Echocardiography of Prosthetic Valves

Slides:



Advertisements
Similar presentations
Edward Evans MD FACC Desoto Heart Clinic Disclosures Medtronic:speaker St. Jude Medical:speaker.
Advertisements

Treatment of Mitral Stenosis.  Medical  Valvotomy  Interventional  Surgical Treatment of Mitral Stenosis.
Mitral Stenosis Meghan York September 23, 2009.
Non-Invasive Assessment of Prosthetic Valves
Echocardiographic Evaluation of Prosthetic Heart Valves
Heart sound.
AORTIC VALVE PROSTHESIS. Basic Types of Artificial Heart Valves Mechanical – made of synthetic material Tissue valve – made from animal tissues called.
There are three major types of Blood flow Doppler patterns: 1) Continuous Wave 2) Pulse Wave 3) Color Flow DOPPLER ULTRASOUND.
Mitral Valve Prosthesis. Types of mechanical valves:  Ball-cage valve (Starr-Edwards is the most common). - a spherical occluder is contained by metal.
Prosthetic Heart Valves Presentation by Brian Meyer.
Artificial Heart Valves By: Brian Volpe. What is it? An artificial heart valve is a mechanism that mimics the function of a human heart valve It’s used.
Artificial Heart Valves. Heart/Heart Valves Heart consists of: Right Atrium and Ventricle Atrium Left Atrium and Ventricle Two Types of Valves: Atrioventricular.
AVR: Choice of Prosthesis Tirone E. David University of Toronto.
43rd PHA Annual Convention
Mostafa Eladawy MD,AFSA Basics of Echocardiography.
MITRAL REGURGITATION. 2D ASSESSMENT LOOK CAREFULLY AT THE MITRAL VALVE APPARATUS.
© Continuing Medical Implementation …...bridging the care gap Valvular Heart Disease Aortic Stenosis.
Echo Conference April 6, 2011 Frances Canet, MD. Causes and Anatomy Assessment of Mitral Stenosis How to Grade Mitral Stenosis Cases and Application Outline.
PROSTHETIC VALVE BOARD REVIEW
Aortic Regurgitation Mohammed AL Ghamdi.
The Björk-Shiley Heart Valve. Normal cardiac function Natural aortic valve mitral valve aortic valve (to body) pulmonary valve (to lungs) tricuspid valve.
Echocardiographic Evaluation of Prosthetic Valves, Part I
Quantitative Assessment of
Ventricular Diastolic Filling and Function
Alsharqia.riyadh Echo meeting Dammam KSA SAYED ABOU EL SOUD MD SBCC.
Common Clinical Scenarios *Younger people *Younger people _Functional murmur vs _Functional murmur vs _ MVP vs _ MVP vs _ AS _ AS *Older people _Aortic.
Catheter estimation of stenotic valves
ECHOCARDIOGRAPHIC ASSESSMENT OF AORTIC REGURGITATION
PROSTHETIC HEART VALVE : Types, Selection, and Functional assessment Dr SHAJUDEEN 2 nd Year DM Resident.
AORTIC REGURGITATION AORTIC REGURGITATION ETIOLOGY LEAFLETS: RHEUMATIC H.D. CALCIFIED PROLAPSE ENDOCARDITIS TRAUMA RHEUMATOID ARTHRITIS ANNULUS, ROOT.
Valve Surgery V.Rohn. Valve Surgery History before the era of ECC 1925 – Suttar – first successful digital commisurolysis of mitral valve 1952 – Hufnagel.
Guidelines for the Echocardiographic Assessment of
Prosthetic Valve Dysfunction
1 Heart Valve Disease Valvular heart disease Valves make sure blood flows in a forward direction –no backward leakage Tricuspid and mitral.
Adult Echocardiography Lesson Two Anatomy Review Harry H. Holdorf.
