15 Schoen Place Pittsford, NY 14534 Phone: 585-267-4860 Non-Blood Contacting Biventricular Mechanical Actuation for the Failing Heart.

Slides:



Advertisements
Similar presentations
Dementia IGHC Tatiana Kuznetsova, Jan A. Staessen University of Leuven, Belgium Diagnostic criteria for subclinical systolic and diastolic LV function.
Advertisements

 By the end of this lecture the students are expected to:  Define cardiac output, stroke volume, end- diastolic and end-systolic volumes.  Define physiological.
Effects of selective heart rate reduction with ivabradine on left ventricular remodelling and function: results from the SHIFT echocardiography substudy.
LVAD is a battery-operated, mechanical pump-type device that's surgically implanted. Goal of LVAD: providing the patient with as close to a normal lifestyle.
Chapter 15 Assessment of Cardiac Output
ECHO/ DOPPLER CARDIOGRAPHY A diagnostic Study that reveals information about: The structure and function of the heart Cardiac hemodynamics of the heart.
ECG Rhythm Interpretation
Cardiovascular Dynamics During Exercise
Small Platform Catheter-Based Left Ventricular Assist Device Support Suppresses Cardioprotective Beta Arrestin- Mediated Signal Transduction Keshava Rajagopal,
Diagnostic Stress Testing
Ventricular Diastolic Filling and Function
Francesco Liistro Cardiovascular Department, Arezzo, Italy Impact of Thrombus Aspiration on Myocardial Tissue Reperfusion and Left Ventricular Functional.
Effects of PG , a Matrix Metalloproteinase Inhibitor to Prevent Left Ventricular Remodeling After Acute Myocardial Infarction Effects of PG ,
ECG Rhythm Interpretation
Safety and effectiveness of granulocyte-colony stimulating factor in advanced chronic heart failure Jacob Joseph, MD Veterans Affairs Healthcare System-
Thomas Wilson 1,2,3, Vijay Krishnamoorthy MD 1,2, Edward Gibbons MD 4, Ali Rowhani-Rahbar MD MPH PhD 2,5, Adeyinka Adedipe MD 6, Monica S. Vavilala MD.
GRK2 Inhibition Reduces Post-Myocardial Infarction Cardiac Fibroblast-Mediated Adverse Remodeling Jennifer L. Philip 1, Xianyao Xu 1,Mei Han 1, Jinju Li.
Date of download: 6/3/2016 Copyright © 2016 American Medical Association. All rights reserved. From: Effect of Perindopril on Large Artery Stiffness and.
Blood Vessels & Circulation. Recap Learned the structure and function of the heart. Visualized movement of blood through arteries, veins, and capillaries.
Conflict of Interest Baxter Research Grant Medtronic Research Grant
Date of download: 7/10/2016 Copyright © The American College of Cardiology. All rights reserved. From: Echocardiography for Percutaneous Heart Pumps J.
Date of download: 9/18/2016 Copyright © ASME. All rights reserved. From: Development of a Non-Blood Contacting Cardiac Assist and Support Device: An In.
Total Occlusion Study of Canada (TOSCA-2) Trial
ABSTRACT Background: Although 2D strain imaging has shown promise to define myocardial dysfunction, it provides only subjective assessment about the extent.
Supplementary Material
Hemodynamic effects of vacuum-assisted closure therapy in cardiac surgery: Assessment using magnetic resonance imaging  Rainer Petzina, MD, Martin Ugander,
Using 3D-Wall Motion Tracking
Using 4D Nongated Echocardiography
Using New High-Resolution 4D Echocardiography
Journal of Nuclear Cardiology | Official Journal of the American Society of Nuclear Cardiology 18Fluorine Sodium Fluoride Positron Emission Tomography,
Ventricular Pacing Alters Twisting Synchrony of the Left Ventricle
Drug Eluting Valvuloplasty Balloon Catheter
No relationships to disclose:
2-Dimensional Strain Studies in Open-Chest Instrumented Pigs
Usefulness of Matrix Metalloproteinase-9 Plasma Levels to Identify Patients With Preserved Left Ventricular Systolic Function After Acute Myocardial Infarction.
