ISE Ion Selective Electrodes Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit.

Slides:



Advertisements
Similar presentations
CHEM 213 Instrumental Analysis
Advertisements

Equilibrium Chemistry
Summary of Potentiometry:
CHEMISTRY ANALYTICAL CHEMISTRY Fall
1 Calibration of pH meters, glass electrodes and conductometers at INPL Calibration of pH meters, glass electrodes and conductometers at INPL Dr. Elena.
Section 11 Potentiometry. Potentiometric Electrodes Potentiometric electrodes measure: Activity not concentration Concepts to review: Activity and affect.
PH pH is a unit of measure which describes the degree of acidity or alkalinity (basic) of a solution. It is measured on a scale of 0 to 14. The formal.
Potentiometry. pH is a Unit of Measurement  pH = Power of Hydrogen (H + )  Defined as the Negative Logarithm of Hydrogen Ion Activity  pH = log (1/H.
Chapter 17 Electrochemistry
© Sauvé 2002 Sébastien Sauvé Department of Chemistry Université de Montréal Metal speciation using ion-selective electrodes.
Chem. 133 – 3/5 Lecture. Announcements Lab –Set 2 Period 2 Labs Set to Finish 3/10 –3/12 will be make up day (for both Period 1 and Period 2 labs) –Set.
Potentiometry and the Detection of Fluoride
Fall 2014, Prof. JB Lee Ion Sensitive FET (ISFET) - What and Why?
PHYSIOLOGY 1 Lecture 11 Membrane Potentials. n Objectives: Student should know –1. The basic principals of electricity –2. Membrane channels –3. Electrical-chemical.
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering pH measurements The probe, from chemistry to voltage to a number + + V.
PH probe
THE NERNST EQUATION RELATES THE MEMBRANE POTENTIAL TO THE DISTRIBUTION OF AN ION AT EQUILIBRIUM  E j = 59mV log C j o / C j i(6.11) 1. A tenfold difference.
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering pH measurements The probe, from chemistry to voltage to a number + + V.
IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards An Overview.
Acids and Bases Biotechnology I. Life Chemistry  Based on water  Cells contain 80-90% water  Proper pH essential to ALL living systems Plants cannot.
Preparation of the buffer solutions, titration of acetic acid and its buffering capacity.
COLORIMETRYCOLORIMETRY Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit Note: A printed description of.
Electrochemistry Chapter 21. Electrochemistry and Redox Oxidation-reduction:“Redox” Electrochemistry: study of the interchange between chemical change.
1 Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: , RTC, December 2004 ABL800 FLEX Calibration.
ECF and ICF show no electrical potential (0 mV).
THE MEASUREMENT OF pH. I GENERAL PRINCIPLES Convenient way to express hydrogen ion concentration, or acidity pH = - log[ H +
Chapter #16 Acid-Base Titration and pH. Chapter 16.1 In the self-ionization of water, two water molecules produce a hydronium ion and a hydroxide ion.
Acids- Base Titration and pH. Aqueous Solutions and the Concept of pH.
1. Blood or Fluid Analysis Using carried out in central labs with problems of time delay costs and potential misslabeling Drive toward more bedside or.
Announcements:. Last lecture 1.Organization of the nervous system 2.Introduction to the neuron Today – electrical potential 1.Generating membrane potential.
Copper Ion Analysis Cory Sennett November 15, 2006 CH EN 4903.
1 Chapter Eighteen Electrochemistry. 2 Electrochemical reactions are oxidation-reduction reactions. The two parts of the reaction are physically separated.
