Hemacytometer and Manual Cell Counts

Slides:



Advertisements
Similar presentations
Principle Concepts of Iv therapy.
Advertisements

Prepared by Yasser El-dahdouh
Significant Figures and Scientific Notation. Significant Figures ► When using our calculators we must determine the correct answer; our calculators are.
RBC Manual Count.
White blood cells.
Week 4: Hemocytometry Hemocytometer RBC and WBC Thoma pipet Unopette
Will the ball float on liquid mercury?
Practical Hematology Manual
STUDY OF NEUBAUER’S CHAMBER
Practical Hematology Manual
Erythrocyte Count Normal ranges:
Lesson Starter Look at the specifications for electronic balances. How do the instruments vary in precision? Discuss using a beaker to measure volume versus.
Using and Expressing Measurements
Chapter 21 Quantitation Cell Counting Technique 1.Hemocytometer 2.Electronic Cell Counter.
Automated CBC Parameters
Automated CBC Parameters
Complete Blood Count.
Chapter 5 Hematology. Figure 5.1 (a) A double ended needle partially pierces the rubber stopper in an evacuated tube. (b) After the left needle pierces.
1. MIKROSCOPE And 2.HEMOCYTOMETER
Leukocytes Manual WBC Counting Clinical Pathology VTHT 2323 Lori VanValkenburg, RVT.
Microbial Count Aim: Count the number of bacterial cells in a provided sample Methods: Total count, Viable count I. Total count This technique involves.
Significant Figures. Accuracy vs. Precision Percentage Error.
Exp. 2 Cell Morphology and Counting
Blood Tests Learning objectives:
Scientific Measurement Chpt 3. Units of Measure –general qualitative – describes matter – ex. Rough, shiny, heavy, blue quantitative – measures matter.
Plant and Mammalian Tissue Culture Counting Cells and Experimental Design.
Precision, Accuracy and Significant Figures Notes.
Microscope parts:.
Scientific Measurements. Measurements Objective Distinguish between accuracy and precision Determine the number of significant figures in measurements.
RBC, WBC, & PLT Counts Nada Mohamed Ahmed, MD, MT (ASCP)i.
RBC, WBC, & PLT Counts.
White Blood Cell (WBCs) Count
Red Blood Cell (RBCs) Count
CHEMISTRY CHAPTER 2, SECTION 3. USING SCIENTIFIC MEASUREMENTS Accuracy and Precision Accuracy refers to the closeness of measurements to the correct or.
UNIT 1 - INTRODUCTION TO PHYSICS Measurements, Accuracy and Errors.
WBC manual count using haemocytometer.  Introduction The blood consist of 3 type of specialized cellular elements:  Erythrocytes -RBCs  Leukocytes.
Chapter 2 © Houghton Mifflin Harcourt Publishing Company Accuracy and Precision Accuracy refers to the closeness of measurements to the correct or accepted.
Measurements and their Uncertainty
Measurement & Data Processing IB Chem. Objective: demonstrate knowledge of measurement & data processing. Warm up: Explain the difference between accuracy.
Practical hematology Assist.Prof. Dr.Sajeda Al-Chalabi 1.
WBC manual count using hemocytometer
Cell count with Hemocytometer
STUDY OF NEUBAUER’S CHAMBER
White blood cells count
STUDY OF NEUBAUER’S CHAMBER Fahareen-Binta-Mosharraf MNS.
COUNTING CELLS USING A HAEMOCYTOMETER
WBC manual count using Hemocytometer
Hemocytometer  For hematology, microbiology, cell culture, and many applications that require use of suspensions of cells it is necessary to determine.
Laboratory Mathematics and Measurements
Practical hematology Assist.Prof. Dr.Sajeda Al-Chalabi.
White Blood Cell (WBCs) Count
Red Blood Cell (RBCs) Count
White Blood Cell (WBCs) Count
WBC manual count using hemocytometer
Blood count Lab 4 Naseem AL-Mthray.
Red Blood Cell (RBCs) Count
RBC Manual Count.
White blood cells count
Manual Red Blood Cell and White Blood Cell Counts
White blood cell (WBC) count
Section 3: Uncertainty in Data
Laboratory Technique LAB.
White Blood Cells Count Using Hemocytometer
Text Section 2.3 Pages
RBC & WBC count Dr. Tamara Alqudah.
Performing a Manual Count for Sperm Concentration
Red Blood Cell (RBC) Count
White blood cell count.
Marine Biotechnology Lab
Scientific Measurements
Presentation transcript:

Hemacytometer and Manual Cell Counts Kimberly Murray University of Phoenix EDTC 560: Applications of Multimedia and Web Page Design June 4, 2007

Quote of the Day “Education is a technique employed to open minds so that they may go from profound ignorance to thoughtful uncertainty” (Mitner, 1999).

Objectives Upon completion of this course, the student will: Identify the dimensions of the Neubauer hemacytometer. Identify sources of error in counting. Calculate manual cell counts.

