Fundamentals of Fluorescence Microscopy

Slides:



Advertisements
Similar presentations
Fluorescence and Confocal Microscopy
Advertisements

CELL BIOLOGY TECHNIQUES
Multiphoton and Spectral Imaging. Multiphoton microscopy Predicted by Maria Göppert-Mayer in 1931 Implemented by Denk in early 1990s Principle: Instead.
Victor Sourjik ZMBH, University of Heidelberg
IMMUNODEPLETION OF CONDENSIN FROM XENOPUS EGG EXTRACTS HSSΔcond.Δmockmock boiled beadsextracts cond. anti- cond. anti- α-tub Smc2 
Fluorescence Fluorescent corals.
Observing the Microbial Cell
Microscopy Outline 1.Resolution and Simple Optical Microscope 2.Contrast enhancement: Dark field, Fluorescence (Chelsea & Peter), Phase Contrast, DIC 3.Newer.
Microscopy Lecture I.
HPLC Detectors UV-Vis Fluorescence Derek Jackson CHM410/1410 October 22, 2009
GFP Fluorescence/Confocal Microscope Kimberly Ching ABE Workshop Summer 2007.
July 2008 Jonathan Sommer.  Has two modes to view samples Optical mode (white light) and Fluorescence mode  Depending on your sample either mode maybe.
Microscopy Boot Camp /08/25 Nikitchenko Maxim Baktash Babadi.
Fluorescence microscopy – Principle and practical consideration Hiro Ohkura.
CALCIUM IMAGING Beata M. Wolska, Ph.D.
Geometrical Optics and Basic Imaging Light Paths of the Bright Field Microscope E. D. Salmon University of North Carolina at Chapel Hill.
Methods: Fluorescence Biochemistry 4000 Dr. Ute Kothe.
Non-radiative energy transfer from excited species to other molecules
Luminescence is rare in nonaromatic hydrocarbons. Possible if highly conjugated due to  –  * transitions. Seyhan Ege, Organic Chemistry, D.C. Heath and.
Luminol Chemoluminescence
Immunofluorescence Microscopy. Immunofluorescence Microscopy: When an antibody, or the antiimmunoglobulin antibody used to detect the antibody is labeled.
Luminescence is rare in nonaromatic hydrocarbons. Possible if highly conjugated due to  –  * transitions. Seyhan Ege, Organic Chemistry, D.C. Heath and.
Immunolabeling & Fluorescent Microscopy Presented by: Sumble Maha Khan ABE Workshop June 13 – 30, 2006.
Light and Fluorescence
Shpol’skii Spectroscopy Analytical potential of fluorescence spectroscopy often limited by unresolved band structure (5-50 nm)Analytical potential of fluorescence.
Study of Protein Association by Fluorescence-based Methods Kristin Michalski UWM RET Intern In association with Professor Vali Raicu.
Microscope.
Prof. Enrico Gratton - Lecture 6 - Part 1 Fluorescence Microscopy Instrumentation Light Sources: One-photon and Multi-photon Excitation Applications in.
Practical aspects of fluorescence microscopy
Illumination and Filters Foundations of Microscopy Series Amanda Combs Advanced Instrumentation and Physics.
DIC, fluorescence and confocal microscopy Department of Mechatronics GIST Yong-Gu Lee References: 1.“Fundamentals of light microscopy and electronic imaging,”
Fluorescence Techniques
Principles of fluorescent probes Anna Dubiel Warsaw,
Detection systems. 1Introduction 2Theoretical background Biochemistry/molecular biology 3Theoretical background computer science 4History of the field.
Plant and Mammalian Tissue Culture Fluorescence Microscope Basics and Beyond - much taken from Nikon’s Microscopy U tutorial -…and.
Epi-illumination is form of Kohler Illumination:
Chapter 15 Molecular Luminescence Spectrometry Three types of Luminescence methods are: (i) molecular fluorescence (ii) phosphorescence (iii) chemiluminescence.
Focal plane of lens full field microscope, magnification defined by image distance/object distance scanning microscope, magnification defined by display.
Powerpoint Templates Page 1 Powerpoint Templates Spectroscopic Microscopy.
Microscopes Compound Bright-Field Light Microscope
Fluorescent Proteins By James Dicarlo and Ingrid Spielman August 12, 2008 Aequorea victoria GFP chromophore.
Advanced Fluorescence & Confocal Microscopy 08/2007 Lecture by Dr. Dirk Lang Dept. of Human Biology UCT Medical School Room Phone:
Use of Ethidium Bromide in gel electrophoresis  Ethidium bromide is an intercalating agent commonly used as a fluorescent tag (nucleic acid stain) in.
Wavelength and Light Properties 575 nm  c  E = h  High Energy Low Energy.
Fluorescence microscopy Principle and applications.
Molecular Cell Biology Light Microscopy in Cell Biology Cooper Modified from a 2010 lecture by Richard McIntosh, University of Colorado.
Science 10 – Unit C BIOLOGY Chapter 1 – The Microscope.
Fluorescence Microscopy Kurt Thorn NIC. Why fluorescence? High contrast –Signal against dark background Highly specific, multi-color labeling –GFP etc.
Designing a Microscopy Experiment Kurt Thorn, PhD Director, Image from Susanne Rafelski, Marshall lab.
Fluorescence Microscopy Aalap Tripathy, 2004P3PS208.
Dr R.Jayaprada CONFOCAL MICROSCOPY.
23W!~ Anti-vWF Immunofluorescence in Mouse Brain Minjung Choi, Katie Stockdale, James O. McNamara Department of Internal Medicine, The University of Iowa,
Imaging.
Fluorescence Microscopy: Small Molecule Probes Kurt Thorn NIC
Microscopy.
Fluorescence and Fluorochromes Peter O’Toole Tel:
Trends in Biotechnology TB4 Technological Advances in Microscopes
Fluorescence and Fluorochromes Peter O’Toole Tel:
Fluorescence 101 Steve Lee MiraiBio Inc. STR 2003.
Microscope.
George Stokes ( ) Max Born ( ) E. Schroedinger ( )
Lesson 10 Fluorescence Staining =Fluorescent labeling
Anti-vWF Immunofluorescence in Mouse Brain
LIGHT MICROSCOPY variations
Förster Resonance Energy Transfer (FRET)
Volume 99, Issue 8, Pages (October 2010)
Immunofluorescence Lab. 6.
The Art of Flow Cytometry
Francesca Pennacchietti, Travis J. Gould, Samuel T. Hess 
Immunofluorescence Microscopy cell Biology Ptactical 3
Presentation transcript:

Fundamentals of Fluorescence Microscopy E. D. Salmon University of North Carolina at Chapel Hill References: Murphy Book; http://micro.magnet.fsu.edu/primer/techniques/ Fluorescence; and www.chroma.com

Basic Concept of Absorption and Emission

Common Fluorophores Have Complex Electronic Structures

Excitation and Emission Spectra

Jablonski Diagram

Basic Features of Fluorescence Excitation occurs in 10-15 sec Emission occurs in 10-12 – 10-8 sec Usually broad excitation spectrum w peak Usually broad emission spectrum w peak Stokes shift is separation of Ex. & Em peaks Iem = Iexeclj Photobleaching: Rate depends on Iex ,environment

Fluorophore Parameters Absorption coefficient at peak absorption Quantum efficiency at peak emission Photostability (e.g. fluorescein has 10,000 excitations before bleaching event) Stokes Shift Widths of excitation and emission spectra Fluorescence is polarized: absorption and emission usually for E vector in plane of conjugated bonds

Quantum Yields Compound Solvent Ex. l (nm) Quantum Yield Acridine Orange Ethanol 366 0.46 Benzene Ethanol 248 0.04 Eosin Water 366 0.16 Fluorescein Water 366 0.92 Rhodamine-B Ethanol 535 0.97 Chlorophyl-A Ethanol 644 0.23

Molecular Fluorescent Probes Specific Fluorescent Dyes (e.g. DAPI) Covalently bind fluorescent dye to purified protein Fluorescent Antibodies (e.g immunofluorescent labeling with primary and fluorescent secondary antibodies) Express in cells Green (C,Y,R) Fluorescent Protein (G, C,Y, R-FP) fused to protein of interest

There are Different Fluorescent Molecules for Different Jobs See Molecular Probes Catalog; Sigma Catalog; CloneTech for GFP

Green Fluorescent Protein (GFP) CloneTech

Multi-Wavelength Fluorescence Imaging

Basic Concept of Epi-Fluorescence Microscopy

Hg-Arc Lamp

Xenon Arc-Lamp Spectra

Arc Lamps for Epi-Fluorescence Lamp Type: XBO 150W/1 XBO 75W/2 HBO 200W/2 HBO 100W/2 HBO 50W/3 Current: DC DC DC DC DC Rate Power (watts): 150 75 200 100 50 Luminous Flux (lumens): 3000 950 10000 2200 1300 Light Intensity (Candella): 300 100 1000 260 150 Avg. Brightness (cd/cm): 15000 40000 40000 170000 90000 Arc Size (w x h in millimeters): 0.50 x 2.20 0.25 x 0.50 0.60 x 2.20 0.25 x 0.25 0.20 x 1.35 Life (Hours): 1200 400 400 200 200