Meihua Zhu, MD, PhD; Muhammad Ashraf, MD
Left Ventricular Dysfunction With Pulmonary Hypertension
Mitochondrial Integrity and Function in the Progression of Early Pressure Overload–Induced Left Ventricular Remodeling by Antoine H. Chaanine, K Sreekumaran.
Intracellular Action of Matrix Metalloproteinase-2 Accounts for Acute Myocardial Ischemia and Reperfusion Injury by Wenjie Wang, Costas J. Schulze, Wilma.
The pathophysiology of myocardial infarction-induced heart failure
Hemodynamic effects of vacuum-assisted closure therapy in cardiac surgery: Assessment using magnetic resonance imaging  Rainer Petzina, MD, Martin Ugander,
ECG Rhythm Interpretation
Therapeutic Effect of Midkine on Cardiac Remodeling in Infarcted Rat Hearts  Shinya Fukui, MD, Satoru Kitagawa-Sakakida, MD, PhD, Sin Kawamata, MD, PhD,
Diastolic heart failure
Hemodynamics and myocardial blood flow patterns after placement of a cardiac passive restraint device in a model of dilated cardiomyopathy  Jennifer A.
Volume 9, Issue 1, Pages (July 2017)
Insights into the mechanism(s) of von Willebrand factor degradation during mechanical circulatory support  Carlo R. Bartoli, MD, PhD, Sujith Dassanayaka,
Cell-based gene therapy modifies matrix remodeling after a myocardial infarction in tissue inhibitor of matrix metalloproteinase-3–deficient mice  Denis.
Mechanism of myocardial ischemia with an anomalous left coronary artery from the right sinus of Valsalva  Carlo R. Bartoli, PhD, William B. Wead, PhD,
Hemodynamic effects of partial ventricular support in chronic heart failure: Results of simulation validated with in vivo data  Deborah Morley, PhD, Kenneth.
Is hyperglycemia bad for the heart during acute ischemia?
Mesenchymal stem cell implantation in a swine myocardial infarct model: engraftment and functional effects  Jay G. Shake, MD, Peter J. Gruber, MD, PhD,
ECG Rhythm Interpretation
Nonsurround, nonuniform, biventricular-capable direct cardiac compression provides Frank-Starling recruitment independent of left ventricular septal damage 
Heart Booster: a pericardial support device
Advancing the science of myocardial recovery with mechanical circulatory support: A Working Group of the National, Heart, Lung, and Blood Institute  Stavros.
Aortic valve graft implantation in rats: a new functional model
James E. Lowe, G. Chad Hughes, Shankha S. Biswas 
Dominique Shum-Tim, MD, Brian W
Volume 9, Issue 1, Pages (July 2017)
Improved left ventricular unloading and circulatory support with synchronized pulsatile left ventricular assistance compared with continuous-flow left.
Myocardial rescue with autologous mitochondrial transplantation in a porcine model of ischemia/reperfusion  Aditya K. Kaza, MD, Isaac Wamala, MD, Ingeborg.
Hironori Hara et al. BTS 2018;3:
Understanding the cardiac substrate and the underlying physiology: Implications for individualized treatment algorithm  John Gorcsan, MD, Frits W. Prinzen,
Durability of left ventricular assist devices: Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) 2006 to 2011  William L.
Early and persistent activation of myocardial apoptosis, bax and caspases: insights into mechanisms of progression of heart failure  Gordon W Moe, George.
Passive ventricular constraint to improve left ventricular function and mechanics in an ovine model of heart failure secondary to acute myocardial infarction 
Himanshu J. Patel, MDa, Edward B. Lankford, MD, PhDb, David J
Blocking the development of postischemic cardiomyopathy with viral gene transfer of the apoptosis repressor with caspase recruitment domain  Subhasis.
Myocardial remodeling with aortic stenosis and after aortic valve replacement: Mechanisms and future prognostic implications  William M. Yarbrough, MD,
Presentation transcript:

15 Schoen Place Pittsford, NY Phone: Non-Blood Contacting Biventricular Mechanical Actuation for the Failing Heart

2 Authors Curtis J. Wozniak MD Aaron M. Campbell Anthony J. Pothoulakis MD Lawrence J. Prochaska PhD Rebecca J. Darner Shawn M. Gargac Mark P. Anstadt MD

3 Background Direct Mechanical Ventricular Actuation (DMVA) is a unique, non-blood contacting, biventricular assist device that is being developed for support of the failing heart (Myotech Circulatory Support System (CSS)). The device is contoured to fit over the heart and attaches to the ventricular myocardium by an atraumatic vacuum seal (Figure 1). A pneumatic drive is then used to deliver positive (systolic) and negative (diastolic) actuating forces to the ventricular surface. DMVA has been shown to be effective in both laboratory and clinical applications for supporting the fibrillating, asystolic or severely failing heart. Current research is focused on utilizing a rabbit model of heart failure. Figure 1. Schematic of DMVA actuating the failing heart

4 Purpose To determine the effects of the Myotech CSS on maladaptive molecular signals and ventricular dynamics that characterize heart failure. Initial experiments have examined cell membrane disruption (a mechanism of cell signaling), matrix metalloproteinases (central to collagen breakdown), myocardial heat shock protein expression ( an indicator of cellular stress), and regional LV wall motion. Figure 2. Cell membrane disruption is important for myocardial cell signaling and varies with mechanical myocardial stress. Albumin crosses these transient membrane disruptions, and therefore, is a useful static measure of this cellular process.

5 Methods New Zealand white rabbits are anesthetized, undergo median sternotomy, and instrumented for hemodynamic monitoring. A transesophageal probe is positioned for echocardiographic interrogation. Esmolol is titrated to achieve a cardiac output of approximately 50 percent baseline. Animals are then either observed as unsupported heart failure (HF) controls or supported with DMVA application. Figure 3. A transesophageal probe placed in the esophagus for ventricular imaging. Figure 4. DMVA actuating the failing heart. Drive lines exit caudal. Aortic and pulmonary flow probes cephalad

6 Results Figure 5. Intracellular albumin was increased in all myocardial regions during HF. However, DMVA support decreased intracellular albumin in all these regions compared to unsupported HF, suggesting DMVA attenuates plasma membrane disruption.

7 Results Figure 6. Micrographs (200X) stained for albumin indicated by pink-red. The right depicts moderate to severe (grade 3 to 4), the left, trace to mild (grade 1 to 2) membrane disruption.

8 Results Figure 7. Mean cardiac output at baseline, failure, and recovery. During recovery, hearts supported by DMVA exhibited improved function.

9 Results Figure 8. A transesophageal echo probe is used to capture real-time echocardiographic images of DMVA actuating the failing heart. End-systolic actuation is shown in this image.

10 Results Figure 9. Attenuation of LV matrix metalloproteinase (MMP) activity following 30 and 120 minutes of DMVA vs. unsupported HF.

11 Results Figure 10. Attenuation of heat-shock protein (HSP-70) expression following 30 minutes of DMVA vs. unsupported HF.

12 Summary and Conclusions A rabbit model of acute heart failure has been successfully developed for assessing DMVA. The primary focus of these experiments is to determine how DMVA affects myocardial wall stress and maladaptive cell signaling characteristic of the failing heart. Current results indicate that DMVA can significantly augment the failing heart while reducing myocardial stress. These findings suggest that DMVA can provide adequate hemodynamic support while favorably altering the maladaptive pathophysiology of heart failure.

13 Future Studies Ongoing studies are further evaluating DMVA’s impact on maladaptive cellular responses which characterize HF including anti- and pro-apoptotic cell signaling, plasma membrane disruptions and collagen turnover. Mitochondrial function and detailed analysis of regional myocardial wall motion are also under investigation. Findings are expected to determine if and how DMVA favorably alters pathologic remolding of the failing heart.

14 Postscript Funding provided in part by Myotech, LLC and NIH/NHLBI (#2 T35 HL ) Dr. Anstadt is a consultant for Myotech, LLC.