The Physics of Diagnostic Ultrasound FRCR Physics Lectures

Slides:



Advertisements
Similar presentations
Mechanical Waves and Sound
Advertisements

Chapter3 Pulse-Echo Ultrasound Instrumentation
Foundations of Medical Ultrasonic Imaging
إعداد : أ. بلسم فهد صوفي 1 Ultrasound in Medicine Ch.3 Ultrasound pictures of the body.
إعداد : أ. بلسم فهد صوفي،،،المصدر:محاضرات د.حنان 1 Ultrasound in Medicine Ch.4 Ultrasound pictures of the body.
Resident Physics Lectures
Ultrasound Dr.mervat mostafa.
Hospital Physics Group
Pegasus Lectures, Inc. Volume II Companion Presentation Frank Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition.
EE 5340/7340, SMU Electrical Engineering Department, © Carlos E. Davila, Electrical Engineering Dept. Southern Methodist University slides can be.
ECE 501 Introduction to BME
Ultrasound Medical Imaging Imaging Science Fundamentals.
Sound Transmission and Echolocation Sound transmission –Sound properties –Attenuation Echolocation –Decoding information from echos.
Ultrasound Imaging Atam Dhawan.
Doppler Effect Physics 202 Professor Lee Carkner Lecture 11.
Phy 212: General Physics II
Sound In Medicine. Sound waves Sound is a mechanical waves travel in air as longitudinal waves in which vibrations in air causes increases (compressions)
Ultrasound – Physics & Advances
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Ultrasound Physics Sound is a mechanical, longitudinal wave that travels in a straight line Sound requires a medium through which to travel Ultrasound.
Ultrasonography.
4.4.1 Wave pulse: a wave pulse is a short wave with no repeated oscillations Progressive wave: a wave that moves through a medium transferring energy as.
Waves.
Waves. The Nature of Waves What is a mechanical wave?  A wave is a repeating disturbance or movement that transfers energy through matter or space 
Chapter 9 Electromagnetic Waves. 9.2 ELECTROMAGNETIC WAVES.
Waves A wave is a rhythmic disturbance that carries energy through matter or space.
Display of Motion & Doppler Ultrasound
George David Associate Professor Ultrasound Physics 04: Scanner ‘97.
Resident Categorical Course
Chapter 13 - Sound 13.1 Sound Waves.
1 Waves and Vibrations Spectra of Science Amole. 2 Waves are everywhere in nature Sound waves, visible light waves, radio waves, microwaves, water waves,
Display of Motion & Doppler Ultrasound
Review of Ultrasonic Imaging
Sound Vibration and Motion.
Chapter 17 Sound Waves: part two HW 2 (problems): 17.22, 17.35, 17.48, 17.58, 17.64, 34.4, 34.7, Due Friday, Sept. 11.
Lecture 21 Nature of Light Reflection and Refraction
Sarah Gillies Ultrasound Sarah Gillies
Doppler Ultrasound Dr Mohamed El Safwany, MD.. Introduction The Doppler Effect refers to the change in frequency that results when either the detector/observer.
Waves Superposition and Standing Waves The Electromagnetic Spectrum Pulse-Echo Techniques Refraction Polarisation Diffraction.
Saudi Board of Radiology: Physics Refresher Course Kostas Chantziantoniou, MSc 2, DABR Head, Imaging Physics Section King Faisal Specialist Hospital &
Ultrasound Physics Reflections & Attenuation ‘97.
DIFFRACTION AND INTERFERENCE. Specification Topics Interference The concept of path difference and coherence The laser as a source of coherent monochromatic.
Chapter 11 Vibrations and Waves.
BIOE 220/RAD 220 REVIEW SESSION 3 February 6, 2011.
1 Linear Wave Equation The maximum values of the transverse speed and transverse acceleration are v y, max =  A a y, max =  2 A The transverse speed.
Artifacts Ultrasound Physics George David, M.S.
Wave Properties. S8P4. Students will explore the wave nature of sound and electromagnetic radiation. d. Describe how the behavior of waves is affected.
Ultrasound Learning Objectives: Describe the properties of ultrasound;
Fundamental Ultrasound Principles Karen Potts Clinical Scientist Review date Jan 2010 Department of Medical Physics Kent & Canterbury.
TISSUE HARMONIC IMAGING (THI)
 Ultrasound waves are longitudinal with high frequencies ( ≈ > 20,000 Hz, though medical Ultrasound is between 1 to 15 MHz.)  When an ultrasound reaches.
1 Waves and Vibrations. 2 Waves are everywhere in nature Sound waves, visible light waves, radio waves, microwaves, water waves, earthquake waves, slinky.
TISSUE HARMONIC IMAGING (THI)
Holt Physics Chapter 12 Waves Periodic Motion A repeated motion that is back and forth over the same path.
Physics 1 What is a wave? A wave is: an energy-transferring disturbance moves through a material medium or a vacuum.
TISSUE HARMONIC IMAGING NUR FASHIHA BINTI AZMAN A DIAGNOSTIC IMAGING AND RADIOTHRAPY /2.
NUR INANI BINTI MOHD RUZLAN A Tissue harmonic imaging is a technique in ultrasonography that provides images of better quality as compared to conventional.
Chapter 10 Digital Signal and Image Processing
Lecture on Ultrasonography
Sound in medicine Lect.10.
Reflections & Attenuation
Chapter 6 Wave Motion.
Ultrasound.
Properties of sound.
Resident Physics Lectures
MCQs (1) – Questions Ultrasound propagates through the medium as a transverse wave Ultrasound velocity is equal to the product of frequency and wavelength.
Review of Ultrasonic Imaging
DIFFRACTION AND INTERFERENCE
Ultrasonic Imaging Using Contrast Agents
Presentation transcript:

The Physics of Diagnostic Ultrasound FRCR Physics Lectures Session 3 & 4 Mark Wilson Clinical Scientist (Radiotherapy) mark.wilson@hey.nhs.uk Hull and East Yorkshire Hospitals NHS Trust

Session 3 Overview Session Aims: Recap Image Artefacts Contrast Agents Introduction to Doppler US Hull and East Yorkshire Hospitals NHS Trust

Recap Hull and East Yorkshire Hospitals NHS Trust

Recap The term Ultrasound refers to high frequency sound waves. Sounds waves are mechanical pressure waves which propagate through a medium causing the particles of the medium to oscillate backward and forward The velocity and attenuation of the ultrasound wave is strongly dependent on the properties of the medium through which it is travelling  = c / f c =  k /  Hull and East Yorkshire Hospitals NHS Trust

Recap Diagnostic ultrasound utilises the pulse-echo principle Each pulse-echo sequence produces one line of the image Several pulse-echo sequences are needed to compose a full image frame. D Source of sound ) ) Sound reflected at boundary ) ) ) Pulse Distance = Speed x Time Echo 2D = c x t Hull and East Yorkshire Hospitals NHS Trust

Recap Reflection Scatter Refraction Attenuation and Absorption Ultrasound waves undergo the following interactions: Reflection Scatter Refraction Attenuation and Absorption Diffraction Hull and East Yorkshire Hospitals NHS Trust

Intensity Reflection Coefficient (R) Recap Reflection z1 z2 pi , Ii pt , It pr , Ir Intensity Reflection Coefficient (R) R = Z2 – Z1 Z1 + Z2 Ii Ir = ( ) 2 Acoustic Impedance z =  k Acoustic Impedance z = c Hull and East Yorkshire Hospitals NHS Trust

Recap Reflection Strength of reflection depends on the difference between the Z values of the two materials Ultrasound only possible when wave propagates through materials with similar acoustic impedances – only a small amount reflected and the rest transmitted Therefore, ultrasound not possible where air or bone interfaces are present Hull and East Yorkshire Hospitals NHS Trust

