Temperature-Related Stability in Plasma Spectrometry Jerry Dulude, Ron Stux, and Vesna Dolic, Glass Expansion,

Slides:



Advertisements
Similar presentations
Introduction to Plasma-Surface Interactions Lecture 6 Divertors.
Advertisements

A Modern Approach to Controlling ICP Spray Chamber Temperature Jerry Dulude, Glass Expansion,
Ion Sources Some characteristics of ion sources (especially in high precision work): It should have high efficiency in generating ions of the element of.
1 Atomic Absorption Spectroscopy Atomic Emission Spectroscopy Lecture 18.
Elemental Scientific | | PC 3 Advantages Thermal stabilization of spray chamber improves long-term stability Reduces effect.
AAS and FES (Ch 10, 7th e, WMDS)
Inductively Coupled Plasma Mass Spectrometry or ICP-MS
INTRODUCTION TO OPTICAL METHODS
1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission.
Flame photometry.
Printed by Approaching a Universal Pneumatic Nebulizer – the next step Jerry Dulude and Ron Stux (USA), Vesna Dolic and Peter Liddell.
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
September 24-25, 2003 HAPL meeting, UW, Madison 1 Armor Configuration & Thermal Analysis 1.Parametric analysis in support of system studies 2.Preliminary.
Printed by Approaching a Universal Pneumatic Nebulizer Jerry Dulude (USA), Vesna Dolic and Peter Liddell (Australia), Glass Expansion.
3mm amplitude test by Cheng-Jiun Ma. Observation Source : Orion SiO Frequency: GHz 256 channels within 64MHz bandwidth Ant 3 & Ant 4 no Tsys correction.
Lecture 2 M.Sc.. AA Spectrometer Components Lamp and FlameDetector Fuel Oxidant Nebulizer Double-Click picture for VIDEO.
X-Ray Microanalysis – Precision and Sensitivity Recall… K-ratio Si = [I SiKα (unknown ) / I SiKα (std.) ] x CF CF relates concentration in std to pure.
Practical Implementation of Adaptation in a Digital Predistorter for RF Power Amplifiers A. Cesari, D.Dragomirescu, P. Lacroix, J.M. Dilhac LAAS-CNRS Toulouse,
Developments on the Front End of Your ICP Optimizing Performance and Reliability of ICP-AES and ICP-MS.
ICP Detection Limits. Detection Limit (DL) or Limit of Detection (LOD) The detection limit is the concentration that is obtained when the measured signal.
PH and ICP Techniques Clemson. pH quality control checks Scoop 2 checks One blank Double scoop 6 samples Tag set with lab number, check, and blank positions.
Reservoir Performance Curves
Agilent 7500 ICP-MS Date: 28JUL2005.
Atomic Emission Spectroscopy
13.1 The Principles of Feedback Mechanisms 13 Feedback Mechanisms.
in tin pest and pest free Monika Leodolter-Dvorak and Ilse Steffan
Zn x Cd 1-x S thin films were characterized to obtain high quality films deposited by RF magnetron sputtering system. This is the first time report of.
Mass spectrometry Ions are analyzed on the basis of their m/z Chlorine has 2 isotopes, 35 Cl and 37 Cl, in the approximate ratio of 3 :1. Electrons are.
Atomic Emission Spectrometry By: Alexa Kunch
Spectroscop1c Analysis Part 6 – Spectroscopic Analysis using Fluorescence and Atomic Absorption Spectrophotometry Chulalongkorn University, Bangkok, Thailand.
A Universal Free-Standing Accessory to Improve ICP Productivity Jerry Dulude, Glass Expansion, Inc., 4 Barlows Landing Rd., Pocasset, MA 02559,
Advantages of a Syringe Driven Sample Introduction Accessory for ICP Spectrometry Jerry Dulude and Scott Bridger Glass Expansion.
Effect of Sample Introduction on ICP-MS Detection of Nonmetal Analytes in a Simulated HPLC Matrix Adam E. Dill and Jon W. Carnahan Department of Chemistry.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 20