Quartz-Halogen (Tungsten Filament) Lamp Spectra Current

Lasers Have Line Spectra

Ploem-Type Epi-Illuminator

Epi-Fluorescence Microscope

Arc Lamp Housing

Lamp Alignment

Alignment of Arc and Mirror Images at Objective Back Focal Plane (Use Centering-Screen or white Card on Stage W/O Objective)

Filter Cubes

Filter Cubes Are Not Inter-Changeable Between Different Manufactures

Basic Design Features

Exciter and Barrier Filters are Designed to Separate Emission Light from Excitation Light

Problems in Filter Design: Example Absorption and Emission Spectra

The Dichromatic Mirror Further Isolates the Emission Light from the Excitation Light Modern Interference-Reflection filter Design Can Give Sharp Cut-Off with High Transmission Efficiency for the Pass Wavelengths. See web-sites for “Chroma Technology” and “Omega Optical”

Multi-Wavelength Immunofluorescence Microscopy

Fluorophores for Triple-Label

Multiple Band-Pass Filters

Multiple Band-Pass Filters

Choose Wide-Band Emission Filters for Single Fluorophore to Maximize Sensitivity

Chroma Technology Corp Chroma Technology Corp. is an employee- owned company that produces the world's finest optical filters and filter sets. The company specializes in the design and manufacture of optical filters and coatings for applications which require the greatest precision in color separation, optical quality and signal purity. For more about us, see our About Chroma page. Welcome to our new website! This site is under construction, so if you don't find what you need please give us a call at (800) 824-7662. Handbook of Optical Filters for Fluorescence Microscopy: Download a copy of our "Handbook of Optical Filters for Fluorescence Microscopy" in Adobe Acrobat PDF format. www.chroma.com

Multi-Wavelength Immunofluorescence Microscopy

Multi-Wavelength Fluorescence Imaging

Multi-wavelength Fluorescence Imaging

Multi-Wavelength Fluorescence Imaging

Ploem-Type Epi-Illuminator

Parameters for Maximizing Sensitivity Use High Objective NA and Lowest Magnification: Ifl ~ IilNAobj4/Mtot2 Use high efficiency filters Use as few optical components as possible Close Field Diaphragm down as far as possible Buy the newest objective: select for best efficiency Match magnification to camera resolution: MMax = 3*Pixel Size of Detector/Optical Resolution E.g.: 3*7 mm/[0.6 *520nm/1.4] = 91X Reduce Photobleaching Use High Quantum Efficiency Detector in Camera

Reducing Photobleaching For fixed specimens use anti-fade compounds: These reduce oxygen effects 95% glycerol works quite well For live specimens, reduce oxygen with: - Oxyrase - Catalase + glucose + glucose-oxidase

Reducing Photobleaching: Anti-Fade Reagents for Fixed Specimens p-phenylenediamine: The most effective reagent for FITC. Also effective for Rhodamine. Should be adjusted to 0.1% p-phenylenediamine in glycerol/PBS for use. Reagent blackens when subjected to light exposure so it should be stored in a dark place. Skin contact is extremely dangerous.G. D. Johnson & G. M. Araujo (1981) J. Immunol. Methods, 43: 349-350 DABCO (1,4-diazabi-cyclo-2,2,2-octane): Highly effective for FITC. Although its effect is slightly lower than p-phenylenediamine, it is more resistant to light and features a higher level of safety.G. D. Johnson et. al., (1982) J. Immunol. Methods, 55: 231-242. n-propylgallate: The most effective reagent for Rhodamine, also effective for FITC. Should be adjusted to 1% propylgallate in glycerol/PBS for use. H. Giloh & J. W. Sedat (1982), Science, 217: 1252-12552. mercapto-ethylamine: Used to observe chromosome and DNA specimens stained with propidium iodide, acridine orange, or Chromomysin A3. Should be adjusted to 0.1mM 2-mercaptotheylamine in Tris-EDTAS. Fujita & T. Minamikawa (1990), Experimental Medicine, 8: 75-82

Use High Quantum Efficiency Camera Detector: e.g. ORCA cooled CCD

Cdc20 Persists At Kinetochores Throughout Mitosis and Exhibits Fast Kinetics: FRAP t1/2 = [4 sec (attached) 25 sec (unattached] Green: GFP-Cdc20 At Kinetochores Red: Phase Contrast Images of PtK1 Tissue Cells