Recap Scatter Reflection occurs at large interfaces such as those between organs where there is a change in acoustic impedance Within most organs there are many small scale variations in acoustic properties which constitute small scale reflecting targets Reflection from such small targets does not follow the laws of reflection for large interfaces and is termed scattering Scattering redirects energy in all directions, but is a weak interaction compared to reflection at large interfaces Hull and East Yorkshire Hospitals NHS Trust

Recap sin (i) c1 = c2 sin (t) Refraction c1 c2 (>c1) Snell’s Law i When an ultrasound wave crosses a tissue boundary at an angle (non-normal incidence), where there is a change in the speed of sound c, the path of the wave is deflected as it crosses the boundary c1 c2 (>c1) Snell’s Law sin (i) i c1 = sin (t) c2 t Hull and East Yorkshire Hospitals NHS Trust

Where  is the attenuation coefficient Recap Attenuation As an ultrasound wave propagates through a medium, the intensity reduces with distance travelled Attenuation describes the reduction in intensity with distance and includes scattering, diffraction, and absorption Attenuation increases linearly with frequency Limits frequency used – trade off between penetration depth and resolution Intensity, I Low freq. High freq. Distance, d I = Ioe- d Where  is the attenuation coefficient Hull and East Yorkshire Hospitals NHS Trust

Recap Absorption Decibel Notation Decibel, dB = 10 log10 (I2 / I1) In soft tissue most energy loss (attenuation) is due to absorption Absorption is the process by which ultrasound energy is converted to heat in the medium Absorption is responsible for tissue heating Decibel Notation Attenuation and absorption is often expressed in terms of decibels Decibel, dB = 10 log10 (I2 / I1) Hull and East Yorkshire Hospitals NHS Trust

Image Artefacts Hull and East Yorkshire Hospitals NHS Trust

Artefacts Image Artefacts When forming a B-mode image, a number of assumptions are made about ultrasound propagation in tissue. These include: Speed of sound is constant Attenuation in tissue is constant Ultrasound pulse travels only to targets that are on the beam axis and back to the transducer Significant variations from these conditions in the target tissues are likely to give rise to visible image artefacts Hull and East Yorkshire Hospitals NHS Trust

Artefacts Range Errors The distance, d, to the target is derived from the time elapsed between transmission of the pulse and receipt of the echo from the target, t In making this calculation the system assumes that t = 2d / c, where the speed of sound is constant at 1540 m/s If the speed of sound in the medium between the transducer and target is greater (or less) than 1540 m/s, the echo will arrive back at the transducer earlier (or later) than expected for a target of that range Fat c = 1420 m/s Tissue c = 1540 m/s Target Displayed at Hull and East Yorkshire Hospitals NHS Trust

Artefacts Refraction Refraction of the ultrasound beam as it passes between tissues with a different speed of sound can result in objects appearing at an incorrect position in the image Target Displayed at Medium 1 c1 Medium 2 c2 > c1 Hull and East Yorkshire Hospitals NHS Trust

Artefacts Attenuation Artefacts During imaging the outgoing pulse and returning echoes are attenuated as they propagate through tissue, so that echoes from deeper targets are weaker than those from similar superficial targets Time Gain Compensation (TGC) is applied to correct for such changes in echo amplitude with target depth Most systems apply a constant rate of compensation designed to correct for attenuation in typical uniform tissue The operator can also make additional adjustments to compensate via slide controls that adjust the gain applied specific depths in the image TGC artefacts may appear in the image when the applied compensation does not match that actual attenuation rate in the target tissue Hull and East Yorkshire Hospitals NHS Trust

Artefacts Acoustic Enhancement Occurs when ultrasound passes through a tissue with low attenuation Echoes from deeper lying tissues are enhanced due to the relatively low attenuation in the overlying tissue This occurs because the TGC is set to compensate for the greater attenuation in the adjacent tissues Image of renal cyst Low attenuation Hull and East Yorkshire Hospitals NHS Trust

Artefacts Acoustic Shadowing Occurs when ultrasound wave encounter a very echo dense (highly attenuating) structure Nearly all of the sound is reflected, resulting in an acoustic shadow This occurs because the TGC is set to compensate for the lower attenuation in the adjacent tissues Image of Gallstone High attenuation Hull and East Yorkshire Hospitals NHS Trust

