Flame Ionization Detector Most common detector Carbon atoms (C-C bonds) are burned in a hydrogen flame. A small portion of carbon atoms are ionized (about.

Slides:



Advertisements
Similar presentations
OFFLINE COMPOSITION MEASURING SENSORS
Advertisements

Fluorescent Lamps.
D e t e c t o r s f o r H P L C.
Gas Chromatography Vaporization of sample Gas-solid
Lecture 8b Gas Chromatography.
GAS CHROMATOGRAPHY ENVE 202 Dr. Aslıhan Kerç.
Lab Methods Day June 25, 2014 Gas Chromatography
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
Gas Chromatography in the detection of Volatile Organic Compounds.
AAS and FES (Ch 10, 7th e, WMDS)
Light and Electronic Transitions. The Big Questions What is light? How is light emitted? What do electrons have to do with light? What are emission spectra?
Emission Spectra and Flame Tests. The Big Questions What is light? How is light emitted? What do electrons have to do with light? What are emission spectra?
Instrumental Analysis
Molecular Fluorescence Spectroscopy
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Experiment 6 Simple and Fractional Distillation.
1 Gas Chromatography Lecture a. Thermal Conductivity Detector (TCD) This is a nondestructive detector which is used for the separation and collection.
Atomic Emission - AES M* → M + hn Thermal excitation M → M*
1 Chapter 24 GC Gas Chromatography. 2 GC Mechanism of separation is primarily volatility. Difference in boiling point, vapor pressure etc. What controls.
ATOMIC ABSORPTION SPECTROSCOPY
BC ILN Atomic Absorption Spectroscopy (AAS) 1 Thompson Rivers University.
Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh
GC Detectors Ideal Detector Characteristics: 1. Universal Response
Biochemical instrumental analysis-2
Atomic Emission Spectroscopy
Atomic Emission Spectrometry
Atomic Absorption Spectrophotometer
가스크로마토그래피 기기의 구성과 기능.
What is plasma? Plasma is an ionized gas comprising molecules,atoms, ions ( in their ground or excited states) electrons and photons. It is electrically.
 Matter is made of atoms and molecules (your book refers to these as particles)
474 PHG 5.
Chapter 27 Gas Chromatography 1. Principles
Atomic Emission Spectrometry By: Alexa Kunch
Chapter 33 Supercritical-Fluid Chromatography
Chemistry of Life A Brief … Overview. Matter Matter occupies space and has weight. It can exist as a solid, liquid, or gas. It may be possible to break.
GAS LIQUID CHROMATOGRAPHY
1 Components of Optical Instruments Lecture Silicon Diode Transducers A semiconductor material like silicon can be doped by an element of group.
1 Atomic Emission Spectroscopy Lecture Advantages of Plasma Sources 1.No oxide formation as a result of two factors including Very high temperature.
1 Atomic Absorption Spectroscopy Lecture Performance Characteristics of Electrothermal Atomizers Electrothermal atomization is the technique of.
Atomic spectroscopy Elemental composition Atoms have a number of excited energy levels accessible by visible-UV optical methods ä Must have atoms (break.
Adv. Inst. Techs.  flame emission (eg flame photometer) known as low temperature emission ( K) ◦ first form of spectroscopy ◦ used in commercial.
Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.
PAS-Intro1 Dated 24th Aug Copyright © Siemens AG 1999 All Rights Reserved Siemens Advanced Engineering Pte Ltd Process Analytical Services Division.
Name : Dr. N Chaamwe 1.
Analytical Separations
Atomic Fluorescence Spectroscopy. Background l First significant research by Wineforder and Vickers in 1964 as an analytical technique l Used for element.
Basic Gas Chromatography. History Separation of dyes by Runge Separation of plant pigments by Tswett Theoretical gc (Martin & Synge)
GAS LIQUID CHROMATOGRAPHY Principles Partition of molecules between gas (mobile phase) and liquid (stationary phase).
Unit 4 Chemistry of Fire. This will be as basic as it gets. Do you remember the Periodic Table?
Ideal Detector Fast Cheap Rugged Responds to all wavelengths of light Can distinguish different wavelengths Sensitive Low LOD.
Instrumental Analysis
Gas Chromatography. In Gas Chromatography (GC), a gaseous mobile phase transports a gaseous solute through a long, thin column containing solid or liquid.
Unit 4 Chemistry of Fire. This will be as basic as it gets. Do you remember the Periodic Table?
K V GOPINATH M Pharm PhD,CPhT Tirumala Tirupati Devasthanams TIRUPATI GAS CHROMATOGRAPHY.
Thermal Conductivity Detector
Gas Chromatography.
M.PRASAD NAIDU Msc Medical Biochemistry, Ph.D Research scholar.
Experiments in Analytical Chemistry
Gas Chromatography Chap 27 Types: Gas-solid chromatography (GSC)
Reading assignment: section 26E(p781) Chapter 26 # 2, 3, 14, 15, 16 Chapter 27 # 7(a,d,f), 22, 23, 24, 25.
GAS CHROMATOGRAPH. GAS CHROMATOGRAPH Principle: Which type of compounds can analyse? Volatile Thermostable.
Light and Electronic Transitions
Mass Spectronomer.
Dnyanasadhana College, Thane.
Gas Chromatography.
Light and Electronic Transitions
NUCLEAR REACTOR MATERIALS
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
32A-4 Chromatographic Detectors
32A-4 Chromatographic Detectors
Presentation transcript:

Flame Ionization Detector Most common detector Carbon atoms (C-C bonds) are burned in a hydrogen flame. A small portion of carbon atoms are ionized (about 1 in 10,000), The ions carry a charge from the flame to the walls of the detector which surrounds the flame.The charge is electronically amplified and sent to a recording device.

Flame Ionization Detector very robust and reliable.

Nitrogen Phosphorus Detector The NPD uses a bead of a compound such as rubidium silicate above a jet of H 2. The bead is heated by an electric current forming a plasma. Nitrogen and phosphorous react in the plasma forming specific ions that carry a small current to the charged collector. The NPD is electronically similar to the FID, but since there is no flame, hydrocarbon ionization does not occur. The response to N is 10 3 – 10 5 greater than response to C The response to P is 10 4 – higher than response to C. The linear dynamic range is 10 4 It is only fairly reliable since the bead burns up over time causing drift in the signal.

Flame Photometric Detector The (FPD) is an element specific detector Commercial instruments are limited to the detection of P and S The analytes are burned in a H 2 flame causing electrons to move to an excited unstable state. When the electrons return to the ground state, they emit a specific wavelength of light 526 nm for P and 394 nm for S These wavelengths are monitored by a photomultiplier, amplified, and turned into an electrical signal. This detector is sensitive to g Linear dynamic range for P of only 10 3 – For S, the response is non-linear.

Flame Photometric Detector

Electron Capture Detector The ECD uses a radioactive source such as Ni 63 which produces Beta particles which react with the carrier gas producing free electrons. These electrons flow to the anode producing an electrical signal. When electrophillic molecules are present, they capture the free electrons, lowering the signal. The amount of lowering is proportional to the amount of analyte present. It is sensitive down to but the dynamic range is only about It’s an excellent detector for molecules containing an electronegative group such as Cl or F etc. (or derivitized molecules).

Electron Capture Detector

Atomic Emission Detector One of the newest gas chromatography detectors Quite expensive compared to other GC detectors The strength of the AED lies in the detector's ability to simultaneously determine elements. It uses microwave energy to excite helium molecules (carrier gas) which emit radiation which breaks down molecules to atoms such as S, N, P, Hg, As, etc. These excited molecules emit distinctive wavelengths which can be separated by a grating and sent to the detector (typically a photodiode array) which produces the electrical signal.

Atomic Emission Detector

Photoionization Detector The photoionization detector (PID) uses a UV lamp (xenon, krypton or argon) to ionize compounds. The ionization produces a current between the two electrodes in the detector. The detector is non-destructive and can be more sensitive than an FID for certain compounds- substituted aromatics and cyclic compounds

Photoionization Detector

Thermal Conductivity Detector Consists of an electrically-heated wire or thermistor. The temperature of the sensing element depends on the thermal conductivity of the gas flowing around it. Changes in thermal conductivity, such as when organic molecules displace some of the carrier gas, cause a temperature rise in the element which is sensed as a change in resistance. TCD’s are often used to measure lightweight gasses or water (compounds for which the FID does not respond). The TCD is not as sensitive as other detectors but it is a universal detector and is non-destructive. Modern detectors called micro-TCD’s have very small cell volumes, and new electronics that produce much higher sensitivities and wider linear ranges. Due to its increased sensitivity, and the fact that it is a universal non-destructive detector, it is again becoming more popular for certain applications.

Thermal Conductivity Detector

Representative Thermal Conductivity Values, 100 o C

GC Detectors Sensitivities and Ranges