Basic fill color R 100 G 160 B 200 Accent 1 ”Logo Blue” R 0 G 39 B 118 Yellow (highlight) R 252 G 217 B 0 Accent 2 R 118 G 118 B 118 OP300 Technical training presentation Full Calibration Procedures
Calibrations of the unit Prerequisites: –s2terminal open and connected to OP300 in normal mode –Cliniview software open and image capture activated How to start: S2terminal must be connected to the unit - select in the control panel - select “Service” in the control panel After calibrations are completed, service indicator must be cleared with s2terminal command
Alignment and QA Tools Pan alignment cone assembly QC PhantomCT Geometry tool Leveling base
Calibrations during installation Necessary calibrations –Preheat calibration –mA calibration –3D collimator calibration –PAN collimator calibration –PAN geometry calibration –3D geometry calibration(s) –3D lasers alignment –PAN lasers alignment –3D pixel calibration –PAN pixel calibration –3D QC program
Acceptance limits Passed Calibration program is successfully done. Move on to next calibration. Not passed Adjustment is still needed. Follow the instructions the image (if any) and take another exposure. Some calibration programs are iterative and needs a few repetitions. Failed System could not decide what adjustment should be done in order for the calibration to succeed. This calibration status is always the result of some error condition. Taking another exposure will not help. The image may give a hint on what the problem is (e.g. no radiation, collimator severely tilted, image data corrupted…).
Preheat calibration Purpose: Preheat calibration Acceptance limits: Passed / Not passed / Failed GUI informs when calibration is done
mA calibration Purpose: Tube current calibration Acceptance limits: Passed / Not passed / Failed GUI informs when calibration is done
3D collimator calibration Purpose: Collimator position calibration Acceptance limits: Passed / Not passed / Failed GUI informs if collimator has to be adjusted 3D Col Loosen Adjustment
Pan collimator calibration Purpose: Collimator position calibration Acceptance limits: Passed / Not passed / Failed GUI informs if collimator has to be adjusted Loosen Adjustment
Pan geometry calibration Purpose: Panoramic geometry position calibration; uses double-cone phantom Acceptance limits: Passed / Not passed / Failed GUI informs if calibration has passed
3D Geometry calibration Purpose: Calibrates 3D geometry, needed only with 3D units; uses geometry phantom Acceptance limits: Passed / Not passed / Failed GUI informs if calibration has passed 3D Geom Std Res 3D Geom Hi Res
3D laser alignment Purpose: To check 3D laser alignment Acceptance limits: 3D calibration phantom is used to check laser alignment 1. Top of FOV 2. Mid-sagittal 3. Bottom of FOV 4. Adjustable phantom holder 3D Lasers
Pan laser alignment Purpose: To check panoramic laser alignment Acceptance limits: Double cone calibration phantom is used to check laser alignment Pan Lasers Pan Lasers TMJ Frankfort Plane Sharp Layer (Cuspid light) Mid-Sagittal
3D pixel calibration Purpose: Performing pixel gain and dead pixel correction Acceptance limits: Passed / Not passed / Failed GUI informs when calibration is passed 3D Pix
Purpose: Panoramic sensor pixel calibration Acceptance limits: Passed / Not passed / Failed GUI informs if calibration has passed Pan pixel calibration
3D QC calibration Purpose: QC phantom installed; checks for correct density calibration of 3D rendering Acceptance limits: Passed / Not passed / Failed GUI informs if calibration has passed 3D QC
Ceph primary collimator calibration Purpose: To check ceph collimator Acceptance limits: Passed / Not passed / Failed GUI informs if calibration has passed Ceph Pri Col
Ceph rotation angle calibration Purpose: Check ceph tracking Acceptance limits: Passed / Not passed / Failed GUI informs if calibration has passed Ceph Rot Angle Collimator height adjustment can be performed simultaneously with this test – measure the height of the exposed portion; it should reach the dotted line (180mm). If not, adjust ceph collimator height at the collimator block as shown
Ceph rotation position calibration Purpose: To check ceph mechanism tracking Acceptance limits: Passed / Not passed / Failed GUI informs if calibration has passed Ceph Rot Pos Ceph Rot Pos
Ceph secondary collimator adjustment (vertical) Purpose: Verify correct position of secondary collimator aperture Acceptance limits: Exposed portion centered vertically in Ceph Rotation Position image Too low Too high Adjustment
Ceph secondary collimator adjustment (horiz) Purpose: Verify correct position of secondary collimator aperture Acceptance limits: Exposed portion centered horizontally in Ceph Rotation Position image, and straight up-and- down Tilted left Tilted right Adjustment
Ceph pixel calibration Ceph Pix Ceph Pix Purpose: Performing pixel gain and dead pixel correction Acceptance limits: Passed / Not passed / Failed GUI informs when calibration is passed
Ceph ear rod adjustment Vertical locking screw B Vertical adjusting screw Horizontal adjusting screw Horizontal locking screw
Ceph laser calibration Purpose: To check ceph laser alignment Acceptance limits: Align laser with ear rod, level with respect to horizontal plane Ceph Lasers
Calibrations needed when parts are replaced THARotating Unit R3200GUI PC3DPanelPan Detector CollimatorR3800R3300 Preheat XXXXXX mA XXXXXX Pan Lasers XXX Pan Collimator XXXX (1)XX Pan Geometry XXXX (1)XX Pan Pixel XXXX (1)XX Pan QC XXXX (1)XX 3D Lasers XXX 3D Collimator XXXXX(1)X 3D Geometry XXXXX(1)X 3D Pixel XXXXX(1)X 3D QC XXXXX (1)X (1)Calibration needed if the collimator is mechanically adjusted during the replacement.