KSVAS LTD1 WORLD CONGRESS OF STERILIZATION COMBINED MEETING WFHSS / AIOS Small Steam Sterilizers BS EN 1360:2004.

Slides:



Advertisements
Similar presentations
ACTIVE LEARNING PROCESS
Advertisements

Preventing Equipment Failures
VALIDATION – ARE WE DOING ENOUGH David McNamara Manager – CSSD Holy Spirit Northside Private Hospital SRACA (NSW) Conference 10 th November 2011.
1 Ann Van Lysebetten CO 2 cooling experience in the LHCb Vertex Locator Vertex 2007 Lake Placid 24/09/2007.
Manual Set-Up for Density by Hydrometer Method ASTM D1298, D287, D6822 and IP160 Feb
Chapter 13 States of Matter 13.2 The Nature of Liquids
“Getting it right, every step of the way”
1 Calibration of Thermometry Instruments Nihal Gunasekara Sri Lanka Bangladesh BEST Programme.
David Robinson Washer Disinfectors – specifications, uses, maintenance and testing.
Thermal Power Station Plant. Introduction 150 MW Thermal power station plant, produce 450t/hr steam at full load The max steam pressure is 150 bar with.
CONDOMS A series of presentations developed by TGPSH/GTZ and PSI-TANZANIA.
Easi-Sterilise Standard Operating Procedures Section 3 Sterilizing.
The Mondial Forni Mondial and Ecomondial steam tube fixed deck ovens, gas or gasoil heated, are composed of independent decks with standard working height.
Student Book © 2004 Propane Education & Research CouncilPage Performing Pressure Tests on Gas Distribution Lines Gas personnel must understand.
1 VACUUM ATMOSPHERES Co NEW PRODUCT ANNOUNCEMENT November 28, 2001.
Unit 206: Domestic hot water systems
The Nature of Liquids. A Model for Liquids According to the kinetic theory, both the particles that make up gases and liquids have motion. While particles.
Copyright © Pearson Education, Inc., or its affiliates. All Rights Reserved. A Model for Liquids What factors determine the physical properties of a liquid?
STEAM CONDENSERS.
Domina PLUS B WORKING DIAGRAMS TROUBLESHOOTING INTERNAL VIEWS WIRING DIAGRAMS INSTALLATION 03 PREVIOUS MODEL.
Plan 2 – Dead-end Seal Chamber Optional Jacket
Clean and Pure steam in the Biopharm industry
HVAC523 Basic Subsystems.
“We Bring Engineering to Life” U.S. Tobacco GAP – Barn Testing Procedure.
NFPA 31 Standard for the installation of Oil- Burning Equipment
Chapter 13: States of Matter
23 Jan 2007 LASA Cryogenics Global Group 1 ILC Cryomodule piping L. Tavian for the cryogenics global group.
© 2009 Michigan State University licensed under CC-BY-SA, original at Water Quality.
ESS Cryogenic Distribution System for the Elliptical Linac CM - CDS requirements Preliminary Design Review Meeting, 20 May 2015, ESS, Lund, Sweden J. Polinski.
Introduction To the pr EN ISO Project norm
SIZING PNEUMATIC SYSTEMS. Introduction Pneumatic systems are sized to meet output power requirements. The air distribution system is sized to carry the.
Module 1, Part 3: Process validation Slide 1 of 22 © WHO – EDM – 12/2001 Validation Part 3: Process validation Supplementary Training Modules on Good Manufacturing.
Ch The Nature of Liquids. A Model for Liquids In kinetic theory, there are no attractions between particles of a gas. Particles of a liquid are.
Boiler Mounting Made by : Group 2 Electrical Eng-B.
Chapter 13: States of Matter
Prepared by : Guided by : Mr. Tanmay Patel.
Hydrogen system R&D. R&D programme – general points Hydrogen absorber system incorporates 2 novel aspects Hydrogen storage using a hydride bed Hydrogen.
Nuclear Thermal Hydraulic System Experiment
Kinetic Theory and a Model for Gases The word kinetic refers to motion. The energy an object has because of its motion is called kinetic energy. According.
Firefighter III Module B Water Supply (4-19) (4-19)
What tests are necessary to make sure your kegs are being cleaned, sanitized and filled correctly By Andrew J. Brewer
Dr.ir. Josephus P.C.M. van Doornmalen Gomez Hoyos 3M Deutschland GmbH (Germany) Eindhoven University of Technology (the Netherlands) Measurements of only.
Manual Density Business Today
WFHSS 2015 Lille/France Klaus Roth SMP GmbH Prüfen Validieren Forschen Hechingerstrasse Tuebingen Germany Validation of hydrogene peroxide Sterilizers.
Chemical Sterilization
Esbe Rotary Motorized valves The ESBE Rotary Motorized valves (mixing valves and actuators) are used to control heating and cooling in buildings. We offer.
LEADER® MULTICHAMBER MEDICAL TUNNEL WASHER LEADER is a professional manufacturer Backed up by 30 years experience in the Asia area of laboratory, pharmaceutical.
Cryogenic scheme, pipes and valves dimensions U.Wagner CERN TE-CRG.
Wet Packs: Strategies for Prevention and Resolution
Heat. Nature of Heat Heat is the transfer of energy (every in transit) from one body to another due to the temperature difference between the two bodies.
Boiler Safety Valve Regulations
Development of Cryo-Module Test Stand (CMTS) for Fermi Lab (R.L.Suthar, Head,CDM, BARC) Cryo-Module Test stand (CMTS) is a very sophisticated equipment.
EQUIPMENT VALIDATION - Autoclave Dr. L. Prabakaran Prof. & Head R. R. College of Pharmacy Bangalore -90.
ME444 ENGINEERING PIPING SYSTEM DESIGN
Laboratory equipment Lecture (5).
14 Refrigerant Recovery, Recycling, and Recharging.
Behavior of gases and Gas Laws
Hot Water Systems Cylinders and Systems.
LEADER ULTRASONIC CO., LTD
Cisterns and Insulation
TASK 3: DIGITAL THERMOMETER
 Define: The stirring heating plate or heated stirring heating plate has been developed to heat and mix fluids contained in laboratory receptacles such.
Laboratory equipment Lecture (9).
Condenser in Power Plants
Boiler Safety Valve Regulations
SCITEQ End closure series
Pneumatic System Conditioning
Presentation transcript:

KSVAS LTD1 WORLD CONGRESS OF STERILIZATION COMBINED MEETING WFHSS / AIOS Small Steam Sterilizers BS EN 1360:2004

KSVAS LTD2 L Kingston. Registered Authorised Person {Sterilizers} Member of :-  BSI - CH198 – Sterilization  CEN - TC 102 wg 2/3 – Steam Sterilizers  ISO - TC 198 wg 3 – Moist Heat Sterilization.  EPL/66 Safety of Measuring, Control and Laboratory Equipment.  BSI LBI 35- 1

KSVAS LTD3 SCOPE OF THE PRESENTATION FOR THE CHANGES WITHIN BS EN 1360  To provide a brief introduction to the update requirements and changes.  To introduce the current standard focusing on what is the requirement of small steam sterilizer: 1)Definition Small steam sterilizer:-Size one which is unable to accommodate sterilization module and has a chamber volume less than 60 litres Note One Sterilization module =300mmx 300mm x600mm

KSVAS LTD4 SCOPE OF THE PRESENTATION FOR THE CHANGES WITHIN BS EN )Sterilization cycle type:- classification of the process based on performance 3)Solid:- product not made from porous materials no recesses which present a greater or equal challenge than a hollow load (B) 4)Hollow load A :-single ended open space ratio of length to diameter of cavity greater than 1 less or equal 750 ( 1 ≤ L/D ≤ 750) or ( L ≤ 1 500mm) or ( 2 ≤ L/D ≤ ) ( L ≤ 3000mm ) which is not a hollow load ( B)

KSVAS LTD5 REVIEW OF THE CHANGES WITHIN BS EN )Hollow load B :- single ended open space ratio of length to diameter of cavity greater than 1 less or equal to 5 (1 ≤ L/D ≤ 5) Where diameter greater than or equal to 5mm(D ≥ 5mm or double ended where ratio length / diameter greater than equal to 2 less than10 (2 ≤ L/D ≤ 10) or (D ≥ 5mm) 6)Temperature measurement reference position:- identified by manufacture condition in useable space of chamber

