Advanced Spectroscopy 1. General Aspects
Revision 1. What is the difference between absorption and emission of radiation? Abs. – uptake; Em. - release Exc. Gnd radiation Absorption Emission
Revision 2. Why are the peak wavelengths in the absorption and emission spectra of the same species identical? the energy gap is the same going up and coming back down
Revision 3. In spectroscopic terms, what is the difference between an atom and a molecule? How do their spectra differ in appearance? atom is totally free of any connection to any other species; doesn’t have be neutral; single wavelength of absorption: LINE spectrum molecule has bonds (not necessarily covalent) to other species; eg Na + in water; range of wavelengths: BAND spectrum
Revision 4. Rank the following regions of the electromagnetic spectrum - ultraviolet, visible and infrared - in terms of increasing energy, frequency and wavelength. Energy Freq. UV Visible IR Wavelength
Revision 5. How do absorbance and transmittance differ? Which is more useful for analytical purposes? Why? transmittance is simple ratio of intensity out/intensity in absorbance is log T absorbance is more useful because it is linear with concentration
Revision State Beer’s Law, and explain the meaning of each term. Under what situations does Beer’s Law not apply? A = abc A = absorbance a = constant specific to species b = pathlength c = concentration low & high absorbances (for most species 1)
Revision 7. Explain how you could determine what concentration range for a given species obeyed Beer’s Law. dilute a standard until you get an absorbance <2 use simple proportions to work out the concentration that would give an absorbance of 0.2 round this conc. to a convenient value make stds 1x, 2x and 4x
Revision 8. Draw a schematic diagram showing the components of a typical absorption spectrophotometer. Radiation source Sample cell Wavelength selector DetectorReadout
1.1 Radiation sources all spectroscopic instruments require one radiation is measured to provide the analytical measurement: spectrum or absorbance the type of source varies particularly between absorption and emission instruments.
Exercise 1.1 What is the most important difference between the radiation source in absorption and emission instruments? absorption: separate lamp source emission: excited sample is source
Role of radiation sources to provide radiation that can be absorbed at specific wavelengths by the analyte allowing a comparison of intensity before and after sample
General requirements it must produce radiation in the wavelength range that the instrument is designed to operate most sources are continuous produce radiation at every wavelength across the range they are designed to work in Exercise 1.2 One absorption instrument that you are familiar with does not use a continuous source. Which one is it? AAS
Desirable characteristics the intensity should be consistent across the range (normally it will taper off at the extremes of the range), the intensity should not fluctuate over time the intensity of radiation should be not be too low the intensity of radiation should be not be too high
Exercise 1.3 a) consistent across the range no missing bits or big spikes b) consistent over time so that measurements don’t drift (no need to re-zero all the time) c) not too low detector inaccuracy d) not too high decompose the sample & detector inaccuracy
Wavelength selectors The role of a wavelength selector To reduce the range of wavelengths reaching the detector to those near the absorption (or emission) wavelength of the analyte.
Why is one needed? absorption of radiation at one wavelength is not affected by the presence of others it is the detector that creates the need for a selector it can’t tell the difference between wavelengths and responds to all of them without a selector: no spectra, since only one measurement is available totally inaccurate absorbance measurements
Illustration of absorbance problem assume the following: visible abs. spectr. ( nm) each 10 nm range has 100 units of radiation from the source (a total of 4000 units) sample absorbs 50% of radiation in nm band
Exercise 1.4 a) How many units of radiation will reach the detector with the sample out? 100 b) How many units of radiation will reach the detector with the sample in? 50 c) What should the % transmittance at be? 50%
Exercise 1.4 d) What is the total intensity reaching the detector with the sample in (given no wavelength selector)? 4000 – 50 = 3950 e) What is the actual %T that the instrument will display? 100 x (3950 ÷ 4000) = 98.8%
without the wavelength selector, the detector is swamped by lots of radiation that has nothing to do with the analyte’s absorption radiation absorbed radiation not absorbed absorption wavelength range
Types of wavelength selectors an ideal wavelength selector would allow one wavelength of radiation only to pass to the detector actually allow through a range of wavelengths (this is known as the bandpass) how wide that range is depends on the design of the selector, and also the experimental conditions required two basic classes of wavelength selector: monochromators filters
Filter sheet of plastic or glass that absorbs most radiation cheap simple no moving parts => portable wide range of wavelengths filter must chosen to match absorption peak
Monochromators a series of optics inside a lightproof box entry and exit slits which allows the radiation of all wavelengths in and a narrow range of wavelengths out
Monochromators dispersing medium is either a prism or a diffraction grating. work by causing the different wavelengths of radiation to change their direction at different angles depending the wavelength results in a band of single wavelengths which are directed towards the exit slit because it is very narrow, only a small range of wavelengths can actually exit and reach the detector
Monochromators to select a wavelength, the prism or grating rotates causing the band of radiation to shift, moving a different wavelength over the exit slit Monochromator setting 550 nm 500 nm
Prisms vs gratings prisms are simpler but less accurate gratings the most commonly used a grooved surface, where the grooves are extremely close together 100’s-1000’s grooves/mm transmission or more commonly reflection better performance in terms of throughput and consistency