Artefacts Reverberation Artefact Reverberation artefacts arise due to reflections of pulses and echoes by strongly reflecting interfaces Occur most commonly where there is a strongly reflecting interface parallel to the transducer face Involves multiple reflections - Initial echo returns to reflecting interface as if it is a weak transmission pulse and returns a second echo (reverberation) Transducer Interface Reverberation Hull and East Yorkshire Hospitals NHS Trust

Attenuation Artefacts Hull and East Yorkshire Hospitals NHS Trust

Attenuation Artefacts Hull and East Yorkshire Hospitals NHS Trust

Contrast Agents Hull and East Yorkshire Hospitals NHS Trust

Contrast Agents Ultrasound Contrast Agents Ultrasound contrast agents are gas-filled micro-bubbles which are injected into the blood stream Micro-bubbles will give increased backscatter signal due to the large acoustic impedance mismatch between the gas-filled bubble and surrounding tissue Hull and East Yorkshire Hospitals NHS Trust

Contrast Agents Ultrasound Contrast Agents Micro-bubble suspension is injected intravenously into the systemic circulation in a small bolus The micro-bubbles will remain in the systemic circulation for a certain period of time Ultrasound waves are directed on the area of interest and when the micro-bubbles in the blood flow past the imaging window they give rise to increased signal Allows detection of blood flow where it would otherwise not be seen Hull and East Yorkshire Hospitals NHS Trust

Ultrasound Contrast Agents Name Capsule Gas Bubble Size LEVOVIST Palmitic acid Air 3-5 m SONOVIST Cyano-acrylate 2 m DEFINITY Lipid Perfluoropropane OPTISON Albumin Octafluoropropane 3.7 m SONOVUE Phospholipids SF6 2-3 m SONAZOID Surfactant Fluorocarbon 3.2 m ALBUNEX 4 m Hull and East Yorkshire Hospitals NHS Trust

Contrast Agents Targeted Contrast Agents Targeted contrast agents are under preclinical development They retain the same general features as untargeted micro-bubbles, but they are outfitted with ligands that bind to specific receptors expressed by cell types of interest Micro-bubbles theoretically travel through the circulatory system, eventually finding their respective targets and binding specifically If a sufficient number of micro-bubbles have bound to the target area, an increased signal will be seen Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound The Doppler Effect The Doppler effect is observed regularly in our daily lives, e.g. it can be heard as the changing pitch of an ambulance siren as it passes by The Doppler effect is the change in the observed frequency of the sound wave (fr) compared to the emitted frequency (ft) which occurs due to the relative motion between the observer and the source Consider three situations - Source and observer stationary - Source moving towards observer - Source moving away from observer Hull and East Yorkshire Hospitals NHS Trust

Source and observer stationary Doppler Ultrasound Source and observer stationary ) ) ) Observer ) ) fr = ft Source The observed sound has the same frequency as the emitted sound (Note: Frequency is the number of cycles per second) Hull and East Yorkshire Hospitals NHS Trust

Source moving towards observer Doppler Ultrasound Source moving towards observer ) ) ) ) ) fr > ft Causes the wavefronts travelling towards the observer to be more closely packed, so that the observer witnesses a higher frequency wave than emitted Hull and East Yorkshire Hospitals NHS Trust

Source moving towards observer Doppler Ultrasound Source moving towards observer ) ) ) ) ) fr < ft The wavefronts travelling towards the observer will be more spread out, so that the observer witnesses a lower frequency wave than emitted Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound The Doppler Effect The resulting change in the observed frequency from that transmitted is known as the Doppler shift The magnitude of the Doppler shift frequency is proportional to the relative velocity between the source and the observer It does not matter if it is the source or the observer is moving The Doppler effect enables Ultrasound to be used to assess blood flow by measuring the change in frequency of the ultrasound scattered from moving blood Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Ultrasound measurement of blood flow Transducer is held stationary and the blood moves with respect to the transducer The ultrasound waves transmitted by the transducer strike the moving blood, so the frequency of the ultrasound experienced by the blood is dependent on whether the blood is stationary, moving towards or away from the transducer The blood then scatters the ultrasound, some of which travels in the direction of the transducer and is detected The scattered ultrasound is Doppler frequency shifted again as a result of the motion of the blood, which now acts as a moving source Therefore, a Doppler shift has occurred twice between the ultrasound being transmitted and received back at the transducer Hull and East Yorkshire Hospitals NHS Trust

