Wald, fMRI MR Physics Massachusetts General Hospital Athinoula A. Martinos Center MR physics and safety for fMRI Lawrence L. Wald, Ph.D.

Slides:



Advertisements
Similar presentations
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Advertisements

Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
Magnetic Resonance Imaging
Imaging Sequences part I
MRI Phillip W Patton, Ph.D..
In Chan Song, Ph.D. Seoul National University Hospital
Richard Wise FMRI Director +44(0)
BE 581 Lecture 3- Intro to MRI.
Statistical Parametric Mapping
Fund BioImag : MRI contrast mechanisms 1.What is the mechanism of T 2 * weighted MRI ? BOLD fMRI 2.How are spin echoes generated ? 3.What are.
Figure 2. Signal level (left) degrades with slice offset and slice thickness when Z2 SEM is used in GradLoc imaging (ROI = FOV/2). To recover the full.
Parameters and Trade-offs
Topics spatial encoding - part 2. Slice Selection  z y x 0 imaging plane    z gradient.
Principles of MRI: Image Formation
Chapter 9 Basic MRI I Mark D. Herbst, MD, PhD. Notice This lecture contained many drawings on the whiteboard, so get these from one of the other students.
Magnetic Resonance Imaging Magnetic field gradients.
Basic Principles MRI related to Neuroimaging Xiaoping Hu Department of Biomedical Engineering Emory University/Georgia Tech
Wald, fMRI MR Physics Massachusetts General Hospital Athinoula A. Martinos Center MR physics for fMRI Lawrence L. Wald, Ph.D.
Contrast and Acquisition
Terry M. Button, Ph.D. Principals of Magnetic Resonance Image Formation.
Bruce Fischl MGH ATHINOULA A. MARTINOS CENTER Harvard Medical School MIT CSAIL/HST some open issues in functional MRI (thanks to Larry Wald for almost.
Chapter 10 Lecture Mark D. Herbst, M.D., Ph.D..
Encoding and Image Formation
Steady-state free precession and other 3D methods for high-resolution FMRI Steady-state free precession and other 3D methods for high-resolution FMRI Karla.
EPI – Echo Planar Imaging Joakim Rydell
Nuclear Magnetic Resonance (NMR) Magnetic Resonance Imaging (MRI)
Basics of Magnetic Resonance Imaging
Psy 8960, Fall ‘06 Fourier transforms1 –1D: square wave –2D: k x and k y Spatial encoding with gradients Common artifacts Phase map of pineapple slice.
FMRI: Biological Basis and Experiment Design Lecture 7: Gradients and k-space FFT examples –Sampling and aliasing Gradient Gradient echo K-space
FMRI: Biological Basis and Experiment Design Lecture 8: Pulse sequences, Take 2 Gradient echo review K-space review Slice selection K-space navigation.
Compressed Sensing for Chemical Shift-Based Water-Fat Separation Doneva M., Bornert P., Eggers H., Mertins A., Pauly J., and Lustig M., Magnetic Resonance.
MRI Image Formation Karla Miller FMRIB Physics Group.
Medical Imaging Systems: MRI Image Formation
Magnetic Resonance Imaging 4
Principles of MRI Physics and Engineering
Imaging Sequences part II
2012 spring fMRI: theory & practice
Medical Imaging Systems: MRI Image Formation
Basic of Magnetic Resonance Imaging Seong-Gi Kim Paul C. Lauterbur Chair in Imaging Research Professor of Radiology, Neurobiology and Bioengineering University.
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
fMRI Methods Lecture2 – MRI Physics
Quiz In a 2D spin warp or FT MR scan, aliasing should only occur
G Practical MRI 1 Gradients.
Contrast Mechanisms in MRI Introduction to Cardiovascular Engineering Michael Jay Schillaci, PhD Managing Director, Physicist Tuesday, September 16 th,
Allen W. Song, PhD Brain Imaging and Analysis Center Duke University MRI: Image Formation.
Contrast Mechanism and Pulse Sequences
FMRI – Week 4 – Contrast Scott Huettel, Duke University MR Contrast FMRI Graduate Course (NBIO 381, PSY 362) Dr. Scott Huettel, Course Director.
Anna Beaumont FRCR Part I Physics
MR Image Formation FMRI Graduate Course (NBIO 381, PSY 362)
MRI Physics: Spatial Encoding Anna Beaumont FRCR Part I Physics.
Declaration of Relevant Financial Interests or Relationships David Atkinson: I have no relevant financial interest or relationship to disclose with regard.
MRI: Contrast Mechanisms and Pulse Sequences
Magnetic Resonance Learning Objectives
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Charged particle. Moving charge = current Associated magnetic field - B.
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Lecture 1: Magnetic Resonance
Chapter 5 Mark D. Herbst, M.D., Ph.D.. The MR Imaging Process Two major functions –Acquisition of RF signals –Reconstruction of images.
Parameters which can be “optimized” Functional Contrast Image signal to noise Hemodynamic Specificity Image quality (warping, dropout) Speed Resolution.
10 spring fMRI: theory & practice
FMRI data acquisition.
MRI Physics in a Nutshell Christian Schwarzbauer
Magnetic Resonance Imaging: Physical Principles
Bioengineering 280A Principles of Biomedical Imaging Fall Quarter 2005 MRI Lecture 5 Thomas Liu, BE280A, UCSD, Fall 2005.
An Optimal Design Method for MRI Teardrop Gradient Waveforms
Eduardo H. M. S. G. de Figueiredo, BSc, Arthur F. N. G
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
Basic MRI I Mark D. Herbst, MD, PhD
(4)ELECTRONIC SUPPORT SYSTEM
Presentation transcript:

Wald, fMRI MR Physics Massachusetts General Hospital Athinoula A. Martinos Center MR physics and safety for fMRI Lawrence L. Wald, Ph.D.

Wald, fMRI MR Physics Outline: Monday Oct 18 (LLW): MR signal, Gradient and spin echo Basic image contrast Monday Oct 25 (LLW): Fast imaging for fMRI, artifacts Wed. Oct 20 (JJ): Encoding the image Wed. Oct 25 (LLW): fMRI BOLD contrast Mon. Nov. 1 (JJ): Review, plus other contrast mechanisms for fMRI (CBV, CBF)

Wald, fMRI MR Physics Why, introduction How: Review of k-space trajectories Different techniques (EPI, Spiral) Problems from B0 Susceptibility artifacts Fast MR Imaging Techniques

Wald, fMRI MR Physics Why fast imaging Capture time course, (e.g. hemodynamic) eliminate artifact from motion (during encode.)

Wald, fMRI MR Physics Magnetization vector durning MR RF Voltage (Signal) time Mz time encode

Wald, fMRI MR Physics Review of Image encoding, journey through kspace Two questions: 1) What does blipping on a gradient do to the water magnetization. 2) Why does measuring the signal amplitude after a blip tell you info about the spatial frequency composition of the image (k-space).

Wald, fMRI MR Physics Aside: Magnetic field gradient Uniform magnet Field from gradient coils Total field B o G x x B o + G x x x z

Wald, fMRI MR Physics Step two: encode spatial info. in-plane x y x B Field (w/ x gradient) BoBo B TOT = B o + G z x Freq. Signal B o along z oo  with gradient without gradient “Frequency encoding” v =  B TOT =  (B o + G x x)

Wald, fMRI MR Physics How does blipping on a grad. encode spatial info? y B Field (w/ z gradient) BoBo BoBo GyGy  y1y1 y2y2 z y y1y1 y2y2 all y locs process at same freq. all y locs process at same freq. spins in forehead precess faster...  y  =  B TOT =  B o  y G y  y  =  y  =  B o  y (G y 

Wald, fMRI MR Physics How does blipping on a grad. encode spatial info? z y y1y1 y2y2 yxyx z x y z oo 90° yx after RFAfter the blipped y gradient... position y 1 position 0position y 2 z z BoBo  y  =  y  =  B o  y (G y 

