Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller

Slides:



Advertisements
Similar presentations
Some structures Dansyl chloride 1,5-I-AEDANS Fluorescein isothiocyante ANS Ethidium bromide 5-[2-[(2-iodoacetyl)amino]ethylamino] naphthalene-1-sulfonic.
Advertisements

Lecture 3 Kinetics of electronically excited states
Methods: Fluorescence Biochemistry 4000 Dr. Ute Kothe.
Data Analysis Issues in Time-Resolved Fluorescence Anisotropy Bill Laws Department of Chemistry University of Montana Support from NSF EPSCoR, NIH, and.
Fluorescence Microscopy Chelsea Aitken Peter Aspinall.
Are You FRETting? Find Out for Sure With FLIM Frequency Domain FLIM for Your Scope Intelligent Imaging Innovations.
FRET(Fluorescent Resonance Energy Transfer)
Lecture 7: Fluorescence: Polarization and FRET Bioc 5085 March 31, 2014.
Lecture 5 FCS, Autocorrelation, PCH, Cross-correlation Enrico Gratton Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics.
Fluorescence Fluctuation Spectroscopy – A tool for the detection of nanometer sized particles in living cells Michael Edetsberger Max F. Perutz Laboratories,
Fluorescence Absorption of light occurs within ~ seconds, leaving a molecule in an excited state What happens next? –If no photon is re-emitted,
Pbio550: Biophysics of Ca2+ signaling ( washington
Two-Focus Fluorescence Correlation  Spectroscopy: A New Tool for Accurate and Absolute Diffusion Measurements Jörg Enderlein et al., ChemPhysChem, 8, 433–443.
Powerpoint Templates Page 1 Powerpoint Templates Spectroscopic Microscopy.
Scanning excitation and emission spectra I Wavelength (nm) )Scan excitation with emission set at 380 nm -λ ex,max = 280 nm 2) Scan emission.
FCS Club Sept 8 th 2006 Schwille, P., F. Meyer-Almes and R. Rigler (1997). "Dual-color fluorescence cross- correlation spectroscopy for multicomponent.
FAT Average lifetime (ps) GFP- Pax GFP-Pax + FAT- mCherry Lifetime (ns) Pax FAT Advanced Fluorescence Microscopy I: Fluorescence (Foster)
Physical Fluorescence Excitation Dr Maria Kiskowski Byrne, Department of Mathematics, Vanderbilt University. Dr Anne Kenworthy, Depts. of Molecular Physiology.
Pursuing the initial stages of crystal growth using dynamic light scattering (DLS) and fluorescence correlation spectroscopy (FCS) Takashi Sugiyama Miyasaka.
Today’s take-home lessons (i.e. what you should be able to answer at end of lecture) FRET – why it’s useful, R -6 dependence; R 0 (3-7 nm), very convenient.
Fluorescence Correlation Spectroscopy
Spectroscopy images GiGi Gs Globals by Enrico Gratton The Globals program originally developed at the LFD for analysis of multiple files from spectroscopy.
Today’s Announcements 1.Next Tuesday: Diffusion (Why moving in a cell is like swimming in concrete.) 2. Homework assigned today Last graded Homework:
Förster Resonance Energy Transfer (FRET)
Date of download: 7/8/2016 Copyright © 2016 SPIE. All rights reserved. Schematic representations of the HP1α and C/EBPα proteins indicating the relative.
Expert Supervision By: Dr. C.K.Nandi Associate professor School Of Basic Sciences I.I.T Mandi Presentation By: Kush Kaushik V16016 M.Sc Chemistry I.I.T.
Bin Wu, Jeffrey A. Chao, Robert H. Singer  Biophysical Journal 