Chapter 15 Preview Lesson Starter Objectives
Electrochemical Methods Dr M.Afroz Bakht. Potentiometry Potentiometry is a method of analysis used in the determination of concentration of ions or substances.
Quality Assurance How do you know your results are correct? How confident are you?
Lesson declared p[H] After finished this lesson student will to able to # Say about p[H]. # Say about pH indicators. # Say about Buffer solution.
Chapter Goals (Membrane Potentials) After studying this chapter, students should be able to 1. explain what is meant by active transport and describe how.
Resting Membrane Potential (Voltage) Dr.Mohammed Alotaibi MRes, PhD (Liverpool, England) Department of Physiology College of Medicine King Saud University.
Electrophysiology 1.
Joshua Dudman :: 0 mV -80 mV.
Ch. 15 Titration And pH. Ionization of Water _____________________: two water molecules produce a hydronium ion and a hydroxide ion by transfer of a proton.
Ch. 4. Measurements (on Site) in Field  pH 와 E H  Basic water quality parameters  Definitions  pH = -log 10 a H+ (logarithmic value of the reciprocal.
Thursday, May 1 st : “A” Day Friday, May 2 nd : “B” Day Agenda  Homework questions/collect  Finish section 15.2  Section 15.2 workday: Practice pg.
Chapter 16 Acid-Base Titration and pH. Aqueous Solutions and the Concept of pH Self-ionization of water – 2 water molecules produce a hydronium ion and.
1 Acid-Base Titration and pH Chapter Self-Ionization of water Two water molecules produce a hydronium ion and a hydroxide ion by transfer of a proton.
Potentiometry.
Objectives Basics of electrophysiology 1. Know the meaning of Ohm’s Law 2. Know the meaning of ionic current 3. Know the basic electrophysiology terms.
IC Controls Quality Water Solutions for pH pH / ORP Conductivity Dissolved Oxygen Chlorine Standards R1.0 © 2004 IC CONTROLS Concerns.
Challenges are what make life interesting; overcoming them is what makes life meaningful. -Joshua J. Marine-
PH THEORY What is it pH?What is it pH?/How is pH Measured?/What Equipment is Required to Measure pH?/How is a pH Measurement Device Calibrated?/Why is.
PH meter Bushra mubarak. pH pH is defined as the negative logarithm of the hydrogen ion activity aH. Unit of measure which describes the degree of acidity.
POTENTIALS AND THERMODYNAMICS OF CELLS (1) POTENTIALS AND THERMODYNAMICS OF CELLS (1)
Instrumental Analysis Ion selective electrodes
Analytical biochemistry laboratory (BIOC 343) L. Nouf Aalshareef KAU- Biochemistry department Second semester 2013.
David Housman for Math 323 Probability and Statistics Class 05 Ion Sensitive Electrodes.
Field Equipment Calibration PH, Temperature, Conductivity, Turbidity, Dissolved Oxygen, Free & Total Chlorine Gregory Langland CCSF.
Chem. 133 – 2/28 Lecture.
Ion Selective Electrode
EMSA 22 Lab Module 1 Field Equipment Calibration
PH & Buffers.
Ion Selective Electrode
Conductivity Lecture.
PH pH is a unit of measure which describes the degree of acidity or alkalinity (basic) of a solution. It is measured on a scale of 0 to 14. The formal.
Instrumental Methods II
Topic 8: Acids and Bases Topic 8.3: The pH scale is an artificial scale used to distinguish between acid, neutral, and basic/alkaline solutions Topic.
Electrode Training HORIBA Instruments, Inc. Pamela Millett Scientific
Determination of I- in Salt (Ion-Selective Electrode)
Exp:5-1: Determination of Iodine in Salt
FACTORS AFFECTING REACTIONS IN CELLS
Presentation transcript:

ISE Ion Selective Electrodes Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit

Ion Selective Electrode Electrode Body Ion Sensitive Area Electrical Connection

Sensing Electrode Reference Electrode Current Flow Meter Electrochemical Measuring System

Combination Probe Sensing Element Reference Element

Sensing Electrode Reference Electrode Current Flow Meter Electrochemical Measuring System

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration What the Meter Tells Us What We Want To Know } Meter Reading Also Affected By All This

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration What the Meter Tells Us What We Want To Know } Must Be Accounted For To Get True Concentration

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration What the Meter Tells Us What We Want To Know Must Maintain Reference Electrode

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration What the Meter Tells Us What We Want To Know Will Be Constant for Specific Ion, Whole Number, 1, 2, 3, etc. + or -

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration What the Meter Tells Us What We Want To Know Must Be Controlled By Making It A Very High Value

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration What the Meter Tells Us What We Want To Know Add Ions ISAB OH - Na +

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration What the Meter Tells Us What We Want To Know Follow The Directions!

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration What the Meter Tells Us What We Want To Know Must Be Controlled

NERNST EQUATION E = E o T Log  C n E = Measured Voltage E o = Reference Constant T = Temperature n = Charge on Ion  = Ionic Strength C = Concentration

NERNST EQUATION E = E o T Log  C n 2.3 T n = s = slope Slope is Direction of “Curve” When Plotting E vs. C y = mx + b

Effect of Temperature Electrode Potential (mv) Concentration (mg/L) 0 o C 50 o C 100 o C s is the Direction (Angle) of the Line Temperature Affects the Slope

Effect of Temperature Electrode Potential (mv) Concentration (mg/L) O O O 0 o C 50 o C 100 o C One Concentration will Result in Different Readings Dependent on Temperature

Effect of Temperature Electrode Potential (mv) Concentration (mg/L) O O Calibration Temperature Sample Temperature X When the Sample Temperature is Different from the Calibration Temperature, an Incorrect Sample Concentration will be Obtained

Effect of Temperature Electrode Potential (mv) Concentration (mg/L) 0 o C 50 o C 100 o C Isopotential Point

Effect of Temperature Electrode Potential (mv) Concentration (mg/L) 0 o C 50 o C 100 o C Isopotential Point At the Isopotential Point, One Concentration will Result in the Same Reading Independent of Temperature

Effect of Temperature Electrode Potential (mv) Concentration (mg/L) 0 o C 50 o C 100 o C } Normal Range of Analysis The Range of Analysis is Far From the Isopotential Point for Most Samples

Effect of Temperature Electrode Potential (mv) Concentration (mg/L) 0 o C 50 o C 100 o C } Normal Range of Analysis Temperature Must be Accounted For in All Analyses

Automatic Temperature Compensation ATC Adjusts Slope in Relation to Temperature Does Not Work! Sounds Good!

Is Accurate OnlyWhen Analysis Is Accurate Only When Analysis is Close to Isopotential Point Only For pH Adjusts Slope in Relation to Temperature Automatic Temperature Compensation ATC

ATC (ExceptForpH)

Effect of Temperature Electrode Potential (mv) Concentration (mg/L) Calibrate With Standards and Samples At The SAME Temperature (Usually Room Temperature)

Calibration Control Electrode Potential (mv) Concentration (mg/L). Standard Solution The Calibration Control Adjusts the “Curve” to the Point of Standard Used

Slope Control Electrode Potential (mv) Concentration (mg/L). Standard The Slope Control Rotates the “Curve” Around the Point Of the First Standard

Slope Control Electrode Potential (mv) Concentration (mg/L). Standard. Second Standard The Slope Control Adjust the “Curve” through the Point of the Second Standard Used

NERNST EQUATION E 1 = E o + s Log  C 1 E 2 = E o + s Log  C 2 If C 2 = 10 x C 1 E 2 = E o + s Log  10 C 1 Then:

NERNST EQUATION E 1 = E o + s Log  C 1 E 2 = E o + s Log  C 2 If C 2 = 10 x C 1 E 2 - E 1 The Slope of an Electrode is the millivolt Change that is seen for a Ten Times Change in Concentration. = s

ISE Checking Electrode Slope Carefully Prepare Two Standards 10 X Concentration Difference Set Meter to Read in millivolts (or Relative mv) Record mv Reading for Each Standard Determine Difference = Slope

ISE Checking Electrode Slope E = E o T Log  C n Electrode Potential (mv) Concentration (mg/L)

ISE Checking Electrode Slope The Slope Value Determined is Affected By: Charge on the Ion of Interest (n) Temperature (T) E = E o T Log  C n Positive or Negative 1, 2, etc.

ISE Checking Electrode Slope The Slope Value Determined is Affected By: Charge on the Ion of Interest (n) Temperature (T) Quality of the Standards E = E o T Log  C n Efficiency of the Electrode

ISE Checking Electrode Slope The Slope of the Electrode Will Change {Loss of Efficiency} The Slope Should be Checked Regularly to Assure Reliable Results Most Meters Give Slope When Calibrating Record Slope for QA/QC Daily (or at least every two weeks) ( In mv or %)

ISE Checking Electrode Slope Ideal Slope Depends on Ion of interest (n) Charge on the Ion of Interest (+ or -) Approximately 59 mv for n=1 or 29 mv for n=2 Usually ± 10 % of Ideal Check Manufacturer for Acceptable Slope Range

ISE Checking Electrode Slope If Determined Slope is Outside of Acceptable Range May Be Due To: Improper Probe Maintenance Poor Quality of Standards Faulty Electrode The Slope Should be Checked Regularly to Assure Reliable Results

ISE Ion Selective Electrodes Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit