Introduction to MALDI-TOF MS

Slides:



Advertisements
Similar presentations
Mass Spectrometry Chem 3500 Mass Spectrometry Y. Zhao Lecture 2 A Glimpse of Mass Spectrometry Began from 1920s: Aston, Dempster, Thompson MS is often.
Advertisements

FC-MS from Teledyne Isco CombiFlash ® a Name You Can Rely On.
Mass Spectrometry Mass spectrometry involves ionizing a compound, separating the ions that result on the basis of mass to charge ratio (m/z) and obtaining.
Module 2: Analytical methods and separation techniques.
Mass spectroscopy. In a typical MS procedure:  1- a sample is loaded onto the MS instrument, and undergoes vaporization.  2- the components of the sample.
12-1 Molecular Mass Spectroscopy Molecular structure Composition of mixtures Molecular mass spectra Ion Source Mass Spectrometers Applications.
Lecture 9. MALDI/TOF. Introduction Matrix-assisted laser desorption/ionization (MALDI) is a soft ionization technique used in mass spectrometry, allowing.
Lecture Three: "MALDI-TOF MS, Up Close, and Personal Review of principles Delayed Extraction Reflector / reflectron Post-source Decay (PSD)
Primary Ionization Techniques for Molecules
Mass Spectroscopy Mass Spectrometry ä Most useful tool for molecular structure determination if you can get it into gas phase ä Molecular weight of.
Chapter 20 Molecular Mass Spectrometry. Introduction... Mass spectroscopy is perhaps one of the most widely applicable of all the analytical tools available.
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform.
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform.
Mass Spectrometry Separation principle: A particle with a certain mass (m) and a certain charge (z) behaves in a predictable manner when exposed to a electromagnetic.
Atomic Mass Spectrometry
PROTEIN IDENTIFICATION BY MASS SPECTROMETRY. OBJECTIVES To become familiar with matrix assisted laser desorption ionization-time of flight mass spectrometry.
Mass Spectrometry 101 An Introductory Lecture On Mass Spectrometry Fundamentals Presented to the Sandler Mass Spectrometry Users’ Group University of California.
MALDI-TOF-TOF with High Resolution Precursor Selection and Multiplexed MS-MS Poster Number ThP 618 Kevin Hayden, Stephen C. Gabeler, Mark Dahl and Marvin.
Development of a TOF Version of the Desktop MiniSIMS Design & Applications A.J. Eccles, B. Cliff, C. Jones, N. Long, P. Vohralik Millbrook Instruments.
Proteomics Informatics – Overview of Mass spectrometry (Week 2)
Mass Spectrometer Mass spectrometere is a very powerful method to analyse the structure of organic compounds.
TOF Mass Spectrometer &
Understanding mass spectroscopy. Mass spectroscopy is a very powerful analytical tool that can provide information on the molecular mass of a compound,
Gaucher’s Disease – A glycolipid storage disorder characterized by the accumulation of glucosylceramide in the spleen, liver, lungs, bone marrow and brain.
Molecular mass spectrometry Chapter 20 The study of “molecular ions” M + e -  M e -
Mass Spectrometry Mass spectrometry involves ionizing a compound, separating the ions that result on the basis of mass to charge ratio (m/z) and obtaining.
Mass Spectrometry I Basic Data Processing. Mass spectrometry A mass spectrometer measures molecular masses. The mass unit is called dalton, which is 1/12.
INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable.
Based on ionization of gas phase molecule followed by analysis of the masses of the ions produced. The Mass Spectrum: Graph of ion intensity versus mass-to-charge.