K-Space Image Correlation Spectroscopy: A Method for Accurate Transport Measurements Independent of Fluorophore Photophysics  David L. Kolin, David Ronis,
Today’s take-home lessons: FRET (i. e
Volume 93, Issue 2, Pages (July 2007)
Volume 85, Issue 3, Pages (September 2003)
Molecular Dynamics in Living Cells Observed by Fluorescence Correlation Spectroscopy with One- and Two-Photon Excitation  Petra Schwille, Ulrich Haupts,
Volume 95, Issue 11, Pages (December 2008)
Quantification of Membrane Protein Dynamics and Interactions in Plant Cells by Fluorescence Correlation Spectroscopy  Xiaojuan Li, Jingjing Xing, Zongbo.
Volume 83, Issue 6, Pages (December 2002)
Förster Resonance Energy Transfer (FRET)
Aleš Benda, Yuanqing Ma, Katharina Gaus  Biophysical Journal 
Volume 98, Issue 12, Pages (June 2010)
Volume 104, Issue 5, Pages (March 2013)
CW and CCW Conformations of the E
Factor Xa Binding to Phosphatidylserine-Containing Membranes Produces an Inactive Membrane-Bound Dimer  Tilen Koklic, Rinku Majumder, Gabriel E. Weinreb,
Francesca Pennacchietti, Travis J. Gould, Samuel T. Hess 
Asako Sawano, Hiroshi Hama, Naoaki Saito, Atsushi Miyawaki 
V. Vetri, G. Ossato, V. Militello, M.A. Digman, M. Leone, E. Gratton 
Volume 114, Issue 12, Pages (June 2018)
Volume 94, Issue 7, Pages (April 2008)
Kinesin Moving through the Spotlight: Single-Motor Fluorescence Microscopy with Submillisecond Time Resolution  Sander Verbrugge, Lukas C. Kapitein, Erwin.
Volume 107, Issue 8, Pages (October 2014)
Quantifying the Interaction between EGFR Dimers and Grb2 in Live Cells
Samuel T. Hess, Watt W. Webb  Biophysical Journal 
Shaohui Huang, Ahmed A. Heikal, Watt W. Webb  Biophysical Journal 
Volume 88, Issue 6, Pages (June 2005)
Saswata Sankar Sarkar, Jayant B. Udgaonkar, Guruswamy Krishnamoorthy 
Characterization of the Photoconversion on Reaction of the Fluorescent Protein Kaede on the Single-Molecule Level  P.S. Dittrich, S.P. Schäfer, P. Schwille 
Saswata Sankar Sarkar, Jayant B. Udgaonkar, Guruswamy Krishnamoorthy 
Volume 114, Issue 3, Pages (February 2018)
Volume 105, Issue 10, Pages (November 2013)
Volume 102, Issue 5, Pages (March 2012)
The Use of pHluorins for Optical Measurements of Presynaptic Activity
Fluorescence Fluctuation Spectroscopy of mCherry in Living Cells
High-Pressure Fluorescence Correlation Spectroscopy
Single-Molecule Three-Color FRET
Polarized Fluorescence Resonance Energy Transfer Microscopy
Shaohui Huang, Ahmed A. Heikal, Watt W. Webb  Biophysical Journal 
Daniela Krüger, Jan Ebenhan, Stefan Werner, Kirsten Bacia 
Yufang Wang, Ling Guo, Ido Golding, Edward C. Cox, N.P. Ong 
The Dual-Color Photon Counting Histogram with Non-Ideal Photodetectors
Probing the Endocytic Pathway in Live Cells Using Dual-Color Fluorescence Cross- Correlation Analysis  Kirsten Bacia, Irina V. Majoul, Petra Schwille 
Orientation of the Myosin Light Chain Region by Single Molecule Total Internal Reflection Fluorescence Polarization Microscopy  Margot E. Quinlan, Joseph.
Laurdan Fluorescence Lifetime Discriminates Cholesterol Content from Changes in Fluidity in Living Cell Membranes  Ottavia Golfetto, Elizabeth Hinde,
Presentation transcript:

Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure and slide acknowledgements: Enrico Gratton

Fluorescence Parameters & Methods 1. Excitation & Emission Spectra • Local environment polarity, fluorophore concentration 2. Anisotropy & Polarization • Rotational diffusion 3. Quenching • Solvent accessibility • Character of the local environment 4. Fluorescence Lifetime • Dynamic processes (nanosecond timescale) 5. Resonance Energy Transfer • Probe-to-probe distance measurements 6. Fluorescence microscopy • localization 7. Fluorescence Correlation Spectroscopy • Translational & rotational diffusion • Concentration • Dynamics

Historic Experiment: 1st Application of Correlation Spectroscopy (Svedberg & Inouye, 1911) Occupancy Fluctuation Gold particles time 1200020013241231021111311251110233133322111224221226122142345241141311423100100421123123201111000111_211001320000010011000100023221002110000201001_333122000231221024011102_1222112231000110331110210110010103011312121010121111211_10003221012302012121321110110023312242110001203010100221734410101002112211444421211440132123314313011222123310121111222412231113322132110000410432012120011322231200_253212033233111100210022013011321131200101314322112211223234422230321421532200202142123232043112312003314223452134110412322220221 Svedberg and Inouye, Zeitschr. F. physik. Chemie 1911, 77:145 Collected data by counting (by visual inspection) the number of particles in the observation volume as a function of time using a “ultra microscope” ! Statistical analysis of raw data required

Particle Correlation *Histogram of particle counts *Autocorrelation Poisson statistics Autocorrelation not available in the original paper. It can be easily calculated today.

Historical Science Investigator Svedberg claimed: Gold colloids with radius R = 3 nm (Stokes-Einstein) characteristic diffusion time Experimental facts: Slit Conclusion: Bad sample preparation The ultramicroscope was invented in 1903 (Siedentopf and Zsigmondy). They already concluded that scattering will not be suitable to observe single molecules, but fluorescence could.

Fluctuations are in the Fluorescence Signal In FCS Fluctuations are in the Fluorescence Signal Diffusion Enzymatic Activity Phase Fluctuations Conformational Dynamics Rotational Motion Protein Folding Example of processes that could generate fluctuations

Generating Fluctuations By Motion What is Observed? 1. The Rate of Motion 2. The Concentration of Particles 3. Changes in the Particle Fluorescence while under Observation, for example conformational transitions Observation Volume Sample Space

Defining Our Observation Volume: One- & Two-Photon Excitation. Defined by the pinhole size, wavelength, magnification and numerical aperture of the objective Approximately 1 um3 Defined by the wavelength and numerical aperture of the objective

1-photon 2-photon Need a pinhole to define a small volume Brad Amos MRC, Cambridge, UK

Data Treatment & Analysis Time Histogram Autocorrelation Autocorrelation Parameters: G(0) & kaction Photon Counting Histogram (PCH) PCH Parameters: <N> & e

Autocorrelation Function Factors influencing the fluorescence signal: kQ = quantum yield and detector sensitivity (how bright is our probe). This term could contain the fluctuation of the fluorescence intensity due to internal processes C(r,t) is a function of the fluorophore concentration over time. This is the term that contains the “physics” of the diffusion processes W(r) describes our observation volume

Fluorescence Fluctuation Calculating the Autocorrelation Function Fluorescence Fluctuation F(t) in photon counts time  Average Fluorescence t t + t

As time (tau) approaches 0 The Autocorrelation Function t3 t5 t4 t2 t1 G(0)  1/N As time (tau) approaches 0 Diffusion

The Effects of Particle Concentration on the Autocorrelation Curve

Why Is G(0) Proportional to 1/Particle Number? A Poisson distribution describes the statistics of particle occupancy fluctuations. In a Poissonian system the variance is proportional to the average number of fluctuating species:

G(0), Particle Brightness and Poisson Statistics 1 0 0 0 0 0 0 0 0 2 0 1 1 1 0 0 0 0 0 0 1 0 0 0 0 0 0 0 1 0 1 0 0 0 1 0 0 1 0 0 Time Average = 0.275 Variance = 0.256 Lets increase the particle brightness by 4x: 4 0 0 0 0 0 0 0 0 8 0 4 4 4 0 0 0 0 0 0 4 0 0 0 0 0 0 0 4 0 4 0 0 0 4 0 0 4 0 0 Average = 1.1 Variance = 4.09 0.296

What about the excitation (or observation) volume shape?

Effect of Shape on the (Two-Photon) Autocorrelation Functions: For a 2-dimensional Gaussian excitation volume: 1-photon equation contains a 4, instead of 8 For a 3-dimensional Gaussian excitation volume:

Additional Equations: 3D Gaussian Confocor analysis: ... where N is the average particle number, tD is the diffusion time (related to D, tD=w2/8D, for two photon and tD=w2/4D for 1-photon excitation), and S is a shape parameter, equivalent to w/z in the previous equations. Note: The offset of one is caused by a different definition of G() : Triplet state term: ..where T is the triplet state amplitude and tT is the triplet lifetime.