KSVAS LTD6 Type All wrapped or non wrapped products Type All wrapped or non wrapped products Type All non wrapped Solid products Type B All wrapped or non wrapped products Type N All non wrapped Solid products Type S All wrapped or non wrapped products BS EN 1360– Structure B N S

KSVAS LTD7 Type All wrapped or non wrapped products Type All wrapped or non wrapped products Type All non wrapped Solid products Type S1 Hollow Load A&B Dryness Solid Load Type S2 Dryness, Porous Load & Solid products Type S3 Specific Medical Devices BS EN 1360– Structure S2 S3 S1

KSVAS LTD8 Operational Services Points to be aware of :  Water Supply Steam Generation.  Designed to function with water free of contaminants  Cause no harm to sterilizer or the sterilizer load

KSVAS LTD9 Operational Services Points to be aware of :  Water Used other than for Steam Generation.  Potable quality  Range specified manufacturer, including 15°C  Cooling Purposes  Vacuum systems  Total hardness value 0,7 mmol/l 2,0 mmol/l

KSVAS LTD10 Operational Services Points to be aware of :  Water reservoir if fitted  Pipe work shall be fitted with valve or other device / automatic to allow draining by operator reservoir and pipework  M anufacturers design shall facilitate cleaning filling and inspection  Indicate sufficient water for operating cycle or shall not be capable of starting cycle  Designed to prevent back siphoning to chamber

KSVAS LTD11 BS EN 1360 PROCESS PERFORMANCE DATA Values Type testing 1 of 2 Sheets Type Tests Sterilization Process Cycle Types BNS1S2S3 Dynamic chamber pressure YES Air leakage YES Empty chamber YES Solid load YES Small Porous items YES Small porous loads YES Full porous load YES

KSVAS LTD12 BS EN 1360 PROCESS PERFORMANCE DATA Values Type testing 2 of 2 Sheets Type Tests Sterilization Process Cycle Types BNS1S2S3 Hollow load B YES Hollow load A YES Multiple wrapping YES Dryness solid load YES Dryness porous load YES Residual air YES Specific medical devices YES

KSVAS LTD13 BS EN 1360 PROCESS DATA limiting values to be recorded StartTime for vacuum pulse Time for pressure pulse Sterilization start time Holding time Sterilization end time Drying start time Drying end time YES 0kPa to 400kPa 50°C to 150°c Sample each channel At least every 2.5sec Sample each channel At least every 2.5sec Accuracy Better than ±5kPa Accuracy Better than ± 1% Sterilizers which have no vacuum phase Pressure Scale 100kPa to 400kPa

KSVAS LTD14 BS 1360 Sterilizers Process Evaluation System If fitted 1.Compare with validated cycle any change pressure or temperature time period that change occurs 2.Change beyond programme limits Shall cause an indicate Fault

KSVAS LTD15 BS 1360 Sterilizers Process Evaluation System If fitted 3.Compare two independent temperature sensors sterilizer temperature instrument chamber and recorder ; or 4.Capable comparing theoretical steam temperature with chamber during the holding period Shall cause an indicate Fault

KSVAS LTD16 BS 1360 Sterilizers Process System Time – Temperature relationships:- STERILIZATION Sterilization Temperature Minimum Holding Time 121°C15 Min 126°C10 Min 134°C3 Min 143°C1 Min °C Minutes

KSVAS LTD17 BS 1360 Sterilizers Process System Time – Temperature relationships:- Equilibration TemperatureEquilibration Time 121° 126°134° 143°15 seconds Not exceeding 30 seconds is acceptable If ? >1K/min < 8K/min °C Seconds 30sec

KSVAS LTD18 BS 1360 Sterilizers Process System Time – Temperature relationships:- Equilibration TemperatureEquilibration Time 121° 126°134° 143°30 seconds Theoretical steam Temperature 10K heating stage Less than8K//min But greater than 1K/min During last 10K of heating stage chamber and load do not differ by more than 2K °C Seconds