Chapter 20 # 4, 5, 12, 13, 16. Fig (p.551) Mass spectrum of ethyl benzene Fragment peaks Unit: amu or dalton Fragment peaks.
Mass Spectroscopy 1 Mass Spectroscopy (Mass Spec) Applying Atomic Structure Knowledge to Chemical Analysis.
Marvin Vestal and Kevin Hayden SimulTOF Systems
The Apparatus…. Ionic target studies…. Neutral target studies…. Queens University Belfast University College London.
MS Intro. MS requires gas-phase ions, why? MS uses magnetic and electric fields to control the path of a compound based on mass to charge ratio (m/z)
Chemistry Topic: Atomic theory Subtopic : Mass Spectrometer.
7-1 Atomic Mass Spectroscopy General Features Mass Spectrometers §Sources Atomic mass spectroscopy determines elements by mass §Mass to charge ratio §Lower.
Temple University MASS SPECTROMETRY FURTHER INVESTIGATIONS Ilyana Mushaeva and Amber Moscato Department of Electrical and Computer Engineering Temple University.
Temple University MASS SPECTROMETRY INTRODUCTION Ilyana Mushaeva and Amber Moscato Department of Electrical and Computer Engineering Temple University.
For all CHEM5161 students: The first day of class for CHEM5161 (Analytical Spectroscopy) will be on TUE Sept 4 (following Labor Day). There will be no.
INF380 - Proteomics-51 INF380 – Proteomics Chapter 5 – Fundamentals of Mass Spectrometry Mass spectrometry (MS) is used for measuring the mass-to-charge.
Chemistry 367L/392N Macromolecular Chemistry Lecture 8.
Mass Analyzers : Time-of-flight Mass Spectrometry CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Joel Kimmel Fall 2007.
LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY
Separates charged atoms or molecules according to their mass-to-charge ratio Mass Spectrometry Frequently.
Mass Spectrometry Relative atomic masses and the mass of individual isotopes can be determined using a mass spectrometer. The principle behind mass spectrometry.
2014 생화학 실험 (1) 6주차 실험조교 : 류 지 연 Yonsei Proteome Research Center 산학협동관 421호
What is Mass Spectrometry? Mass spectrometry could be considered as an analytical technique that involves the study in the gas phase of ionized molecules.
Chapter 11 Structure Determination: Mass Spectrometry, Infrared Spectroscopy, and Ultraviolet Spectroscopy.
Yonsei Proteome Research Center Peptide Mass Finger-Printing Part II. MALDI-TOF 2013 생화학 실험 (1) 6 주차 자료 임종선 조교 내선 6625.
An introduction to Mass spectrometry. What is mass spectrometry? Analytical tool measuring molecular weight (MW) of sample Only picomolar concentrations.
Peptide Mass Finger-Printing Part II. MALDI-TOF
Mass Spectrometry 101 (continued) Hackert - CH 370 / 387D
Dissociation of Molecular Ions Studied by
MALDI-TOF MS Bergdís Björk Bæringsdóttir Daníel Arnar Tómasson
Acknowledgements Slides and animations were made by Dr. Jon Karty Mass Spectrometry Facility Indiana University, Bloomington.
The Covalent Structure of Proteins
Mass Spectrometry 101 Hackert - CH 370 / 387D
Mass Spectrometry Mass spectrometry involves ionizing a compound, separating the ions that result on the basis of mass to charge ratio (m/z) and obtaining.
Time of Flight Analyzers
Mass Spectrometry Obaid M. Shaikh.
Starter Complete the quiz and hand in..
An Introduction to the ProteinChip® Reader
Mass Spectroscopy. Mass Spectroscopy Mass Spectrometry Most useful tool for molecular structure determination if you can get it into gas phase Molecular.
Time-of-Flight Mass Analyzers
Mass Spectrometry.
Sample Spectra 1/1/2019.
Figure 1 Schematic representation of a typical MALDI-MSI workflow
Mass Spectrometry THE MAIN USE OF MS IN ORG CHEM IS:
M.Prasad Naidu MSc Medical Biochemistry, Ph.D,.
Presentation transcript:

Introduction to MALDI-TOF MS Sandler Mass Spectrometry User’s Group University of California San Francisco May 20, 2003

Features of MALDI-TOF MS Soft ionization - analyze intact biomolecules and synthetic polymers Broad mass range - analyze a wide variety of biomolecules Simple mixtures are okay Relatively tolerant of buffers and salts Fast data acquisition Easy to use and maintain, no water or gas hook ups required High sensitivity, superior mass resolution and accuracy

MALDI: Matrix Assisted Laser Desorption Ionization Sample plate hn 1. Sample (A) is mixed with excess matrix (M) and dried on a MALDI plate. 2. Laser flash ionizes matrix molecules. 3. Sample molecules are ionized by proton transfer from matrix: MH+ + A  M + AH+. AH+ Variable Ground Grid Grid +20 kV

Time-of-flight mass analyzer Source Drift region (flight tube) + + detector + + V Ions are formed in pulses. Small ions reach the detector before large ones. Measures the time for ions to reach the detector.

Calibration of the mass scale The mass-to-charge ratio of an ion is proportional to the square of its drift time. t = Drift time L = Drift length m = Mass K = Kinetic energy of ion z = Number of charges on ion

Voyager-DE STR MALDI TOF Sample Linear Extraction plate Reflector detector grids Attenuator Timed ion detector selector Reflector Prism Laser Collision cell Camera Pumping Pumping

MALDI TOF Hardware Laser, Attenuator and Prism Nitrogen laser at 337 nm, 3 ns wide pulses, 20 Hz. Laser attenuator varies the intensity of the laser hitting the sample. Prism deflects the laser beam into the ion source. Sample Plate and Sample Stage An accelerating voltage is applied to the sample plate in the range 15-25 kV. Variable Voltage Grid A grid 1-2 mm above the sample plate with an additional voltage to fine- tune ion acceleration Ground Grid Grounded surface defines end of acceleration region Grounded Aperture Entrance to flight tube

MALDI TOF Hardware Vacuum System High vacuum is required to avoid ion collisions Flight tube A field free region where ions drift at a velocity inversely proportional to the square root of their mass/charge. Linear Detector Measures the ion abundance in linear mode (no reflector used) and sends a signal to the digitizer.

Ions are detected with a microchannel plate + e - primary ion L D 1000V 100V L >> D

Microchannel Plate (MCP)

High current detector schematic Used in linear mode to enhance signal from high mass molecules Fast scintillator Microchannel plate Condenser Photomultiplier tube signal 1kV 15kV

MALDI TOF Hardware Reflector A single stage gridded ion mirror that subjects the ions to a uniform repulsive electric field to reflect them. It is tilted by 1° in the DE-STR to focus the ions on to the detector Collision Cell Gas cell for collision induced dissociation (CID) to enhance fragmentation in PSD analysis Reflector Detector Measures ions reflected by the mirror. In the DE-STR this is a 6-10 mm pore size micro-channel plate. Timed Ion Selector A velocity selector that allows a single precursor ion of a selected mass and their fragment ions to pass to the detector. A Bradbury-Neilson gate is used.

Voyager-DE STR MALDI TOF Sample Linear Extraction plate Reflector detector grids Attenuator Timed ion detector selector Reflector Prism Laser Collision cell Camera Pumping Pumping

The problem: Peaks are inherently broad in MALDI-TOF spectra (poor mass resolution). The cause: Ions of the same mass coming from the target have different speeds. This is due to uneven energy distribution when the ions are formed by the laser pulse. Sample + matrix on target Ions of same mass, different velocities + + +

Can we compensate for the initial energy spread of ions of the same mass to produce narrower peaks? Delayed Extraction Reflector TOF Mass Analyzer

Delayed Extraction (DE) improves performance Step 1: No applied electric field. Ions spread out. + Ions of same mass, different velocities Step 2: Field applied. Slow ions accelerated more than fast ones. 0 V. Step 3: Slow ions catch up with faster ones. 20 kV.

What is a reflector TOF analyzer? A single stage gridded ion mirror that subjects the ions to a uniform repulsive electric field to reflect them. Detector Ion Source Reflector (Ion Mirror) The reflector or ion mirror compensates for the initial energy spread of ions of the same mass coming from the ion source, and improves resolution.