Ultrasound measurement of blood flow Doppler Ultrasound Ultrasound measurement of blood flow 2 ft v cos Doppler frequency shift, fd = fr – ft = c Target direction  fr = received frequency ft = transmitted frequency c = speed of sound v = velocity of blood  = angle between the path of the ultrasound beam and the direction of the blood flow (angle of insonation) Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Ultrasound measurement of blood flow The detected Doppler shift also depends on the cosine of the angle  between the path of the ultrasound beam and the direction of blood flow The operator can alter  by adjusting the orientation of the transducer on the skin surface Desirable to adjust  to obtain the highest Doppler frequency shift cos 1  90 Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Ultrasound measurement of blood flow If the angle of insonation of the ultrasound beam is known it is possible to use the Doppler shift frequency to estimate the velocity of the blood using the Doppler equation In diseased arteries the lumen will narrow and the blood velocity will increase c fd v = 2 ft cos Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Ultrasound measurement of blood flow The flow (Q) remains constant Q = A1 V1 = A2 V2 A = Area V = Velocity Narrowing in artery V1 A1 A2 V2 Hull and East Yorkshire Hospitals NHS Trust

Break Hull and East Yorkshire Hospitals NHS Trust

Session 4 Overview Session Aims: Continuous Wave Doppler US Pulsed Wave Doppler US Harmonic Imaging Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Continuous Wave and Pulsed Wave Doppler Doppler systems can be either continuous wave or pulsed wave Continuous wave (CW) systems transmit ultrasound continuously Pulsed wave (PW) systems transmit short pulses of ultrasound The main advantage of PW Doppler is that Doppler signals can be acquired from a known depth The main disadvantage of PW Doppler is that there is an upper limit to the Doppler frequency shift which can be detected Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Continuous Wave (CW) Doppler In a CW Doppler system there must be separate transmission and reception of ultrasound – transducer with two separate elements The region from which Doppler signals are obtained is determined by the overlap of the transmit and receive ultrasound beams Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Pulsed Wave (PW) Doppler In a PW Doppler system it is possible to use the same transducer element for both transmit and receive The region from which Doppler signals are obtained is determined by the depth of the gate and the length of the gate, which can both be controlled by the operator Transducer Gate depth Gate length Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Ultrasound signal received by transducer The received ultrasound signal consists of the following four types of signal: echoes from stationary tissue echoes from moving tissue echoes from stationary blood echoes from moving blood The task for the Doppler system is to isolate and display the Doppler signals from blood, and remove those from stationary and moving tissue Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Ultrasound signal received by transducer Doppler signals from blood tend to be low amplitude (small reflected echo) and high frequency shift (high velocity) Doppler signals from tissue are high amplitude (large reflected echo) and low frequency shift (low velocity) These differences provide the means by which signals from true blood flow may be separated from those produced by surrounding tissue Amplitude Tissue Blood Frequency Hull and East Yorkshire Hospitals NHS Trust

