MRI Artifacts.

Slides:



Advertisements
Similar presentations
Pulse Timing parameters & Weighting
Advertisements

BOLD Imaging at 7T Mark Elliott CfN Symposium 4/9/2008.
Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
Magnetic Resonance Imaging
Imaging Sequences part I
Proton Spin In absence of a magnetic field, protons spin at random
MRI Phillip W Patton, Ph.D..
In Chan Song, Ph.D. Seoul National University Hospital
Richard Wise FMRI Director +44(0)
M R I Pulse Sequences Jerry Allison Ph.D..
BE 581 Lecture 3- Intro to MRI.
PHYSICS OF MAGNETIC RESONANCE
Magnetic Resonance Imaging - MRI
MR Sequences and Techniques
Statistical Parametric Mapping
Chapter 14: Artifacts Mark D. Herbst, MD, PhD. Artifact Something on the image that does not represent something in the patient –Many causes Equipment.
Parameters and Trade-offs
Topics spatial encoding - part 2. Slice Selection  z y x 0 imaging plane    z gradient.
Principles of MRI: Image Formation
Chapter 9 Basic MRI I Mark D. Herbst, MD, PhD. Notice This lecture contained many drawings on the whiteboard, so get these from one of the other students.
Magnetic Resonance Imaging
Noll Spin-Warp Imaging For each RF pulse:For each RF pulse: –Frequency encoding is performed in one direction –A single phase encoding value is obtained.
Functional Brain Signal Processing: EEG & fMRI Lesson 12 Kaushik Majumdar Indian Statistical Institute Bangalore Center M.Tech.
Terry M. Button, Ph.D. Principals of Magnetic Resonance Image Formation.
Chapter 10 Lecture Mark D. Herbst, M.D., Ph.D..
EPI – Echo Planar Imaging Joakim Rydell
Basics of Magnetic Resonance Imaging
Magnetic Resonance Imaging Mary Holleboom ENGR 302 May 7, 2002.
Compressed Sensing for Chemical Shift-Based Water-Fat Separation Doneva M., Bornert P., Eggers H., Mertins A., Pauly J., and Lustig M., Magnetic Resonance.
Tissue Contrast intrinsic factors –relative quantity of protons tissue proton density –relaxation properties of tissues T1 & T2 relaxation secondary factors.
MRI: an Introduction By Mohammad Ali Ahmadi Pajouh Amirkabir University of Technology Biomedical Eng. Dep.
MRI Image Formation Karla Miller FMRIB Physics Group.
Medical Imaging Systems: MRI Image Formation
Magnetic Resonance Imaging 4
Principles of MRI Physics and Engineering
Imaging Sequences part II
The comparison of MRI imaging In 3.0 T & 1.5T By a.r.shoaie.
Medical Imaging Systems: MRI Image Formation
Resident Categorical Course
RT 4912 Review (A) Rex T. Christensen MHA RT (R) (MR) (CT) (ARRT) CIIP.
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
Quiz In a 2D spin warp or FT MR scan, aliasing should only occur
Rad T 265 MRI Lecture. No Magnetic Field = No Net Magnetization Protons align with a magnetic field…
G Practical MRI 1 Gradients.
Contrast Mechanism and Pulse Sequences
Magnetic Resonance Imaging
Basic MRI Chapter 1 Lecture. Introduction MRI uses radio waves and a magnetic field to make images MRI uses radio waves and a magnetic field to make images.
Anna Beaumont FRCR Part I Physics
V.G.Wimalasena Principal School of Radiography
MR Image Formation FMRI Graduate Course (NBIO 381, PSY 362)
MRI: Contrast Mechanisms and Pulse Sequences
Magnetic Resonance Learning Objectives
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
DTI Acquisition Guide Donald Brien February 2016.
Charged particle. Moving charge = current Associated magnetic field - B.
Real time shimming (RTS) for compensation of respiratory induced field changes P van Gelderen, JA de Zwart, P Starewicz, RS Hinks, JH Duyn Introduction.
CT and MR artifacts.
Artifacts and Suppression Techniques
Chapter 5 Mark D. Herbst, M.D., Ph.D.. The MR Imaging Process Two major functions –Acquisition of RF signals –Reconstruction of images.
Parameters which can be “optimized” Functional Contrast Image signal to noise Hemodynamic Specificity Image quality (warping, dropout) Speed Resolution.
MRI Artifacts Ray Ballinger, MD, PhD
FMRI data acquisition.
MRI ARTEFACTS.
Magnetic Resonance Imaging
Sunday Case of the Day Physics
Where Mt is the magnetization at time = t, the time after the 90o pulse, Mmax is the maximum magnetization at full recovery. At a time = one T1, the signal.
Magnetic Resonance Imaging: Physical Principles
Parallel Imaging Artifacts in Body Magnetic Resonance Imaging
MRI ARTEFACTS PG Student : Dr Aditi Shah PG Guide: Dr Rajendra Chavan
T2 Relaxation Time T2 relaxation time is defined as the time needed to dephase up to 37% of the original value. T2 relaxation refers to the progressive.
Presentation transcript:

MRI Artifacts

Introduction There are numerous kinds of artifacts that can occur in MRI. Some effect the quality of the MRI exam. Others may be confused with pathology. Some artifacts can be mitigated by the MR Tech while others require an engineer.

