THE DICOM 2013 INTERNATIONAL CONFERENCE & SEMINAR March 14-16Bangalore, India Analysis of Errors in DICOM Image for Rapid Prototyping Application Manmadhachary.A.

Slides:



Advertisements
Similar presentations
PRINCIPLES OF CT.
Advertisements

Reconstruction from Voxels (GATE-540)
Computers and Computed Tomography
Image Reconstruction.
The Field of Digital Radiography
Introduction Radiostereometry (RSA) is well established research method for assessing movement of orthopaedic implants, bone fractures, and.
The use of Rapid Prototyping to assist medical applications I. Gibson, L.K. Cheung, S.P. Chow, W.L. Cheung, S.L. Beh, M. Savalani, S.H. Lee.
Parameters and Trade-offs
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Topics spatial encoding - part 2. Slice Selection  z y x 0 imaging plane    z gradient.
Automatic Segmentation and Volume Estimation of Lung Cancer Nodules using Computed Tomography Timothy R. Tuinstra, PhD Candidate.
MEGN 536 – Computational Biomechanics Prof. Anthony J. Petrella Basics of Medical Imaging Introduction to Mimics Software.
Advanced Biomedical Imaging
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
CT Physics V.G.Wimalasena Principal School of radiography.
Computed Tomography III
Surface Reconstruction from 3D Volume Data. Problem Definition Construct polyhedral surfaces from regularly-sampled 3D digital volumes.
tomos = slice, graphein = to write
CT scanning (f) Candidates should be able to show an understanding of the principles of CT scanning. (g) Candidates should be able to show an understanding.
Lean Body Mass Assessment: Interpretations of Computed Tomography
X-ray Based Craniofacial Visualization and Surgery Simulation Junjun Pan 1, Jian J Zhang 1, Yanning Zhang 2, Hong Zhou 3
Tissue Contrast intrinsic factors –relative quantity of protons tissue proton density –relaxation properties of tissues T1 & T2 relaxation secondary factors.
Conventional and Computed Tomography
…….CT Physics - Continued V.G.WimalasenaPrincipal School of radiography.
COMPUTED TOMOGRAPHY (I) RAD 365 CT - Scan
Basic principles Geometry and historical development
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
Digital Image Characteristic
Demetrios J. Halazonetis Cone-beam Computed Tomography in Orthodontics: Limitations and future directions.
An Interactive Segmentation Approach Using Color Pre- processing Marisol Martinez Escobar Ph.D Candidate Major Professor: Eliot Winer Department of Mechanical.
Statistical analysis of pore space geometry Stefano Favretto Supervisor : Prof. Martin Blunt Petroleum Engineering and Rock Mechanics Research Group Department.
X-Ray Medical Imaging Physics – IB Objectives
3D Desktop Printing (printer) By G.Seshu Kumar (10A25A0503)
Computed Tomography Physics, Instrumentation, and Imaging
Module B Computed Tomography Physics, Instrumentation, and Imaging.
Information Needed for Segmenting Patient-Specific Abdominal Aortic Aneurysms Using Serial CT Scan Images with Contrast By: Jaime E. Zelaya, Jr. MD/PhD.
PRINCIPLES OF CT Dr Mohamed El Safwany, MD. Intended learning outcome The student should learn at the end of this lecture principles of CT.
Factors affecting CT image RAD
IAEA International Atomic Energy Agency Optimization of Protection in Computed Tomography (CT)-What can radiographers do? IAEA Regional Training Course.
1 POLY JET PROCESS - An effective RP Technique for Sheet Metal Works.
Geant4 DICOM Interface Susanna Guatelli INFN Genova, Italy 8 th March 2004 How to model human anatomy in a Geant4 application.
High resolution X-ray analysis of a proximal human femur with synchrotron radiation and an innovative linear detector M.Bettuzzi, R. Brancaccio, F.Casali,
Creating Medical Prototypes from MRI data Jillian Urban Mentor: John T. Tester, Ph.D. Mechanical Engineering.
3D Digital Cleansing Using Segmentation Rays Authors: Sarang Lakare, Ming Wan, Mie Sato and Arie Kaufman Source: In Proceedings of the IEEE Visualization.
CT IMAGE RECONSTRUCTION  Hounsfield envisioned dividing a slice into a matrix of 3-dimensional rectangular boxes (voxels) of material (tissue). Conventionally,
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
MEGN 536 – Computational Biomechanics
Quality Assurance.
A SEMINAR ON 1. CONTENTS Prototyping Rapid prototyping process Stereo lithography  Machine  Process  Material used Benefits Application Problems with.
By Prof. Stelmark. Digital Imaging In digital imaging, the latent image is stored as digital data and must be processed by the computer for viewing on.
Copyright © 2012, 2006, 2000, 1996 by Saunders, an imprint of Elsevier Inc. Chapter 26 Three-Dimensional Digital Imaging.
Squared and Cubed Conversion Factors
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Radiology. The Cathode… Provides a source of electrons and directs the electrons towards to anode. The cathode has a coiled wire filament that emits electrons.
Lesson Plan: Drafting and Design J6-2. What is 3D solid modeling? How do 3D solid modeling programs work?
The Optimization of Reconstruction Method Reducing Partial Volume Effect in PET/CT 3D Image Acquisition Department of Nuclear Medicine, Samsung Medical.
MEGN 536 – Computational Biomechanics
-42 Impact of signal non-repeatability on spectral CT images
CT Multi-Slice CT.
CT PRINCIPLES AYMAN OSAMA.
UNIVERSITI SAINS MALAYSIA 2016
Digital Processing Techniques for Transmission Electron Microscope Images of Combustion-generated Soot Bing Hu and Jiangang Lu Department of Civil and.
CS 698 | Current Topics in Data Science
ATCM 6317 Procedural Animation
Basic principles Geometry and historical development
Basic Principles of CT Chapter 1.
Procedural Animation Lecture 10: Volume simulation
The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT.
Computed Tomography (C.T)
Presentation transcript:

THE DICOM 2013 INTERNATIONAL CONFERENCE & SEMINAR March 14-16Bangalore, India Analysis of Errors in DICOM Image for Rapid Prototyping Application Manmadhachary.A Department of Mechanical Engineering, National Institute of Technology, Warangal, Andhra Pradesh, , India.

Outline General procedure for creating medical models Introduction of Partial volume effect Partial volume effect due to matrix size Partial volume effect due to pixel size Partial volume effect due to voxel size Partial volume effect due to Field of View (FOV) Partial volume effect due to layer thickness and interval thickness Dumb-bell effect Results Conclusion References March 2013 DICOM International Conference & Seminar 2

General Procedure for Creating Medical Models  MRI/CT data acquisition  Creation of a CAD model  Conversion of the CAD model to STL format  Slicing the STL file into thin cross-sectional layers  Construction of the model one layer atop another  Cleaning and finishing the model  Preplanning surgery on RP Model and use as a prosthetic and implantation March 2013 DICOM International Conference & Seminar3 RP is a technology that produces models and prototype parts from 3D CAD model data, CT and MRI scan data, and model data created from 3D object digitizing systems

Partial Volume Effect or Volume Averaging  The DICOM image contains number of pixels.  Pixel value represents the proportional amount of x-ray energy that passes through anatomy and strikes the detector.  The information contained in each pixel is averaged so that one density number (or Hounsfield unit [HU]) is assigned to each pixel.  If an object is smaller than a pixel, its density will be averaged with the information in the remainder of the pixel.  This phenomenon is referred to as the partial volume effect or volume averaging. It results in a less accurate image.  Partial volume effect arises due to matrix size, field of view (FOV), voxel size, Layer Thickness and Interval Thickness. March 2013 DICOM International Conference & Seminar4

Partial Volume Effect due to Matrix Size  Generally 256x256, 512x512, 1024x1024 matrix sizes are used.  Based on this matrix size the image is constructed, while tissue size is constant  DICOM image errors depend on length of the square. Because after converting.stl file from this DICOM images, each square is divided into two triangles. These triangles height gives the error of RP model  Example: 512 x 512 mm 2 area of the tissue is considered for constructing the image  If selecting 512x512 matrix. In this each square the tissue is selected with 1 mm 2. Here error is 1mm  If selecting 256x256 matrix. In this each square the tissue is selected with 2 mm 2. Here error is 2mm  If selecting 1024x1024 matrix in this each square the tissue is selected with 0.5 mm 2 from this error is observed 0.5 mm  From the above example to minimize the error we have to choose 512x512 or 1024x1024 matrix size March 2013 DICOM International Conference & Seminar5

Partial Volume Effect due to Matrix Size

HU Selected AreaTissue Area Partial Volume effect is reduced by decreasing the pixel size In the RP process the RP model error depends on pixel size because these pixels are divided into two triangles in the “.stl” file. Partial Volume Effect due to Pixel size March 2013 DICOM International Conference & Seminar

Partial Volume Effect due to Field of View (FOV):  Field of View (FOV): area of the tissues covered to construct the image is called Field of View (FOV)  Based on this area the image is divided into number of pixels  Field of View (FOV)= Matrix size X Pixel size  Example: Suppose while taking the CT Scan of the mandible height of 10cm, a radiologist considers the full skull height of 30cm.  If the image is constructed in both cases by 512x512 matrix, for 10cm FOV the pixel size assigned is 0.19 mm. similarly for same matrix size for 30cm FOV the pixel size assigned is 0.58 mm.  If Pixel size increases accuracy of the RP model decreases  From the above observation we have to choose less FOV to construct the image in order to capture the required coverage area. March 2013 DICOM International Conference & Seminar8