Orders of magnitude (for 1 μM solution, small molecule, water) Volume Device Size(μm) Molecules Time milliliter cuvette 10000 6x1014 104 microliter plate well 1000 6x1011 102 nanoliter microfabrication 100 6x108 1 picoliter typical cell 10 6x105 10-2 femtoliter confocal volume 1 6x102 10-4 attoliter nanofabrication 0.1 6x10-1 10-6

The Effects of Particle Size on the Autocorrelation Curve 300 um2/s 90 um2/s 71 um2/s Diffusion Constants Fast Diffusion Slow Diffusion Stokes-Einstein Equation: and Monomer --> Dimer Only a change in D by a factor of 21/3, or 1.26

FCS inside living cells Two-Photon Spot Correlation Analysis Dsolution Dnucleus = 3.3 Coverslip objective Measure the diffusion coefficient of Green Fluorescent Protein (GFP) in aqueous solution in inside the nucleus of a cell.

Autocorrelation Adenylate Kinase -EGFP Chimeric Protein in HeLa Cells Examples of different Hela cells transfected with AK1-EGFP Fluorescence Intensity Examples of different Hela cells transfected with AK1b -EGFP Qiao Qiao Ruan, Y. Chen, M. Glaser & W. Mantulin Dept. Biochem & Dept Physics- LFD Univ Il, USA

Autocorrelation of EGFP & Adenylate Kinase -EGFP EGFP-AK in the cytosol G(t) EGFP-AKb in the cytosol EGFPsolution EGFPcell Time (s) Normalized autocorrelation curve of EGFP in solution (•), EGFP in the cell (• ), AK1-EGFP in the cell(•), AK1b-EGFP in the cytoplasm of the cell(•).

Autocorrelation of Adenylate Kinase –EGFP on the Membrane Clearly more than one diffusion time A mixture of AK1b-EGFP in the cytoplasm and membrane of the cell.

Autocorrelation Adenylate Kinaseb -EGFP Plasma Membrane Cytosol D D Diffusion constants (um2/s) of AK EGFP-AKb in the cytosol -EGFP in the cell (HeLa). At the membrane, a dual diffusion rate is calculated from FCS data. Away from the plasma membrane, single diffusion constants are found.

Multiple Species Case 1: Species vary by a difference in diffusion constant, D. Autocorrelation function can be used: (2D-Gaussian Shape) ! fi is the fractional fluorescence intensity of species i. G(0)sample is no longer g/N !

Antibody - Hapten Interactions Binding site Binding site carb2 Digoxin: a cardiac glycoside used to treat congestive heart failure. Digoxin competes with potassium for a binding site on an enzyme, referred to as potassium-ATPase. Digoxin inhibits the Na-K ATPase pump in the myocardial cell membrane. Mouse IgG: The two heavy chains are shown in yellow and light blue. The two light chains are shown in green and dark blue..J.Harris, S.B.Larson, K.W.Hasel, A.McPherson, "Refined structure of an intact IgG2a monoclonal antibody", Biochemistry 36: 1581, (1997).

Binding titration from the autocorrelation analyses: Anti-Digoxin Antibody (IgG) Binding to Digoxin-Fluorescein triplet state Digoxin-Fl•IgG (99% bound) Autocorrelation curves: Digoxin-Fl•IgG (50% Bound) Digoxin-Fl Binding titration from the autocorrelation analyses: Kd=12 nM S. Tetin, K. Swift, & , E, Matayoshi , 2003

Two Binding Site Model IgG•2Ligand-Fl IgG + 2 Ligand-Fl IgG•Ligand-Fl 50% quenching Kd IgG•Ligand No quenching IgG•2Ligand [Ligand]=1, G(0)=1/N, Kd=1.0

Digoxin-FL Binding to IgG: G(0) Profile Y. Chen , Ph.D. Dissertation; Chen et. al., Biophys. J (2000) 79: 1074

Case 2: Species vary by a difference in brightness assuming that The quantity G(0) becomes the only parameter to distinguish species, but we know that: The autocorrelation function is not suitable for analysis of this kind of data without additional information. We need a different type of analysis