Wald, fMRI MR Physics How does blipping on a grad. encode spatial info? The magnetization vector in the xy plane is wound into a helix directed along y axis. Phases are ‘locked in’ once the blip is over. y

Wald, fMRI MR Physics  y  =  y  =  B o  y (G y  The bigger the gradient blip area, the tighter the helix y small blip medium blip large blip GyGy

Wald, fMRI MR Physics What have you measured? Consider 2 samples: uniform water 1 cm no signal observed signal is as big as if no gradient

Wald, fMRI MR Physics 10 mm kxkx kyky 1/10 mm 1/5mm 1/2.5mm 1/1.2mm = 1/Resolution Measurement intensity at a spatial frequency...

Wald, fMRI MR Physics kxkx kyky 1 / Res x 1 / FOV x FOV x = matrix * Res x Fourier transform

Wald, fMRI MR Physics Sample 3 points in kspace More efficient! t kxkx 1/10 mm 1/5mm 1/2.5mm 1/1.2mm = 1/Resolution t GyGy GyGy Frequency and phase encoding are the same principle!

Wald, fMRI MR Physics Conventional “Spin-warp” encoding “slice select” “phase enc” “freq. enc” (read-out) RF t S(t) t GzGz t GyGy GxGx t t kxkx kyky one excitation, one line of kspace... a1a1 a2a2

Wald, fMRI MR Physics Image encoding, “Journey through kspace” The Movie…

Wald, fMRI MR Physics kxkx kyky 1 / Res x 1 / FOV x FOV x = matrix * Res x Fourier transform

Wald, fMRI MR Physics Conventional “Spin-warp” encoding “slice select” “phase enc” “freq. enc” (read-out) RF t S(t) t GzGz t GyGy GxGx t t kxkx kyky one excitation, one line of kspace... a1a1 a2a2

Wald, fMRI MR Physics “Echo-planar” encoding RF t S(t) (no grads) t GzGz t GyGy GxGx t kxkx kyky one excitation, many lines of kspace... etc... T2*

Wald, fMRI MR Physics Bandwidth is asymmetric in EPI kxkx kyky Adjacent points in k x have short  t = 5 us (high bandwidth) Adjacent points along k y are taken with long  t (= 500us). (low bandwidth) The phase error (and thus distortions) are in the phase encode direction.

Wald, fMRI MR Physics RF t GzGz t GyGy GxGx t esp = 500 us for whole body grads, readout length = 32 ms esp = 270us for head gradients, readout length = 17 ms Characterization of EPI performance length of readout train for given resolution or echo spacing (esp) or freq of readout… ‘echo spacing’ (esp)

Wald, fMRI MR Physics What is important in EPI performance? Short image encoding time. Parameters related to total encoding time: 1) echo spacing. 2) frequency of readout waveform. Key specs for achieving short encode times: 1) gradient slew rate. 2) gradient strength. 3) ability to ramp sample. Good shimming (second order shims)

Wald, fMRI MR Physics Susceptibility in MR The good. The bad. The ugly.

Wald, fMRI MR Physics Enemy #1 of EPI: local susceptibility gradients B o field maps in the head

Wald, fMRI MR Physics Enemy #1 of EPI: local susceptibility gradients B o field maps in the head

Wald, fMRI MR Physics What do we mean by “susceptibility”? In physics, it refers to a material’s tendency to magnetize when placed in an external field. In MR, it refers to the effects of magnetized material on the image through its local distortion of the static magnetic field B o.