A UNIVERSAL PROGRAMMABLE TEMPERATURE SPRAY CHAMBER Jerry Dulude, Glass Expansion,
Adv. Inst. Techs.  flame emission (eg flame photometer) known as low temperature emission ( K) ◦ first form of spectroscopy ◦ used in commercial.
ADVANCE IN AUTOMOBILES HYDROGEN FUELLED ENGINES BY C.SUBRAMANIAN, 10MECH50, III-MECH, VCET,MADURAI.
Solution of a System of ODEs with POLYMATH and MATLAB, Boundary Value Iterations with MATLAB For a system of n simultaneous first-order ODEs: where x is.
Printed by Techniques for Increasing Productivity in ICP Spectrometry Jerry Dulude, Glass Expansion, Inc., 4 Barlows Landing Rd.,
ATOMIC SPECTROMETRY 1. Flames 2. Electrothermal Atomizers 3. Plasmas.
Gas Chromatography Lecture 38.
SPRAY CHAMBER TEMPERATURE: A CRITICAL (YET OFTEN MISSING) PARAMETER IN ICP EXPERIMENTS Jerry Dulude, Ron Stux, and Vesna Dolic, Glass Expansion,
Robust Heavy Ion Fusion Target Shigeo KAWATA Utsunomiya Univ. Japan U.S.-J. Workshop on HIF December 18-19, 2008 at LBNL & LLNL.
Advanced Analytical Chemistry – CHM 6157® Y. CAIFlorida International University Updated on 9/26/2006Chapter 3ICPMS Interference equations Isobaric.
Innovations in ICP Sample Introduction
Atomic Fluorescence Spectroscopy. Background l First significant research by Wineforder and Vickers in 1964 as an analytical technique l Used for element.
Printed by A New Versatile Programmable Temperature Spray Chamber for ICP Jerry Dulude and Ron Stux (USA), Vesna Dolic (Australia),
The Heavy Ion Fusion Virtual National Laboratory Neutralized Transport Experiment (NTX) P. K. Roy, S. S. Yu, S. Eylon, E. Henestroza, A. Anders, F. M.
A STAND-ALONE DEVICE TO SPEED ICP ANALYSES Jerry Dulude, Glass Expansion, Inc., 4 Barlows Landing Rd., Pocasset, MA 02559,
Homeostasis & Fluid Compartments By Sokha Kheam. Learning Objectives Explain the basic organization of the body, Define the fluid compartment of the body,
FREQUENCY-AGILE DIFFERENTIAL CAVITY RING-DOWN SPECTROSCOPY
1 Feature of Energy Transport in NSTX plasma Siye Ding under instruction of Stanley Kaye 05/04/09.
Studies on 2.45 GHz microwave ion sources Abhishek Nag IISER, KOLKATA Presented By: G.O. Rodrigues IUAC, New Delhi Supervised By:
Abstract Analysis of organic solvents by ICP is often seen as problematic and challenging due to the nature of the solvents used. Commonly encountered.
MISS. RAHIMAH BINTI OTHMAN
Date of download: 6/24/2016 Copyright © 2016 SPIE. All rights reserved. Different options for blood glucose monitoring in clinical chemistry using IR-spectroscopy.
이 장 우. 1. Introduction  HPLC-MS/MS methodology achieved its preferred status -Highly selective and effectively eliminated interference -Without.
Seok-geun Lee, Young-hwa An, Y.S. Hwang
COMPARISON OF EXCITATION CONDITIONS IN
Live Cell Imaging Chamber
Strategies for Eliminating Interferences in Optical Emission Spectroscopy Best practices to optimize your method and correct for interferences to produce.
Basic Design of PID Controller
The Analysis of Soils and Waters in Accordance with U. S
Corial 200R 11/17/2018 Simplicity, performance, and upgradability in a system designed for R&D environments RIE capabilities over a variety of materials.
Easy-to-use Reactive Ion Etching equipment
Strong and Weak Acids and Bases
Пламена Спектроскопија
Easy-to-use Reactive Ion Etching equipment
Coordination & Regulation
Injector Setup for G0 and HAPPEX & Lessons Learned
Presentation transcript:

Temperature-Related Stability in Plasma Spectrometry Jerry Dulude, Ron Stux, and Vesna Dolic, Glass Expansion,

