DESIGNING FOR RADIATION PROTECTION
TUBE HOUSING REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING One meter is 3.3 feet Body parts should not rest on tube housing
Control panel should indicate Condition of exposure When x-ray tube is being energized kVp, mA or mAs Visible or audible signal of exposure
SID Tape measure or laser lights indicate the distance Must be accurate with 2% of the indicated SID
COLLIMATION PBL BEAM ALIGNMENT X-ray beam and light should be within 2% of SID PBL not required anymore Beam should line up with image receptor Proper alignment of beam to film (indicator light)
FILTRATION 2.5 kVp 1.5 mm between kVp .5 mm below 50 kVp (mammo) See question on page 569 (refer to chart 31-3 on page 461)
Reproducibility Linearity Operator shield
MOBILE RADIOGRAPHY Lead apron assigned to portable Exposure switch should allow operator to be 2 meter from tube (6+)feet
FLUOROSCOPY Source to skin distance – 38 cm Mobile SSD – 30 cm When intensifier is in parked position—no fluoro Intensifier serves as a primary protective barrier and must be 2 mm Pb equivalent. Filtration should be at least 2.5 mm Al equivalent—Tabletop, patient cradle or other material factored in for total filtration Collimation—unexposed border should be visible on TV monitor
FLUOROSCOPY Dead man type exposure switch Bucky opening covered automatically by.25 mm lead Protective curtain mm Pb equivalent Timer (audible) when fluoro time has exceeded 5 minutes
FLUOROSCOPY Intensity (R ) should not exceed 2.1 R per minute for each mA at 80 kVp DAP DOSE RESPONSE PRODUCT DOSE AND VOLUME OF TISSUE IRRADIATED DAP INCREASES WITH INCREASING FIELD SIZE
PROTECTIVE BARRIERS
DESIGN CRITERIA Location of x-ray table Where is the primary beam directed? Surrounding environment (controlled area vs. uncontrolled area) RF room Dedicated room Use factor # of exams in a room
Primary Protective Barrier Anywhere the primary beam is directed ( dedicated chest rooms) Lead bonded to sheet rock of wood paneling Concrete, concrete block, brick 4 inches of masonry = 1/16 inch of lead Image intensifier considered a primary protective barrier
SECONDARY BARRIERS Secondary radiation (scatter, leakage) Patient is source of scatter Barrier does not have to be leaded gypsum board 4 thicknesses of 5/8 th inch drywall glass ½ to 1 inch thickness lead acrylic Control booth Lead aprons (5mm of lead attenuates____%_at _____kVp
Factors that affect thickness of barrier Distance Occupancy-levels Control vs uncontrolled workload Use factor
USE FACTOR Amount of time x-ray beam is directed at wall/floor Wall given a use factor of ¼ Floor given a factor of 1 Secondary barrier use factor of 1 Dedicated chest room-use factor of 1
FINALLY Barriers are designed with kVp usage in mind so most barriers are thicker than needed Exposure to outside of room is calculated to result in a DL of 100mrem per week but do not factor in patient and image receptor interception. DL is actually 1/10 th of the recommended DL
Exposure switch Mounted of fixed to control panel No long cords
TLD, OSL