Prosthetic Heart Valve Sounds & Murmurs Chapter 18 Are G. Talking, MD, FACC Instructor Patricia L. Thomas, MBA, RCIS.
3/98medslides.com1 Aortic Valve Homografts A Cinical Perspective Michael E. Staab, MD Rick A. Nishimura, MD Joseph A. Dearani, MD Thomas A. Orszulak, MD.
Date of download: 6/3/2016 Copyright © The American College of Cardiology. All rights reserved. From: Pregnancy in Patients With Pre-Existing Cardiomyopathies.
Lawrence Scotten1 David Walker1 Peter Harbott2 Jurgen de Hart3
Date of download: 6/23/2016 Copyright © The American College of Cardiology. All rights reserved. From: Choice of Prosthetic Heart Valve in Adults: An Update.
Division of Cardiology Department of Internal Medicine Tae Kyung Yu
Date of download: 6/29/2016 Copyright © The American College of Cardiology. All rights reserved. From: The Emerging Role of Exercise Testing and Stress.
Thanks ………… Echocardiographic Evaluation Of Prosthetic Cardiac Valves Dr Gaurav Kumar Chaudhary MD,DM( Cardiology) Assistant Professor Department of Cardiology.
Twelve Months and Beyond: Long-Term Results of the Direct Flow Medical Repositionable and Retrievable Pericardial Valve for Percutaneous Aortic Valve Replacement.
Date of download: 7/8/2016 Copyright © The American College of Cardiology. All rights reserved. From: Choice of prosthetic heart valve for adult patients.
© free-ppt-templates.com 2017 AHA/ACC Focused Update of Valvular Heart Disease Guideline of 2014 DR. OMAR SHAHID TR CARDIOLOGY SZH.
Original slides courtesy of Dr. Alex Morss
Basic Echo As I understand it… Dr Claudia Wong.
Cardiothoracic Surgery
Imaging and Quantification of Aortic Regurgitation after TAVI
AORTIC STENOSIS.
Aortic Valve Homografts A Cinical Perspective
Valve Durability Considerations: Fluid Dynamic Principles, Assessment Techniques, Predictors, and Minimum Standards Jorge H. Jimenez, Ph.D. Georgia Institute.
Clinical, Echocardiographic, and Pathologic Features of Aortic Wall Dehiscence of Porcine Bioprosthetic Valves: A Cause of Rapidly Progressive Mitral.
DIASTOLIC DYSFUNCTION and DIASTOLIC HEART FAILURE
PULMONARY REGURGITATION- SEVERITY ASSESSMENT - fazil bishara
RICK A. NISHIMURA, M. D. , FLETCHER A. MILLER, M. D. , MARK J
Homograft Replacement of the aortic valve:Ten-year results
Adult Echocardiography Lecture 11 Prosthetic Valves
ADULT ECHOCARDIOGRAPHY Lesson Seven The Mitral Valve
Interpreting Cardiac Echo Reports
Late Follow-Up from the PARTNER Aortic Valve-in-Valve Registry
ADULT ECHOCARDIOGRAPHY Lesson Eight The Tricuspid Valve
ADULT ECHOCARDIOGRAPHY Lesson Eight The Tricuspid Valve
ADULT ECHOCARDIOGRAPHY Lesson Seven The Mitral Valve
Timing and indication of surgery- Mitral valve
Figure 2 Echocardiographic methods to estimate left atrial pressure
Ανεπάρκεια Μιτροειδούς : Σταυρόλεξο για δυνατούς λύτες Ποσοτικοποίηση με Ηχωκαρδιογράφημα και ο ρόλος των δοκιμασιών φόρτισης Μαρία Μπόνου Διευθύντρια,
AORTIC VALVE Aortic Valve is located at the junction of LV outflow tract and ascending Aorta. Aortic valve consists of 3 components – annulus, cusp and.
Presentation transcript:

Echocardiography of Prosthetic Valves Dr. Tehrani

Different Types of Valves Homografts (allograft) Cadaveric human aortic and pulmonary valves Heterograft (Xenograft) Prosthetic Valves Bioprosthetic valves Pig aortic valve Bovine pericardial (other) Mechanical Urethane ball in a cage Single or multiple discs

Homografts (allograft)

Homografts Homograft Valves Harvested soon after death w/ the endothelium still viable Preparation for implantation Storage in ABX Cryopreservation (more recently) No anticoagulation Low incidence of endocarditis Failure due to gradual aortic incompetence

Homografts Position Mitral Fitted w/ stent, not proved successful (high failure rate at 5 years) Stentless grafts not an option for MVR Aortic Stentless Subcoronary Root Replacement What should I say about Tricuspid and pulmonic positions???

Echocardiography of Stentless Aortic Homografts Fig. 38.1 weyman, and 10-A Doppler flow characteristics similar to native valve. Only 2-D evidence: Increased Echo intensity, and Thickness of aortic annulus.

Stentless Heterografts (Xenograft)

Stentless Heterografts (Xenograft) Same utility as allografts for AVR: Subcoronary implantation, and Root replacement Advantage over allografts is wider availability Durability is at least as good as allografts Insert Pic from Doty’s Lecture

Prosthetic Valves

Bio-Prosthetic and Mechanical Prosthetic Prosthetic Valves Bio-Prosthetic and Mechanical Prosthetic All Prosthetic valves have a sewing ring anchored to the native tissue with sutures The occluding portion of the valve: Tissue leaflets  Bio-Prosthetic Single or multiple discs/ Urethane ball in a cage  Mechanical Prosthetic

Bio-prosthetic Valves

Carpentier-Edwards bioprosthesis Bioprosthetic Valves Two types, of occluding mechanism Porcine aortic valve (the valve size of the biggest pig is limiting) Hancock, and Carpentier-Edwards bioprosthesis Carpentier-Edwards bioprosthesis

Bioprosthetic valves Bovine Pericadium leaflets are shaped to size. More choices Echocardiographiaclly these two valves types are indistinguishable. Fig. 38-2, and 3, Actual pictures Ionescu-Shiley (1976)

Bioprosthetic valves Mitral Position 38-25A, 38-32, two diff views of mitral bioprosthetic 2-D ECHOCARDIOGRAPHIC APPEARANCE

Bioprosthetic valves Aortic Position 38-34, 38-27, aortic bioprosthetic 2-D ECHOCARDIOGRAPHIC APPEARANCE

Bioprosthesis Used extensively in a variety of sites: Aortic Mitral Tricuspid Advantage: Low thrombogenicity => No anticoagulation

Bioprosthesis Disadvantages: Less durable than mechanical prosthesis Mitral position worse Due to greater backpressure gradient Dysfunction: Leaflet thickening, and Ca++ Fracture, tears, or progressive stenosis In vivo, roughly 10% of normal bioprosthetic valves have some leakage.

Overview of Various Devices Bio-Prosthetic Valves Mechanical Prosthetic Valves

Mechanical Vlaves Ball-and-Cage Valves Tilting disc Prosthesis Single disk Bileaflet

Mechanical Prosthesis The occluding mechanism dictates both: The echocardiographic appearance of the valve, and The flow pattern through the valve To assess performance, the type of valve implanted must be known

Ball-and-Cage Valves First implanted by starr and Harken in 1960. Figure 38-4

Opening and closure of the ball-valve Ball-and-Cage Valves Fig. 38-28 Opening and closure of the ball-valve

Ball-and-Cage Valves Axisymmetric flow around the valve. Stagnant flow in the shadow of the ball. Figures 38-5,

Ball-and-Cage Valves Doppler assessment at the margins of the ball Fig. 38-6 Doppler assessment at the margins of the ball

M-Mode assessment of Ball-Cage Valve Ball-and-Cage Valves 38-29, m-mode M-Mode assessment of Ball-Cage Valve

Ball-and-Cage Valves Durable Mitral position Satisfactory profile with the largest size (34 or 32 mm diameter devices) Can affect the interventricular septum Aortic position Small prosthesis required, which can be associated with significant gradient Regurgitation limited to closure backflow.

