Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 5 Advanced X ray machines &Mammogram.

Slides:



Advertisements
Similar presentations
DIGITAL RADIOGRAPHY.
Advertisements

Advanced Biomedical Imaging Lecture 3
Technique Guidance Systems
Introduction to x-rays
Image Reconstruction.
Formation Characteristics
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Conventional Radiography
IMAGE FORMATION  Introduction  The Invisible and Visible Image  Image Characteristics.
RADIOLOGY. NEXT GENERATION SCIENCE / COMMON CORE STANDARDS ADDRESSED! CCSS.ELA-Literacy.RST Determine the central ideas or conclusions of a text;
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
MAMMO QC – covered in week 8
Advanced Biomedical Imaging
Medical Imaging X-Rays I.
Imaging Systems X-Rays. Imaging Systems: Shared Elements 1.Where did the energy come from? 2.What happens when the energy interacts with matter? 3.How.
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Instrumentation Part 1.
Part No...., Module No....Lesson No
Computed Tomography RAD309
Radiation Sources in medicine diagnostic Radiology
X-Rays and CT Scans.
Robert L. Metzger, Ph.D., C.H.P. Roland Wong, Sc.M., D.A.B.M.P.
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
Direct Digital Radiography or Direct Capture Radiography Bushong Ch. 27.
X-Ray Production & Emission
Basics of X-ray and Mammography Systems:
QUALITY ASSURANCE ADVANCED IMAGING EQUIPMENT CHAPTER 10.
X-Ray Production & Emission
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Quality Control Rad T 110.
Why Digital Mammography. Why Cogdell Memorial Hospital?  Cogdell Memorial Hospital delivers the highest quality patient care available –Dedicated staff.
INSTRUMENTATION. ELECTRICAL REQUIREMENTS AND EFFICIENCY  Effective voltage  High frequency inverter technology (over 10,000Hz)  Constant potential.
X-Ray Medical Imaging Physics – IB Objectives
Medical Image Analysis Medical Imaging Modalities: X-Ray Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan,
DMI 261 Radiation Biology AndProtection. Unit 2 RADIATION Electromagnetic Spectrum –The frequency range of electromagnetic radiation and the photon wavelength.
FLUROSCOPY. What is Fluoroscopy? Fluoroscopy is a method of using low intensity X-ray beams to continuously visualize the area of interest in "real.
Quality Control.
Quality Assurance, Quality Control & Equipment Bushong Ch. 19 & 20
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
HABIS X-RAY PRODUCTION AND EXPOSURE FACTORS X-RAY PRODUCTION AND EXPOSURE FACTORS PREPARED BY PREPARED BY Dr fahad albadr radiology chairman radiology.
Dr Mohamed El Safwany, MD.
Lecture (2). 2 1.Explain how to produce x-rays and discuss its properties 2.List the basic components of the radiographic machine and identify primary.
Mammography Questions and Answers
Chapter 2 The X-ray Beam.
X - RAYS IN DIAGNOSTICS D. Krilov
Factors affecting CT image RAD
Computed Radiography By Prof. Stelmark. Presently, an acceleration in the conversion from screen-film radiography (analog) to digital radiography (DR)
X-rays. Electromagnetic Spectrum to 10 nm 400 to 700 nm to nm 10 to 400 nm 700 to 10 4 nm X-ray radiation was discovered by Roentgen.
PRINCIPLES OF TECHNIQUE AND EXPOSURE
Radiographic Quality Visibility and Sharpness
X-rays discovered in 1895 by Wilhem Roentgen Roentgenology – Branch of medicine dealing with x-ray or gamma rays in diagnosis and treatment. Radiology.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Image Production and Evaluation NOTE: QA AND QC ARE USED INTERCHANGABLY IN APPLETON AND LANGE.
Rad T 275 Mammography. Types of Mammography Diagnostic –Patients with symptoms or risk factors Screening –Asymptomatic patients.
RAD 254 Chapter 19 Mammography
DIGITAL IMAGING.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
By Prof. Stelmark. Digital Imaging In digital imaging, the latent image is stored as digital data and must be processed by the computer for viewing on.
Direct Digital Radiography or Direct Capture Radiography
Basis of Mammography F. Milano Dept. Clinical Physiopathology University of Florence
Presented by: Aaditee. S. Kulkarni Omkar. R. Haldonkar Pranali. N. Kulkarni Ganesh M. Nair.
History of digital radiology 1980 Kinos L.T. ET all developed a portable radiographic X-ray camera made of tantalum and aluminum. Digital radiography.
Radiology. The Cathode… Provides a source of electrons and directs the electrons towards to anode. The cathode has a coiled wire filament that emits electrons.
Exposure Factors or Prime Factors
X-RAY PRODUCTION AND EXPOSURE FACTORS
Part No...., Module No....Lesson No
X-Radiation.
Computed Radiography Feb.
Resident Physics Lectures (Year 1)
Resident Physics Lectures (Year 1)
Image Production and Evaluation
Presentation transcript:

Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 5 Advanced X ray machines &Mammogram

is revolutionizing medicine in the same way that digital cameras changed photography. Computed Radiography (CR) Advanced X ray machines

CR uses instead of ordinary cassette, an imaging plate coated with storage phosphor to capture x-rays as they pass through the patient. CR uses instead of ordinary cassette, an imaging plate coated with storage phosphor to capture x-rays as they pass through the patient. imaging plate imaging plate When irradiated, the enhanced phosphor absorb and store x-ray energy. When irradiated, the enhanced phosphor absorb and store x-ray energy. This trapped energy comprises a latent image This trapped energy comprises a latent image

CR cassette is then placed in a digitizer where the phosphor plate is scanned causing release of trapped energy into visible light. CR cassette is then placed in a digitizer where the phosphor plate is scanned causing release of trapped energy into visible light.digitizer This light is captured and converted into an electrical signal, which is converted into digital image This light is captured and converted into an electrical signal, which is converted into digital image Digital image can be displayed on laser-printed films or softcopy workstations. Digital image can be displayed on laser-printed films or softcopy workstations.