Doppler signal processing Doppler Ultrasound Doppler signal processing Demodulation Separation of the Doppler frequencies from the underlying transmitted signal Transducer Demodulator High-pass filtering Removal of the tissue signal Signal processor High-pass filter Frequency estimation Calculation of Doppler frequency and amplitudes Frequency estimator Display Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Demodulation The Doppler frequencies produced by moving blood are a tiny fraction of the transmitted ultrasound frequency E.g. If transmitted frequency is 4 MHz, a motion of 1 m/s will produce a Doppler shift of 5.2 kHz, which is less the 0.1% of the transmitted frequency The extraction of the Doppler frequency information from the ultrasound signal received from tissue and blood is called demodulation In PW Doppler, need the PRF to be at least twice the maximum Doppler shift frequency in order to avoid ‘aliasing’ (not a problem in CW) Aliasing is an artefact introduced by under-sampling in which high frequency components take the alias of a low frequency component Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound High-pass Filtering Amplitude Amplitude Tissue Blood Blood Frequency Frequency Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Frequency Estimation A spectrum analyser calculates the amplitude of all the frequencies present within the Doppler signal In the spectral display the brightness is related to the amplitude of the Doppler signal component at that particular frequency Frequency shift Time Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Aliasing Original signal Aliased signal Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Aliasing Aliased signal Increased PRF Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Duplex Imaging Duplex imaging combines Doppler information with a real-time B-mode image This produces a 2D representation of the direction and velocity of the blood flow on a grey-scale image In a typical display blood flowing towards the transducer is coded as red and blood flow away from the transducer is coded blue Red – towards transducer Blue – away from transducer Green - Variance Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Doppler Displays – Spectral Doppler Available in CW and PW Doppler Detailed analysis of distribution of flow Examine change in flow with time Frequency shift or Velocity Time Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Doppler Displays – Colour Doppler Available in PW Doppler only Superimposed Doppler information on underlying B-mode image Overall view of flow in region The sign (direction), mean Doppler shift (mean velocity) and variance (turbulence) of Doppler spectrum are usually colour-coded and displayed Red – towards transducer Blue – away from transducer Green - Variance Hull and East Yorkshire Hospitals NHS Trust

Doppler of Common Carotid Artery Doppler Ultrasound Doppler of Common Carotid Artery Hull and East Yorkshire Hospitals NHS Trust

Blockage in Carotid Artery Doppler Ultrasound Blockage in Carotid Artery Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Renal Colour Doppler Hull and East Yorkshire Hospitals NHS Trust

Doppler Ultrasound Doppler Displays – Power Doppler In a Colour Doppler image the magnitude of the frequency shift colour encodes the pixel value and assigns a colour depending on blood flow direction This Doppler signal processing places a restriction on the motion sensitivity since the signals received must be extracted to determine the velocity (magnitude of Doppler shift) and direction (phase shift) Power Doppler encodes the strength of the Doppler shifts (amplitude, intensity, power) with colours and ignores directional (phase) information In Power Doppler the magnitude of the Doppler signal is displayed rather than the Doppler frequency shift (e.g. the density of the red blood cells is depicted rather than their velocity) Power Doppler therefore exhibits increased sensitivity to slow flow rates at the expense of directional and quantitative flow information Hull and East Yorkshire Hospitals NHS Trust

Suspicious dark lesion Doppler Ultrasound Suspicious dark lesion Hull and East Yorkshire Hospitals NHS Trust

Power Doppler – Circle of Willis Doppler Ultrasound Power Doppler – Circle of Willis Suspicious dark lesion Hull and East Yorkshire Hospitals NHS Trust

Suspicious dark lesion Doppler Ultrasound Prostate Cancer Suspicious dark lesion Hull and East Yorkshire Hospitals NHS Trust

Harmonic Imaging Hull and East Yorkshire Hospitals NHS Trust

Harmonic Imaging Introduction Harmonic imaging of tissue is useful in suppressing weak echoes caused by artefact (acoustic noise) which cloud the image and make it difficult to identify anatomical features. These echoes (often called clutter) are particularly noticeable in fluid filled areas (e.g. heart or cyst) and are a common problem when imaging large patients. Harmonic imaging is possible by utilising harmonic frequencies of the ultrasound pulse. Hull and East Yorkshire Hospitals NHS Trust

Harmonic Imaging The Ultrasound Pulse Ideally the US pulse would rise and fall very quickly and contain only one frequency (or wavelength) In reality the US pulse contains a finite range of frequencies (or wavelengths) Actual Ideal Frequency Frequency Hull and East Yorkshire Hospitals NHS Trust

