EE 4BD4 Lecture 25 Electrosurgery Biomedical Device Technology: Principles and Design, Charles C. Thomas Publisher 2007 Anthony K. Chan 1.

Slides:



Advertisements
Similar presentations
Effect of Electricity on Living Tissue
Advertisements

The L-Network L-networks are used to match the output impedance of one circuit to the input of another. Rsource < Rload, 1< Q < 5 Rsource > Rload, 1
BIOPOTENTIAL AMPLIFIERS
Chapter 1 Problems ECET 214 Prof. Park NJIT.
7th Lecture Dimitar Stefanov. Recapping Three types electrodes are used for sensing of EMG signals: 1.indwelling (intramuscular) electrodes (single fiber.
Heating Control Devices
Surgical Diathermy Machine
Kenny Low Dec 2009 Urology registrar Thursday teaching
Shortwave and Microwave Diathermy part(2). Induction Electrodes (Drum Electrode) One Or More Monopolar Coils Rigidly Fixed In A Housing Unit May Use More.
Variable Capacitance Transducers The Capacitance of a two plate capacitor is given by A – Overlapping Area x – Gap width k – Dielectric constant Permitivity.
Oscillators with LC Feedback Circuits
Electrical Surgical Unit
Lecture 20-1 Alternating Current (AC) = Electric current that changes direction periodically ac generator is a device which creates an ac emf/current.
ECE 442 Power Electronics1 Class E Resonant Inverter.
Lecture 4 Defibrillator
OSCILLATORS.
 Distortion – the alteration of the original shape of a waveform.  Function of distortion analyzer: measuring the extent of distortion (the o/p differs.
General Licensing Class G5A – G5C Electrical Principles Your organization and dates here.
Oscillators 2. LC Oscillators.
28. Alternating Current Circuits
What is a Watt? Everything you ever wanted to know about Electrosurgery. Fahad bamehriz, MD.
Diathermy.
Receivers.
Dr M A Panneerselvam, Professor, Anna University
Created by Jimmie Fouts Houston County Career Academy for.
Section 2 AC Circuits. Chapter 12 Alternating Current.
1 AC Electricity. Time variation of a DC voltage or current 2 I V Current Voltage time t.
UNIT-1 Rectifiers & Power Supplies. Rectifier A rectifier is an electrical device that converts alternating current (AC), which periodically reverses.
Tesla’s Alternating Current Dr. Bill Pezzaglia Updated 2014Mar10.
Without reference, identify principles about Electrosurgical Units with at least 70 percent accuracy.
Electrical Principles 1 1 G5 - ELECTRICAL PRINCIPLES [3 exam questions - 3 groups] G5A - Reactance; inductance; capacitance; impedance; impedance matching.
1. Which characteristic of electromagnetic (EM) energy is common to all energies in the spectrum? Wavelength, frequency, amplitude, or velocity. 2. Describe.
EE 4BD4 Lecture 24 Defibrillator Biomedical Device Technology: Principles and Design, Charles C. Thomas Publisher 2007 Anthony K. Chan 1.
BASICS OF ELECTROSURGERY
Electrosurgery & Gyn Surgery: Get the Point Across
Computer controlled I NSTANT R ESPONSE ™ system provides safer and more precise electrosurgery. TM.
Inductance and AC Circuits. Mutual Inductance Self-Inductance Energy Stored in a Magnetic Field LR Circuits LC Circuits and Electromagnetic Oscillations.
Power Amplifiers Topics Covered in Chapter : Classes of Operation 31-2: Class A Amplifiers 31-3: Class B Push-Pull Amplifiers 31-4: Class C Amplifiers.
Instrumentation & Power Electronics
II-System configuration Arc discharge: this is a high power thermal discharge of very high temperature ~10,000 K. It can be generated using various power.
EKT 214 ANALOGUE ELECTRONIC CIRCUIT II
Electro surg i cal unit  PURPOSE ESU are used for surgical cutting and for controlling bleeding by causing coagulation “ hemostasis ” at the surgical.
GCSE REVISION NOTES. Dangers of Electricity An electric current can cause: –Electric shock, muscle spasms –The heart may be stopped –Burning Do NOT touch.
OSCILLATORS Dr.S.SUJA Associate Professor
Electrosurgical Unit.
Power Supply Design Howie Pfeffer Mu2e Extinction Technical Design Review 2 November 2015.
EE 2353 HIGH VOLTAGE ENGINEERING
Electrosurgery Units ELECTROSURGICAL UNITS (ESUs)
Done by: Khalid Hijjawi Mohammad Massad.  Is an electrical instrument that measures electrical resistance, the opposition to an electric current. The.
SAFE USE OF DIATHERMY RACHEAL AFOLABI, RN, RM, H. Dip. PERI-OPERATIVE NURSING, H.Dip. HEALTH SYSTEM MANAGEMENT.
EE 2353 HIGH VOLTAGE ENGINEERING Faculty Name :A.JAIBUNISHA Faculty Code : EE 58 Designation : LECTURER Department : EEE.
Function Generators. FUNCTION GENERATORS Function generators, which are very important and versatile instruments. provide a variety of output waveforms.
SHANTILAL SHAH ENGINEERING COLLEGE ELECTRICAL DEPARTMENT.
BME 312 BIOMEDICAL INSTRUMENTATION II LECTURER: ALİ IŞIN LECTURE NOTE 7 Electrosurgical Devices BME 312-BMI II-L7- ALİ IŞIN 2014 FACULTY OF ENGINEERING.
Electro surgical unit.
Biomedical Instruments Design Biopotential Amplifiers
Wireless Power Driven Car or Train
Electrosurgical/Surgical Diathermy Units
Variable Capacitance Transducers
Harmonic Distortion Analyzer, Wave Analyzer and Function Generator
GROUND-FAULT GROUND-CHECK MONITOR
An {image} series circuit has {image} , {image} , and {image}
SURGICAL DIATHERMY.
General Licensing Class
SURGICAL DIATHERMY.
Diathermy and its safe use
Biomedical Instruments Design Biopotential Amplifiers 1.
Electronic Fundamental Muhammad Zahid
Fundamental Use of Surgical Energy
Created by Jimmie Fouts Houston County Career Academy for
Presentation transcript:

EE 4BD4 Lecture 25 Electrosurgery Biomedical Device Technology: Principles and Design, Charles C. Thomas Publisher 2007 Anthony K. Chan 1

Overview An electrosurgical unit (ESU) delivers high frequency electric current, 100kHz to 5 MHz, to tissue through a small “active” electrode, with the current returning to the device through a large “passive” electrode To provide a mechanism for both cutting tissue and cauterization resulting in “bloodless” surgery Can also supply this current through endoscopic or laparoscopic equipment Degree of heating depends on tissue resistivity and current density 2

3

Modes of Operation Desiccation – low RF current causes destruction and of cells – may have steam and bubbles with tissue turning brown Cutting – separating electrode about 1 mm from tissue and maintaining 100 V between electrodes. RF current jumps gap creating plasma with cells exploding – 500 kHz continuous sine wave with high power output Fulguration – touch tissue,withdraw several mm, no cells exploding but heat causes necrosis charring tissue, current goes deeper into tissue, medium power 4

5

Electrosurgical Active Electrode (“Pencil”) 6

7

Bipolar Operation 8

Return Electrode for Monopolar Maximum current density to avoid tissue damage is 50 ma/cm 2 Need large surface electrode (e.g. 100 cm 2 ) Ordinary flat plate under patient caused burns Now gel electrode pads 9

Standards for Return Electrode ANSI/AAMI HF18 states overall tissue-return electrode resistance shall be below 75 Ω No part of tissue in contact with return electrode shall have more than 6 O C temperature increase when the ESU is activated continuously for up to 60 sec with output up to 700 mA 10

Return Electrode Monitor REM checks contiuity of electrode cable using low frequency (140 kHz 3ma typical) signal High resistance (>20 Ω) triggers alarm REQM Return Electrode Quality Monitor uses a double return path and checks cable plus electrode contact using same waveform Resistance > 135 Ω or < 5 Ω indicate poor contact or electrolyte/pad bridging 11

REM and REQM 12

ESU Functional Building Blocks Spark Gap ESU generator (from 1920’s) uses step up transformer to 2 – 3 kV As signal rises capacitor C1 charges and gas inside the spark gap (discharge tube) ionizes until an arc is generated (same as closing a switch in resonance circuit formed by C1, L1 and spark gap impedance. Fundamental frequency of arcing current is resonant frequency of L1/C1 Voltage amplitude will decay until arc is extinguished Proper choice of L1/C1 gives RF damped sinusoid that occurs twice in one period of 60 Hz Taps on secondary give different voltage levels L3, L4 and C4 stop RF from entering power supply Spark gaps were used till 1980s but are now replaced by solid state circuitry 13

14

Functional Block of ESU 15

Output Stage Output can go to 1000 watts, 9000 V (peak to peak open circuit) and 10 amps 16