Lower radiation dose Higher contrast. Better image quality. Rapid result No need to repeat examination. Post-processing capabilities… Enables workstation display & archiving. Advantages of computed Radiography (CR)

It is performed by a system consisting of: A digital image receptor A digital image processing unit An image management system Image and data storage devices A communications network A display device with viewer operated controls Digital Radiography (DR)

Digital cassette

Include mammography, US & MRI. Mammography is used to detect breast pathology & cancer. US is used to differentiate solid from cystic lesions which have similar appearance on mammography. MRI is used for evaluation of silicon implants and assessment of stage of breast cancer. Mammogram imaging modalities

Approximately 1 woman in 8 will develop breast cancer over a lifetime. Breast cancer screening programs rely on mammography because it is a low-cost, low-radiation-dose procedure with sensitivity to detect early-stage breast cancer. It demonstrates both micro calcifications (high contrast) & much low contrast areas of tissues on same film. Mammography

Breast is composed of fatty & glandular tissues. There is a small x-ray attenuation differences between them which decrease with high energy. The best differentiation between tissues obtained at low x-ray energy but this increases patient dose and exposure time. So breast imaging needs a special designed x- ray tube. So breast imaging needs a special designed x- ray tube.

System design Because of the risks of ionizing radiation, techniques that minimize dose and optimize image quality are essential, and have led to: –Specialized x-ray equipment –Specialized x-ray tubes –Compression devices –Antiscatter grids –Phototimers

Photo timer: Photo timer: Radiation sensor used to control the exposure. It sets the optimal kV from a short test exposure.

Breast is composed of fatty & glandular tissues. There is a small x-ray attenuation differences between them which decrease with high energy. So the best differentiation between tissues obtained at low x-ray energy Target

Conventional x ray tube Tungsten (Z=74) E K =70Kev E L =12Kev Ch. Rad.= 58kev Mammography Molybdenum (Z=42) E K =20Kev E L =2.5Kev Ch. Rad.= 17.5kev Photoelectric Target

Focal spot Small focal spot is used for best contrast ( mm) The problem is heat dissipation ( tube cooling problem).

Filter Inherent filtration must be kept low; beryllium (Z = 4) is used for the tube port as it has low z so low U & less filtration. Added tube filters of the same element as the target reduce the low- and high-energy x-rays in the spectrum and allow transmission of characteristic x-ray energies. (Mo/Mo).

Uniformity of x ray beam:  Heal effect: x ray travels toward anode edge have more target to cross and attenuated more than those travel toward cathode edge so intensity is different.  Anode heel effect: thickest part of breast at cathode side end where beam is more intense. This decreases the equipment bulk near patient’s head for easier positioning

Anode heel effect

Molybdenum (Mo)Target MolybdenumFilter Mo ch, rad tungsten target, Rhodium filterFor large dense breast, implant mm, tube cooling prob, 3Focal spot, Film λ min. filtration (z=4) due to low UBeryllium window, Not glass usedAir gap MovingGrids, GR 65cmSID 25-35kvTube voltage single screenScreen photoelectric effectMain source of contrast 0.5-1msvEffective dose

Factors affecting doses in mammography: Beam energy: ↑ Kv requires ↓ mAS & ↓ lower dose, ↓contrast So low Tube voltage is used (25-35Kev) Target & filter material: – Mo is used → emit characteristic x-ray ( kev) – Rhodium is used for thick or dense breast → (23 kev) – Tungsten →dense breast.

Grids: Breast dose (MGD) increased by 2-3 but image contrast improves by factor of 2. Breast thickness & tissue composition: Large & dense breast are more difficult to penetrate so ↑ energy x-ray beam but it ↑ average glandular dose. Small breast & of more adipose tissue → ↓ AGD Breast compression to reduce overlap tissues & scatter, more contrast, less motion, lower tissue radiation dose.

Magnification: 1.5 to 2 times in mammography image small breast but ↑ AGD, best achieved with small focal spot. achieved by moving breast away from image and close to tube –decreased scatter –Increased resolution, – but ↑ dose to breast

Screen film comb & film processing conditions: Film processing are important as image must detect small object & object with ↓ contrast Single-screen and single emulsion film, it has better resolution (SR). AGD is limited to 3 mGy or 300 mRad per film for a compressed breast thickness of 4.2 cm.

Decrease dose in mammography Small breast Compression Increase kv Increase contrast Decrease kv Beryllium window Grids / air gap Film gamma 3, low speed Increase resolution Small focal spot Single screen and single emulsion film Film processing

Questions 1. Tabulate the differences between conventional and mammographic x ray machine? 2. Mention the main difference between conventional, computerized and digital radiography? 3. Define heel effect & mention its importance?