The significance of the exit slit how wide the exit slit is determines the range of wavelengths that come through known as the slit width usually the exit slit is adjustable as the slit width decreases, the range of wavelengths that are passed by the monochromator decreases (and vice versa)
The significance of the exit slit the actual slit width is not important in itself it is not equal to the range of wavelengths that pass through it the important measure is spectral bandwidth (or bandpass: the wavelength interval of radiation leaving the monochromator eg: 500 nm; 1 nm the radiation leaving the exit slit would range from to nm the bandpass affects the appearance of the spectrum
Positioning the wavelength selector can go either before or after the sample cell better afterwards in some instruments, it must go before determining factor is the energy of the beam from the source whole UV beam may decompose the sample =>the selector is placed before the sample Source Sample selector Source selector Sample
if before sample, light from surrounds can enter and reach detector known as stray light: any radiation that reaches the detector that is not from the source light-seal doors over the sample compartment are required if after sample, wavelength selector will block most external light
Sample holders shouldn’t absorb where analyte is absorbing
Detectors respond only to the total intensity of radiation cannot distinguish between different wavelengths high sensitivity high signal-to-background ratio constant response across the range of wavelengths rapid response linear response (i.e. output is proportional to radiant intensity) minimal response to no radiation (known as dark current)
Evolution of detectors human eye photographic plates and film electrical electronic most generate a current from the radiation energy
1.5 Instrument configurations Scanning or non-scanning the ability to record a spectrum requires automatic wavelength changes and many of them two key requirements: measure the intensity at wavelengths that are very close together vary the wavelength without human assistance
Exercise 1.5 An instrument using a filter cannot be a scanning instrument. True Cannot measure close together wavelengths An instrument using a monochromator must be a scanning instrument. False A monochromator doesn’t have to have a motor
use a stepping motor which rotates the grating or prism by very small angles other ways exist
Exercise 1.6 Is the filter photometer scanning or non-scanning? non-scanning
Single or double beam absorption instruments require two intensity measurements going into the sample (“before”) passing through before computers, two ways of measuring the “before”
Single beam only one sample holder “before” measurement is taken using a blank at the start if the wavelength is changed, the instrument needs re- zeroing Double beam two sample holders and a split optical system “before” intensity measured continually really double-path, not double-beam
Rotating chopper (see below) Mirror Semi-transparent mirror transparent sector mirrored sector Design of Chopper Source Detector
Double-beam should double-beam configurations have two of everything: beams, sources, detectors etc cost no way that the two systems could be made exactly equal extra optics (mirrors) mean that less radiation goes through the system (known as throughput) more bits and pieces which can get out of alignment necessary to have two matched cells – difficult/expensive
Single-beam (no PC) does not have double-beam problems or requirements without a computer it can’t record spectra when the wavelength changes, re-zero take the sample out and put the reference back in Exercise 1.8 Why is it necessary to re-zero the instrument when the wavelength changes? source output and detector response vary
Single-beam (with PC) advent of desktop computers revolutionised spectrometer design allows the spectrum baseline to be measured at the start stored in memory a single-beam instrument can scan double-beam instruments with PCs do exist – they seem to be overkill (one or the other!)
Exercise 1.9 Is the filter photometer single– or double-beam? single
Dispersive or non-dispersive means “dividing up” a component (eg grating) that divides up the radiation by wavelength doesn’t apply to filter-based instruments ND1: no wavelength selector at all employed for pollution monitoring in harsh environments achieve some measure of selectivity by clever use of reference materials non-scanning mostly in the infrared region
Dispersive or non-dispersive ND2: use a mathematic function Fourier transform scanning they are very fast simpler internal configuration also mostly in the infrared region ND3: a detector that is capable of distinguishing between different wavelengths of radiation only in the X-ray region
Exercise 1.10 Is the filter photometer dispersive or non-dispersive? non-dispersive
Single- or multi-channel refers to the number of detectors scanning using a stepping-motor monochromator takes time and the optics can become misaligned alternative is numerous detectors, each responsible for a range of wavelengths a dispersing medium is still needed no exit slit or motor parts some are not capable of producing a spectrum, only numerous wavelength measurements others can a continuous bank of very small detectors
Single- or multi-channel Radiation source Sample cell Dispersing element Array of detectors
Exercise 1.11 Is the filter photometer single or multi-channel? single
Transmission or reflectance does not apply to emission instruments familiar with spectroscopy involving a semi-transparent liquid or solid how much radiation passes through the sample not all materials will transmit radiation if opaque, unabsorbed light is reflected (bounces off) Exercise 1.12 Give examples of materials which transmission spectroscopy would be unsuitable for. paint, fabric, some plastics
Transmission or reflectance Reflectance spectroscopy measures radiation that bounces off the surface of an opaque material direct reflection by a mirror-like surface (called specular) scattered at a variety of angles by a rough surface (called diffuse) basic principles still apply: certain wavelengths will be absorbed, and an absorption spectrum obtained
Transmission I in I out Reflectance I in I out
Exercise 1.13 Is the filter photometer transmission or reflectance? transmission
Exercise 1.14 Now categorise all the instruments on the provided sheet.