of syncope or bradycardia requiring CIED placement, AV node ablation, 100% paced on EKG Device function Best way: comprehensive evaluation (interrogation) At minimum: evaluation of EKG or rhythm strip, discussion with patient has aicd fired? syncope? cardiology follow up? When was it placed? On average – PM 5-8 years AICD 3-6 years

CXR Interpretation Female patient with single lead ventricular pacemaker – tip in the apex of the RV -likely valve repl too – sternal wires visible

CXR Interpretation Female patient with single lead atrial pacemaker (tip in right atrial appendage)

CXR Interpretation Female with two lead pacemaker with leads in RA and RV

CXR Interpretation Male with Bi-V pacemaker. White arrow is RA lead. Black arrow is RV lead. While chevron is LV lead in the posterior coronary vein at the LV.

CXR Interpretation Patient with ICD. Note the thickened coil representing the defibrillator lead in the RV.

Preparation is Everything! Likelihood of Electromagnetic Interference (EMI) Need for preoperative CIED reprogramming Asynchronous pacing Suspension anti-tachyarrhythmia functions Suggest the use of Bipolar/ Harmonic electrocautery Assure presence of external pacing/ defibrillation capabilities before, during, and after the procedure Evaluating the possible effect of anesthetic technique on CIED function Consider positioning of the patient Specific anesthetic techniques are unlikely to alter CIED function BUT anesthesia-induced changes (such as HR, BP, rhythm, or ischemia) can lead to unexpected device function. Positioning. Bovie is monopolar

EMI above umbilicus? NO YES Pacemaker ICD From: Neelankavil et al. 2013. EMI above umbilicus? NO YES Pacemaker ICD No reprogramming or magnet necessary. Have magnet available. Deactivate ICD – magnet/ reprogram Dependent? Yes- Magnet/ reprogram to asynch No- Consider reprogram/ magnet if source is <15 cm from generator. Have magnet avail. Pacemaker Dependent? No- No reprogram necessary Yes- Reprogram to asynch.

Intra-operative Management If a magnet is placed or the device is reprogrammed, external defibrillation should be immediately available! Place pads as far away from generator as possible A-P placement is preferred

Intraoperative Management Monitor patient appropriately, monitor function of device, and monitor for signs of EMI Assure the cautery grounding pad is positioned so that the current pathway doesn’t cross through or near the device This may mean that sites other than the thigh should be used Surgeon should avoid cautery near the device Short, intermittent bursts of cautery at the lowest possible energy level is ideal Risk of EMI is much greater with monopolar than bipolar cautery ASA gives the example of placing the grounding pad on the contralateral shoulder for a head and neck case!

Algorithm for Emergent Cardioversion or Defibrillation Surgeon terminates all sources of EMI Magnet placed preop Remove Magnet! Observe for appropriate CIED response Prepare for external defib or cardioversion Device reprogrammed preop Re-enable therapies if programmer immediately available This algorithm has been adapted from ASA practice advisory – for a patient with ICD facing VTACh in the clinical setting

Post-Operative Care Continuous monitoring Pacing & defibrillation available Regardless of the anesthetic approach to the CIED, electrocautery within 6 inches of the device can damage to the internal circuitry and post-operative interrogation is recommended If there is any question, device should be interrogated to assess function

Case Study Conclusion Electrocautery removed from the field, device representative called to the room Interrogated device, confirmed defibrillation Unsure why magnet failed Manual reprogramming of defibrillator Surgery proceeded without further incident Device interrogated postoperatively and returned to preoperative settings

The Future of CIED’s

NanostimTM Leadless PM Placed via groin access, battery life equivalent to current PPM. No pocket or leads req’d. Can communicate with interrogator similar to current devices.

Subcutaneous ICD Boston Scientific http://www.bostonscientific-international.com/templatedata/imports/HTML/CRM/A_Closer_Look/pdfs/ACL_SQRX_Inhibit%20Tachy%20Therapy%20with%20Magnet_20121016.pdf

Thank you! Questions?

References American Society of Anesthesiologists Committee on Standards and Practice Parameters. (2011). Practice advisory for the perioperative management of patients with cardiac implantable electronic devices: Pacemakers and implantable cardioverter-defibrillators. Anesthesiology, 114, 247-261. doi: 10.1097/ALN.0b013e3181fbe7f6 Neelankavil, J. P., Thompson, A., Mahajan, A. (2013). Managing cardiovascular implantable electronic devices (CIED’s) during perioperative care. APSF Newsletter, 28, 31-35. Jacob, S., Panaich, S. S., Maheshwari, R., Haddad, J. W., Padanilam, B. J., John, S. K. (2011). Clinical applications of magnets on cardiac rhythm management devices. Europace, 13, 1222-1230. doi: 10.1093/europace/eur137 Lanzman, R. S., Winter, J., Blondin, D., Furst, G., Scherer, A., Miese, F. R., Abbara, S., Kropil, P. (2011). Where does it lead? Imaging features of cardiovascular implantable electronic devices on chest radiograph and CT. Korean J Radiol, 12 (5), 611-619. doi: 10.3348/kjr.2011.12.5.611. Schulman, P. M., Rozner, M. A. (2013). Use caution when applying magnets to pacemakers or defibrillators during surgery. Anesthesia & Analgesia, 117, 422-427. doi:10.1213/ANE.0b013e31829003a1 Rooke, G. A., Bowdle, T. A. (2013). Perioperative management of pacemakers and implantable cardioverter defibrillators: It’s not just about the magnet. Anesthesia & Analgesia, 117, 292-294. doi: 10.1213/ANE.0b013e31829799f3 Rozner, M. (2004). Pacemaker misinformation in the perioperative period: Programming around the problem. Anesthesia & Analgesia, 99, 1582-1584. doi: 10.1213/01.ANE.0000140244.35896.D7