In The Name of Allah The Most Beneficent The Most Merciful 1.

Slides:



Advertisements
Similar presentations
Automatic External Defibrillator
Advertisements

Current And Resistance
Chapter 16 Electric Energy and Capacitance
BIOPOTENTIAL AMPLIFIERS
AUTOMATED EXTERNAL DEFIBRILLATION
Automated External Defibrillators Saving Cardiac Arrest Victims.
Automated External Defibrillation
NFPA ELECTRIC VEHICLE SAFETY FOR EMERGENCY RESPONDERS Module II : Basic Electrical Concepts And Hazards Module II : Basic Electrical Concepts And Hazards.
1 At the Heart of Saving Lives – Shockingly Obvious! Creating a Heartsafe Environment Steve Jelfs, International Clinical Manager, Defibrillation.
ECE 442 Power Electronics1 Step Up Converter Close the switch to store energy in the inductor L Open the switch to transfer the energy stored in L to the.
ECE 501 Introduction to BME
Lecture 4 Defibrillator
Capacitors in a Basic Circuit
Defibrillators Kaarel Koovit.
Defibrillation 4/19/2017
Defibrillators.
Defibrillator Dr. Sanjeev Chatni. Definition An electrical device used to counteract fibrillation of the heart muscle and restore normal heartbeat by.
DEFIBRILLATOR MACHINE BAsic TECHNICAL TRAINING
The New CCC-CPR for Cardiac Arrest
CPR/AED – Assisting A Person in Cardiac Arrest Lessons Provided by the Office of Healthy Schools – March 2015.
ENTC 4350 DEFIBRILLATOR & PACE MAKERS. DEFIBRILLATOR.
Thursday, Oct. 13, 2011PHYS , Fall 2011 Dr. Jaehoon Yu 1 PHYS 1444 – Section 003 Lecture #14 Thursday, Oct. 13, 2011 Dr. Jaehoon Yu EMF and Terminal.
Wednesday, Feb. 29, 2012 PHYS , Spring 2012 Dr. Jaehoon Yu 1 PHYS 1444 – Section 004 Lecture #12 Wednesday, Feb. 29, 2012 Dr. Jaehoon Yu Electric.
Chapter 6 AED.
Different Waveforms, Different Results Understanding the Differences Between Biphasic Technologies.
Dr. Ali Saad, BMT department1 Dr. Ali Saad, Biomedical Engineering Dept. College of applied medical sciences King Saud University BMT414 Biomedical Engineering.
P4: Radiation for Life Lesson 4: Uses of Electrostatics (part 2)
S1-3-09/10/12 Define electric current, voltage and resistance and solve related problems. KEY WORDS CellCoulombVoltage VoltVoltmeterCurrent AmpereAmmeterResistance.
CARDIAC MONITORING & RHYTHM RECOGNITION. How to monitor the ECG (1): Monitoring leads 3-lead system approximates to I, II, III3-lead system approximates.
CPR and Automated External Defibrillation (AED)
Automatic external defibrillator ©2004 Heartsine Technologies, Inc. Basic samaritan ® PAD Product Training Toolkit.
Ohm’s law CURRENT, POTENTIAL DIFFERENCE, AND RESISTANCE.
Electricity and Magnetism Electrostatics Capacitance and Capacitive Circuits.
C H A P T E R 20 Electric Circuits Kirchhoff's Rules There are two KIRCHHOFF'S RULES. 1. Junction rule 2. Loop rule These are useful in circuit.
Chapter 17: Electric Potential 1.  As in earlier chapters on mechanics we learned that energy is conserved; it is neither created nor destroyed but is.