Wald, fMRI MR Physics Susceptibility effects occur near magnetically dis-similar materials Field disturbance around air surrounded by water (e.g. sinuses) Field map (coronal image) 1.5T BoBo Ping-pong ball in water…

Wald, fMRI MR Physics B o map in head: it’s the air tissue interface… Sagittal Bo field maps at 3T

Wald, fMRI MR Physics Susceptibility field (in Gauss) increases w/ B o 1.5T 3T 7T Ping-pong ball in H 2 0: Field maps (  TE = 5ms), black lines spaced by 0.024G (0.8ppm at 3T)

Wald, fMRI MR Physics Other Sources of Susceptibility You Should Be Aware of… Those fillings might be a problem…

Wald, fMRI MR Physics Local susceptibility gradients: 2 effects 1)Local dephasing of the signal (signal loss) within a voxel, mainly from thru- plane gradients 2)Local geometric distortions, (voxel location improperly reconstructed) mainly from local in-plane gradients.

Wald, fMRI MR Physics 1) Non-uniform Local Field Causes Local Dephasing Sagittal B o field map at 3T 5 water protons in different parts of the voxel… z slowest fastest x y z 90° T = 0 T = TE

Wald, fMRI MR Physics Local susceptibility gradients: thru-plane dephasing in grad echo EPI Bad for thick slice above frontal sinus… 3T

Wald, fMRI MR Physics Thru-plane dephasing gets worse at longer TE 3T, TE = 21, 30, 40, 50, 60ms

Wald, fMRI MR Physics Problem #2 Susceptibility Causes Image Distortion in EPI Field near sinus To encode the image, we control phase evolution as a function of position with applied gradients. Local suscept. Gradient causes unwanted phase evolution. The phase encode error builds up with time.  =  B local  t y y

Wald, fMRI MR Physics Susceptibility Causes Image Distortion Field near sinus y y Conventional grad. echo,   encode time  1/BW

Wald, fMRI MR Physics Susceptibility in EPI can give either a compression or expansion Altering the direction kspace is transversed causes either local compression or expansion. choose your poison… 3T whole body gradients

Wald, fMRI MR Physics Susceptibility Causes Image Distortion Field near sinus z Echoplanar Image,   encode time  1/BW Encode time = 34, 26, 22, 17ms 3T head gradients

Wald, fMRI MR Physics EPI and Spirals kxkx kyky GxGx GyGy kxkx kyky GxGx GyGy

Wald, fMRI MR Physics EPISpirals Susceptibility:distortion,blurring,dephasing Eddy currents:ghostsblurring k = 0 is sampled:1/2 through1st Corners of kspace:yesno Gradient demands:very highpretty high

EPI and Spirals EPI at 3T Spirals at 3T (courtesy Stanford group)

Wald, fMRI MR Physics B0B0 Nasal Sinus

Wald, fMRI MR Physics B0B0 Nasal Sinus + mouth shim

Wald, fMRI MR Physics Effect of Ear & Mouth Shim on EPI From P. Jezzard, Oxford B0B0

Wald, fMRI MR Physics With fast gradients, add parallel imaging Acquisition: SMASH SENSE Reconstruction: Folded datasets + Coil sensitivity maps Reduced k-space sampling { Folded images in each receiver channel

Wald, fMRI MR Physics (iPAT) GRAPPA for EPI susceptibility Fast gradients are the foundation, but EPI still suffers distortion 3T Trio, MRI Devices Inc. 8 channel array b=1000 DWI images iPAT (GRAPPA) = 0, 2x, 3x

Wald, fMRI MR Physics 4 fold acceleration of single shot sub- millimeter SE-EPI: 23 channel array 23 Channel array at 1.5T With and without 4x Accel. Single shot EPI, 256x256, 230mm FOV TE = 78ms Encoding with RF…

Extending the phased array to more channels: 23 channel “Bucky” array for 1.5T

Wald, fMRI MR Physics 9 Fold GRAPPA acceleration 3D FLASH 23 Channel array at 1.5T Can speed up encoding by an order of magnitude! 9 minute scan down to 1 minute… 3D Flash, 1mm x 1mm x 1.5mm, 256x256x128

Wald, fMRI MR Physics Whys stop at 32? 96 Channels on its way… Receiver array: Andreas Potthast Helmut array: Graham Wiggins

Wald, fMRI MR Physics Questions, comments to: Larry Wald