Photon Counting Histogram (PCH) Sources of Non-Poissonian Noise Aim: To resolve species from differences in their molecular brightness Molecular brightness ε : The average photon count rate of a single fluorophore PCH: probability distribution function p(k) where p(k) is the probability of observing k photon counts Single Species: Note: PCH is Non-Poissonian! Sources of Non-Poissonian Noise Detector Noise Diffusing Particles in an Inhomogeneous Excitation Beam* Particle Number Fluctuations* Multiple Species*

PCH Example: Differences in Brightness Increasing Brightness frequency (en=1.0) (en=2.2) (en=3.7) Increasing Brightness Photon Counts

Single Species PCH: Concentration 5.5 nM Fluorescein Fit: e = 16,000 cpsm N = 0.3 550 nM Fluorescein Fit: e = 16,000 cpsm N = 33 As particle concentration increases the PCH approaches a Poisson distribution

Photon Counting Histogram: Multispecies Binary Mixture: Molecular Brightness Concentration Snapshots of the excitation volume Intensity Time

Photon Counting Histogram: Multispecies Sample 2: many but dim (23 nM fluorescein at pH 6.3) Sample 1: fewer but brighter fluors (10 nM Rhodamine) Sample 3: The mixture The occupancy fluctuations for each specie in the mixture becomes a convolution of the individual specie histograms. The resulting histogram is then broader than expected for a single species.

! Resolve a protein mixture with a brightness ratio of two Alcohol dehydrogenase labeling experiments Singly labeled proteins Mixture of singly or doubly labeled proteins + ! Both species have same color fluorescence lifetime diffusion coefficient polarization kcpsm kcpsm

PCH in cells: Brightness of EGFP Excitation=895nm The molecular brightness of EGFP is a factor ten higher than that of the autofluorescence in HeLa cells Chen Y, Mueller JD, Ruan Q, Gratton E (2002) Biophysical Journal, 82, 133 .

Brightness and Stoichiometry EGFP2 Intensity (cps) EGFP Brightness of EGFP2 is twice the brightness of EGFP Chen Y, Wei LN, Mueller JD, PNAS (2003) 100, 15492-15497

Caution: PCH analysis and dead-time effects PCH analysis assumes ideal detectors. Afterpulsing and deadtime of the photodetector change the photon count statistics and lead to biased parameters. Improved PCH models that take non-ideal detectors into account are available: Hillesheim L, Mueller JD, Biophys. J. (2003), 85, 1948-1958

Distinguish Homo- and Hetero-interactions in living cells ECFP: EYFP: Apparent Brightness B A + ε 2ε single detection channel experiment distinguish between CFP and YFP by excitation (not by emission)! brightness of CFP and YFP is identical at 905nm (with the appropriate filters) you can choose conditions so that the brightness is not changed by FRET between CFP and YFP determine the expressed protein concentrations of each cell!

PCH analysis of a heterodimer in living cells The nuclear receptors RAR and RXR form a tight heterodimer in vitro. We investigate their stoichiometry in the nucleus of COS cells. RAR RXR We expect: Chen Y, Li-Na Wei, Mueller JD, Biophys. J., (2005) 88, 4366-4377

Two Channel Detection: Each detector observes Cross-correlation Sample Excitation Volume Beam Splitter Increases signal to noise by isolating correlated signals. Corrects for PMT noise Detector 1 Detector 2 Each detector observes the same particles

Removal of Detector Noise by Cross-correlation 11.5 nM Fluorescein Detector 2 Detector after-pulsing Cross-correlation

Calculating the Cross-correlation Function Detector 1: Fi time  t + t t Detector 2: Fj time

Cross-correlation Calculations One uses the same fitting functions you would use for the standard autocorrelation curves. Thus, for a 3-dimensional Gaussian excitation volume one uses: G12 is commonly used to denote the cross-correlation and G1 and G2 for the autocorrelation of the individual detectors. Sometimes you will see Gx(0) or C(0) used for the cross-correlation.

Two-Color Cross-correlation The cross-correlation ONLY if particles are observed in both channels Sample Red filter Green filter Each detector observes particles with a particular color The cross-correlation signal: Only the green-red molecules are observed!!

Two-color Cross-correlation Equations are similar to those for the cross correlation using a simple beam splitter: Information Content Signal Correlated signal from particles having both colors. Autocorrelation from channel 1 on the green particles. Autocorrelation from channel 2 on the red particles.

