Rapid Actinide Analysis for Large Soil Samples

Slides:



Advertisements
Similar presentations
Ra in Water using MnO2 Resin: New Developments
Advertisements

Analysis of Actinide Elements from Large Environmental Samples
A Grace company Avoiding Matrix Interferences in Ion Chromatography Using Solid Phase Extraction Alltech Associates, Inc. Brandstraat 12, 9160 Lokeren.
A NALYTICAL S EPARATIONS G ROUP Megan Bennett, Ashlee Crable, Sherry Faye, Narek Gharibyan, Julie Gostic, and Chris Klug Subgroup Leader: Ralf Sudowe.
Magnesium metal is added to nitrogen gas
The Separation of Beryllium from Spectral Interfering Elements in Inductively Coupled Plasma – Atomic Emission Spectroscopic Analysis Daniel R. McAlister.
Water and Solutions. The importance of water Important component of solutions Water quality, composition, and pH can drastically affect an experiment.
New Applications of Rapid Column Extraction for Methods at SRS Sherrod L. Maxwell III Savannah River Site.
Analysis of Actinide Elements from Large Samples Henrieta Dulaiova Guebuem Kim Bill Burnett Florida State University and E. Philip Horwitz PG Research.
New Extraction Methods for Urine and Fecal Samples
 What is the molarity of a bleach containing 9.5 grams of NaOCl per liter of bleach?  How many grams of CaCl 2 would be dissolved in 0.75 L of water.
D. P. DiPrete, C. C. DiPrete, and R. A. Sigg
Technical Updates on UTEVA, TEVA and TRU Resins
Standard Addition Used with difficult matrices such as blood Difficult to get an accurate comparison of the amount of analyte in sample vs. amount of analyte.
Optimizing Radiochemical Methods at SRS
Rapid Separation Methods for Bioassay Samples S. L. Maxwell, III and D. J. Fauth Westinghouse Savannah River Site.
1) In the lab, students follow the chemical reaction above using 30 grams of Barium chloride. If 39 grams of silver chloride are actually produced, what.
Mg + N 2  Mg 3 N 2 Pb + Ag +  Pb 2+ + Ag Magnesium metal is burned in nitrogen Lead foil is immersed in silver nitrate A solution of ammonium sulfate.
Solution Stoichiometry
The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.
The Determination of 224 Ra, 226 Ra, & 228 Ra in Water Samples Michael Fern Eichrom Technologies, Inc.
Unit 10B Review Reg Chem When a solution sits out over a long period of time and water evaporates the concentration of the solution __________.
Determination of Sr-89/Sr-90 in primary cooling water
Standard Addition Used with difficult matrices such as blood Difficult to get an accurate comparison of the amount of analyte in sample vs. amount of analyte.
Analysis of Actinide Elements from Large Samples Henrieta Dulaiova Guebuem Kim Bill Burnett Florida State University and E. Philip Horwitz PG Research.
Radioactive Materials Analytical Laboratory Eichrom Technologies Inc. User Group Meeting May 2001 J.M. Giaquinto, L.D. Bible, D.C. Glasgow, J.M. Keller.
Actinide Analysis from Large- Volume Seawater Samples Bill Burnett & Guebuem Kim Department of Oceanography Florida State University and Phil Horwitz PG.
AP/IB Chemistry Chapter 4: Aqueous Solutions and Solution Stoichiometry.
Rapid Actinide Analysis for Large Soil Samples-Update Sherrod L. Maxwell, III Westinghouse Savannah River Company.
MnO 2 Resin: Concentration of Radium Isotopes Deok-Soo Moon Bill Burnett Department of Oceanography Florida State University Tallahassee, Florida.
Lecture 7: Digestion of Samples Digestion of water samples with Kjeldhl method for determination of N Destruction and extraction of soil and sludge for.
Less Manganese lost over C Manganese in Sea Water Excessive Background signal from NaCL Large loss of analyte.
00J ppt Rapid Extraction Methods for the Process Laboratory S. L. Maxwell, III V. D. Jones S. T. Nichols J. Satkowski M. A. Bernard Westinghouse.
Elution and Mounting Techniques for Americium using Eichrom TRU Resin Mitch Abbate Radiochemistry Technical Director Severn Trent Services St. Louis May.
Ulrika Nygren, June 2002 Determination of actinides in emergency preparedness using Diphonix or Actinide resin in combination with gamma spectrometry Ulrika.
General Engineering Laboratories Use of Eichrom Resins for Bioassay Pa-231 Bob Timm - GEL Tim Chandler - GEL Bill Burnett - FSU Mike Schultz - PerkinElmer.
Developments in Environmental Radiochemistry* Environmental Radioactivity Measurement Facility Department of Oceanography Florida State University Tallahassee,
Distribution of radionuclides in soils in surroundings of Bratislava, capital of the Slovak Republic. II. Man-made radionuclides Mátel Ľ., Rosskopfová.
The Determination of 226Ra in Water Samples
Working Through Mistakes, Miscues, and Mishaps in Actinide Separations Shane Knockemus US EPA / NAREL Montgomery, AL May 4, 2004.
Updates from Eichrom’s Tech Support Lab Anil Thakkar May 4 th 2004 Charleston, SC.
New Extraction Methods for Actinides in Urine at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site.
1) How many moles of water are in an 82.3g sample?
Recent Improvements in Rapid Column Separation Methods at SRS S. L. Maxwell, III, D. J. Fauth, R. C. Henderson, J. J. Smiley, S. R. Johnson Westinghouse.
Th, Pu & U Separation Using TEVA and UTEVA Resin
Recent Improvements in Rapid Column Separation Methods at SRS
New Fecal Method for Plutonium and Americium at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site.
Optimizing Peak Separation for Simultaneous Pu and U Measurements Shane Knockemus US EPA / NAREL May 3, 2005.
 Extraction of OM  Rock-Eval (Pyrolysis)  Elemental analysis  GC, HPLC & MS  SIRMS.
Rapid actinide measurements in environmental samples Dr Phil Warwick Geosciences Advisory Unit Southampton Oceanography Centre.
Determination of 210 Pb in mineral waters using Eichrom Resin Alžbeta Ďurecová, Dagmar Bursová, František Ďurec, Daniela Borošová State Institute of Public.
New Fecal Method for Plutonium and Americium at SRS Sherrod L. Maxwell III and David Fauth Savannah River Site.
Optimizing Radiochemical Methods at SRS (Ni-63, I-129, Actinides) Sherrod L. Maxwell, III Westinghouse Savannah River Company.
One-Column Separation and Analysis of Isotopic Am and Pu in Soil Samples Nidal M. Jadalla, Eberline Services, Albuquerque, NM Anil Thakkar, Eichrom Technologies,
Pyrosulfate fusion vs. TEVA / TRU forcing the issue Shane Knockemus U.S. EPA / NAREL Montgomery, AL November 11, 2002.
WARM UP 1. Write the equation for the neutralization reaction between sulfuric acid (H 2 SO 4 ) and ammonium hydroxide (NH 4 OH).
An Improved Method for Determination of Ra-228 * Jamie Christoff & Bill Burnett Department of Oceanography Florida State University “The Double-Pass Approach”
Anil H.Thakkar Eichrom Industries, Inc.. Darien, IL
CERIMETRI SMK Negeri 13 Bandung.
The Determination of 226Ra in Water Samples
Strontium Extraction from Urine Samples
Radioactive Waste Processing at the Savannah River Site
Ra in Water using MnO2 Resin: Update
Technical Updates on TRU and TEVA Resins
Rapid Emergency Methods for Actinides and Sr-89/90
Aqueous Solutions Pt. 2.
Solution Concentrations
LO The student can design, and/or interpret data from, an experiment that uses titration to determine the concentration of an analyte in a solution.
National Air and Radiation Environmental Lab
Unit 11 – Solutions and Ions in Aqueous Solutions
Presentation transcript:

Rapid Actinide Analysis for Large Soil Samples Sherrod L. Maxwell, III Westinghouse Savannah River Company

Background Need for improved analysis of large soil samples for actinides large sample size - lower detection limits total dissolution-refractory actinides minimize problems with column load solutions removal of soil matrix interferences consistency in tracer recoveries good alpha peak resolution

Recent Work Using Actinide Resin (Dipex) and Diphonix Resin Actinide Resin and Diphonex Resin W.C. Burnett et al, "Preconcentration of Actinide Elements from Soils and Large Volume Water Samples Using Extraction Chromatography". Journal of Radioanalytical & Nuclear Chemistry. Vol. 226, (May 1997), 121-127. W.C. Burnett et al, Efficient Preconcentration and Separation of Actinide Elements from Large Soil and Sediment Samples, Analytical Chemistry, Vol. 72, pp. 4882-4887(2000) Diphonex Resin S.L. Maxwell and S. Nichols, “Actinide Recovery Method for Large Soil Samples”, Radioactivity and Radiochemistry Journal (January, 2001)

Actinide Resin/Diphonix Resin-Soil Eliminated soil matrix but still had some disadvantages: Actinide Resin/Diphonix Resin (Burnett) high phosphate -HEDPA solvent destruction (Fenton’s Reagent)- large TRU column Fe in Fenton’s reagent -can’t stack TEVA+TRU cartridges Diphonix (Maxwell)-requires microwave dissolution of resin microwave work takes time/ can be tedious using HEDPA strip instead of microwave on fecal samples Pre-concentration steps take time and can have losses Need a simple, user-friendly, effective approach

New Soil Method Cerium fluoride precipitation for soil matrix removal CeF3 used for high iron water samples M. Schultz et al, “Analysis of Am, Pu and Th in Large Volume Water Samples in the presence of High Concentrations of Iron”, International Radiochemical Analysis Conference, Maidstone, Kent, UK, Sept. 2002 Rapid elimination of soil matrix problems Rare earth removal for large soil samples anyway-no extra separation time

Benefits Faster Total dissolution (5-10 grams) After drying, blending, overnight furnace: 1 day sample preparation + 1 day column extraction Previous method 4-5 days Total dissolution (5-10 grams) High recoveries/clean spectra Eliminates interferences (Ex. Th, Po) Th-228 on Pu-238, Am-241, etc Po-210 affecting U-232

Soil Sample Preparation Total dissolution and matrix removal (1 day) After drying and blending, heat at 550C Ash with nitric acid/HF to remove silicon, then fusion Fusion in zirconium crucibles for 20 minutes at 700C 10 grams sodium peroxide + 5 grams sodium hydroxide (or as needed for larger samples) Hydroxide precipitation Iron carrier with TiCl3 reductant Barium to eliminate carbonate interference Additional Ce carrier sometimes added here for very sandy samples to enhance precipitation

Soil Sample Preparation, contd. Redissolve in dilute HCl and perform cerium fluoride precipitation (1 mg to ~5 mg Ce) with TiCl3 present Rinse precipitate with 0.25M HCl- 6MHF with 0.02M rongalite present Soil load solution: 5 mLs 3M HNO3-0.25M boric acid 6 mL 7M HNO3 7.5 mL 2M AL(NO3)3

Transfer to crucible after silicon removal

10 grams sodium peroxide + 5 grams sodium hydroxide 700C for 20 minutes

Just add water after cooling briefly

Dilute to 1 liter with water with iron carrier, TiCl3, and barium present

After centrifuging, acidify and add TiCl3, Ce and HF.

Place on ice, then centrifuge

Rinse with dilute HCl, HF with rongalite present

Redissolve in acid, boric acid and aluminum nitrate

TEVA +TRU Resin-Stacked Cartridges

2.0mL TRU-Resin (3mL if needed) Cerium fluoride / Alpha spectrometry Actinides in Soil 1) Redissolve in 18.5 mL 3M HN03 - 0.8M Al(NO3)3- 0.07M boric acid 2) Add 0.5 mL 1.5M Sulfamic Acid + 1.25 mL 1.5M Ascorbic Acid 3) Add 1 mL 3 M Sodium Nitrite Rinse Beaker rinse: 3mL 5MHN03 Separate cartridges: TEVA: 5 mL 3MHN03 Collect/acidify - with 2 mL16M HN03 then to TRU TEVA:10 mL5M HNO3 8 mL3M HN03 Th Elution 20mL 9MHCI Pu (and/or Np) Elution 20mL 0.10MHCl - 0.05MHF - 0.03M TiCl3 Remove TRU cartridge: 1) Elute Am with 15mL 4M HCI 2) Add 3M HNO3 rinse from TEVA 3) 10 mL 6M HNO3 -remove any Po-210 4) 15 mL 4M HCl-0.2M HF-remove Th 5) Elute U with 15mL 0.1M ammonium bioxalate Add 0.5 mL 30 wt% H2O2 2mL TEVA Resin (50-100 um) Cerium fluoride Alpha spectrometry 2.0mL TRU-Resin (3mL if needed) (50-100 um) Cerium fluoride / Alpha spectrometry

(warm and rinse original beaker) Am/RE Removal on TEVA 1) Evaporate 4M HCl with 5mL con.HNO3, 50 uL of 1.8M H2SO4, then ash with nitric acid and hydrogen peroxide 2) Redissolve in 5 mL of 4M NH4SCN, warm gently. Rinse Beaker rinse: 3mL 4M NH4SCN, warm 10 mL 1.5 M NH4SCN to column Am Elution 25 mL 1M HCl (warm and rinse original beaker) 2mL TEVA Resin (50-100 um) Cerium fluoride Alpha spectrometry

Final Am Clean-up on TEVA

QAP 0309-SOIL SRS EML Ratio Pu-238 14.9 14.6 1.021 Pu-239 31.6 30.4 1.039 Am-241 18.3 18.4 0.995 U-234 117.7 127.3 0.925 U-238 119.9 127.1 0.943 Results in Bq/kg 5 gram sample analyzed

QAP 0403-SOIL SRS EML Ratio Pu-238 0.826 0.82 1.007 Pu-239 19.8 22.82 0.868 Am-241 13.2 13.0 1.015 U-234 86.7 87.22 0.994 U-238 93.7 89.73 1.044 Results in Bq/kg 5 gram sample analyzed

Summary Large sample size Total dissolution fast fusion, multiple samples Removal of soil matrix interferences Cerium fluoride precipitation Reduced sample preparation time by 2 days Good tracer recoveries and alpha peak resolution Reduced spectral interferences