Glass Expansion 1 Causes of Signal Drift in ICP-OES Optical drift RF instability Sample related baseline shift Matrix variations Drift in sample intro temperature Profiling & thermostat Feedback loop Background correction Internal std. or matrix matching ???????? Cause Solution

Glass Expansion 1 IsoMist™ Programmable Temperature Spray Chamber (PTSC)

Glass Expansion 1 IsoMist Characteristics Programmable from -10 to 60C in 1 degree increments Maintains temperature to within 0.1 degree Built-in Peltier device No external plumbing Temperature measured near the chamber surface

Glass Expansion 1 IsoMist on Optima 2100DV

Glass Expansion 1 PE 2100DV Standard Parameters Power: 1450 watts Coolant gas: 15LPM Aux. gas: 0.2LPM Quartz injector, 1.8mm bore Neb. Gas: 0.7LPM SeaSpray concentric glass nebulizer (1ml/min) IsoMist Programmable Temperature spray chamber with Twister Baffled Cyclonic

Glass Expansion 1 Intensity vs. Temperature-PE 2100DV (1ml/min uptake)

Glass Expansion 1 Temperature Plots on PE 2100DV 31% increase from 25 to 35C or 3% per degree C 31% increase from 10 to 25C or 2% per degree C AVG INTENSITY

Glass Expansion 1 Temperature Plots on PE 2100DV AVG INTENSITY Transport Efficiency

Glass Expansion 1 Temperature Plots on PE 2100DV Transport Efficiency AVG INTENSITY Avg. suppression of all lines Interference std: 1000ppm Na + K, 500ppm Ca

Glass Expansion 1 Temperature Plots on PE 2100DV Avg. suppression of all lines Transport Efficiency AVG INTENSITY Avg. suppression of ion lines Interference std: 1000ppm Na + K, 500ppm Ca

Glass Expansion 1 IsoMist on the Varian Vista MPX

Glass Expansion 1 Varian Vista MPX Standard Parameters Power: 1300 watts Coolant gas: 15LPM Aux. gas: 1.5LPM Quartz injector, 2.3mm bore Neb. Gas: 0.8LPM SeaSpray concentric glass nebulizer (0.4ml/min) IsoMist Programmable Temperature spray chamber with Twister Baffled Cyclonic

Glass Expansion 1 Effect of Spray Chamber Temperature on Intensity (Varian Vista MPX)

Glass Expansion 1 Effect of Temperature on Intensity Varian Vista MPX 10% increase from 20 to 25C or 2%/degree C

Glass Expansion 1 Effect of Spray Chamber Temperature Varian Vista MPX Avg. Intensity Transport Efficiency

Glass Expansion 1 Varian Vista MPX Parameters Power: 1300 watts Coolant gas: 15LPM Aux. gas: 1.5LPM Quartz injector, 1.8mm bore Neb. Gas: 0.8LPM SeaSpray concentric glass nebulizer (0.4ml/min) IsoMist Programmable Temperature spray chamber with Twister Baffled Cyclonic

Glass Expansion 1 Temperature vs. Intensity (Varian Vista MPX- 1.8mm bore)

Glass Expansion 1 Temperature vs. Intensity (Varian Vista MPX) 10% decrease from 20 to 40C or 0.5% per degree C 29% increase from 10 to 40C or 1% per degree C 1.8mm bore2.3mm bore

Glass Expansion 1 Low Sample Uptake Applications Biological applications Intracellular fluids Neonatal samples Forensic applications Speciation (HPLC-ICP-MS) Radioactive wastes reduction Low DL’s desirable

Glass Expansion 1 Effect of Temperature on Normalized Intensity (20 uL/min uptake rate) PE Optima 2100DV

Glass Expansion 1 Effect of Temperature on Intensity 20ul/min uptake rate 133% increase from 21 to 40C or 7% per degree C Avg. Intensity Estimated Transport Efficiency

Glass Expansion 1 Spray Chamber Temperature Impact High impact with respect to intensity Interdependent with other plasma parameters Provides another dimension for optimization Must remain constant during the experiment

Glass Expansion 1 Constant temperature benefits (1ml/min sample uptake) Higher accuracy Higher productivity Greater reproducibility Better experiments