Experimental Concerns: Excitation Focusing & Emission Collection We assume exact match of the observation volumes in our calculations which is difficult to obtain experimentally. Excitation side: (1) Laser alignment (2) Chromatic aberration (3) Spherical aberration Emission side: (1) Chromatic aberrations (2) Spherical aberrations (3) Improper alignment of detectors or pinhole (cropping of the beam and focal point position)

Two-Color Fluctuation Correlation Spectroscopy Uncorrelated Ch.2 Ch.1 Correlated Interconverting For two uncorrelated species, the amplitude of the cross-correlation is proportional to:

Does SSTR1 exist as a monomer after ligand binding while SSTR5 exists as a dimer/oligomer? Collaboration with Ramesh Patel*† and Ujendra Kumar* *Fraser Laboratories, Departments of Medicine, Pharmacology, and Therapeutics and Neurology and Neurosurgery, McGill University, and Royal Victoria Hospital, Montreal, QC, Canada H3A 1A1; †Department of Chemistry and Physics, Clarkson University, Potsdam, NY 13699 Somatostatin Fluorescein Isothiocyanate (FITC) Somatostatin Texas Red (TR) Cell Membrane R1 R5 Three Different CHO-K1 cell lines: wt R1, HA-R5, and wt R1/HA-R5 Hypothesis: R1- monomer ; R5 - dimer/oligomer; R1R5 dimer/oligomer

SSTR1 CHO-K1 cells with SST-fitc + SST-tr Green Ch. Red Ch. Very little labeled SST inside cell nucleus Non-homogeneous distribution of SST Impossible to distinguish co-localization from molecular interaction

Monomer A = 0.22 Dimer B = 0.71 G12(0) G1(0) G12(0) G1(0) Minimum Maximum = 0.71 G12(0) G1(0)

Experimentally derived auto- and cross-correlation curves from live R1 and R5/R1 expressing CHO-K1 cells using dual-color two-photon FCS. R1 R1/R5 The R5/R1 expressing cells have a greater cross-correlation relative to the simulated boundaries than the R1 expressing cells, indicating a higher level of dimer/oligomer formation. Patel, R.C., et al., Ligand binding to somatostatin receptors induces receptor-specific oligomer formation in live cells. PNAS, 2002. 99(5): p. 3294-3299

Molecular Dynamics What if the distance/orientation is not constant? Fluorescence fluctuation can result from FRET or Quenching FCS can determine the rate at which this occurs This will yield hard to get information (in the ms to ms range) on the internal motion of biomolecules

10 -3 -2 -1 1 40 80 120 160 200 t (s) G( ) B) A) . A) Cameleon fusion protein consisting of ECFP, calmodulin, and EYFP. [Truong, 2001 #1293] Calmodulin undergoes a conformational change that allows the ECFP/EYFP FRET pair to get cl ose enough for efficient energy transfer. Fluctuations between the folded and unfolded states will yield a measurable kinetic component for the cross - correlation. B) Simulation of how such a fluctuation would show up in the autocorrelation and cross correl ation. Red dashed line indicates pure diffusion.

In vitro Cameleon Data Ca2+ Saturated Crystallization And Preliminary X-Ray Analysis Of Two New Crystal Forms Of Calmodulin, B.Rupp, D.Marshak and S.Parkin, Acta Crystallogr. D 52, 411 (1996) Are the fast kinetics (~20 s) due to conformational changes or to fluorophore blinking?

Dual-color PCH analysis (1) Fluorescence F(t) <FA> FA Time t Cross-Correlation Dual-Color PCH FB Fluorescence F(t) <FB> Time t

Brightness in each channel: eA, eB Signal A Brightness in each channel: eA, eB Average number of molecules: N Tsample Signal B Tsample

Dual-color PCH analysis (2) Signal A Tsample Signal B Tsample Single Species: Brightness in each channel: eA, eB Average number of molecules: N

Resolve Mixture of ECFP and EYFP in vitro fluctuations Dual-Color PCH 2 channels 1 species model c2 = 17.93 2 species model c2 = 1.01 Chen Y, Tekmen M, Hillesheim L, Skinner J, Wu B, Mueller JD, Biophys. J. (2005), 88 2177-2192 ECFP & EYFP mixture resolved with single histogram. Note: Cross-correlation analysis cannot resolve a mixture of ECFP & EYFP with a single measurement!