Sources of Artifacts Hardware Issues e.g. calibration, power stability Software problems e.g. programming errors Physiological phenomena e.g. blood flow Physics limitations e.g. Gibbs and susceptibility

Types of Artifacts Chemical Shift Artifact Aliasing Black Boundary Artifact Gibbs or Truncation Artifact Zipper Artifact Motion Artifacts (Phase direction) Entry Slice Phenomenon Field inhomogeneity Slice-overlap Artifact

Types of Artifacts (continued) Magic Angle Effects Moire Fringes RF Overflow Artifact Central Point Artifact Quadrature Ghost Susceptibility Artifact Zero-fill (Zebra Artifact) Eddy Current Artifacts Diastolic Pseudogating Gadolinium Pseudolayering

Chemical Shift Artifact Frequency-encoding direction The different resonant frequency of fat & water is transformed into spatial difference. Common in vertebral bodies, orbits, solid organs surrounded by fat. Worst at higher field strength, less with stronger gradients.

Chemical Shift Artifact

Aliasing or "Wrap-around " Occurs when the field of view (FOV) is smaller than the body part being imaged causing the region beyond to project on the other side of the image. Caused by undersampling in the phase or (rarely) frequency direction. May occur in end slices of a 3D acquisition.

Aliasing or "Wrap-around "

Aliasing

Aliasing or "Wrap-around " Correction: Increase the FOV (decreases resolution). Oversampling the data in the frequency direction (standard) and increasing phase steps in the phase-encoded direction – phase compensation (time or SNR penalty). Swapping phase and frequency direction so phase is in the narrower direction. Use surface coil so no signal detected outside of FOV.

Black Line Artifact An artificially created black line located at fat-water interfaces such as muscle-fat interfaces. Occurs at TE when the fat and water spins located in the same pixel are out of phase, cancelling each other’s signal. Particularly noticeable on GE sequences. Both freq and phase direction. At 1.5 Tesla, occurs at 4.5 ms multiples, starting at about 2.3 ms.

Black Line Artifact

Black Line Artifact Mitigation: Use in-phase TE’s Fat suppression Increase bandwidth or matrix size.

Gibbs or Truncation Artifact Bright or dark lines that are seen parallel & next to borders of abrupt intensity change. May simulate a syrinx on sagittal image of spinal cord. Related to the finite number of encoding steps used by the Fourier transform. Mitigation: More encoding steps lessen the intensity and narrows the artifact.

Gibbs or Truncation Artifact 128x256 256x256

Zipper Artifacts Most are related to hardware or software problems beyond the radiologist control. May occur in either frequency or phase direction. Zipper artifacts from RF entering room are oriented perpendicular to the frequency direction.

Zipper Artifacts

Motion Artifacts Bright noise or repeating densities usually oriented in the phase direction. Extend across the entire FOV, unlike truncation artifacts that diminish quickly away from the boundary causing them. Examples: Arterial pulsations, CSF pulsations, swallowing, breathing, peristalsis, and physical movement.

Motion Artifacts Mitigation: Arterial and CSF pulsation artifacts can be reduced with flow compensation and cardiac gaiting. Spatial presaturation can reduce some swallowing and breathing artifacts and arterial pulsations.

Motion Artifacts Mitigation (cont.): Surface coil localization can reduce artifacts generated at a distance from the area of interest.

Motion Artifacts

Slice-overlap (cross-slice) Artifacts Loss of signal seen in an image from a multi-angle, multi-slice acquisition. Same mechanism as spatial presaturation for reduction of motion and flow artifacts. Example: Two groups of non-parallel slices in the same sequence, e.g., L4-5 and L5-S1. The level acquired second will include spins that have already been saturated.

Slice-overlap (cross-slice) Artifacts

Slice-overlap Artifacts

Slice-overlap Artifacts Correction: Avoid steep change in angle between slice groups. Use separate acquisitions. Use small flip angle, i.e. GE sequence

Cross-talk Artifact Result of imperfect slice excitation, i.e. non-rectangular, of adjacent slices causing reduction in signal over entire image. May be reduced by using gap, interleaving slices and optimized (but longer) rf pulses.