KSVAS LTD19 BS 1360 Sterilizers Process Cycle HIGHEST & LOWEST VALUES PRESSURE TEMPERATURE DATA AND LIMITING VALUES TO BE RECORDED AMBIENT PRESSURE -1BAR START TIMEEND TIME TOP OF PULSE TIME HOLDING TIME START DRYING END TIME TIME SECONDS TEMPERATURETEMPERATURE 134°C

KSVAS LTD20 P (bar) Fractionated vacuum Plateau Drying 15 min min TYPE B cycle 20 min ! ECO-B

KSVAS LTD21 BS 1360 Sterilizers Testing System Air Leakage Test 300sec600sec Ambient atmospheric pressure P1 P2 P3 f0 f1f2 f3 △p p 3 - p 2 △p△p 10 = Time in seconds Requirement Shall not cause rate pressure rise to exceed 0.13kPa/min{1.3mbar/min} when tested

KSVAS LTD22 BS STERILIZER Leak Test Reports FAIL PASS

KSVAS LTD23 BS 1360 Sterilizers Performance ATTAINMENT STERILIZATION CONDITIONS Requirement:- All temperatures measured in useable space and load Are not more than 4K above sterilization temperature selected Do not differ from each other by more than 2K are not lower than the sterilizer selected temperature   

KSVAS LTD24 BS 1360 Sterilizers Performance TYPE TEST PROCEDURE Test equipment requirements One external pressure sensor Sensors connect to chamber Type tests shall be repeated Twice (three tests in all) Thermometric measurement Eight sensors shall be used   

KSVAS LTD25 BS 1360 Sterilizers Testing Process Cycle Key 1 Capsule 2 Sealing 3 Indicator system 4 Connector 5 Open end 6 Tube Process Challenge Device & Chemical Indicator system for Hollow Load A

KSVAS LTD26 BS 1360 Sterilizers Steam Testing Steam pressure kPa (bar) Bore mm ± 0,02up to 400 (3)0,8up to 500 (4)0,6up to 800 (7)0,4 Dimensions in millimetres Bulkhead fitting

KSVAS LTD27 BS EN 1360 Porous load Non- Condensable Gases Apparatus Measurement Non- condensable gases Exact position of the capillary tube inlet to sample from the sterilizer chamber has been found to vary in position for collection of sample For measurement of non – condensable gases

KSVAS LTD28 BS 1360 Sterilizers Steam Testing Steam pressure kPa (bar) Bore mm ± 0,02up to 400 (3)0,8up to 500 (4)0,6up to 800 (7)0,4 Steam Service supply main steam

KSVAS LTD29 BS 1360 Sterilizers Steam Testing Apparatus for sampling Steam Condensate for chemical analysis Steam Service supply main steam

KSVAS LTD30 BS 1360 Sterilizers Definition hollow space A Cavity with length to diameter greater than 1 and greater than 5 for all objects Hollow A objects Objects which are not hollow Within standard cavity to diameter less than 1 Not hollow Less than 1 1 Greater than 5

KSVAS LTD31 BS 1360 Sterilizers Definition hollow space B Ratio of the length of cavity to diameter greater than 1 and less than than 5 for both objects Hollow B objects Example 8mm x 15mm

KSVAS LTD32 BS EN 1360 Porous load x Small porous load test single wrapped Volume of more than 54 litres Diameter of at least 35 cm Small load textiles cotton sheets 450mm x 300mm Thermal Sensors x positions Reduced test pack WORKS / INSTALLATION TEST x x x x

KSVAS LTD33 BS EN 1360 Porous load x x x x Small porous load test single wrapped Volume of more than 54 litres Diameter of at least 35 cm Small load textiles cotton sheets 450mm x 300mm Thermal Sensors x positions Reduced test pack TYPE TEST

KSVAS LTD34 SMALL STERILIZERS Moist Heat Sterilization  The future may be a combined standard applicable to all moist heat sterilizers  EN  EN 285  In accordance with the Vienna agreement

KSVAS LTD35 The Vienna Agreement Harmonisation of standards CEN & ISO  An agreement made between CEN and ISO to ensure harmonisation of standards published by the two organisations.

KSVAS LTD36 BS 1360 Small Steam Sterilizers Thank you Question Time JUNE 2008