Harmonic Imaging Spatial Pulse Length (SPL) The spatial pulse length (in mm) is defined as  n Where  is the wavelength and n is the number of cycles A wider range of wavelengths and more cycles produces a longer SPL Use a higher frequency and shorter US pulse to give smaller SPL and improved resolution For Diagnostic US typical SPL values range from 0.3 to 1.0 mm SPL Hull and East Yorkshire Hospitals NHS Trust

Harmonic Imaging Bandwidth Amp Bandwidth describes the spread of US frequencies the transducer can transmit/receive The frequency content may be specified in terms of the Q factor: Q factor = F0 / (F2 – F1) An increased bandwidth and a decreased SPL reduces the Q Factor High Q = pure ultrasound pulse Freq F0 is the centre frequency and the lower and upper frequencies (F2 and F1) are at half the peak amplitude (reduction of 3 dB) Hull and East Yorkshire Hospitals NHS Trust

Harmonic Imaging Side Lobes Ideally US pulse energy appears as a single front travelling in forward direction Some energy travels off in different directions called side lobes The energy transmitted in side lobes can reduce image quality (contrast and resolution) side lobes Main lobe Transducer Hull and East Yorkshire Hospitals NHS Trust

Harmonic Imaging Non-linear Propagation In the description of propagation of sound waves given in the first session the wave propagated with a fixed speed determined by the properties of the medium This is a good approximation to reality when the amplitude of the wave is small, but at higher pressure amplitudes the effects on non-linear propagation become noticeable The speed at which each part of the wave travels is related to the properties of the medium and to the local particle velocity, which enhances or reduces the speed In the high pressure (compression) parts of the wave this results in a slight increase in speed In the low pressure (rarefaction) parts of the wave this results in a slight decrease in speed Hull and East Yorkshire Hospitals NHS Trust

Fundamental Frequency Harmonic Imaging Non-linear Propagation As the wave propagates into the medium the compression parts catch up with the rarefaction parts The compression parts become taller and narrower in amplitude, while the rarefaction become lower in amplitude The rapid changes in pressure in the compression parts of the wave appear in the pulse spectrum as high frequency components, these are multiples of the fundamental frequency F0 known as Harmonics Fundamental Frequency F0 Second Harmonic 2F0 Third Harmonic 3F0 Frequency Hull and East Yorkshire Hospitals NHS Trust

Harmonic Imaging Harmonic Imaging In harmonic imaging an US pulse is transmitted with fundamental frequency F0 but due to non-linear propagation the returning echoes also contain energy at harmonic frequencies 2 F0 and 3 F0. The imaging system ignores the fundamental frequency component of the echo and forms an image using only the 2nd harmonic component. The effective ultrasound beam (the harmonic beam) is narrower than the conventional beam because non-linear propagation occurs most strongly in the highest amplitude parts of the transmitted beam (i.e. near beam axis). Weaker parts of the beam such as side lobes and edges of the main lobe produce little harmonic energy and are suppressed in relation to the central part of the beam. Harmonic imaging also reduces acoustic noise from weak echoes and reverberations. Hull and East Yorkshire Hospitals NHS Trust

Harmonic Imaging Transmit Pulse Received Echo Filtered Echo F0 F0 2F0 Transducer Bandwidth Transmit Pulse F0 F0 Frequency 2F0 2F0 Hull and East Yorkshire Hospitals NHS Trust

Harmonic Imaging Harmonic Imaging Harmonic imaging can only be performed with a wide bandwidth transducer which can respond to both the fundamental frequency and its 2nd harmonic. The received echoes are passed through a filter which removes frequencies around F0 and only allows through those near to 2F0. As ‘acoustic noise’ echoes are mainly at the fundamental frequency, they are suppressed giving a clearer image. To achieve good separation of the received 2nd harmonic frequencies from the fundamental frequencies, the frequency spectra of the pulse and received echoes must be made narrower than in normal imaging. Reduction of the frequency range results in an increase in pulse length and a reduction in axial resolution. Hull and East Yorkshire Hospitals NHS Trust

The End Hull and East Yorkshire Hospitals NHS Trust