Fabrication of a Microelectrode Array Biosensor Based on a Modified Enzyme-Chitosan Biocomposite Lorenzo D’Amico October 1, 2008.

Slides:



Advertisements
Similar presentations
NEW SENSING DEVICES FOR IMPROVING THE QUALITY OF LIFE Interest in FP 6 call: 'Nanotechnologies and nanosciences, knowledge-based multifunctional materials.
Advertisements

Conclusions Future work Methods Background Introduction Lily Stanley, Juan Du, and Xuan Gao Department of Physics, Case Western Reserve University Nanowire.
EEE 529 Microsystems Amperometric Biosensors
On-line Glucose Measurement in Disposable Bioreactors Michael Hartlep and Wolfgang Künnecke INTRODUCTION The use of disposables in biopharmaceutical processes.
Hydrogen Peroxide Detection Using Biogenic and Synthetic MnO 2 Shu Feng, Jim Nurmi, Paul Tratynek Satya Chinni, Brad Tebo Department of Environmental and.
New electrode materials for H 2 O 2 New electrode materials for H 2 O 2 based sensors and biosensors Francesco Ricci 1, Carla Gonçalves 2, Giuseppe Palleschi.
Abstract Sibley School of Mechanical and Aerospace Engineering, Cornell University Researchers from Cornell University and the Cornell Lab of Ornithology.
Determination of Partition Coefficient of NAD + on Nafion 117 membranes Importance: Partitioning process affects actual NAD + and NADH concentrations,
The Detection of Hydrogen Peroxide and Glucose for the Electrode Modified with Ferrocene Yi-Sheng Wang ( 汪乙生 ), Ting-Li Lin ( 林庭立 ), Hau Lin ( 林浩 ) Department.
Application of Silicon Nanowire in Biosensor Student: HongPhan ID: Professor: Cheng-Hsien Liu.
Charge-Based Biosensor Using Carbon Nanotube Transistors Array Presenter: Jui-Ping Chiang.
Development of an Electrochemical Micro Flow Reactor (EMFR) for electrocatalytic studies of methanol oxidation and fuel cell applications. Nallakkan Arvindan*,
Polypyrrole Nanoribbon Based Nano Gas Sensors Sergio Hernandez, Nicha Chartuprayoon and Nosang V. Myung Department of Chemical and Environmental Engineering,
ELECTRICAL POROUS SILICON MICROARRAY FOR DNA HYBRIDIZATION DETECTION M. Archer*, D. Persaud**, K. D Hirschman**, M. Christophersen* and P. M Fauchet* *Center.
Distributed Microsystems Laboratory ENose Toolbox: Application to Array Optimization including Electronic Measurement and Noise Effects for Composite Polymer.
Carbon Nanotubes as Biosensors ME 695 Yang 11/30/2004.
Cyclic Voltammetry for the Detection of Dopamine in vivo
Communication: Why Undergraduate - measure of success? Graduate student – measure of success? After school – measure of success? Your future success will.
Nanofabrication Using Viral Biotemplates for MEMS Applications NSF Grant Konstantinos Gerasopoulos, Ekaterina A. Pomerantseva, Xiaozhu Fan, Adam.
PREPARATION OF ZnO NANOWIRES BY ELECTROCHEMICAL DEPOSITION
Jon Jay, Dr. Kim Pierson Department of Physics & Astronomy, University of Wisconsin-Eau Claire Investigation of Recently Developed Photovoltaic Material.
Crosstalk between Shear Stress and Human Brain Microvascular Endothelial Cells Sungkwon Kang 1 Advisors: Yong Woo Lee 2, Pavlos P. Vlachos 1,2 1 Department.
Preparation of the Carbon Paste Electrode Modified with Ferrocene and Its Applications to Detection of Hydrogen Peroxide and Glucose Yi-Sheng Wang ( 汪乙生.
High Spatial Resolution 3D Probes for Neurobiology Applications PSD Conference ’08 2 nd September Deborah E. Gunning EPSRC Overseas Life Science Interface.
Presented by: Jay Leitch. What are Glucose Biosensor?   Immobilized biological components fixed on ordinary analytical detectors   Selectively detects.
Reproducible Enzyme Assembly and Catalytic Activity in Reusable BioMEMS Rubloff Research Group Accomplishments.
Aquious Halide Concentration Determination via Leucigenin Quenching and Indirect CE Christopher Hampton, Dr. Eamonn F. Healy, Chemistry Department, St.
Quasi-simultaneous Electrochemical and Electrophysiological Measurements at the Same Sensor: Probing the Chemical Environment and Bioelectrical Activity.
BioSensors Yang Yang 9/28/2004. Outlines BioMEMS Enzyme-coated carbon nanotubes Microcantilever biosensor with environmentally responsive hydrogel Cantilever.
NOVEL NON-CONDUCTING FILMSFOR INTERFERENCE-FREE ELECTROCHEMICAL SENSORS M. BADEA a, A. CURULLI b*, G. PALLESCHI a, S. KACIULIS c, A. MEZZI c a Università.
