Qualification of Capillary Electrophoresis Instrument M. Koupparis.

Slides:



Advertisements
Similar presentations
Dr. Birgit Schmauser, BfArM, Bonn
Advertisements

OFFLINE COMPOSITION MEASURING SENSORS
Analytical Method Development and Validation
Cruces-Blanco, C., Gamiz-Gracia, L., Garcia-Campana A.M., Applications of Capillary Electrophoresis in Forensic Analytical Chemistry Trends in Analytical.
Calibration methods Chemistry 243.
Review of Analytical Methods for Nutrient Parameters focusing on Nitrates An Overview By Michael Jackman Chemistry Expert.
Yu. I. BARANOV and W. J. LAFFERTY Optical Technology Division Optical Technology Division National Institute of Standards and Technology, Gaithersburg,
Conductivity Testing of Unsaturated Soils A Presentation to the Case Western Reserve University May 6, 2004 By Andrew G. Heydinger Department of Civil.
Microspectrophotometry Validation. Reasons for Changing Instruments Reduced reliability. Limited efficiency. Limited availability and cost of replacement.
ANALYTIC METHODS II. PART Jana Švarcová.  Chromatography  Electrophoresis  Potentiometry  Titration  Spectrophotometry.
Chapter 26 Other Methods. Ion-Exchange Chromatography The mechanism of separation will be the exchange of ions from the column to the solution. Water.
PRINCIPLE AND INSTRUMENTATION OF CAPILLARY ELECTROPHORESIS
PHARM 462 PART / /31 Good Manufacturing Practices (GMP) VALIDATION of ANALYTICAL TEST METHODS.
World Health Organization
Figure 1. Influence of sample solvent on peak shape. The sample is dissolved in buffer with (a) 0%, (b) 30%, (c) 50%, and (d) 70% acetonitrile. Sample.
Selection of measuring instruments
Atomic Absorption Spectroscopy (AAS)
Detecting Melamine in the Food Supply Michael Parks CHEM 4101, Fall /04/09.
Classification of Instruments :
Validation of Analytical Method
Qualification and Validation Basic Principles of GMP.
Simultaneous quantification of bergenin, catechin, and gallic acid from Bergenia ciliata and Bergenia ligulata by using thin-layer chromatography 张 慧.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
1.1 General description - Sample dissolved in and transported by a mobile phase - Some components in sample interact more strongly with stationary phase.
Analytical considerations
Introduction to Analytical Chemistry
Supercritical Fluid Chromatography Introduction : Supercritical Fluid Chromatography (SFC) is a hybrid of gas and liquid chromatography that combines some.
Aquious Halide Concentration Determination via Leucigenin Quenching and Indirect CE Christopher Hampton, Dr. Eamonn F. Healy, Chemistry Department, St.
Supercritical Fluid Chromatography Introduction : Supercritical Fluid Chromatography (SFC) is a hybrid of gas and liquid chromatography that combines some.
Simultaneous determination of inorganic anions and cations by capillary electrophoresis with indirect UV detection I. Haumann, J. Boden, A. Mainka, U.
CHAPTER 21 SUPERCRITICAL-FLUID CHROMATOGRAPHY, CAPILLARY ELECTROPHORESIS, AND CAPILLARY ELECTROCHROMATOGRAPHY Introduction to Analytical Chemistry.
Accuracy Based Generation of Thermodynamic Properties for Light Water in RELAP5-3D 2010 IRUG Meeting Cliff Davis.
MS Calibration for Protein Profiles We need calibration for –Accurate mass value Mass error: (Measured Mass – Theoretical Mass) X 10 6 ppm Theoretical.
Quality Assurance How do you know your results are correct? How confident are you?
High Throughput Enantiospecific Separations Using Multiplexed Capillary Electrophoresis with Absorbance Detection Jeremy Kenseth, Andrea Bastin, and Brett.
CALIBRATION METHODS. For many analytical techniques, we need to evaluate the response of the unknown sample against the responses of a set of standards.
Factors Affecting Stability of Quinidine Melissa Usry and Sharmistha Basu-Dutt Department of Chemistry, University of West Georgia, Carrollton, GA
Capillary Electrophoresis (CE) PHAR Lecture Objectives By the end of the lecture, students should be able to: 1.Illustrate the CE instrumental.
Created with MindGenius Business 2005® Capillary Electrophoresis Capillary Electrophoresis Advantages Only needs nL sample High speed and resolution, virtually.
VALIDATIONS: TOWARDS RELAIBLE QUALITY OUTCOMES DR. SHIVPRAKASH MANAGING DIRECTOR SYNCHRON RESEARCH SERVICES PVT. LTD., INDIA.
Chapter 21: Molecules in motion Diffusion: the migration of matter down a concentration gradient. Thermal conduction: the migration of energy down a temperature.
B. Neidhart, W. Wegscheider (Eds.): Quality in Chemical Measurements © Springer-Verlag Berlin Heidelberg 2000 H. Albus Basic Course Experiments to Demonstrate.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 23 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
Principles Separation is carried out by applying a high potential (10-30 kV) to a narrow (25-75 pm) fused silica capillary filled with a mobile phase.
19-1 Supercritical Fluid Chromatography Theory Instrumentation Properties of supercritical fluid §Critical temperature àAbove temperature liquid cannot.
Introduction to Chromatography. Introduction Chromatography permit the scientist to separate closely related components of complex mixtures. In all chromatographic.
CAPILLARY ELECTROPHORESIS Prepared by Dr: Sherif A.Abdel-Gawad.
Analytical methods Houdková Chocholáčová. Analytical methods צ Absolute analytical methods (direct evaluation of the analytical signals with regard to.
Unit 4 Assignment 3 Practical techniques. Task 1 (P4) You will need to complete the following experiments. You will need to hand in a portfolio of your.
Welcome To my presentation.
Lecture 8 Peak Parameters and Quantitative chromatography
Chapter 1: The Nature of Analytical Chemistry
SEMINAR ON PRESENTED BY BRAHMABHATT BANSARI K. M. PHARM PART DEPARTMENT OF PHARMACEUTICS AND PHARMACEUTICAL TECHNOLGY L. M. COLLEGE OF PHARMACY.
Instrumental Analysis (I) HPLC Tutorial 8. Graded presentation Students in groups of 4-5 individuals are asked to prepare a presentation (weight=5% of.
Instrumental Analysis (I)  HPLC Tutorial #7 PHCMt561 – 5 th Sem. Pharm.
EQUIPMENT and METHOD VALIDATION
Field Equipment Calibration PH, Temperature, Conductivity, Turbidity, Dissolved Oxygen, Free & Total Chlorine Gregory Langland CCSF.
Capillary electrophoresis. Principle of electrophoresis Electrophoresis is the process of moving charged molecule in solution by applying an electric.
Fundamentals of Analytical Chemistry Eighth Edition.
Tutorial 11 Modes and Methods in Capillary Electrophoresis
EMSA 22 Lab Module 1 Field Equipment Calibration
Internal Standard method (experiment lab-4)
Режа: 1. Метрология тушунчаси 2. Валидация
SPECTROPHOTOMETRY Applied Chemistry.
Internal Flow: General Considerations
ANALYTICAL METHOD VALIDATION
Liquid Chromatography - Method Development and Validation
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Measurements & Error Analysis
Presentation transcript:

Qualification of Capillary Electrophoresis Instrument M. Koupparis

Principle of Capillary Electrophoresis Αρχή Τριχοειδούς Ηλεκτροφόρησης

Internal Structure of Capillary Εσωτερική Δομή Τριχοειδούς

Capillary Electrophoresis (CE) Τριχοειδής Ηλεκτροφόρηση

Species Movement in CE Τριχοειδής Ηλεκτροφόρηση Κίνηση Σωματιδίων

Parameters to be tested in Qualification of CE Instrument Temperature module –Stability –Accuracy –Cooling system Voltage module –Stability –Accuracy Detector module –Noise (Drift) –Wavelength accuracy –Linearity Injection module –Accuracy (hydrodynamically) –Linearity –Precision

Temperature Module (1) The correct thermostating of CE is essential for reproducible results of migration time and peak area. The generation of heat inside the capillary is unavoidable in electrodriven separations depending on the applied voltage and the resulting electric current. This heat (Joule heating) results in an increase of the buffer temperature causing a radial temperature profile and can also include perturbations to the electro-osmotic velocity profile.

Temperature Module (2) Both effects lead to peak broadening. The viscosity of the electrolyte and the stability of solutions and samples are also dependent on the temperature. In turn, the stability of migration times and peak areas is affected. Therefore stability and accuracy of the temperature have to be tested in an OQ procedure.

Temperature Module (3) Additionally, the thermostating system should be tested to ensure heat transport away from the surface area of the capillary. Temperature stability can be easily measured if the used CE software comes with the possibility to record the temperature profile during a separation run.

Temperature Module (4) Otherwise, especially for older CE instruments, an external thermometer with a record function is required to ensure a relatively high measurement frequency (>1 Hz). An external thermometer is always required to assess accuracy. The measurement with an external thermometer can be difficult to implement in CE systems which use a liquid cooling system.

Temperature Module (5) Stability parameter –During a run without sample injection the temperature is recorded over 10 min testing at a low, a high and an often used value –Acceptance criterion: ± 0.1 o C Accuracy parameter –During the temperature stability testing an external thermometer measures the temperature in the thermostated area. –Acceptance criterion: ± 2 o C

Temperature Module (6) Cooling parameter –Depending on the capillary length a buffer and an adequate voltage is chosen. –On the basis of the arising current and the resultant power per unit length (W/m) one can decide whether the cooling system works sufficiently on not. –Acceptance criterion: > 0.6 m (buffer 0.1 M phosphate pH 7.0, 30 kV/m) →3 W/m <0.6 m (buffer 0.05 M phosphate buffer pH 7.0, 50 kV) →5 W/m

Temperature Module (7) In order to get reproducible electropherograms it is essential that the temperature is kept constant during separation. It is less important to accurately adjust the temperature to a certain value. Therefore the acceptance criterion for stability is set tight at ± 0.1 o C. On the contrary for temperature accuracy it is set at a wider interval of ± 2 o C.

Voltage Module (1) The voltage is a central parameter in CE. Hence, the stability and the accuracy have to be tested –They are directly responsible for the reproducibility of peak migration times. The stability of the voltage is tested during a run without sample injection (provided that voltage data during separation can be recorded) over 10 min testing at a high and an often used value. –Acceptance criterion: ± 0.5%

Voltage Module (2) The direct accuracy measurement of the voltage proves to be difficult as the electrodes are not easily accessible for an independent voltage measurement during a separation. The voltage accuracy can be more easily obtained indirectly using effective mobilities of the sample analytes.

Voltage Module (3) The following equation presents the relationship between the effective mobility (μ eff ) and the applied voltage (V). Total length (L), effective length (l), migration times of an EOF marker (t EOF ) and one of the sample peaks (t s ) are required to calculate the effective mobility.

Voltage Module (4) The EOF is measured with the widely used EOF marker acetanlide. A sample containing acetanilide, 3,5- dihydroxide benzoic acid and nicotinic acid is separated. The effective mobilities are determined and compared to a reference. Acceptance criterion ± 4%.

Detector Module (1) The well-established concept for HPLC detectors can be largely employed for CE systems, with wider acceptance criteria due to the shorter optical path length. Noise parameter –During a run without sample injection the absorbance is recorded over 10 min testing at an often used value and the standard deviation of the absorbance is assessed. –Acceptance criterion: < 5x10 -5 AU

Detector Module (2) Wavelength Accuracy parameter –The distinctive spectra (in the area less than 300 nm) of three sample substances (acetaminophen, 3,5- dihydroxybenzoic acid and nicotinic acid) are measured and evaluated by comparison with reference spectra. –Acceptance criterion: ± 3 nm –The CE systems use also as internal wavelength calibration a holmium oxide filter comparing the maxima in the region of nm.

Detector Module (3) Linearity parameter –Using different concentrations of the three sample substances (0.1 – 5x10 -3 mol/L) a linear regression is modelled between peak areas and concentration determining the coefficient of determination as criterion. –Acceptance criterion: R 2 > 0.99 –A plot of residuals should also computed beside the linear regression, in order to properly recognize non-linearity and trends.

Injection Module (1) The common mode of injection in CE is the hydrodynamic, because of the higher repeatability (in comparison with the electrokinetic mode). Accuracy (hydrodynamically) parameter –Four weighted vials are filled with water and weighted again. One hundred injections of 50 mbar over 1 min from one vial into another vial at 20 o C are performed. –In the end all vials were weighted and two vials are used to compute the evaporation. –The injected amount is computed according to the Hagen – Poiseuille equation.

Injection module (2) The volumetric flow rate (V f ) is calculated by the volume (V) and the time flowing through a cylindrical tube and depends on the internal radius (r) of the tube, the pressure difference (Δp) between the two tube ends, the viscosity (η) of the streaming solution and the total length (l) of the tube. For the test bi-distelled water is pressed through the capillary at a frequently used pessure of 50 mbar thermostated at 20 o C and collected in the outer viakl. In this case the viscosity of water is mPa s.

Injection module (3) Acceptance criterion: ± 25% For example, the injected amount should be 7.7 mg ± 25% for a capillary length of 0.6 m injected 100 times for 1 min with 50 mbar. The weight of an inlet vial corrected for the evaporation should be reduced by this amount.

Injection module (4) Linearity parameter –Using different injection amounts ( mbar.s) a linear regression is modelled between the peak area and the product of time and pressure determining the coefficient of determination as criterion. –Acceptance criterion: R 2 > 0.99

Injection module (5) Precision parameter –The RSDs of peak areas and migration times are obtained from six consecutive runs of one sample. –Acceptance criterion Migration time RSD < 1.5% Peak area RSD < 2%

Holistic approach for PQ (1) Performing a run without injection, temperature stability and accuracy, as well as voltage stability and noise can be evaluated at the same time. Injection linearity and precision can be tested together using one injection setting of the linearity test for the precision test. Then three measurements of the linearity test and three additional ones with the same setting evaluated for the test of injection precision.

Holistic approach for PQ (2) Furthermore the wavelength accuracy can be obtained from the test of every parameter recording a spectrum of a suitable sample substance.