Tilting Disc Prosthesis All essentially similar consisting of Circular prosthetic material, and One or two hinged and mobile disc(s) Disc attachment to the ring is eccentric Closure occurs by backpressure on the largest portion of the disk

Single Disc Prosthesis Single Disc devices: Hall-Medtronics monostrut Bjork-Shiley Opening arch is 55-70 degrees Flow orifice: Major and minor flow orifices Streamlines of flow passing through the sewing ring and then laterally out and around the prosthetic disc 38-7-c, 38-8-a, comment that it is identical of BS-standard

Single Disc Prosthesis Bjork-Shiley Standard Convex-concave Many other variations in the market All of these devices have a zone of stagnation behind the disc  thrombus formation 38-7-a, and 38-8-b

Single Disc Prosthesis Bjork-Shiley in the Mitral position 38-25-b, echo of BS tilting disc

Single Disc Prosthesis Leak around: Central strut Dominant jet Between the occluding disc and sewing ring. Two smaller peripheral jets Normal hemodynamics Reg.Frac. approx. 12% Tachycardia, and low output Reg.Frac. upto 37% Figure 38-10- a and b

Single Disc Prosthesis

Single Disc Prosthesis Dysfunction Gradual ingrowth of fibrous tissue (panus) Flow obstruction Intermittent sticking of the valve with associated flash pulmonary edema

Bileaflet Mechanical Prosthesis St. Jude prosthesis The most commonly used. Two equal sized semi-circular leaflets attached by a midline hinge. Discs can tilt in excess of 80 degrees, resulting in larger: Orifice area Figure 38-11

Bileaflet Mechanical Prosthesis St. Jude prosthesis The most commonly used. Two equal sized semi-circular leaflets attached by a midline hinge. Discs can tilt in excess of 80 degrees, resulting in larger: Orifice area 38-11 Regurgitant back flow

Bileaflet Mechanical Prosthesis Fig. 38-14 (a&b) Regurgitation occurs at the disc margins The regurgitant jets converge toward the center of the valve

Bileaflet Mechanical Prosthesis 38-25-c, 38-33, echo of st.judes St. Jude valve in the mitral position.

Imaging of Prosthetic Valves

Special Problems of 2-D Imaging Artificial Valves Echocardiographs are calibrated to measure distance based on the speed of sound in tissue. Prosthetic valves have different acoustic properties than tissue. Hence, distortion of: Size Location, and Appearance, of the prosthesis. Insert this and the following slides in the section for bioprosthetic valves, right before x? the dopplar features

Special problems of artificial valves Intense reverberation, and Shadowing Less gain leads to less: Reverberation, and Shadowing, as well as Better visualization of non-biologic components of the valve HOWEVER  Decreased definition of cardiac structures

Special problems of artificial valves First image at normal settings, then Reduce the gain to interrogate the leaflets of Bio-prosthetic valves. Utilize multiple views.

Prosthetic Valve Pathology Prosthetic Valve Stenosis Aortic Mitral Prosthetic Valve Regurgitation

G E N E R A L L Y Prosthetic Stenosis (and Regurgitation) is: A question of degree, Not a question of whether.

Prosthetic Valve Stenosis Determinants of gradients across normal prosthetic valves include: Valve type, i.e., Manufacturer Valve size Flow through the valve Wide range of “Normals”

Aortic Prosthesis Gradients as a Function of Valve TYPE and SIZE Dependence on: Valve type, and Size No.21 Fig. 38-16 No.27

Gradient as a Function of Valve Type Pg 508, Table 1 Normal Dopplar data in patients with various types of prosthetic valves in the Aortic Position

Gradient as a Function of Valve Size Pg 508, Table 2 Valve specifications and doppler echocardiographic data in 67 St. Jude medical valves in the Aortic position Chafizadeh ER, Circ. 83:213, 1991

Gradient as a Function of Flow Valve type, i.e., Manufacturer Valve size No.21 Flow through the valve

