A TRACEABILITY TOOL FOR ELEMENTAL SOLUTION STANDARDS Gregory C. Turk NIST.

Slides:



Advertisements
Similar presentations
Radiochemical Methods and Data Evaluation
Advertisements

Calibration Techniques
Instrumental Analysis
Lessons Learned from Measuring Cell Response by Quantitative Automated Microscopy FDA Workshop, Potency Measurements for Cellular and Gene Therapy Products,
Copyright Alan Rowley Associates Steps to an Accurate Result Select a method and validate it as suitable for the purpose envisaged. Establish that.
Introduction The use of qNMR for purity measurement has been steadily growing in recent years. The assessment of the purity of calibration materials and.
Result validation. Exercise 1 You’ve done an analysis to the best of your ability using the correct procedure. Is your answer correct? possibly, hopefully.
Chemical Analysis Qualitative Analysis Quantitative Analysis Determination “Analyze” a paint sample for lead “Determine” lead in a paint sample.
World Health Organization
Chapter 3 – Volumetric Analysis Week 1, Lesson 1.
AOSC 634 Air Sampling and Analysis Lecture 1 Measurement Theory Performance Characteristics of instruments Nomenclature and static response Copyright Brock.
Instrumental Methods: Intro  Types of Instrumental Methods  Fundamental Components of an Instrument  Instruments Measure Voltages and Currents! (“Machines”
Types of experimental error
Senior Lecturer in Pathology
CHEM-3245 Quantitative Analysis
Concentration and Dilution of Solutions. HOW TO MAKE A SOLUTION “The Procedure” We’ll make a Kool-Aid Solution!!!
CALIBRATION METHODS.
Validation of Analytical Method
Approach of a UK auditor Paul Mudway Mudway Health, Safety & Environment.
NIST Standard Reference Material Project: Pure DNA Standard for Cytomegalovirus SoGAT XX Warsaw, Poland June 13,2007 Marcia Holden.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.
Atomic Absorption Spectrometer Varian SpectrAA Model 220FS.
Solutions & Solubility
B. Neidhart, W. Wegscheider (Eds.): Quality in Chemical Measurements © Springer-Verlag Berlin Heidelberg 2000 A. RíosTraceability / Trackability 1 It is.
Quality Assurance How do you know your results are correct? How confident are you?
Williams, A.: Measurement Uncertainty© Springer-Verlag Berlin Heidelberg 2003 In: Wenclawiak, Koch, Hadjicostas (eds.) Quality Assurance in Analytical.
ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your measurements (Blank, 100ppb, 500ppb, 1000ppb, etc)
CALIBRATION METHODS. For many analytical techniques, we need to evaluate the response of the unknown sample against the responses of a set of standards.
Advanced Analytical Chemistry – CHM 6157® Y. CAIFlorida International University Updated on 9/26/2006Chapter 3ICPMS Interference equations Isobaric.
Instrumental Methods: Intro
Use of Reference Materials in Medical Testing SLAB 10th Anniversary Celebrations Accreditation Conference 10 th November 2015 Dr Deepani Siriwardhana Senior.
CHEM 213 Instrumental Analysis Lab Lecture – Fluorescence and Calculations.
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
Metrology for Chemical Analysis
Introduction The Equipment The Terms The Process Calculations
Traceability Ivo Leito Materials:
Temperature profiles for a nature gas/air flame.
Solutions & Solubility Solution Preparation. Solution Preparation from a solid  Standard Solution = a solution for which the precise concentration is.
QUANTITATIVE ANALYSIS Determining amount of ions present in samples.
Industrial Technology Institute Test Method Validation & Verification H.P.P.S.Somasiri Principal Research Scientist / SDD-QAD /QM Industrial Technology.
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
Precision, Uncertainty and Accuracy Not very precise!
1 Analytical Forensic Metrology Analytical Forensic Metrology 4 th International Conference on Forensic Research & Technology September 29, 2015 Jerry.
Uncertainty and error in measurement
Chapter 1: The Nature of Analytical Chemistry
SEMINAR ON PRESENTED BY BRAHMABHATT BANSARI K. M. PHARM PART DEPARTMENT OF PHARMACEUTICS AND PHARMACEUTICAL TECHNOLGY L. M. COLLEGE OF PHARMACY.
Chapter 5 Quality Assurance and
Chem. 31 – 6/1 Lecture. Announcements Website – now has links to past classes where you can fine example exams Homework and Quiz –quiz tomorrow –corrected.
THEORY OF SAMPLING MMEA Certainty Seminar Markku Ohenoja 1 Markku Ohenoja / Control Engineering Laboratory
T RACEABILITY OF R ECORDS. M EI B ETH S HEPHERD 3 rd Party Assessors Consultants.
1 NIST Electrolytic Conductivity Standard Reference Materials ® Traceability and Stability Issues Kenneth W. Pratt National Institute of Standards and.
Quality is a Lousy Idea-
An in-depth look at trace metallic impurity determination by ICP-OES
James Byrd, Marta Kozak 28 Apr 2011
Serpil KILIÇ, Murat KILIÇ
Chem. 31 – 8/30 Lecture.
Principles of Instrumental Analysis
Preparation of Solutions
Instrumental Methods: Intro
Chapter 0: The Analytical Process (Example)
Strategies for Eliminating Interferences in Optical Emission Spectroscopy Best practices to optimize your method and correct for interferences to produce.
iCAP OES Analysis of Trace Elements in Hair
Chapter 3 – Volumetric Analysis
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
Analytical Method Validation
Chapter 5 Quality Assurance and Calibration Methods
Solution Concentration
DETERMINATION OF GLUCOSE IN BLOOD STREAM
Solution Concentration
Presentation transcript:

A TRACEABILITY TOOL FOR ELEMENTAL SOLUTION STANDARDS Gregory C. Turk NIST

The NIST SRM 3100 Series -- Spectrometric Solutions NIST produces and certifies 69 Spectrometric Solution SRMs Typical mass fraction is 10 ± 0.03 mg/g Serve as national standards

An important link in the calibration infrastructure for elemental analysis

Traceability “Property of the result of a measurement or the value of a standard whereby it can be related to stated references, usually national or international standards, through an unbroken chain of comparisons all having stated uncertainties.”

Traceability of the value of a standard to a NIST SRM requires a comparison measurement NIST SRM 3113 Co Calibration Standard and the uncertainty of the traceable value must include the uncertainty of the comparison.

We have designed a “Traceability Tool” for Elemental Solution Standards a spreadsheet –an experimental design with replication of all critical measurements and preparation steps »to assess measurement uncertainty –a ratio-based method »interactively designs the correct “spike” –optional drift correction and diagnostics –calculates the traceable value –full uncertainty budget

What you need to use this tool An instrument that gives a signals linearly proportional to the mass fraction of the analyte and an internal standard mixed in a solution –Designed for ICP-OES, but not required

Plus 10 g of SRM 4 units of Test Solution Rough setup solutions for analyte and IS Stock solution of IS One 250 mL bottle Eight 30 mL bottles for spiked solutions Eight more bottles for diluted working solutions A 1 mg balance A digital pipette

The tool does NOT specify Any instrumental operating parameters –Wavelengths, background correction points, integration times ……… Choice of internal standard

ICP-OES is a great double-pan balance for comparing elemental solution standards. spectroscopic elemental selectivity simultaneous detection of analyte and internal standard emission –highly correlated noise excellent precision –better than 0.03 % in most cases –thus adds negligible uncertainty to traceable values

Ratio-Based Measurement ICP SRM Internal Standard Working SRM Test Solution Spiked Test Solution Spiked SRM Working Test Solution

Start select internal standard, analyte and internal standard emission lines determine sensitivities prepare Internal Standard Spike Solution weigh and spike aliquots of SRM and Test Samples prepare Working Solutions measure Working Solutions, repeat 5 times copy measurements to spreadsheet, review results Finish

Example Cobalt Standard g of SRM mg/g g water = mg/g

Step 1. Experiment Related Information

Step 1 (continued). Instrument Related Information

Step 2. Sensitivity Test HP-ICP Comparison of Example (lot 1) to NIST SRM 3113 (lot ), 10/26/2001.  measure instrument response for the Sensitivity Solutions of the analyte and internal standard ug/g HP-ICP response Sensitivity (response/(µg/ g)) Target working mass fraction Sensitivity ratio Co (Analyte) Sensitivity Solution E Sc (Int. Std.) Sensitivity Solution 52.20E

Step 3. Solution Preparation

Step 4. Paste Data

Correlation Diagnostic

Drift Diagnostic

Results (without drift correction)

Results (with drift correction)

Uncertainty Budget (without drift correction)

Uncertainty Budget (with drift correction)

Summary of Results

The uncertainty of the traceable value is identical to the uncertainty of the SRM. SRM 3113:9.996 ± mg/g Test Solution: ± mg/g

For more information For a copy of the traceability tool: – able=104-5.htm NIST Traceability Policy – Salit, et al., Anal. Chem. 73, (2001). Salit and Turk, Anal. Chem. 70, (1998). –Drift Correction Procedure

What’s next? Other traceability tools? –Metals? –______