Sources of Measurement Imprecision. Possible Areas for Measurement Errors Definition of what is to be measured “WIDTH or Dimater or Critical Dimension”

Slides:



Advertisements
Similar presentations
Lecture Notes Part 4 ET 483b Sequential Control and Data Acquisition
Advertisements

SEM & TEM in Polymer Characterization
Lecture 11. Microscopy. Optical or light microscopy involves passing visible light transmitted through or reflected from the sample through a single or.
Scanning Electron Microscope (SEM)
Introduction Secondary electron secondary electron detector The electron beam interaction with near surface specimen atoms will make a signal which results.
SMAW Equipment (Shielded Metal Arc Welding- commonly called “stick welding”, or “arc welding”)
SEM Magnification Calibration. Magnification Errors Proper calibration of the SEM scans (magnification) is primary to metrology. SEM Magnification requires.
Science Afternoon at NIST Scanning Electron Microscopy (SEM) Brad Damazo, Prem Kavuri, Bin Ming, Kate Klein and Andras Vladar Nanometer Scale Metrology.
Electron Optics Two essential components: 1)Electron source (gun) 2)Focusing system (lenses) Add scanning apparatus for imaging Electron gun Cathode Anode.
Cutnell/Johnson Physics 8th edition
Lecture 5.1 Scanning Electron Microscopy (SEM)
Lab meetings Week of 6 October
Scanning Electron Microscope Jamie Goings. Theory Conventional microscopes use light and glass lenses SEM uses electrons and magnetic lenses to create.
Scanning Electron Microscope د. عمر عبد القادر.
Main detector types Scintillation Detector Spectrum.
Surface Emissions Specimen current X-rays Cathodoluminescence Pole Piece, etc SE3  ≈ 1 nm for metals up to 10 nm for insulators.
MiniSIMS Secondary Ion Mass Spectrometer Dr Clive Jones Millbrook Instruments Limited Blackburn Technology Centre, England
By Stephen, Victor, Thomas and Tim. What Is a Transformer? A device designed to transfer energy from one electrical circuit to another.
Introduction to scanning electron microscopy
SENSORS AND TRANSDUCERS
SEM microscope By: Doug, Holly & Oleg. Scanning Electron Microscope vs. Optical Microscope Advantages Continuously variable magnification High resolution.
BY SANTANU PRAMANIK(09369) HITESH KUMAR GUPTA(09320) CHANDAN SINGH(09260) SCANNING ELECTRON MICROSCOPE MATERIAL SCIENCE ASSIGNMENT.
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
Introduction to Operation of the XL- 30 Scanning Electron Microscope Tutorial # 2 Introduction to Operation and Imaging.
lecture 2, linear imaging systems Linear Imaging Systems Example: The Pinhole camera Outline  General goals, definitions  Linear Imaging Systems.
NE 110 – Introduction to NDT & QA/QC
Operation of the S4700 FESEM
SEM Metrology Today the SEM is used in many fields as a metrology tool. Measurements are made from: –the screen image –directly from the micrographs –video.
Remember?  An electron is moving downward with a velocity, v, in a magnetic field directed within the page, determine direction of force.
Scanning Electron Microscopy
ISAT 303-Lab3-1  Measurement of Condition: Lab #3 (2005):  List of parameters of condition: –Linear distance, angular displacement, vibration, displacement,
SEM (SCANNING ELECTRON MICROSCOPE) Özgen Buğdaycı Elif Topçuoğlu Yavuz Duran Hacettepe University
College of Nanoscale Science & Engineering By: Elroy Tatem Advisors: Dr. Cherrice Traver Dr. Bradley Thiel (U Albany) Modeling of Dynamic Secondary Electron.
LASER AND ADVANCES IN METROLOGY
Define Problem Select Appropriate Methods Obtain and store sample Pre-treat sample Perform required measurements Compare results with standards Apply necessary.
LECTURER PROF.Dr. DEMIR BAYKA AUTOMOTIVE ENGINEERING LABORATORY I.
Scanning Electron Microscope (SEM)
ELECTROMAGNETIC INDUCTION
EKT 451/4 SEM 2 Chapter 6 SENSOR & TRANSDUCER SHAIFUL NIZAM MOHYAR
Scanning Electron Microscope (SEM)
Magnification Calibration Interlaboratory SEM Study: Part 1.
NANO 225 Micro/NanoFabrication Electron Microscopes 1.
NANO 230 Micro/Nano Characterization
Scanning capacitance microscopy
08/03/09 SEM signal generation
SEM Scanning Electron Microscope
Techniques for Nuclear and Particle Physics Experiments By W.R. Leo Chapter Eight:
Electromagnetic Induction. Motion of a magnet in a coil or loop creates (induces) voltage If coil is connected to complete circuit, current flows Relative.
Optimization of planar pixel detector. T. Habermann Planar pixel detectors L W H ground U.
SARDAR PATEL INSTITUTE OF TECHNOLOGY E.NO : Guide By:- V.N.Thakkar.
Scanning Transmission Electron Microscope
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
EKT 451 CHAPTER 6 Sensor & Transducers.
1 Transducers EKT 451 CHAPTER 6. 2 Definition of a Transducers   Transducer is any device that converts energy in one form to energy in another.  
Electric Pressure Transducer
Electromagnetic Testing (ET). Electromagnetic Testing Electromagnetic testing is a general test category that includes Eddy Current testing (ECT), Alternating.
Presentation on SEM (Scanning of Electron Microscope) Represented by:-Ravi Kumar Roll:- (BT/ME/1601/006)
Introduction to scanning electron microscopy
Chapter 17 Electric Potential
Laboratory equipment Lecture (3).
Introduction to Scanning Electron Microscope by Sameer S
NANO 230 Micro/NanoFabrication
TEM and SEM.
Chapter 17 Electric Potential
Generators.
Do it with electrons !.
Semiconductor Detectors
Computed Tomography (C.T)
Computed Tomography (C.T)
Computed Tomography (C.T)
Presentation transcript:

Sources of Measurement Imprecision

Possible Areas for Measurement Errors Definition of what is to be measured “WIDTH or Dimater or Critical Dimension” Environmental Influences –Vibration, Stray Fields Instrument Maintenance Specimen Charging Electron Detector Position and Type Accelerating Voltage Affects Sample Contamination Sample Dimensional Changes Errors due to specimen tilt Resolution of the measurement system

Possible Areas for Measurement Errors (cond.) SEM Magnification calibration Raster rotation Lens Hysteresis Temperature variations Conductive coating Noise on the signal Repeatability of low voltage operation Interpretation of the signal profile Choice of measurement algorithm Acoustic Coupling

Environmental Variables Influencing the Measurement Electromagnetic Interference –Tool Design instrument shielding electron source/column design Mechanical Vibrations –Robust design –Supplementary vibration isolation Acoustic Coupling –E-beam column sensitivity –Dampened enclosures Temperature changes

Material Variables Influencing the Measurement Dielectrics –polymers, ceramics Conductors –metals, doped silicon Edge Roughness Surface Roughness Wall angle Wall shape Contact hole issues

Human Variables Operator –Training –Skill level –Experience Automation –Minimizes operator variability Interpretation of the data –Where are the problems? –Are there any problems? –Is the instrument working properly?

GOAL The goal of this portion of the tutorial/ course is to make you aware of some of the metrology pitfalls so that you might avoid them.

Resolution of the Measurement System

Pixel Resolution (512 Pixels; 115 mm field width)

Effect of Pixel Size on the Measurement NIST RM 8090 sample measured in a commercial CD-SEM system 0.2 micrometer pitch structures and 50 micrometer pitch structures Instrument operating the same in both instances, the sample is the same only the magnification (= pixel size) differs. –512 pixels

Measurement Data SRM 2090 Prototype.2 µm lines –Maximum = µm –Minimum = µm –Mean = µm –3σ µm (1σ µm = 100,000x pixel size) 50 µm lines –Maximum = µm –Minimum = µm –Mean = µm –3σ µm (1σ µm = 5,000x pixel size)

Edge Detection

Types of Measurement Pitch measurement –No need to know the position of the “true” edge. –Self compensating measurement. –Requires that the lines have the same edge shape. –Requires that the same definition of “edge” be applied to both measured edges. Width Measurement –Position of the “true” edge is needed.

Types of Measurement Why is the type of measurement important? What is the effects of the electron beam interaction? How does the signal mechanism used make a difference?

SE vs. BSE Metrology SE images of the same structure consistently measure larger than the BSE image Experimental verification is difficult and good samples are needed. Further modeling must account for this difference.

Signal to Noise Ratio

SEM Signal and Contrast Signal is what we choose to collect. Signal can be: – Secondary electrons – Backscattered electrons – Transmitted electrons –etc. This signal is collected by an appropriate detector.