Sample 2 inlet Buffer inlet Pneumatic micropump The vitamin C filtration Micro-valve outlet Sample 1 inlet MIP morphine sensing electrode Pt Ag Multiple.
A Study of Statistical Analysis of the Main Effects and Interaction Effects of Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with.
The Optimum Operating Conditions for Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with Ferrocene Pao-Tsai Kuo ( 郭寶財 ), Chung-Min.
The Operating Potential on the Sensitivity of Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with Ferrocene Pao-Tsai Kuo ( 郭寶財.
Microwave Assisted ZnO Nanorod Growth for Biosensing This material is based upon work supported by the National Science Foundation.
The Effect of Stirring Rate on the Detection of Hydrogen Peroxide for Carbon Paste Electrode Modified with Meldola’s Blue Chi-Wen Lo ( 羅濟玟 ), Chih-Ying.
Biosensors and Carbon Nanotubes
The Detection of Glucose for the Carbon Paste Electrode Modified with Ruthenium Hexacyanoferrate Kuo-Hsiang Liao ( 廖國翔 ), Chih-Ying Wu ( 巫致穎 ), Hau Lin.
Amperometric Glucose Biosensor Based on AFM Nano-indented-Electrode Ensemble Gymama E. Slaughter, Virginia State University, DMR The VSU’s Center.
The Operating Conditions for the Sensitivity of Detection of Hydrogen Peroxide for the Electrode Modified with Copper Hexacyanoferrate Chia-Cheng Hsiao(
Application of the Screen Printed Planar Electrode Modified with Ruthenium Hexacyanoferrate to Glucose Biosensor Kuo-Hsiang Liao( 廖國翔 ), Chung-Min Lien(
The Effect of Ratio of Meldola’s Blue to Carbon Powders and Carbon Paste on Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with.
Electrode Array GK-12 Meeting November 2nd 2010
BIOFABRICATION OF MEA GLUCOSE SENSORS Dry in air Ready for in vitro glucose detection (B) Chitosan biopolymer is electrodeposited on target electrode sites;
The pH on the Sensitivity of Detection of Hydrogen Peroxide for the Small Electrode Modified with Meldola’s Blue Chi-Wen Lo ( 羅濟玟 ), Chung-Min Lien ( 連崇閔.
National Science Foundation Outcome: Researchers at University of Maryland have created a 3D biological templated current collector with improved solar.
The pH on the Detection of Hydrogen Peroxide for Electrode Modified with Chromium Hexacyanoferrate Chen-Hsun Hu ( 胡真熏 ), Ting-Li Lin ( 林庭立 ), Hau Lin (
© SSER Ltd..
Development of Microfluidic Glucose Sensors BME 273: Kristen Jevsevar, Jason McGill, Sean Mercado, Rebecca Tarrant Advisors: Jennifer Merritt, Dr. John.
The Detection of Hydrogen Peroxide and Glucose for the Electrode Modified with Ruthenium Hexacyanoferrate Kuo-Hsiang Liao ( 廖國翔 ), Chung-Min Lien ( 連崇閔.
Advanced Biosensors Based on Nano-scaled Arrays and Films of Electro- and Biocatalysts Arkady A. Karyakin Chemistry faculty of M.V. Lomonosov Moscow State.
Biosensors Dr. Bhavesh Patel Principal
By Jason Tyser Professor Debashis Dutta /images/fuelcell.jpg.
A Study of Electrical Impedance Property of an L 2 ePt Electrode 1 Interdisciplinary Program, Bioengineering Major, Graduate School, Seoul National University,
Date of download: 6/28/2016 Copyright © 2016 SPIE. All rights reserved. Optogenetic tools and light tissue penetration: (a) schematic representation of.
Introduction Lei, Yang; Luo, Ning. A highly sensitive electrochemical biosensor based on zinc oxide nanotetrapods for L-lactic acid detection. Nanoscale,
/ CHEMICAL CROSS TALK STUDIES ON A MICROPROBE ARRAY BASED CONTINUOUS GLUCOSE MONITORING SENSOR Sanjiv Sharmaa,
Biosensors Definition
Objective Functions for Optimizing Resonant Mass Sensor Performance
RESULTS AND DISCUSSION
Cell Delivery Mechanisms for Tissue Repair
Chen-Hsun Hu (胡真熏) , Chih-Ying Wu (巫致穎) , Hau Lin (林浩)
Immobilization Of Biomolecules On Biosensors
s Changes: Cite figures under EIS
Joy Qiu ♦ Dr. Greg Swain ♦ HSHSP 2014
Detection of Beta Carotene and Lutein
Cell Delivery Mechanisms for Tissue Repair
DNA conjugation and reversibility on chitosan surfaces
Wearable/disposable sweat-based glucose monitoring device with multistage transdermal drug delivery module by Hyunjae Lee, Changyeong Song, Yong Seok Hong,
Fig. 2 Investigating the interactions between the n-type polymer and the enzyme, which lead to efficient electrical communication. Investigating the interactions.
Presentation transcript:

Fabrication of a Microelectrode Array Biosensor Based on a Modified Enzyme-Chitosan Biocomposite Lorenzo D’Amico October 1, 2008

 Amgen Scholars, 10 week summer research program at UCLA funded by the Amgen Foundation  Harold Monbouquette’s Biotechnology laboratory in the Chemical and Biomolecular Engineering Department Our ghetto Faraday cage (Amgen did not pay for equipment)

Immobilized oxidase Analyte H2O2H2O2 O 2 + H 2 0 Byproduct 2 e - ~0.4 – 0.7 V vs. Ag/AgCl Electrode surface Current Electroenzymatic detection: general concept Amperometric biosensors convert a biochemical signal into an electric signal Hydrogen peroxide is produced by the immobilized enzyme in the presence of analyte and is subsequently oxidized at the electrode surface, resulting in an electron flow.

 Using MEMS technology, sensors are mass produced and manufacturing costs dramatically reduced (~200 probes per wafer);  Probe shaft is ~200 µm wide, 9 mm long;  At the probe tip is the microelectrode array. Each platinum site has a surface area of ~ cm 2.  This design offers an opportunity to make the sensor capable of detecting multiple analytes, or using one site as the reference electrode. Developing a fabrication strategy with good spatial control is a current obstacle for researchers working in this area. Micromachined Probe Design Credit: Monbouquette et al.