Hemacytometer Counting Chamber A precision tool designed to standardize counting. Used to count cells in peripheral blood or other body fluids.

Basic hemacytometer formula Number of cells counted X dilution factor total volume of the cells counted total volume = # primary squares x 0.1 (depth) This formula can be used with any counting chamber as long as one knows the dimensions for the chamber

Hemacytometer Components Cover slip –NBS standardized Moat Fill trough Chambers or Counting grid Diagrams by K. P. Murray, 2007

Sample is placed between cover slip and chamber The design of the hemocytometer is to have the support for the cover slip at a specific height, resulting in a known depth of exactly 0.1mm. Cover slip –NBS standardized Moat 0.1 mm Depth National bureau of standards must be met for both the chamber and cover slip. Sample Diagrams by K. P. Murray, 2007

Do Not Use Disposable Cover Slip Sample floats coverslip Do not use a regular cover slip especially for viscous fluids. The sample will float disposable coverslips and the depth will be greater than 0.1mm. Diagrams by K. P. Murray, 2007

Neubauer Hemacytometer Dimensions Each side (chamber) has a ruled area 3 mm wide and 3 mm long Divided into nine squares The large squares are subdivided into 16 The middle square is subdivide into 25 Each of the 25 is again subdivided into 16 When the dimensions are known, one will always be able to calculate the area counted.

Dimensions of the Neubauer Note: Center primary square has different divisions Primary squares 1.0 mm Secondary Squares 0.2mm 0.25mm Tertiary squares 0.0125 mm Diagrams by K. P. Murray, 2007

Hemacytometer Quality Control (QC) Ensures accuracy Performed by duplicating the counting area exactly. For example: If squares 1,3,5,7,9 are counted. Then 10,12,14,16,18 are counted on the other side. QC is duplicating the counts if you count squares 1,2,3 then on the other side you count 10, 11, 12 Diagrams by K. P. Murray, 2007

Area Counted Depend on Cell Type WBC WBC . . Platelet RBC RBC RBC RBC RBC . . WBC WBC Diagrams by K. P. Murray, 2007

RBC Count Method Count the five inner tertiary squares on both sides of the chamber on Peripheral Blood Check difference in counts Calculate RBC count using the basic hemacytometer formula . . RBC RBC RBC RBC RBC . . Diagrams by K. P. Murray, 2007

RBC Count Method If a Body Fluid count is less than 100 cells count all 9 squares on both sides to provide greater accuracy. Adjust the hemacytometer calculation accordingly . . . . Diagrams by K. P. Murray, 2007

WBC Count Method Count the four outer secondary squares on both sides of the chamber Check difference in counts Calculate WBC count using the basic hemacytometer formula WBC WBC WBC WBC Diagrams by K. P. Murray, 2007

WBC Count Method If WBC count is less than 1000 cells count all 9 squares on both sides to provide greater accuracy. Adjust the hemacytometer calculation accordingly Diagrams by K. P. Murray, 2007

Platelet (PLT) Area Counted Diagrams by K. P. Murray, 2007

Neubauer Hemacytometer Calculation Cells X dilution = cells/L squares X 0.1 Note: 0.1 mm depth is constant in Neubauer Hemacytometer

Unopette – used today for manual counts Components Reservoir Capillary pipette Protective sleeve that also is used to open the reservoir. Pre-measured diluents in vials. Capillary pipette is specific for the diluent used. The capillary tubes can not be interchanged. Premeasured diluents in vials. And a specific capillary pipette for the diluent needed. The capillary tubes can not be interchanged. Picture by K. P. Murray, 2007

Unopette System For RBC, WBC and PLT counts RBC vial Isotonic or physiologic saline 1/200 dilution WBC only vials Turk’s solution 1/20 dilution WBC/PLT vials Ammonium oxalate --11.45 g Sorenson’s buffer--1.0g Thimerosal -- 0.1g (preservative) Distilled H2O qs to 1 liter 1/100 dilution RBC can not resist osmotic pressures, as the WBC can, therefore RBC counts need isotonic solutions Manual RBC counts in PB rarely done Always performed on other body fluids

Unopettes Test name Sample diluent Type of diluent Final dilution WBC Note: different final dilutions Test name Sample diluent Type of diluent Final dilution WBC 25 L 0.475 ml TURKS- 3% acetic acid 1:20 RBC 10 L 1.99 ml Isotonic saline 1:200 PLATELET 20 L 1.98 ml 1% ammonium oxalate 1:100 Eosinophils 0.775 ml Phloxine B 1:32

Loading Chamber using a Unopette Picture by K. P. Murray, 2007

“Wet Box” prevents solution from drying out and altering count Settle cells to the same optical plane WBC 2-3 minutes PLT 10 minutes (take the longest to settle) RBC Picture by K. P. Murray, 2007

Acceptable differences between chambers of the hemacytometer No more than 10 cell variation between the 4 squares for the WBC count. The difference between the highest and the lowest number should not be greater than 25 cell for the 10 tertiary squares (5 squares per side) the RBC count. Diagrams by K. P. Murray, 2007

How to Tell If Sample has Uneven Distribution Difference between individual squares exceeds 10% Difference between each chamber count exceeds 20 % REPEAT COUNT with a freshly loaded hemacytometer, if the count does not match the above criteria.

