Position Sensing Glove Barbara Tuday Nathan Schreppler Lauren McIlwain Michael Nolan Team 6:
For the last 56 years ILC Dover Inc. has been a leader in engineered softgoods products in industry as well as government. They are commonly known for NASA’s spacesuits.
During a hand therapy session, physical therapists must manually measure the angle of each joint in a patient’s hand and wrist. Background
Problem Statement Our goal is to design, develop, produce, and test a prototypic, low-cost, and functional glove that will sense key finger, thumb, and wrist positions using compliant advanced technologies.
Key Joint Positions What joints do we actually need to measure? Thumb Fingers Wrist
Key Motions What motions do we need to measure? Flexion and Extension
Key Motions cont. Abduction & Adduction Pronation & Supination (Wrist Rotation) Total of 41 angles to be measured
Problem Constraints Safety United States Patents FDA 510 K Regulations
Customers ILC Dover Inc. Engineers Manufacturers Physical Therapists Patients Hospitals Retailers
Product Breakdown Overall Glove SensorsWiring Glove Material DAQ
Product Wants - Overall
Product Wants - Sensors
Sensor Metrics and Target Values MetricTarget Value Accuracy+/- 2.5 degrees Cost$ Cycles to Failure50,000 Resistance35,000 ohms Radius of Curvature0.1 inches No. of Parts Needed21 Ease of Flexibility (0-5)4 Frequency50 records per second Less Threatening (0-5)4
Product Wants - Wire
Wire Metrics and Target Values MetricTarget Values No. of Flexures before Failure 50,000 Radius of Curvature0.1 in Ease of Flexibility (0-5)5 Diameter0.025 in No. of Washes500
Concept Sensors Peratech (1 st choice) Spectra Symbol (used in prototype) Wire Calmont (1 st choice) Silver wire (used in prototype) DAQ LabView Material Knit, lightweight comfort glove
Modeling Gloves were modeled using acceptable substitutes for our materials Attachment Methods Wiring Various Securing Devices
Modeling – Attachment Methods
Modeling – Wiring
Modeling – Securing Devices
LabView Program Created for individual sensor
LabView Program cont. Convert From Voltage to Degrees Analog In Show Array on A Graph Display Max. & Min. Degrees Obtained
Wiring Sensors
Testing Attachment Method Glue Sensors Directly to Glove
Testing Attachment Method cont. Sensors Held to Glove by Patches
Initial Testing To initially test the sensors, a spring was used as a finger model Cloth was placed around the spring to represent the glove The sensor was attached and tested
Labview Expansion Labview was expanded in order to capture the results of several sensors simultaneously Able to connect five sensors at a time
Testing Position Due to the length of our sensors, a slanted configuration was favorable for close-fitting locations Sensors on knuckles could remain aligned with the joint
Calibrating Sensors Output data was collected for several angles on specific joints Data was not repeatable between tests and oftentimes varied greatly We had to try to understand where the error was coming from
Sensor Error We analyzed the situation and came up with two areas of possible error: Instability of sensor connections Absence of a stiff backing to provide support for the sensors
General Sensor Criteria After experiencing difficulty with the sensors we realized that these problems could arise with future sensors We decided to create tests that could analyze the behavior of these and future sensors
Behavior Testing of Individual Sensor Tests were performed, keeping the electrical connection tightly secured, for the following three types of backing on the sensors: No Backing Thick Backing (0.01”) – provided by Spectra Symbol Medium Backing (0.008”) – 3 layers of Scotch Packaging tape
Behavior Testing of Individual Sensor Bend Radius Test
Behavior Testing of Individual Sensor Tension Test
Behavior Testing of Individual Sensor Twist Test
Behavior Testing of Individual Sensor Pressure Test
Stabilizing Connection Since all the tests with the stabilized connections came out with satisfactory results, we felt that the previous problems were at the sensor connection, so we added a connection stiffener The backing on the sensor could help the stability
Testing New Sensors The new backing was then tested for repeatability with and without the connection stiffener This testing showed the backing offers some extra support which improves the repeatability
Calibrating New Sensors The new sensors were then calibrated Blocks were cut to the angles desired so that the calibration procedure could be easily repeated
Results The results from the final calibration came out better then the initial calibration These results, however, were still not accurate enough to meet the needs of the desired product
Cost of Materials
Future Work ILC should use the testing methods devised in this project first test the new Peratech sensors The sensors should then be tested on the actual glove If problems arise, the sensor connection should be addressed as done in this project
Thank You!! Questions?
Schedule
Total Cost
Final Prototype
Initial Calibration
Final Calibration
Final Calibration Data
Sensors Results