Indicies of Valve Stenosis which are Less Flow Depenent Contour of jet velocity Doppler velocity index Effective orifice area Valve resistance

A-Contour of the jet velocity With prosthetic obstruction there is: Late peaking of the velocity, More rounded contour, Prolonged ejection. Pg 508, fig. 1

B-Doppler Velocity Index DVI= Pk VelLVOT/Pk Veljet Flow independent 0.2 – 0.27 cutoff for critical stenosis Caveat: Pressure recovery To be discussed …

C-Effective Orifice Area Continuity Eqn. Caveat: Pressure recovery To be discussed …

D-Valve resistance At cutoff of 280 dynes.sec.cm5, best at differentiating AS, from NL (Zoghbi et al.)

Special Caveats Re: Overestimation of Gradients Two scenarios: The velocity upstream from the valve is not negligible in application of the Bernoulli Eqn. Usually in AV when proximal velocity on the LVOT is > 1.5 m/s

…Overestimation of Gradients Central acceleration with the St. Jude valve: Increase of velocities (and gradients) is created at the level of the valve through the smaller central orifice. Most significant with: High flow states Small valves The only other valve where this is holds true is the Ball-Cage valve

…Overestimation of Gradients Central acceleration with the St. Jude valve: CW Doppler records these high velocities. Catheter-derived gradients show pressure recovery at 30mm downstream from the valve.

Indicies which are Less Flow Dependent, BUT… Contour of jet velocity Doppler velocity index Effective orifice area Valve resistance Clearly, both of These Parameters Will be Affected by The Pressure Recovery Phenomenon.

Prosthesis-Patient Mismatch

Mismatch Rahimtoola 1978: “Mismatch is present when the effective prosthetic valve area, after insertion into the patient, is less than that of a normal human valve.” By definition: Some such “mismatch” will almost always be present.

Mismatch Literature identifies the above as a cut-off for mismatch

Mismatch This is the EOA that The patient physiologically Needs. Next locate the published in-vivo EOA of the valve used.

Mismatch Not the company reported data JACC Review Article, 10/2000

Prosthetic Valve Pathology Prosthetic Valve Stenosis Aortic Mitral

Mitral Prosthesis Stenosis Parameters used for assessment of function: PHT/Area by PHT Effective Orifice Area by continuity Mean gradient

Mitral Prosthesis Stenosis A-PHT/Area by PHT Not expected to yield accurate valve area The empiric constant of 220 validated for the geometry of rheumatic MS Useful in longitudinal follow-up of valve Fx Should not be used when diastolic filling period is short (fusion of E and A) Tachycardia Long first degree block

Mitral Prosthesis Stenosis B-Effective Orifice Area by continuity Eqn. One underlying assumption is absence of significant AI or MR Physiologic prosthetic MR 10-30% (Medtronic-Hall, significant central MR, specific design feature  less thrombogenic)

Mitral Prosthesis Stenosis C-Mean gradient, function of: Size Type of prosthetic Flow Pg.512, Table 3 Croocked, sraighten it out Heart rate (should also be reported when evaluating MVA)

Prosthetic Valve Pathology Prosthetic Valve Stenosis Aortic Mitral Prosthetic Valve Regurgitation

Prosthetic Valve Regurgitatoin Two issues: Physiologic v.s. Pathologic regurgitation TTE v.s. TEE for assessment of regurgitation

Prosthetic Valve Regurgitation Physiologic Regurgitation Early onset and brief duration Reflects backflow from closing movement of occluding device Tilting disc and bileaflet valves have additional late backflow leakage Intended to reduce risk of thrombosis

Aortic Prosthesis Regurgitation Criteria similar to grading native valve AI: Jet width PHT < 350 Holodiastolic flow reversal Regurgitant fraction>40%

Mitral Prosthesis Regurgitation TTE of limited value in assess MR due to acoustic shadowing of the LA Doppler findings suggestive of severe MR E wave > 1.9 m.s PISA Short isovolumetic relaxation time TVILVOT/TVIPr-MV < 0.4