A reflector focuses ions to give better mass resolution + + 0 V. +20 kV

Reflector

Resolution & mass accuracy on mellitin 2000 4000 6000 8000 Counts 2840 2845 2850 2855 Mass (m/z) Resolution = 14200 Resolution = 4500 Resolution = 18100 15 ppm error 24 ppm error 55 ppm error

Fundamentals of Post Source Decay (PSD) PSD refers to a method of detecting and measuring the masses of fragment ions that are formed from a selected precursor ion. Fragment ions are mainly formed by unimolecular decomposition after the precursor ions are fully accelerated (after they exit the source—hence post-source decay) Fragment ions are separated and detected in the reflector.

Decomposition occurs in the flight tube Linear detector Reflector detector Laser Decay can occur at any point along here Reflector Source

Internal energy of precursor ions Only a small fraction of the precursor ions have enough energy to fragment during their lifetimes. No of ions Internal energy For peptides the efficiency of PSD fragmentation is amino acid composition and sequence dependent.

Increasing PSD Fragmentation There are two ways to increase the amount of fragmentation: both act to increase the precursor ions’ internal energy. Use higher laser intensity Use a collision cell

PSD fragment ion velocities are the same as their precursors + + All three of these species travel at the same velocity in the flight tube until they reach the reflector. + Why? Velocity is determined by initial acceleration. Initial energy = 20 keV. Bond energies = ~ 10 eV, so breaking a bond has a very minor effect on velocities.

Timed Ion Selector (TIS) The TIS is a Bradbury-Neilson gate, which is a type of velocity selector. It allows only selected precursor ions and their fragments to pass through to the reflector. Gate closed: alternating potentials on wires Gate open: wires at ground potential + - Ions + -

Timed Ion Selector operation TIS off “Gate open” TIS on “Gate closed”

Effect of the timed ion selector

Before fragmentation KE = 1/2 Mv2 KE = kinetic energy (= zeV) M = mass The intact molecular ion has translational kinetic energy equal to: KE = 1/2 Mv2 where: KE = kinetic energy (= zeV) M = mass v = velocity

Post source fragmentation The translational kinetic energy of a fragment ion is where KEM = precursor kinetic energy KEm = fragment kinetic energy M = precursor mass m = fragment mass

Precusor and PSD fragment ions take different paths in the “normal” reflector Reflector detector Intact precursor ion + + Fragment ion formed by PSD 0 V. +20 kV Reflector

How are PSD fragment ions that are traveling at the same speed as the precursor ion but contain reduced kinetic energy made to arrive at the detector so that they are focused? By varying the “steepness” of the voltage gradient in the reflector across the fragment ion mass range.

PSD mirror ratio setting Consider an ion (MH+) that can decompose into two fragments, A and B. Either of the following reactions can occur: MH+ AH+ + B MH+ A + BH+ Assume MH+ = 1,000 Da, AH+ = 700 Da, and BH+ = 300 Da

At mirror ratio = 1.00 MH+ ( 1,000) correctly focused AH+ BH+ MH+ MH+ ( 1,000) correctly focused AH+ (700) Poorly focused BH+ (300) Poorly focused

At mirror ratio = 0.7 BH+ AH+ MH+ MH+ ( 1,000) not focused AH+ (700) correctly focused BH+ (300) Poorly focused

At mirror ratio = 0.3 MH+ ( 1,000) not focused AH+ (700) not focused BH+ AH+ & MH+ MH+ ( 1,000) not focused AH+ (700) not focused BH+ (300) correctly focused

Resolution decreases as the fragment ions penetrate less into the mirror MR=0.61 MR=0.71 MR=0.80 (MR= mirror ratio)

A PSD spectrum is taken in “stitches”

PSD Spectrum of Angiotensin I, MH+ = 1296.7 Da Composite of the focused mass regions from several spectra acquired with different mirror ratios

Characteristics of CID (collison induced dissociation) Immonium ion signals are enhanced with collision gas; use routinely below fragment mass 200. Collisions can induce fragmentation of ions that do not decompose under normal PSD conditions. Side chain fragmentation may allow one to distinguish between Leu and Ile.

Effect of CID on immonium ions

CID to distinguish between Ile and Leu (GRF Lys-C peptide KLLQDILSR; MH+ = 1085.667)