Dan Wagner Hydrologist Arkansas Water Science Center

Slides:



Advertisements
Similar presentations
FlowTracker Bias Caused by Flow Disturbance David S. Mueller Office of Surface Water U.S. Geological Survey The use of trade or brand names does not imply.
Advertisements

Practice Insight Instructional Webinar Series Eligibility Manager
Maine Water Science Center
ADCP Measurements: Laptop to NWIS via SWAMI
U.S. Department of the Interior U.S. Geological Survey Effective Review of FlowTracker Measurements OSW WebinarMike Rehmel August 27, 2013 (Please Mute.
Quickfilter Pro Software Demonstration for QF4A512 The following slides will illustrate how you can design and verify a filter design in minutes! BEGIN.
Maintaining Gage Datum: Levels at Gaging Stations.
RD Instruments Home of the ADCP Measuring Water in Motion and Motion in Water THETIS Waves Data QA RD Instruments Home of the ADCP.
Overview of ADCP “Best Practices” Report
Syringe Pumps. Topics  Diagrams  Principles of Operation  Applications  Safety  Operation  Preventive Maintenance  Common Failure Modes  Troubleshooting.
Inter-Warehouse Transfers An Enhancement For iSeries 400 DMAS from  Copyright I/O International, 2004, 2005, 2007, 2010 Skip Intro.
All critical-flow devices determine flow rate by measuring one or more water levels and computing flow from a calibration equation. Water Level Sensors.
Starter The radius of the moon is 1,737,000 meters. Write this in scientific notation. The diameter of a carbon atom is meters. Write.
Review and Rating Discharge Measurements David S. Mueller Office of Surface Water March 2010.
1 4 th International Symposium on Flood Defence: Managing Flood Risk, Reliability and Vulnerability, Toronto, Ontario, Canada, May 6-8, 2008 J. Chen, J.
Discharge Measurements Created by: Jake Jacobson, ID WSC Modified by: Jeff Kitchen, CA WSC.
Get Acquainted with the Latest Versions of LC & SMBA USGS, TRDI, and SonTek Versions David S. Mueller, OSW April, 2013.
Wildland Fire Management Information (WFMI) Weather Lesson 5: Station Information April 2014.
QARTOD II Currents and Waves In-Situ Currents: Breakout Group Report Out QARTOD II February 28 – March 2, 2005.
Digital Notetaker Walter Jordan Tom Warsaw Senior Design Project EECC 657.
Principle of Index-Velocity Method and its Application Randy Marsden Teledyne RD Instruments.
Acquisition and Interpretation of Water-Level Data Travis von Dessonneck.
GPS. Meet the Bad Elf GPS Pro The Bad Elf GPS Pro offers the first Apple approved GPS with an informative user interface.
ChannelMaster Horizontal ADCP
Garmin GPS III Plus Data Collection. Objectives Collect: - Waypoints -Average Position Waypoints -Reference Waypoints - Multiple Tracks in One Track Log.
Page 1 IV) Systems for Hydrology – Wading Measurements FlowTracker Handheld ADV ® The World Leader for Water Velocity Measurement.
Water Quality Monitoring and Parameter Load Estimations in Lake Conway Point Remove Watershed and L’Anguille River Watershed Presented by: Dan DeVun, Equilibrium.
David S. Mueller U.S. Geological Survey Office of Surface Water
Excel ® Registry Tool Diane Powers Manager, AIMS Center University of Washington Psychiatry & Behavioral Sciences
INTERNATIONAL INSTITUTE FOR GEO-INFORMATION SCIENCE AND EARTH OBSERVATION The Global Positioning System Bart Krol / Jeroen Verplanke.
LGC Website and Customer On-line Tools LGC RESOURCE 2014.
HydrometJanuary AMS Short Course on Instrumentation1 HydrometeorologicalMeasurements Melanie A. Wetzel Desert Research Institute University of.
Working with GPS Stephen M. Kiama (MSc) [Geo-information Specialist]
SonTek FlowTracker® Basics of Operation
ICC Features Features Supporting unlimited projects per ICC –Advanced technical support Friendly registration utility 5 password-levels exceptional workspace.
U.S. Department of the Interior U.S. Geological Survey Alabama Water Science Center StreamStats: By Kernell Ries and J.
1 Team Leader TKS Job Aid. 2 Viewing the On-line Presentation If you are viewing this presentation via Internet Explorer for best results resize the “Notes”
INSTRUCTIONS We are looking to develop a detailed understanding of what information about an Platform needs to be displayed somewhere in the user interface,
LC and SMBA Updates Office of Surface Water Hydroacoustics Webinar January 6 and 9, 2008 David S. Mueller.
Surveying Techniques I. The USGS supplies 1:24,000 scale maps for all the U.S. But detailed topography at larger scales is rare and/or.
Excavator X62 Installation. This Webinar is LISTEN only Questions will be answered in writing at the end of the Webinar. QUESTIONS?
MODERN SURVEY (FAMILARISATION WITH EQUIPMENTS). Modern equipments EDM – Electronic distance measurement eqp. EDM – Electronic distance measurement eqp.
Introduction to WinRiver II
Motor-driven pumps  Sigma/ 1  Sigma/ 2  Sigma/ 3.
NWS FLDWAV ANALYSIS TOOL
U.S. Department of the Interior U.S. Geological Survey IMPROVING THE QUALITY AND EFFICIENCY OF ADCP STREAMFLOW MEASUREMENTS David S. Mueller U.S. Geological.
Slide #1 Computerised Transit Computer System Training Session Trader Application Users.
Harry Williams, Cartography1 Surveying Techniques II. GPS Despite the fact that Global Positioning Systems use very modern technology, the basic concept.
BASINS 2.0 and The Trinity River Basin By Jóna Finndís Jónsdóttir.
1 Sponsor Claim Process  Click on the Sponsor Claims menu in order to process claims.  The submenu will display: Sponsor Claims will display the most.
Data Logging (Part 1).
Navigation NAU 102 Lesson 19. New Advances Flux Gate Magnetic Compass Ring Laser Gyrocompasss GPS Compass & DGPS Compass.
Acoustic Doppler Technology for Streamflow Data Collection
By Larry R. Bohman. 2 Our data collection infrastructure is already in place! Uses: Why USGS needed a policy… Not published before about 1990, but number.
Brian Haggard Director Arkansas Water Resources Center Funding provided by ANRC through Beaver Water District.
SIO 218A, Spring 2014 Petia Yanchulova. cPIES IES Data Processing Manual, URI, 2007.
Importance of Testing Hydroacoustic Instruments: USGS Case Studies Kevin Oberg National Coordinator for Hydroacoustics USGS Water Mission Area.
Child Care Subsidy Program Online Billing Provider Training Spring 2016.
Comprehensive Continuous Improvement Plan(CCIP) Training Module 4 Funding Application.
Field Methods 2 When a GPS only won’t do: –Forestry canopies –Canyons (no reception or can’t reach) –Underwater –Small areas when you don’t have a sub-
Indicator and Warning.
U.S. Geological Survey Office of Surface Water.
Powertap User Guide.
Microsoft Word 2010 Lesson 1.
Surveying Techniques II. GPS
NextGen Utility Billing
By Sanjay and Arvind Seshan
Comparing NFIE RAPID models with measured river flood hydrographs.
The Global Positioning System
Module 1b – ICIS Permitted Features
Presentation transcript:

OTT ADC Acoustic Digital Current Meter: Description of meter and Results of USGS testing Dan Wagner Hydrologist Arkansas Water Science Center Hydroacoustics Webinar May 24, 2010

Outline Description of ADC Results of USGS Testing New USGS Firmware / enhancements In this presentation, I’ll describe the ADC, discuss results of USGS testing of the ADC, and show enhancements that have been made to the ADC with a new “USGS Firmware,” currently in Beta testing.

ADC History Released in USA December 2007 Testing by HIF, HAWG, and WSCs from 2008-2010 New USGS Firmware currently in Beta Testing USGS testing was performed by HIF, HAWG, and Pennsylvania, Illinois, Florida, Arkansas, Alaska, Oregon, and Maine WSCs. New Firmware version

System Components Figure courtesy of OTT 2 6 MHZ transducers Thermistor Pressure sensor- 0 to 5m (16.4 ft) depth Meter will work in 0.15’ of water (2”) Figure courtesy of OTT

Handheld Display Mini-USB to USB cable for downloading w/ PC (no serial cables!) Robust sensor cable connection Big keys, beep when key pressed Separate function keys for toggling through menus Sealed, battery accessible only by opening case (send to factory) Rechargeable battery not accessible unless sent back to factory-still a concern amongst some users, would like to see an accessible battery compartment like with a laptop, where the rechargeable battery could be replaced. Battery life should be 20+ hours of use, 500 charging cycles (basically, days without charging and several years’ worth of charges) Figure courtesy of OTT

Handheld Display Operation Depending on the display screen, commands appear above the function keys-press the appropriate key to enter the command Figure courtesy of OTT