Capacitors. What is a capacitor? Electronic component Two conducting surfaces separated by an insulating material Stores charge Uses –Time delays –Filters.
ELECTRODES FOR STIMULATION. Cardiac Pacemakers and Defibrillators  The heart is a natural example of a critical Bioelectromagnetic system.  An electrocardiogram.
EE 4BD4 Lecture 24 Defibrillator Biomedical Device Technology: Principles and Design, Charles C. Thomas Publisher 2007 Anthony K. Chan 1.
Copyright © 2000 Agilent Technologies, Inc.
Limmer, First Responder: A Skills Approach, 7th ed. © 2007 by Pearson Education, Inc. Upper Saddle River, NJ Chapter 9 Automated External Defibrillation.
describes the relationship between current, voltage, and resistance greater the voltage across a device with resistance, the greater the current through.
NEED FOR CARDIAC PACEMAKER Rhythmic beating of heart originates from SA node If SA node doesn’t function properly results in decreased heart rate & change.
Resources to Complete AED Certification
Chapter Three - AED. The Heart’s Electrical System - The heart’s electrical system sends out signals that tell the heart to pump blood. The signal travels.
Using an Automatic External Defibrillator (AED) Korean Red Cross Important Definitions  Ventricular Fibrillation (VF): The most common abnormal rhythm.
SPH3U: Electricity Electric Potential Difference.
The Students will be able to: Learn how defibrillation works. Identify the general steps in using an AED. Learn Precautions when using an AED. Practice.
Bell Ringer Grab the following  1 wire  1 light bulb  1 battery  Using just this material. Light the light bulb.  Draw what you did.
Electrocardiogram Amplifier Design Using Basic Electronic Parts
Defibrillators WHO. “Defibrillator, External, Automated; Semiautomated.” From the publication: Core Medical Equipment. Geneva, Switzerland, 2011.
Chapter 9 CAPACITOR.
1 Automated External Defibrillator. 2 Sudden Cardiac Arrest Sudden cardiac arrest is the most common cause of death in Australia. Sudden Cardiac Arrest.
H. Reihani MD Emergency Department MUMS Dr H Reihani.
Chapter 2 Cardiac Emergencies. Cardiac Emergencies Objectives 1. Identify the common cause of a heart attack 2. List signs and symptoms of a heart attack.
22nd April 2009 ECG Recording and Basic Interpretation.
Defibrillation 10/24/2017
BME 311: BIOMEDICAL INSTRUMENTATION I Lecturer: Ali Işın
DEFIBRILLATOR & PACE MAKERS
ELECTRICAL SAFETY FOR MEDICAL EQUIPMENT By m.naga sundhari
Electric Current Refers to the rate of movement of electric charge (when electrons move from one place to another) Ampere (A) – unit of electric current.
Chapter 4 Automated External Defibrillation
Control of Heart Contractions
AED- Automated External Defibrillation
CPR and Automated External Defibrillation (AED)
LIFEPAK® 1000 Defibrillator Orientation Guide
CARDIAC DEFIBRILLATORS
CARDIAC DEFIBRILLATORS
Current and Resistance
CURRENT ELECTRICITY: Components.
Automated External Defibrillation
Presentation transcript:

In The Name of Allah The Most Beneficent The Most Merciful 1

ECE 4550: Biomedical Instrumentation Lecture: Defibrillators Engr. Ijlal Haider University of Lahore, Lahore 2

NEED FOR A DEFIBRILLATOR  Ventricular fibrillation is a serious cardiac emergency resulting from asynchronous contraction of the heart muscles.  Due to ventricular fibrillation, there is an irregular or rapid heart rhythm. Fig. Ventricular fibrillation Fig. Normal heart beat

 Ventricular fibrillation can be converted into a more efficient rhythm by applying a high energy shock to the heart.  This sudden surge across the heart causes all muscle fibres to contract simultaneously.  The instrument for administering the shock is called a DEFIBRILLATOR.  Possibly, the fibres may then respond to normal physiological pacemaking pulses. NEED FOR A DEFIBRILLATOR

5 TYPES OF DEFIBRILLATORS InternalExternal

TYPES OF DEFIBRILLATORS a) Internal defibrillator Electrodes placed directly to the heart Eg.-Pacemaker b) External defibrillator Electrodes placed directly on the heart Eg.-AED

DEFIBRILLATOR ELECTRODES  Types of Defibrillator electrodes:- a) Spoon shaped electrode Applied directly to the heart. b)Paddle type electrode Applied against the chest wall c)Pad type electrode Applied directly on chest wall

DEFIBRILLATOR ELECTRODES

Fig.- Pad electrode DEFIBRILLATOR ELECTRODES

 Electrodes have insulated handles  Designed to prevent the spread of jell from electrodes to handles for the safety and ease of operator

PRINCIPLE OF DEFIBRILLATION  Energy storage capacitor is charged at relatively slow rate from AC line.  Energy stored in capacitor is then delivered at a relatively rapid rate to chest of the patient.  Simple arrangement involve the discharge of capacitor energy through the patient’s own resistance.

12

PRINCIPLE OF DEFIBRILLATION

 The discharge resistance which the patient represents as purely ohmic resistance of 50 to 100Ω approximately for a typical electrode size of 80cm2.  This particular waveform Fig 13.9(b) is called ‘ Lown’ waveform.  The pulse width of this waveform is generally 5-10 ms.

15

16  Peak current of 50 A  Time 5 sec  Resistance 50 ohm  Power delivered = 320 J  To overcome losses across capacitor and inductor  25% additional energy required  Energy to be stored is 400 J  Capacitor of 16 micro farad  Voltage required is 7000V

17  For Internal electrodes only 50 J is needed

Classes of discharge waveform Monophasic pulse or waveform Bi-phasic pulse or waveform

Classes of discharge waveform  There are two general classes of waveforms: a)mono-phasic waveform Energy delivered in one direction through the patient’s heart a)Biphasic waveform Energy delivered in both direction throuth the patient’s heart

Classes of discharge waveform Fig:- Generation of bi-phasic waveform

Classes of discharge waveform  The biphasic waveform is preferred over monophasic waveform to defibrillate.why????? A monophasic type, give a high-energy shock, up to 360 to 400 joules due to which increased cardiac injury and in burns the chest around the shock pad sites. A biphasic type, give two sequential lower-energy shocks of joules, with each shock moving in an opposite polarity between the pads.

Classes of discharge waveform  Dual Peak Waveform  10 msec  Truncated Waveform  Time can be adjusted

AUTOMATIC EXTERNAL DEFIBRILLATOR

 AEDs require self-adhesive electrodes instead of hand held paddles.  AED is a type of external defibrillation process. AUTOMATIC EXTERNAL DEFIBRILLATOR  AED is a portable electronic device that automatically diagnoses the ventricular fibrillation in a patient.  Automatic refers to the ability to autonomously analyse the patient's condition.  The AED uses voice prompts, lights and text messages to tell the rescuer what steps have to take next.

ELECTRODE PLACEMENT OF AED Anterior electrode pad Apex electrode pad Fig. anterior –apex scheme of electrode placement

WORKING OF AED  turned on or opened AED.  AED will instruct the user to:- Connect the electrodes (pads) to the patient. Avoid touching the patient to avoid false readings by the unit. The AED examine the electrical output from the heart and determine the patient is in a shockable rhythm or not.

When charged, the device instructs the user to ensure no one is touching the victim and then to press a red button to deliver the shock. when device determined that shock is warranted, it will charge its internal capacitor in preparation to deliver the shock. WORKING OF AED  Many AED units have an 'event memory' which store the ECG of the patient along with details of the time the unit was activated and the number and strength of any shocks delivered.

28 Courtesy Texas Instruments

Testing Defibrillators A dummy load is connected across the defibrillator electrodes Discharge is observed 50 ohm resistance is taken as human resistance connected in series with a low value resistor. Neon light is connected with low value resistor and lights when defibrillator is discharged

Testing Defibrillators Some defibrillators have built-in test system If they do not have, it can be designed and mounted on the trolley

31 Current requirements normally range up to 20 A. Voltage ranges from 1000V to 6000V Time of discharge is kept from 5 to 10 msec Current is dependent on the body (chest) resistance For reference visit: ic-waveform/ ic-waveform/ And various other resources on internet