PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY.

Slides:



Advertisements
Similar presentations
Advanced GAmma Tracking Array
Advertisements

1 Continuous Scintillator Slab with Microchannel Plate PMT for PET Heejong Kim 1, Chien-Min Kao 1, Chin-Tu Chen 1, Jean-Francois Genat 2, Fukun Tang 2,
Study of plastic scintillators for fast neutron measurements
Recent news about SiPM based applications R&D in DESY Nicola D’Ascenzo University of Hamburg - DESY.
Chapter 8 Planar Scintigaraphy
Quartz Plate Calorimeter Prototype Ugur Akgun The University of Iowa APS April 2006 Meeting Dallas, Texas.
Computed Tomography II
SmartPET: A Small Animal PET Demonstrator using HyperPure germanium Planar Detectors R. Cooper 1 st year PhD Student
Fysisk institutt - Rikshospitalet 1. 2 Overview Gamma camera Positron emission technology (PET) Computer tomography (CT) Proton therapy Electrical impedance.
Signal Analysis and Processing David Scraggs. Overview Introduction to Position Resolution Induced Charges Wavelet Transform Future Work Discussion.
1 A Design of PET detector using Microchannel Plate PMT with Transmission Line Readout Heejong Kim 1, Chien-Min Kao 1, Chin-Tu Chen 1, Jean-Francois Genat.
Kirov A S, MSKCC Overview of Geant4 Use and Issues in Imaging: Emission Tomography (PET and SPECT) Assen S. Kirov Department of Medical Physics Memorial.
Scintillator based muon upgrade / BELLE Super B Factory Workshop In Hawaii Jan 2004, Honolulu, Hawaii 1.Scintillator strip option 2.Geiger photodiodes.
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
8/18/2015G.A. Fornaro Characterization of diffractive optical elements for improving the performance of an endoscopic TOF- PET detector head Student: G.
RF background, analysis of MTA data & implications for MICE Rikard Sandström, Geneva University MICE Collaboration Meeting – Analysis session, October.
High spatial resolution measurement of depth-of-interaction of a PET LSO crystal Aliz Simon a László Balkay b, István Chalupa b, Gábor Kalinka a, András.
Very High Resolution Small Animal PET Don J. Burdette Department of Physics.
Optimization of LSO for Time-of-Flight PET
The Transverse detector is made of an array of 256 scintillating fibers coupled to Avalanche PhotoDiodes (APD). The small size of the fibers (5X5mm) results.
Report on SiPM Tests SiPM as a alternative photo detector to replace PMT. Qauntify basic characteristics Measure Energy, Timing resolution Develop simulation.
The PEPPo e - & e + polarization measurements E. Fanchini On behalf of the PEPPo collaboration POSIPOL 2012 Zeuthen 4-6 September E. Fanchini -Posipol.
The Design of Barrel Electromagnetic Calorimeter (BEMC) October Lu Jun-guang.
Department of Radiology UNIVERSITY OF PENNSYLVANIA 1 Investigation of LaBr 3 Detector Timing Resolution A.Kuhn 1, S. Surti 1, K.S. Shah 2, and J.S. Karp.
Evaluation of Silicon Photomultiplier Arrays for the GlueX Barrel Calorimeter Carl Zorn Radiation Detector & Medical Imaging Group Jefferson Laboratory,
February 14, 2013 Study of a Cherenkov TOFPET modulePeter Križan, Ljubljana Samo Korpar, Rok Dolenec, Peter Križan, Rok Pestotnik, Aleš Stanovnik J. Stefan.
Fundamental Limits of Positron Emission Tomography
SiPM: Development and Applications
The Scintillator ECAL Beam Test at FNAL K. Kotera, Shinshu-u, 1st October 2009 CALICE Scintillator ECAL group; Kobe University, Kyungpook University, the.
First Results from a Test Bench for Very High Resolution Small Animal PET Using Solid-State Detectors Klaus Honscheid for The CIMA Collaboration The Ohio.
Single Photon Emission Computed Tomography
PET/SPECT Phantom. Side View of Phantom Image Resolution Intrinsic resolution FWHM Intrinsic resolution FWHM Field of view Field of view Measurement:
Pedro Arce Introducción a GEANT4 1 GAMOS tutorial Compton Camera Exercises
Increase in Photon Collection from a YAP:Ce Matrix Coupled to Wave Lenght Shifting Fibres N. Belcari a, A. Del Guerra a, A. Vaiano a, C. Damiani b, G.
16. January 2007Status Report On Compton Imaging Projects 1 Status Of Compton Imaging Projects Carried Out In The CIMA Collaboration HPD Brain PET Meeting.
R&D status of the Scintillator- strip based ECAL for the ILD Oct LCWS14 Belgrade Satoru Uozumi (KNU) For the CALICE collaboration Scintillator strips.
Enhancing InBeam PET with single Photon (Compton) Detection CERN September 2nd 2008 VALENCIA GROUP, IFIMED José M. Benlloch (PET hardware, speaker) José.
Timing Studies of Hamamatsu MPPCs and MEPhI SiPM Samples Bob Wagner, Gary Drake, Patrick DeLurgio Argonne National Laboratory Qingguo Xie Department of.
Medical applications of particle physics General characteristics of detectors (5 th Chapter) ASLI YILDIRIM.
Impact of Axial Compression for the mMR Simultaneous PET-MR Scanner Martin A Belzunce, Jim O’Doherty and Andrew J Reader King's College London, Division.
SiPM for CBM Michael Danilov ITEP(Moscow) Muon Detector and/or Preshower CBM Meeting ITEP
00 Cooler CSB Direct or Extra Photons in d+d  0 Andrew Bacher for the CSB Cooler Collaboration ECT Trento, June 2005.
By Matthew Kauer First Year Report – 15 June 07 Measurement of 2b2ν Half-Life of Zr96 and Lightguide Studies for SuperNEMO Calorimeter Matthew Kauer UCL.
Peter Dendooven LaBr 3 and LYSO monolithic crystals coupled to photosensor arrays for TOF-PET Physics for Health in Europe Workshop February 2-4, 2010,
Development of RPC based PET B. Pavlov University of Sofia.
Acquisition time6 min1 min 12 s Collimator height25 mm (Anger)12 mm (HiSens) Detector1 layer, 1 pixel / hole3 layers, 1 pixel / hole3 layers, 4 pixels.
Development of a Single Ion Detector for Radiation Track Structure Studies F. Vasi, M. Casiraghi, R. Schulte, V. Bashkirov.
RPC February 2010 SPATIAL RESOLUTION OF HUMAN 3D RPC-PET SYSTEM 1 LIP, Laboratório de Instrumentação e Física Experimental de Partículas, Coimbra,
ArgonneResult_ ppt1 Comparison of data and simulation of Argonne Beam Test July 10, 2004 Tsunefumi Mizuno
Upgrade of the MEG liquid xenon calorimeter with VUV-light sensitive large area SiPMs Kei Ieki for the MEG-II collaboration 1 II.
PET Imaging Positron Emission Tomography
Simulations in Medical Physics Y. TOUFIQUE*, R.CHERKAOUI EL MOURSLI*, M.KACI**, G.AMOROS**, *Université Mohammed V –Agdal, Faculté des Sciences de Rabat,
M.Taguchi and T.Nobuhara(Kyoto) HPK MPPC(Multi Pixel Photon Counter) status T2K280m meeting.
P. Lecoq CERN2 February ENVISION WP2 Meeting CERN Group contribution to ENVISION WP2 Paul Lecoq CERN, Geneva.
Performance of scintillation pixel detectors with MPPC read-out and digital signal processing Mihael Makek with D. Bosnar, V. Gačić, L. Pavelić, P. Šenjug.
Ultra fast SF57 based SAC M. Raggi Sapienza Università di Roma
PET Activities at UiO NFR Review E. Bolle*
AGATA Advanced Gamma Tracking Array
for a scintillator based MTT
A new-concept Absorption Calorimeter for the ELI-NP facility
Detailed simulations of a full-body RPC-PET scanner
Summary of the Compton-PET project
Development of a High Precision Axial 3-D PET for Brain Imaging
KM2A Electron Detector Optimization
Geant4 Simulation :MCP PET
MCP PET Simulation (7) – Pixelated X-tal
First demonstration of portable Compton camera to visualize 223-Ra concentration for radionuclide therapy Kazuya Fujieda (Waseda University) J. Kataoka,
md-NUV PET project meeting
Status Report on MCP PET Simulation
Presentation transcript:

PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

2 Outline Measurements Single crystal studies (GEANT4) PET scanner (GATE)

3 Experimental Setup MPPC and crystal well aligned, optical photons well contained, high reproducibility Reaching energy resolutions of % FWHM (LSO and LFS) 3x3 mm pixels MPPC LFS Reflective Paint Reflective Foil 3x3x15 mm 3 Source Na-22

4 Energy Resolution Poisson stat.8% for LSO/LFSnegligible MPPC (3x3 mm, 3600 pixels) coupled to LFS (3x3x15 mm) Good separation between compton continuum and photo-electric peak. Mean at 2000 pixels. Saturation not an issue: signal width ~120 ns, peak amplitude ~400 p.e. Linearity checked with different sources blue sensitivity

5 Linearity Na-22 Cs-137 (back scatter peak) Linearity confirmed in the energy region of interest.

6 Simulation Studies Single Crystal + MPPC Detailed studies to optimize single crystal dimensions GEANT4 simulation Good description of the measurements in terms of energy resolution and photon yield. PET System Design of a full PET system Spatial resolution (image reconstruction using FBP) Optimize crystal dimensions (DOI) GATE (GEANT4 Application for Tomographic Emission) toolkit based on GEANT4 First results very promising.

7 Single Crystal + MPPC GEANT4 Simulation Important parameters: LSO/LFS: scintillation yield (23k), refraction index (1.82), emission spectrum, absorption length (50m) Optical grease BC-630 (10µm): refraction index (1.5) MPPC: quantum efficiency Teflon wrapping: reflectivity (0.90), refraction index (1.34) Compare to measurements: number of pixels signal width signal shape 1.5 cm 3x3mm 2 Compton-Effect Photo-Effect (LY: 100 photons/MeV) 30 reflections: P(absorbed)=45% (98% reflectivity) crystal MPPC

8 Scintillation Yield and Photon Detection Efficiency Incorporated published distributions. PDE Hamamatsu:

9 A Single Event Number of photons produced : Absorbed in the crystal : 9 Absorbed at boundary : 9013 Detected : 1911 Leakage (optical grease) : 7 z of absorptions at boundary: Crystal length: 1.5 cm Reflectivity (Teflon): 90 % MPPC interaction z of reflections at boundary: optical grease

10 Efficiency ~ 50 % Interaction in crystal?Photo- or Compton-Effect? 5 cm > total length > 3 cm

11 Energy Resolution R = 95% R = 90% Energy resolution is slightly decreasing for longer crystals due to absorption in the Teflon wrapping. The larger the reflectivity the smaller the effect.

12 Time Resolution Hit-time distribution for a single event: Triggering at low threshold, i.e. the very first photons which arrive: Crystal length has only a small influence on the timing. t [ns] Hits t(0.1/e) FWHM

13 Measurements vs. Simulation Reasonable agreement with data. MC R = 90%Tuning of the reflectivity MC

14 Crystal Length and DOI Depth of interaction (DOI) information improves resolution considerably. 1 layer2 layers Gamma ray LOR Simulation of a full PET scanner

15 PET System - GATE Simulation of a full PET system to optimize camera design (high granularity, DOI, TOF). Image reconstruction using Filtered Back Projection (STIR) Determination of spatial resolution in compliance with NEMA NU standard*. Dead material, movement, noise, dead time, dose, phantom, scan time, efficiency Simulation of a simple PET system: 2x387 crystals (LSO, 3x3x15 mm 3 ) * National Electrical Manufacturers Association

16 PET System – First Results Tune energy and time resolution of single crystal+MPPC element according to measurements. Crystal+MPPC FWHM 11% FWHM 600 ps coincident events Sinogram point source x=2cm next steps: granularity, FBP (STIR)

17 Conclusions and Outlook According to simulations no strong dependence of the energy and time resolution on the crystal length. DOI studies with simulation of full PET scanner. Two systems with two layers of crystals of different length. Total length 3.5 to 4.5 cm, First layer 1 to 1.5 cm.

18 BACKUP FOILS

19 Design Criteria Camera design (matrix of crystals, each crystal read out individually by an MPPC) guided by the following criteria: spatial resolution (DOI: two layers of crystals) time resolution efficiency exposure to radiation, scan-time costs Prototype: Two matrices of 6x6 crystal double-layers Study uniformity of channels, multi-channel r/o, calibration Image reconstruction of a point source LFS 3x3x15 mm 3

20 Sinogram Integration along all LORs at fixed 

21 Sinogram - Example Each point in the sinogram corresponds to a LOR.

22 Filtered Back Projection (Analytic 2D image reconstruction) Imaging system: p = Hf + n p: observation, H: system model, f: image, n: error Solution: unfolding, inverse problem ill-posed, regularization 2D central-section theorem: (Radon transform closely related to Fourier transform. ) The Fourier transform of a 1D projection is equivalent to a section at the same angle through the center of the 2D Fourier transform of the object. Ramp-filter:

23 A Commercial Example GEMINI TF PET/CT (Philips) diameter of 90 cm, axial field of view of 18 cm 4 x 4 x 22 mm 3 LYSO (lutetium-yttrium oxyorthosilicate) crystals coincidence timing resolution: 585 ps resolution near the center: 4.8 mm (FWHM) single high-quality image: 1-2 min whole-body scan: min (depending on patient size) available for $1.85M