Display Screen Note the use of familiar symbols—PC-like progress bar, cell-phone like “signal quality” bars for correlation, battery symbol (9,1,2) Note that the ADC tells you a target depth and the actual depth so that the user can position the sensor at the appropriate position in the vertical without even having to look at the wading rod (3,4) Also the ADC tells you the distance of your vertical from the start point and shows a graph of the same (6,7) Velocity is in live streaming mode unless you turn that off…it’s nice for checking for QA/QC probs behind cobbles and boulders Figure courtesy of OTT

Location of Sample Volume Sample volume still 10 cm away from transducers, but in front of meter instead of to the side. Sample volume much larger than that of the FlowTracker…this coupled with the higher acoustic frequency means the ADC should have much less SNR issues in low backscattering environments. ADV ADC

Measuring Principle Two 6 MHz acoustic transducers Large sample volume located in front of sensor Large sample volume reduces potential signal-to- noise (SNR) issues, reduces error with flow angles Two sample volumes Improves data precision Internal cross correlation calculation verifies measurement accuracy and identifies potential boundary issues 4 in 2 in Sensor head Sample volume 2 Sample volume 1 0.4 in Each transducer works both as a transmitter and receiver Large sample – which is controlled by the receiving time window of the timing controller Large sample volume reduces signal to noise issues because scatters, or particles in the water column, contribute a stronger echo Two sample volumes – velocity measurement from the sample volumes are averaged together to produce a single in-line velocity measurement. Using two sample volumes improves precision by averaging measurements collection from each and makes it possible to offset the sample volume from the centerline of the probe Robust design that reduces mechanical impact Figure courtesy of OTT

Measuring Principle 1. Transmit – Receive - Digitize Break Cross Correlation Check for similarities Calculation of time shift An ultrasonic burst is sent into the media from beam 1 and beam 2 simultaneously All particles in the measurement path reflect a small amount of the ultrasonic signal Depending on the shape and size a particular signal results The multitude of the reflected signals result in a reflection pattern which is then saved to the DSP After a certain amount of time a second burst is sent into the medium A second reflection pattern is generated. This pattern is distorted due to slightly different reflections. This is because some particles have been turning around and have another reflection or some particles are no longer in the measurement range. The DSP checks both received reflection patterns for similarities using the cross correlation method All existing signal differences are rejected so that two similar but offset signal patterns are left for velocity evaluation Dt Calculation of flow velocity Slide courtesy of OTT

Depth Sensor Calibrate by zeroing to atmospheric pressure on site (raise out of water to calibrate) Get depth for each vertical by first lowering meter to bottom, rod offset computed automatically Hand-held display shows target depth and actual depth for velocity measurement in vertical depending on the velocity method selected by the user. Depth sensor can be turned off with new firmware The depth sensor is easy to use, but some testers didn’t like it, and wanted to be able to enter the depths manually as they do with other current meters. The new USGS firmware allows the user to turn off the depth sensor prior to starting the discharge measurement. One application of the depth sensor is in making mid-section measurements from a bridge with a long rod (see picture). While not commonly used in the USA, this practice is common in Europe.

QA/QC Figure courtesy of OTT

QA/QC checks With new USGS firmware, a Quick QA/QC check is performed in office (similar to ADV beam check, can print results) Pre-measurement QA/QC check is logged with measurement file and appears on printout Prior to the recent release of the new USGS firmware, the ADC had no logged QA/QC check.

Screen shots of the handheld display during QA/QC and depth sensor calibration procedures

Qreview Software-Data Download Select from files on ADC on left hand side…right hand side shows ADC files currently in selected directory in PC. Nice windows format

Measurement Review This is the new & improved look of Qreview…note the color-coded %Q/vertical and the QA/QC coloration on the Depth window. Use the edit function to delete verticals, change vertical depths (if they were incorrectly entered)

QA/QC check as it appears in QReview

Arkansas Test Sites Location of Arkansas’ test sites, in the NW part of the state. Waters typically clear with low suspended solids (and typically low SNRs when measured with ADV), bedrock lined channels w/ boulders, cobbles, gravel. N

Vmean, Test Measurements Site Vmean,ADV Vmean,ADC % error Mud Creek 0.03 0.014 -53.3 Richland Creek 0.39 0.45 +15.4 Brush Creek 0.46 0.42 -8.7 Baron Fork 0.49 0.44 -10.2 Flint Creek nr. West Siloam Springs 0.60 0.58 -3.3 Flint Creek nr. Springtown 0.80 0.77 The largest percent differences in Arkansas’ test measurements were at low velocities…and the absolute differences were small.