Cross-talk Artifact

Magic Angle Effects Seen most frequently in tendons and ligaments that are oriented at a 55o angle to the main magnetic field. Normal dipolar interactions between the H+’s in water molecule aligned in tendons shortens T2, causing loss of signal. The dipolar interactions go to zero at about 55o increasing the signal.

Magic Angle Effects

Entry slice (Inflow) artifact Unsaturated spins in blood or CSF entering the initial slices results in greater signal than reduces on subsequent slices. May be confused with thrombus. Can use spatial saturation to reduce. Mechanism for TOF angiography.

Entry slice (Inflow) artifact

Field inhomogeneity Types: Main magnetic field RF coil inhomogeneity Dielectric effects – worst at 3T+ May cause variation in intensity across image May cause non-uniform fat suppression

Field inhomogeneity – Bo

Field inhomogeneity- RF coil

Field inhomogeneity- Dielectric

Field inhomogeneity Mitigation: Shimming, area of interest in near isocenter Use STIR for Fat sat vs. Chess. Caution with Gad. Coil – Use volume vs. surface coil, allow space between coil and body. Dielectric – use phased array coils, software compensation

RF Overflow Artifacts (Clipping) Causes a nonuniform, washed-out appearance to an image. Occurs when the signal received from the amplifier exceeds the dynamic range the analog-to-digital converter causing clipping. Autoprescanning usually adjusts the receiver gain to prevent this from occurring.

RF Overflow Artifacts

Moire Fringes Moire fringes are an interference pattern most commonly seen when doing gradient echo images. One cause is aliasing of one side of the body to the other results in superimposition of signals of different phases that add and cancel. Can also be caused by receiver picking up a stimulated echo. Similar to the effect of looking though two window screens.

Moire Fringes

Central Point Artifact A focal dot of increased or decreased signal in the center of an image. Caused by a constant offset of the DC voltage in the amplifiers.

Central Point Artifact

Central Point Artifact Correction: Requires recalibration by engineer Maintain a constant temperature in equipment room for amplifiers.

Quadrature ghost artifact Another amplifier artifact caused by unbalanced gain in the two channels of a quadrature coil. Combining two signals of different intensity causes some frequencies to become less than zero causing 180 degree “ghost.”

Quadrature ghost artifact Image courtesy of http://www.cis.rit.edu/htbooks/mri/

Susceptibility Artifacts Variations in the magnetic field strength that occurs near the interfaces of substance of different magnetic susceptibility such as ferromagnetic foreign bodies. Causes dephasing of spins and frequency shifts of the surrounding tissue.

Susceptibility Artifacts Worst with long echo times and with gradient echo sequences. Worst at higher magnetic field strength. Less with fast/turbo spin echo sequences.

Susceptibility Artifacts

Susceptibility Artifact 2

Zebra Artifacts Band-like, usually oblique stripes. Data in the K-space array will be missing or will be set to zero by the scanner or an electrical spike may occur as from static. The abrupt change from signal to no signal or normal signal to high signal results in artifacts in the images.

Zebra Artifacts

Eddy Current Artifacts Varying magnetic field from gradients can induce electrical currents in conductors such as the cryostat causing distortion of the gradient waveforms. Particularly a problem with echo-planar imaging that uses strong, rapidly changing gradients.

Eddy Current Artifacts Image courtesy of http://www.mr-tip.com/

Eddy Current Artifacts Mitigation: Precompensation- A “distorted” gradient waveform is used which corrects to normal with the eddy current effects. Shielded gradients – Active shielding coils between gradient coils and main gradients.

Diastolic Pseudogating Change in intensity of blood in large vessel such as the aorta from slice to slice when there is synchronization of the cardiac cycle and the pulse sequence, i.e., repetition rate = heart rate (TR=1/HR) Synchronization of the cardiac cycle and the pulse sequence results in high signal in the artery during diastole when blood is relatively stationary and loss of signal during systole when flow is high.

Diastolic Pseudogating

Gadolinium “Pseudolayering” Three density layers in the bladder after Gd Low conc. Gd top layer = dark Med conc. Gd middle layer = bright High conc. Gd lower layer near ureters = dark T2 shortening overshadows normal T1 effects at high concentrations.

Gadolinium “Pseudolayering” Left image from A. Elster’s book

MRI Project Short bond paper Submit an MRI protocols on March 14-16 on the following examination: Neuroscanning Spine scanning Body scanning Neck scanning Pediatric scanning Contrast examinations Short bond paper

Protocols Give 2 protocols per examination It should include: Patient preparation Indication Sequence/Routine Related Imagery Coil used Contrast media