Partial Volume Effect due to Voxel Size Voxel size: It represents a volume element. Voxel is the volume of pixel size area and X-Ray beam attended (Z – Axis direction) distance or slice thickness. The voxel is a cube of data. All of the data within the voxel are averaged together to result in one HU. So Voxel size also plays a vital role in partial volume effect. March 2013 DICOM International Conference & Seminar9 Pixel Slice thickness Volume element

In the voxel element Z axis–will be longer than either the X or Y dimensions. Therefore, the slice thickness will play an even larger role in volume averaging than matrix size. Partial Volume Effect due to Volume Element Size 2 mm slice thickness voxel element Two 1 mm slice thickness voxel elements March 2013 DICOM International Conference & Seminar

Partial Volume Effect due to Layer Thickness and Interval Thickness From the above observation to minimize the errors in RP model, we have to choose smaller layer thickness and interval thickness.

Dumb-bell effect 12 A smaller (or larger) threshold value than the true value ends up adding (subtracting) an additional layer to (from) the original boundary it is called dumb – bell effect.

Method March 2013 DICOM International Conference & Seminar13 Linear measurements of mandible3D model Measurements are measured on CATIA

Results parameter Selected values in CT machine Bicondylar width BCoW in mm Body length BCrW in mm Ramus right with (WRMH) in mm Ramus left with (WLMH) in mm Actual Dimensions Measured with vernier Matrix Size256X X X Layer Thickness0.6 mm mm mm Interval Thickness 0.6 mm mm mm Field of View30 cm cm threshold value in MIMICS software 220 HU HU HU

Conclusions  While taking CT image the following parameters are considered to produce good quality RP model Select 512x512 matrix size to construct the DICOM image Partial Volume effect is reduced by decreasing the pixel size, selecting smaller volume element and chooseing smaller layer thickness and interval thickness Choose less FOV to construct the image and capture the required coverage area.  For CT scanning to reduce the Partial volume effect, for a matrix size of 512×512, pixel size being 0.463mm × mm, voxel size 0.463mm × 0.463mm × 0.6 mm, layer thickness is 0.6mm and interval thickness is 0.6 mm. similarly to reduce the Dumb-bell effect, the threshold value for bone tissue 1000 HU, in MIMICS software 226 HU.  In this study, it is observed that a dimensional error of ±0.4 mm occurrs by comparing human anatomy and 3D model generated in MIMICS software using DICOM images. March 2013 DICOM International Conference & Seminar15

References 1.Goldman. L. W, “Principles of CT and CT Technology”, Journal of Nuclear Medicine Technology, 2007, vol 35, p.p 115– Mildenberger. P, Eichelbery. M and Martin. E, “Introduction to the DICOM Standard”, European Radiology, 2002, vol 12, P.P Milovanovic. J, Trajanovic. M, “Medical Applications of Rapid Prototyping”, Journal of Mechanical Engineering, 2007, Vol. 5, No 1, P.P 79 – Tukuru. N, Gowda K. P, Ahmed. S. M and Badami. S, “Rapid Prototype Technique in Medical Field”, Research Journal of Pharmacy and Technology, Oct.-Dec. 2008, Vol 1, P.P Herlin. G, Koppe.M, Béziat. J. L, Gleizal. A, “Rapid prototyping in craniofacial surgery: Using a positioning guide after zygomatic osteotomy - A case report” Journal of Cranio-Maxillo-Facial Surgery, 2011, vol 39, P.P 376 – Lethaus. B, Poort. L, Bockmann. R, Smeets. R, Tolba. R, Kessler. P, “Additive manufacturing for microvascular reconstruction of the mandible in 20 patients”, Journal of Cranio-Maxillo-Facial Surgery, 2012, vol 40, 43 – Choi. J.Y, Choi. J. H, Kim N. K, Kim.Y, Lee J.K, “Analysis of errors in medical rapid prototyping models”, International Journal of Oral & Maxillofacial surgery, 2002, vol 31, P.P 23–32. 8.Winder.J and Bibb. R, “Medical Rapid Prototyping Technologies: State of the Art and Current Limitations for Application in Oral and Maxillofacial Surgery”, American journal of Oral and Maxillofacial Surgery, 2005, vol 63, P.P Fischer. C. E. H, Vaandrager. J. M, Zonneveld. F. W, Andersen. B. P, “Precision and accuracy of CT-based measurements of masticatory muscles in patients with hemifacial microsomia”, journal of the International Association of Dentomaxillofacial Radiology, 2004, vol 33, P.P 12– Mallepree. T, Bergers. D, “Accuracy of medical RP models”, Rapid Prototyping Journal, 2009, vol 15, P.P 325– 332. March 2013 DICOM International Conference & Seminar16

Author Contacts Manmadhachary.A, Research Scholar, Department of Mechanicla Engineering, National Institute of Technology, Warangal, Andra Pradesh,India. E - MAIL:- March 2013 DICOM International Conference & Seminar17 Thank you for your attention !