Vmean, Test Measurements Site Vmean,ADV Vmean,ADC % error Illinois River nr. Savoy 0.84 0.90 +7.1 Bear Creek 0.85 -1.2 Middle Fork White River 1.22 1.14 -6.6 Osage Creek 1.47 1.45 -1.4 Yocum Creek 1.97 1.95 -1.0 White River nr. Fayetteville 2.04 2.05 +0.49

Percent Difference of Velocity Measured with ADC Relative to Velocity Measured with ADV Plotting percent difference of mean velocity measured with the ADC relative to that measured with the ADV versus the mean velocity measured with the ADC for Arkansas’ test measurements produces a trend similar to other tests (see next slide)

Same as previous slide but for all test measurements Same as previous slide but for all test measurements. Keep in mind that most of the measurements were compared with measurements made with SonTek FlowTracker ADVs, and as with most comparisons, we don’t really know the “TRUE” velocity or discharge. Figure courtesy of Kevin Oberg, OSW

Basically, the same plot as previous slide but showing % difference of ADC from ADV vs. Discharge instead of velocity. Figure courtesy of Kevin Oberg, OSW

Regressing the mean velocities from ADC test measurements against those from reference measurements produces a very linear relationship, as we would expect. Figure courtesy of Kevin Oberg, OSW

Depths measured with the ADC’s depth sensor often differed slightly from those recorded by technicians in ADV or Pygmy meter measurements. Keep in mind that wading rods have error and sometimes bases are not screwed on all the way…also the technician is making a judgement call on anything less than a 0.1’ resolution in depth, possibly introducing error. Figure courtesy of Kevin Oberg, OSW

Initial HIF tow-tank testing indicated that at low velocity, measurements with the ADC were biased low. OTT has since put much effort into correcting the low velocity bias with new firmware fixes and recent tests at HIF indicate that these fixes have improved the low velocity comparisons.

Notice that the percent error of ADC velocity doesn’t exceed 5% between -50 and +70 degrees…the ADC handles flow angles well!

Same as previous slide, but horizontal angles. Point is, Figure courtesy of Kirk Thibideaux, HIF

ADC Pros/Cons-prior to USGS firmware Depth Sensor Sample volume in front of meter Rugged Construction, cabling Rechargeable battery 2 beams-improved performance in environments with low SNR, angles Live velocity reading-can help with meter placement Depth Sensor Sample volume in front of meter Rechargeable battery (?) Can’t edit erroneous stations until reviewing the measurement in Qreview Lack of beam check (“bucket test”) and/or quick field QA/QC check that is logged with measurement Lack of QA/QC warnings during measurement (i.e., for 0.2/0.8 velocity discrepancies, bad correlation resulting from bad boundary conditions -The depth sensor was welcomed by some users, despised by others. The new USGS firmware allows it to be turned off. -The rechargeable battery seems to perform very well. Some users remain skeptical of its long-term performance and the fact that the ADC has to be sent to the OTT factory in Loveland, Colorado for battery replacement. -The sample volume being in front of the meter was helpful in some stream bottom conditions but faces the same issues as a FlowTracker does when placed less than 10 cm from an obstruction or a vertical wall. -The new USGS firmware didn’t address the issue of being able to delete an erroneous station (i.e., technician enters the wrong depth for a vertical, with the ADV, they can toggle back to that vertical and make the change during the discharge measurement; with the ADC, these changes can only be made in Qreview) -Prior to the new USGS firmware, the ADC had no pre-deployment beam check or field QA/QC check that was logged with the measurement. Also, the old firmware lacked QA/QC warnings during the discharge measurement that would alert the technician to possible QA/QC issues.

Software Enhancements Enhanced quality assurance Including pre-deployment quality check and field quality check Ability to enter rated Q Expanded character limits for station name and number and addition of measurement number Computation and recording of averaged flow angles On-screen warning messages, a.k.a., Measurement Guidance Provide user with options to improve measurement results Option to turn off depth sensor Enhanced measurement review Detailed discharge and quality check reports Warning and potential quality issue logs One-second time series data OSW compiled a list of desired enhancements to the ADC based on tester’s comments. This is a list of the software enhancements that are part of the new “USGS Firmware.”

New ADC QA/QC Warnings The QA/QC warnings in the new firmware are similar to those provided in the familiar ADV Figure courtesy of OTT

Questions? HIF does not currently stock the ADC, but plans to move towards carrying the ADC in their warehouse after software improvements to help low velocity bias are tested.