WBC Count Reporting Report WBC Number (period) one decimal X 103/ L Number. decimal X 103/ L Example: 10.9 X 103/ L WBC has one decimal place, if sample is peripheral blood. For other body fluids (BF), report whole number, no scientific notation or abbreviations.

RBC Count Reporting Report RBC Number (period) two decimals X 106/ L Number. two decimals X 106/ L Example: 4.13 X 103/ L RBC has two decimal places, if peripheral blood. For other body fluids (BF), report whole number, no scientific notation or abbreviations.

Platelet Count Reporting Report PLT Whole number (no decimal) X 103/ L Example: 428 X 103/ L Platelet count does not have a decimal point. Other body fluids do not report platelet counts. Verbal reporting always give the full number i.e. “Dr. the platelet count on Ms Smith is 428, 000”

Platelet Count Reporting Verbal reporting always give the full number. For example, “Doctor Jones, the platelet count on Ms. Smith is 428,000”. DO NOT just say the count is 428. The doctors or the nurse may think only four hundred (transfusion time!!!), rather than a normal count of 400 thousand. Verbal reporting always give the full number i.e. “Dr. the platelet count on Ms Smith is 428, 000”

Reference Units Notice the similarities and differences between units Conventional units Example SI units WBC X 103/ L 4.5 X 103 / L X 109 / L 4.5 X 109 / L RBC X 106/ L 5.21 X 106/ L X 1012 / L 5.21 X 1012 / L HGB g/dL 15.4 g/dL g/L 154 g/L MCV fl 88 fl MCH pg 29 pg MCHC % 33 % 330 g/L PLT 182 X 103/ L 182 X 109 / L

Sources of Error Failure to mix sample prior to dilution Pipeting technique Uneven distribution in chamber Failure to remove excess blood from outside of pipette Squeezing out some fluid from reservoir [Unopette] Nature of sample: Clotted not mixed thoroughly Errors caused by apparatus:   1. chipped pipette tips 2. obscure markings on pipettes 3. non-optically plane cover glasses (wrong cover slips) 4. dirty glassware 5. inaccurate rulings on chamber( use certified by the Bureau of Standards) 3) Errors caused by personal technique: Pipetting- not wiping the outside of tip 1. not thoroughly mixing blood 2. inadequate shaking (thoma pipettes) 3. failure to discard first 4 drops 4. not loading chamber properly (under or overfilling, trapped air bubbles) Moving cover slip 5. counting cells inaccurately (skipping cells, counting cells twice, counting on wrong borders) 6. calculation error 7. clerical error (3) Inherent errors in hemocytometry include: 1. "field errors" - relates to the random distribution of cells on the counting chamber 2. statistical error - occurs when total number of cells is too low to give statistical confidence in result (this error is reduced when larger numbers of cells are counted)

Sources of Error Cell aggregates (clots) Bacteria ( may be mistaken for plt) Dirt, talc, debris Drying of the chamber taking too long to count Falsely increase counts

Fill problems Correct fill Coverslip Air bubbles Overfill Underfill Diagrams by K. P. Murray, 2007

Problem 1 A blood sample was diluted1:20. On a Neubauer hemacytometer 140 WBC counted in the four outer squares on side A. 147 WBC counted on the same four squares on side B. What is the WBC count?

Problem 1 answer WBC 140 and 147 ‘match’ and average 143.5 Depth = 0.1 (always with a Neubauer) 143.5 cells x 20 (dilution factor) 4 (squares) x 0.1(depth) WBC = 7,175 WBC =7.2 x 10 3 /dL

Problem #2 A tech counted both side of the hemacytometer on an undiluted spinal fluid sample. 22 WBC were counted and no RBC. What is the WBC count? How is it reported?

Problem #2 solution Both side area = 18 mm2 {(1x1) (9+9)} Depth = 0.1 Both side count = 22 Dilution = 1 Answer: WBC = 12 cells / dL NOTE: no decimal or scientific notation because it is a body fluid and one does not report a fraction of a cell.

References Barry T. Mitzner, (1999), D.V.M, Southeast Vetlab, Inc., Retrieved on July 12, 2003 from http://www.spcollege.edu/hec/vettech/VTDE/Hematology%20methods%20for%20the%20offi/ppframe.htm Murray, K.P.,(2007) Hematology Lab Photographs (personal collection) Murray, K.P.,(2007) Hematology Diagrams and Animations (personal collection) END