SEM Signal and Contrast All systems have noise, the higher the signal relative to the noise the better the signal quality and image. The signal quality can be expressed by the signal-to-noise ratio:

SEM Signal and Contrast Therefore, as the mean number of counts increases, the S/N ratio improves. Signal integration is primary to effective SEM and especially SEM metrology. Field emission instruments and digital frame averaging are two improvements that have helped to a great extent to improve the S/N ratio and thus SEM imaging and measurements.

Environmental Influences

Stray fields –AC Interference Vibration –Instrument induced –Site induced –Operator induced

AC Interference Mains frequency produced by AC current flowing in a loop Field Strength –Generally proportional to the current and inversely proportional to the area of the loop

Sources of AC Interference Coils of large electric motors or transformers AC Voltage regulators Main trunk power cables Mistakes in building wiring –Neutral return erroneously attached to the ground wire on a piece of equipment which is also grounded independently ground strap –Creates a large loop

Sources of AC Interference Two possible current paths between any two nodes in house wiring Instrument itself –internal wiring faults –loose connectors –incorrectly installed accessories –mechanical pump motors –TV Monitors –Improperly shielded computers –Ion pumps

Methods of Dealing with Stray Fields

AC Interference Solutions Field effects diminish with at least the third power of the distance –Attack the source first –Move source –Rotate source –Move instrument Shielding –Shield source –Shield instrument permalloy, Mu Metal

AC Interference Solutions There are consulting companies who will come to your lab and develop a plan to solve these problems. –field canceling coils etc. REMEMBER: these problems affect all types of instruments - because the the SEM is primarily an imaging system it is able to show these problems to you.

The Influence of Vibration

The SEM should be isolated from all forms of vibration –air flow –acoustically induced –mechanical Vibration will be included in the measurement

Measurement where tilt is present

Line Tilt/Line Rotation Errors For precise measurements the specimen should be: –Viewed flat (0 0 tilt) –Lines should be physically rotated to yield a symmetrical profile (may not be necessary in digitally addressable scan systems) Instrument should be calibrated at that working distance and the final lens current measured. Any further focusing should be done with “Z”

Effects of Raster Rotation

Electronic raster rotation rotation is often used to straighten structures on the viewing screen prior to measurement. Raster rotation can distort the image. If raster rotation is routinely used by the operator it should be tested.

Effects of Electronic Scan Shift

The SEM generally does not do flat field scanning. The scan is relatively linear at the high magnifications used for submicrometer work. Moving the scan to the far portions of the field by electronic scan shift can induce a distortion. This will affect the measurements across the field.

Stage Drift

SEM stages can move large distances at a rapid speed. Once the stage stops there is the potential for drift during the measurement process. If the image is frame averaged this will result in larger than expected measurements.

Stage Drift Temperature variations of only a couple of degrees can cause the stage mechanism to expand. Measurements may vary because of this expansion. Due to: –Stage motor heating –Room temperature variations –Etc.

Sample Drift Improperly adhered samples can move when irradiated by the electron beam. Can look like vibration or drift. Sample can become charged and fly off the stub – and usually lands where you do not want to to be.

Specimen Contamination

Contamination Specimen contamination remains one of the major SEM issues to be overcome. Contamination results from many places: –Vacuum system –Specimen surface –Stage lubricants –Specimen handling –Processing chemistry –Specimen itself

Specimen Cleaning (Contamination Removal)

Specimen Cleaning NIST has been investigation cleaning techniques for the removal of electron beam induced contamination. There are tools now available that will “scrub” the specimen chamber and the sample to remove contamination. Oxygen plasma is a successful technique and others are being pursued.

Specimen Damage

Irradiation of the sample with the electron beam can: –Electrically damage fragile devices. –Induce dimensional change in sensitive samples. –Damage the structures through melting or heating. Low accelerating voltage mode results in the depositing of most of the energy in the structure - not the substrate.

Specimen Charging

A full understanding of the effects of charging on image formation and metrology in an SEM has not been fully developed. Charging is a function of a number of parameters –accelerating voltage –material type –surface tilt –beam current –magnification Because of instrumental design differences, a sample that will charge in one instrument may not charge in another.

Specimen Charging Surface Tilt –Beam perpendicular to the sample gives the smallest electron yield –Charging potential increases –Increase tilt and the electron collection improves and the potential for charge build-up decreases Magnification –The beam current density increases as the magnification increases. more electrons/area scanned

Operator Factors Instrument Operation –focus –astigmatism –misalignments –misinterpretations Instrument maintenance –contamination