Current Neuroscience Studies Using This Microelectrode Biosensor Design Glutamate detection in a freely moving rat Platinum electrode surfaces are functionalized with glutamate oxidase and modified with polymers which act to block interferents like dopamine and ascorbic acid. The devices are implanted into brain tissue of a freely moving rat and extracellular glutamate levels are continuously monitored. Sensors are capable of detecting increases in glutamate concentrations (spikes in the graph) induced by pinching the rat’s tail. Credit: Monbouquette et al. Pt electrode e - + Nafion + _ Glutamate oxidase 0.7 V vs. Ag/AgCl H2O2H2O2 Glutamate Polypyrrole Dopamine Ascorbic Acid + _ (analyte) I = I H2O2

Native glucose oxidase Nanoencapsulated glucose oxidase *Image provided by the Lu group, Chemical and Biomolecular Engineering Department at UCLA Nanoencapsulation of glucose oxidase  The original motivation to modify the enzyme was to increase the electrophoretic ability of the molecule;  By making the molecule more negative, it was thought that its deposition could be guided by applying positive voltage to target electrode sites;  Preliminary tests of this hypothesis did not seem promising. However, the modified enzyme was applied to a biofabrication process developed by the Bentley research group at the University of Maryland. *

Chitosan Solution Bulk pH is low, chitosan is soluble Electrode Surface pH gradient 1/2H 2 H+H+ Apply voltage Signal Directed Assembly of Chitosan pH = pK a  The deposited chitosan film offers a scaffold on to which enzyme can adsorbed;  If the electrodeposition of chitosan can be limited to target electrode sites, then indirect spatial control over enzyme deposition could be achieved.

Overview of Biofabrication Process  Probe tip immersed in enzyme solution Probes are removed from wafer, packaged and chemically cleaned in preparation for surface modification Chitosan is electrodeposited to target electrode sites, -2V for 4 min Tips are immersed in modified glucose oxidase solution overnight Control probes, -2V for 4 min in PBS instead of chitosan solution Tips are immersed in native glucose oxidase solution overnight Probes are tested for glucose detection in vitro

Results In vitro glucose detection tests entailed injecting glucose solutions of increasing concentration, ending with an injection of hydrogen peroxide. The 4 channels (red, green, blue and black) represent data collected from the four electrode sites on the probe. The figure above depicts the current response of a sensor that did not have chitosan electrochemically deposited to electrode surfaces. There seems to be little more than noise resulting from glucose injections, suggesting that enzyme function is absent. Experiments much like this demonstrated the need for an immobilization mediator in fabricating these biosensors. Ch 1 Ch 3 Ch 4 Ch 2 Current response of sensor without chitosan during in vitro glucose test

Results, cont’d In these particular set of experiments, one channel (control) was not charged during the chitosan deposition. The current response of one such probe is depicted above. Control channels in many of these probes exhibited a significantly lower current response to glucose injections in vitro. However, although the response is lower in the control, the channel is generating an electrical signal; Ideally, this would not occur. Evaluation of the spatial control achieved during biofabrication with chitosan and modified enzyme

Results, cont’d Experiments were designed to compare the analytical capabilities of sensors fabricated with modified and native glucose oxidase. At the right, the current response of a sensor functionalized with native enzyme (top) and that of a sensor with modified enzyme (bottom) are shown. Clearly, the signal is higher for the electrode surfaces containing chitosan and modified enzyme. Current response of MEA biosensors during In vitro glucose tests

Results, cont’d A direct comparison of the device calibration curves further demonstrates that using nanoencapsulated glucose oxidase improves analytical capabilities of the MEA biosensor design. MEA biosensors functionalized with modified enzyme exhibited greater sensitivity to glucose injections on average during in vitro tests. Calibration curves of sensors fabricated with modified and native glucose oxidase

Conclusion Encapsulating glucose oxidase in a negatively charged polymer was insufficient in addressing the spatial control issues associated with electrochemical fabricating MEA biosensors; Using Chitosan in a two step biofabrication process mediated the immobilization of both native and modified glucose oxidase to the electrode surfaces on the probes, as well as granted some degree of spatial control over enzyme deposition; Use of nanoencapsulated glucose oxidase increased device sensitivity There remains a need to optimize several steps in the overall biofabrication process. Amgen Foundation University Research Center, Center for Academic and Research Excellence Dr. Harold Monbouquette and Vanessa Tolosa The UCLA Amgen Scholars site coordinators: Dr. Tama Hasson and Dr. Patricia Phelps and Michael Flaxman Acknowledgements